K Number
K063334

Validate with FDA (Live)

Date Cleared
2006-12-08

(32 days)

Product Code
Regulation Number
878.4040
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Shanghai Ganghai Model GIKO 1400 Healthcare Particulate Respirator And Surgical Mask is A NIOSH Approved N95 Single use, disposable, LATEX FREE NON-sterile product. It is A NON-Fiberglass, fluid Resistant MASK intended To be worn by operating Room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from TRANSFER of MICROORGANISMS BODY FLUIDS AND PARTICULATE Material. This device Also meets CDC Guidelines for TB Exposure Control.

Device Description

The Shanghai Ganghai Type N95 Healthcare Particulate Respirator and Surgical Mask Model GIKO 1400 is of a duckbill style. It constructed from a white nonwoven material used in the inner and outer shell. The polypropylene melt blown filter media is layered between the inner and outer shell. The headband is made of elastic rubber. The inside nosepiece utilizes a closed cell foam and the outside nosepiece which conforms to the nose is made of plastic coated aluminum.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study details for the Shanghai Ganghai Type N95 Healthcare Particulate Respirator and Surgical Mask Model GIKO 1400, based on the provided 510(k) summary:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria / Identified RiskPerformance Standard / Mitigation MeasureReported Device Performance
Particulate Filtration EfficiencyNIOSH 42 CFR 84 - 0.3 micron diameter challenge, 95% efficiencyNIOSH Approved (TC-84A-4282) - Meets 95% efficiency against 0.3 micron particulate.
Fluid ResistanceASTM F1862-00 Standard Test Method for Resistance of Medical Face Mask to Penetration by Synthetic BloodResistant to synthetic blood (Nelson Laboratories conducted).
Breathing ResistanceNIOSH 30 CFR 11 section 11.140-9Tested as per NIOSH 30 CFR 11 section 11.140-9. (Implicitly passed, as it stated "passed" generally).
Flammability16 CFR Part 1610Flammability testing conducted by Nelson Laboratories (Implicitly passed, as it stated "passed" generally).
Material BiocompatibilityCytotoxicity Agar Overlay, Sensitization Buehler MethodTested (Implicitly passed, as it stated "passed" generally).
Bacterial Barrier (Inadequate barrier for bacteria)NIOSH TC-84A-4282 (N95 approval implies bacterial filtration)NIOSH approved (TC-84A-4282).
Air Exchange (Inadequate air exchange)NIOSH TC-84A-4282 (N95 approval implies adequate breathability)NIOSH approved (TC-84A-4282).
TB Exposure ControlCDC Guidelines for TB Exposure ControlMeets CDC Guidelines for TB Exposure Control.

2. Sample Size Used for the Test Set and Data Provenance

The document does not explicitly state the specific sample size for each individual test (e.g., how many masks were tested for particulate filtration, fluid resistance, etc.). The testing was performed by NIOSH (for N95 certification) and Nelson Laboratories (for other tests). The data provenance is derived from these testing bodies. It's a retrospective evaluation of the device against established standards rather than a prospective clinical trial.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

This information is not applicable in the context of this device. The "ground truth" here refers to established regulatory standards and performance criteria (e.g., 95% filtration efficiency, resistance to synthetic blood). These are objective measurements against published standards, not subjective assessments by human experts.

4. Adjudication Method for the Test Set

This information is not applicable. As mentioned above, the "ground truth" is based on objective testing against predefined standards, not on expert consensus requiring adjudication.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This is not applicable. This document describes a medical device (respirator mask), not an AI-powered diagnostic tool, and therefore no MRMC study with human readers and AI assistance would have been performed or reported here.

6. If a Standalone (i.e. algorithm only, without human-in-the-loop performance) was done

This is not applicable. This document describes a medical device (respirator mask), not an algorithm.

7. The Type of Ground Truth Used

The ground truth used is primarily objective performance standards and regulatory guidelines established by entities like NIOSH, ASTM, and CDC. This includes:

  • NIOSH 42 CFR 84: For particulate filtration and breathing resistance.
  • ASTM F1862-00: For resistance to synthetic blood.
  • 16 CFR Part 1610: For flammability.
  • Biocompatibility Standards: For cytotoxicity and sensitization.
  • CDC Guidelines for TB Exposure Control.

