K Number
K063167

Validate with FDA (Live)

Manufacturer
Date Cleared
2007-02-06

(111 days)

Product Code
Regulation Number
868.1400
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The EMMA Emergency Capnometer Monitor measures, displays and monitors carbon dioxide concentration and respiratory rate during anesthesia, recovery and respiratory care. It may be used in the operating suite, intensive care unit, patient room, clinic, emergency medicine and emergency transport settings for adult and pediatric patients.

The EMMA Emergency Capnometer Analyzer measures and displays carbon dioxide concentration and respiratory rate during anesthesia, recovery and respiratory care. It may be used in the operating suite, intensive care unit, patient room, clinic, emergency medicine and emergency transport settings for adult and pediatric patients.

Device Description

The EMMA Emergency Capnometer is a miniature mainstream infrared gas analysis bench with an integrated user interface. The complete carbon dioxide analyzer is contained within a transducer that is attached to the breathing circuit via the EMMA Airway Adapter.

AI/ML Overview

The provided 510(k) summary for the EMMA Emergency Capnometer focuses on establishing substantial equivalence to predicate devices rather than providing detailed acceptance criteria and a standalone study with specific performance metrics. Therefore, many of the requested details are not explicitly present in the document.

Here's an analysis based on the available information:

Acceptance Criteria and Device Performance

The documentation does not provide specific quantitative acceptance criteria (e.g., a target accuracy range for CO2 concentration or respiratory rate) with corresponding reported device performance. Instead, it states a general conclusion about equivalence or superiority.

The key statement is from section 12: "The EMMA Emergency Capnometer demonstrated performance, safety and effectiveness equivalent or superior to its predicates in all characteristics."

Acceptance CriteriaReported Device Performance
Not explicitly stated as quantitative targets. The implicit acceptance criteria are that the device's performance, safety, and effectiveness are at least equivalent to or superior to the predicate devices."Demonstrated performance, safety and effectiveness equivalent or superior to its predicates in all characteristics."

Study Information

  1. Sample size used for the test set and the data provenance:

    • Sample Size: Not specified.
    • Data Provenance: The study involved "calibrated gas samples and legally marketed anesthesia and ventilation devices." This suggests a laboratory or bench testing environment. There is no information regarding country of origin or whether the data was retrospective or prospective in a clinical setting.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Number of Experts: Not mentioned.
    • Qualifications of Experts: Not mentioned. The ground truth was likely established by the "calibrated gas samples" (known concentrations) and possibly measurements from the "legally marketed anesthesia and ventilation devices" (which would have their own validated measurement capabilities).
  3. Adjudication method for the test set:

    • Not applicable/Not mentioned. The study appears to be a direct comparison against a known standard (calibrated gas) and established medical devices, rather than a subjective assessment requiring expert adjudication.
  4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Was an MRMC study done? No. This device is an emergency capnometer, a medical measurement device, not an AI-powered diagnostic tool requiring human reader interpretation. The study described is a device-to-device performance comparison.
  5. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    • Yes, implicitly. The testing described ("Testing in direct comparison to predicates throughout the operating range was conducted using calibrated gas samples and legally marketed anesthesia and ventilation devices") is a standalone performance test of the device itself, without human interpretation of its outputs being part of the primary evaluation.
  6. The type of ground truth used:

    • Ground Truth Type: A combination of "calibrated gas samples" (representing a known, accurate CO2 concentration) and measurements from "legally marketed anesthesia and ventilation devices" (which are themselves considered accurate and reliable benchmarks). This is akin to a reference standard or benchmarking against established, validated measurement systems.
  7. The sample size for the training set:

    • Not applicable/Not mentioned. This device does not appear to be an AI/machine learning product requiring a training set in the conventional sense. Its development and validation are based on established engineering principles for gas analysis.
  8. How the ground truth for the training set was established:

    • Not applicable/Not mentioned, as there is no mention of a training set for an AI/ML model.

Summary of Device Performance and Equivalence Claim:

The 510(k) submission for the EMMA Emergency Capnometer establishes substantial equivalence primarily through direct comparative testing against its predicate devices (Tidal Wave Model 610, Novametrix Medical Systems Inc. and VEO Multigas Monitor for Pocket PC, Phasein AB). The "testing vs. predicates" section (11) indicates that the device was compared using "calibrated gas samples and legally marketed anesthesia and ventilation devices" across its operating range. The conclusion (12) states that the EMMA Emergency Capnometer "demonstrated performance, safety and effectiveness equivalent or superior to its predicates in all characteristics." This implies that the device's accuracy in measuring CO2 concentration and respiratory rate was found to be at least as good as, if not better than, the predicate devices when tested under controlled conditions.