8. The Sample Size for the Training Set

This is not applicable. This document describes a medical device for which "training sets" in the AI sense are not relevant. The device is manufactured according to specifications and tested against established performance criteria.

9. How the Ground Truth for the Training Set Was Established

This is not applicable. As explained above, there is no "training set" in the context of this medical device. The "ground truth" for evaluating the mask's performance is based on the aforementioned objective performance standards and regulatory guidelines.

{0}------------------------------------------------

510(K) Summary

KOC3334

Submitter Joseph Z. Zdrok & Associates 7380 36 TH Court Vero Beach, FL. 32967

DEC - 8 2006

Contact Joseph Z. Zdrok President 772-778-0240 Telephone 772-778-0240 FAX izzdrok@aol.com E-Mail

Nov.6, 2006 Date

Trade Name

Shanghai Ganghai Purifying Products Co. Ltd. Type N-95 Healthcare Particulate Respirator and Surgical Mask Model GIKO 1400

Common Name: Healthcare Particulate Respirator and Surgical Mask

Classification Device Class - Class II CFR Section - 21 CFR 878.4040

Substantial Equivalency:

The Model GIKO 1400 Heathcare Particulate Respirator and Surgical Mask is found to be substantially equivalent to the Gerson Model 2130 Type N95 Mask . (510(K) No. K050193. Both products have been tested and approved by NIOSH as N-95 respirators.

{1}------------------------------------------------

USE

Description:

The Shanghai Ganghai Type N95 Healthcare Particulate Respirator and Surgical Mask Model GIKO 1400 is of a duckbill style. It constructed from a white nonwoven material used in the inner and outer shell. The polypropylene melt blown filter media is lavered between the inner and outer shell. The headband is made of elastic rubber . The inside nosepiece utilizes a closed cell foam and the outside nosephere which conforms to the nose is made of plastic coated aluminum.

Description of Device Requirements

The Shanghai Ganghai Type N95 Healthcare Respirator and Surgical Mask Model GIKO !400 is approved by NIOSH as per 42 CFR 84. The certification number assigned is TC-84A-4282 for a type N95 Particulate Respirator. The Type N95 must meet the prescribed test criteria which specifies the use of 0.3 micron diameter challenge and requiring a 95% efficiency. This mask is resistant to synthetic blood as per ASTMF1862-00 Standard Test Method for Resistance of Medical Face Mask to Penetration by Synthetic Blood, conducted by Nelson Laboratories. Breathing resistance was tested as per NIOSH 30 CFR11 section 11.140-9. Flammability testing as per16CFR Part1610 was conducted by Nelson Laboratories.

Intended Use:

The Shanghai Ganghai Type N95 Healthcare Particulate Respirator and Surgical Mask Model GIKO 1400 is intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material.

This device also meets CDC Guidelines for TB Exposure Control.

Limitations:

This product does not eliminate the wearer from any risk of contracting any type of disease or infection. The mask should be changed immediately if contaminated with blood or body fluids.

Comparison of Predicate Device

The outside cover stock color of the previously cleared device is white. The Shanghai Ganghai Model GIKO 1400 Healthcare Particulate Respirator and Surgical Mask Model is also white. The headband color of the cleared device is yellow and the Shanghai Ganghai Model GIKO 1400 is white.

{2}------------------------------------------------

The Shanghai Ganghai Model GIKO 1400 Healthcare Particulate Respirator and Surgical Mask incorporates a highly efficient filter media and is 95% efficient against a .3 micron particulate which was scientifically established as the most penetrating particle size. The legally marketed Gerson device previously cleared 510 (k) K050193 NIOSH T C -84A-4123 is manufactured from similar materials.

Risks to Health

The FDA Guidance Document was used to identify the risks to health associated with the use of surgical masks. The device was tested as per the mitigation measures listed below.

Identified RiskMitigation Measures
Inadequate fluid resistanceASTM F1862
Inadequate barrier for bacteriaNIOSH TC-84A-4282
Inadequate air exchangeNIOSH TC-84A-4282
Flammability16CFR Part 1910
Biocompatibility for patient/usercontacting materialsCytotoxicity Agar Overly SensitizationBuehler Method

Safety/Effectiveness

The device has a filtration equivalent to the previously cleared Gerson 2130 Healthcare Particulate Respirator and Surgical Mask Respirator. It is NIOSH approved and meets the CDC guidelines for TB Exposure Control. The GIKO 1400 was subjected to the tests listed and passed.

Conclusion:

Since the basic construction is used in the FDA cleared device as in the new device and approved by NIOSH, the GIKO 1400 Shanghai Ganghai Healthcare Particulate Respirator and Surgical Mask is substantially equivalent to the Gerson Model 2130 Healthcare and Particulate Respirator and Surgical Mask

{3}------------------------------------------------

Physical Characteristics

MATERIALS

PartGIKO 1400Gerson2130
Inside ShellNonwovenNonwoven
FilterNonwovenNonwoven
Outside ShellNonwoven (White)Nonwoven (White)
HeadbandElastic (White)SyntheticRubber (Yellow)
Inside NosepieceClosed Cell FoamClosed Cell Foam
Outside NosepiecePlastic Coated AluminumPlastic Coated Aluminum

Physical Measurements

Physical AttributesGIKO 1400Gerson 2130
Mask Thickness0.038 Inches0.037 Inches
Surface Area27.0 Square inches27.0 Square inches
Headband Length24 Inches Total ContinuousLoop12 Long 10 Short Inches
Facepiece Length7.0 Inches7.0 Inches

Mask Style

.

------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------and count of the County of Concess of Con-STITCC------------------------------------------------------------------------------------------------------------------------------------------------------------------------------THE TIME THE PERSONAL PROPERTY PLANE PROPERTY PLANE PROPERTY PROPERTY PRODUCTION PRODUCTION CONTRACT COLLEGIAL PRODUCTION CONTRACT COLLEGIAL PRODUCTION CONTRACT CONTRACT CALLA 1 4 0iersondel very by the starter development to be and the many of the many of the may be theAnd Amore of AmericanANNUAL CARAN
A STATUTE A BREAK A BREAK A BELLER AND A BREAK A1tJuckof a would to would be a la comment of a long to any andProperty of the been of the first of the Artist ofSTATUTE AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND A BREACH AND A BREACH AND A BREACH AND A BREACH AND A BREACH AND A BREACH AND A BR- D:1'nek------------------------------------------------------------------------------------------------------------------------------------------------------------------------------and and and a desimilation and experienced and

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle. The logo is black and white.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

DEC - 8 2006

Shanghai Gangkai Purifying Products Corporation, Limited C/O Mr. Joseph Z. Zdrok Consultant 7380 36th Court Vero Beach, Florida 32967

Re: K063334

Trade/Device Name: Shanghai Ganghai Type N95 Model GIKO 1400 Healthcare Particulate Respirator and Surgical Mask Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: II Product Code: MSH Dated: November 28, 2006 Received: November 29, 2006

Dear Mr. Zdrok:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{5}------------------------------------------------

Page 2 -- Mr. Zdrok

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Syuta Y. Mohani Dmd.

Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

Image /page/6/Picture/0 description: This image shows the logo of the U.S. Food and Drug Administration. The text "U.S. Food and Drug Administration" is prominently displayed, along with the text "CENTER FOR DEVICES AND RADIOLOGICAL HEAL". At the bottom of the image, there is a navigation bar with links to "FDA Home Page", "CDRH Home Page", "Search", "CDRH A-Z Index", and "Contact CDRH".

Manchester United

Indications for Use 51 16k) Number (if known): K 06834 Anghat Type Not Choose ( Device Name . Indications for I is he Shanghai Garghis Model GIKO HOO HEALTheARe PARTicuLate Respirator And SURGICAL MASK is A NIOSA AppRoved NAS Single use, disposable, LATEX FRE NON-steriLe product. It is A NON-Fiberghass, ALuid Resistant MASK intended To be worn by operating Room personnel during surgical procedures to protect both the surgical patient and t operating room personnel from TRANSACR of HICROORDANISMS DODY ALLINGS AND PARTICULATE Interial . This device ALso meets CDC Guidelines for TB Exposure Control .

Prescription Use (Pari 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use.
(21 CFR 801 Subpart C)

Anas Glesiology, General Hospital

I Control, Dental Devices

(FLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE

Hill

http://www.fda.gov/cdrh/ode/INDICATE HTML

§ 878.4040 Surgical apparel.

(a)
Identification. Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.(b)
Classification. (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.