{0}------------------------------------------------

K063167

510(k) SUMMARY

  • 510(k) Owner Name and Address: 1. PHASEIN AB Svärdvägen 15 182 33 Danderyd Sweden Telephone: 46-8-544-98-150 Fax: 46-8-544-98-169

  • Contact Person: 2. David Weissburg Weissburg Associates Madison, Wisconsin Telephone: 1-608-770-0223

    1. Date: 28 December 2006
  • 6 2007 FEB

  • Trade Name: EMMA Emergency Capnometer 4.

  • Common Name: Carbon Dioxide Gas Analyzer ട.

    1. Classification Names:
    • a. Carbon dioxide gas analyzer (21 CFR 868.1400, Product Code CCK)
    1. Substantially equivalent to:
    • a. Tidal Wave Model 610, Novametrix Medical Systems Inc. (K963327)
    • VEO Multigas Monitor for Pocket PC, Phasein AB. (K051857) ﻗ
  • Device description: The EMMA Emergency Capnometer is a miniature mainstream 8. infrared gas analysis bench with an integrated user interface. The complete carbon dioxide analyzer is contained within a transducer that is attached to the breathing circuit via the EMMA Airway Adapter.

  • ക Intended Use:

The EMMA Emergency Capnometer Monitor measures, displays and monitors carbon dioxide concentration and respiratory rate during anesthesia, recovery and respiratory care. It may be used in the operating suite, intensive care unit, patient room, clinic, emergency medicine and emergency transport settings for adult and pediatric patients.

The EMMA Emergency Capnometer Analyzer measures and displays carbon dioxide concentration and respiratory rate during anesthesia, recovery and respiratory care. It may be used in the operating suite, intensive care unit, patient room, clinic, emergency medicine and emergency transport settings for adult and pediatric patients.

    1. Comparison to predicates: The EMMA Emergency Capnometer combines carbon dioxide measurement capabilities and enhanced portability. The EMMA Emergency Capnometer uses the same basic technology concepts of the predicate devices, while adding improvements derived from advanced electronics and miniaturization. The intended use of the EMMA Emergency Capnometer and its predicates are equivalent. All the devices utilize disposable single-patient-use airway adapters to interface with gases in the breathing circuit. Labeling and materials used are equivalent.
    1. Testing vs. predicates: Testing in direct comparison to predicates throughout the operating range was conducted using calibrated gas samples and legally marketed anesthesia and ventilation devices.
    1. Conclusions from testing: The EMMA Emergency Capnometer demonstrated performance, safety and effectiveness equivalent or superior to its predicates in all characteristics.

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is an abstract image of an eagle with its wings spread.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Phasein AB C/O Mr. David Weissburg Weissburg Associates 4213 Winnequah Drive Madison. Wisconsin 53716

FEB 6 2007

Re: K063167

Trade/Device Name: EMMA Emergency Capnometer Regulation Number: 21 CFR 868.1400 Regulation Name: Carbon Dioxide Gas Analyzer Regulatory Class: II Product Code: CCK Dated: January 3, 2007 Received: January 10, 2007

Dear Mr. Weissburg:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{2}------------------------------------------------

Page 2 - Mr. Weissburg

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely vours,

Chiu Lin, Ph.D.

Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number: _

EMMA Emergency Capnometer Device Name:

Indications for Use:

The EMMA Emergency Capnometer Monitor measures, displays and monitors carbon dioxide concentration and respiratory rate during anesthesia, recovery and respiratory care. It may be used in the operating suite, intensive care unit, patient room, clinic, emergency medicine and emergency transport settings for adult and pediatric patients.

The EMMA Emergency Capnometer Analyzer measures and displays carbon dioxide concentration and respiratory rate during anesthesia, recovery and respiratory care. It may be used in the operating suite, intensive care unit, patient room, clinic, emergency medicine and emergency transport settings for adult and pediatric patients.

Prescription Use (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE

IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

K063167

Page 1 of

§ 868.1400 Carbon dioxide gas analyzer.

(a)
Identification. A carbon dioxide gas analyzer is a device intended to measure the concentration of carbon dioxide in a gas mixture to aid in determining the patient's ventilatory, circulatory, and metabolic status. The device may use techniques such as chemical titration, absorption of infrared radiation, gas chromatography, or mass spectrometry.(b)
Classification. Class II (performance standards).