(109 days)
Indications:
- Correction of revision of unsuccessful osteotomy, arthrodesis, or previous joint replacement.
- Tumor resections.
- Revision of previously failed total joint arthroplasty.
- Trauma.
The Compress® Segmental Femoral Replacement System (Short Spindle and Anchor Plug) components are intended for uncemented use only.
The Compress® Seqmental Femoral Replacement System (Short Spindle and Anchor Plug) utilizes a spring-enhanced and stress-sharing design; to generate stress on the host bone which helps prevent bone atrophy and promotes bone growth. It is intended to be used in conjunction with a knee or hip implant. The device consists of the three main components, the Anchor Plug, Spindle, and the Taper Adapter.
This document is a 510(k) summary for the Compress® Segmental Femoral Replacement System (Short Spindle and Anchor Plug) by Biomet Manufacturing Corp. It is for a medical device, which typically relies on demonstrating substantial equivalence to a predicate device rather than on performance criteria established through clinical studies for new indications.
Based on the provided text, here's an analysis of the acceptance criteria and study information:
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criteria | Reported Device Performance |
|---|---|
| Mechanical Testing | Successful mechanical testing indicated that the device is functional within its intended use and substantially equivalent to the predicate device in performance characteristics. |
| Technological Characteristics | The technological characteristics of the device are identical to the predicate device, including the spring-enhanced, stress-sharing design. |
| Dimensional Specifics | All dimensional specifics are similar to those of the predicate device. |
| Compatibility | Used in conjunction with the same hip or knee components as the predicate device. |
| Materials & Standards | Made from the same materials conforming to the same standards as the predicate device. |
Note: This submission relies on demonstrating substantial equivalence to a predicate device (Compress® Segmental Femoral Replacement System - K043547). The acceptance criteria are largely based on matching the design, materials, and mechanical performance of the predicate device. There are no specific quantifiable clinical performance metrics or acceptance thresholds provided beyond "successful mechanical testing" and "substantially equivalent."
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size for Test Set: Not applicable in the context of clinical testing for this 510(k). The evaluation was primarily based on non-clinical (mechanical) testing and a comparison to a predicate device.
- Data Provenance: The document explicitly states "Non-Clinical Testing: Successful mechanical testing on Compress® Segmental Femoral Replacement System (Short Spindle and Anchor Plug)..." This indicates laboratory-based testing rather than patient data. No country of origin is specified for this mechanical testing, and it is by nature retrospective (testing a finished product).
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications
- Not applicable. As no clinical data was used to establish performance against a "ground truth" for the test set, no experts were needed for this purpose. The substantial equivalence was evaluated by the FDA based on the provided technical documentation and comparison to the predicate.
4. Adjudication Method for the Test Set
- Not applicable. There was no clinical test set requiring adjudication.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
- No. The document explicitly states: "Clinical Testing: None provided as a basis for substantial equivalence." Therefore, no MRMC study was conducted or reported.
6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study
- Not applicable. This is a physical orthopedic implant, not an algorithm or AI-driven device.
7. Type of Ground Truth Used
- Mechanical Performance and Design Equivalence: For the non-clinical testing, the "ground truth" would be the established performance characteristics and design specifications of the predicate device (K043547) and relevant industry standards for orthopedic implants. The new device's mechanical test results were compared against these.
8. Sample Size for the Training Set
- Not applicable. This is a physical orthopedic implant; there is no "training set" in the context of machine learning or algorithms. Its design and manufacturing are based on established engineering principles and the predicate device's specifications.
9. How the Ground Truth for the Training Set Was Established
- Not applicable. As there is no training set, there is no ground truth to establish for it. The design and performance are validated against engineering principles and comparison to the existing predicate device.
{0}------------------------------------------------
510(k) Summary
| Preparation Date: | 1/8/2007JAN 19 2007 |
|---|---|
| Applicant/Sponsor: | Biomet Manufacturing Corp.56 East Bell DriveP.O. Box 587Warsaw, Indiana 46581-0587 |
| Contact Person: | Jing Xie, Ph.D.Biomet Manufacturing Corp. |
| Proprietary Name: | Compress® Segmental Femoral Replacement System (Short Spindle andAnchor Plug) |
| Common Name: | Segmental Femoral Stem Component |
| Classification Name: | Hip Joint, Metal/Polymer, Semi-Constrained, Cemented Prosthesis(21 CFR §888.3350)Knee Joint, Femorotibial, Metal/Polymer Constrained, Cemented Prosthesis(21 CFR §888.3510)The compatible Compress® Proximal Femoral Components, OSS® Systemcomponents, compatible heads, liners, shells, screws, knee femoral andtibial components included in this submission have the followingclassifications: |
| Subsequent Product Codes: | JDI, Hip Joint, Metal/Polymer, Semi-Constrained, Cemented Prosthesis(21 CFR §888.3350)JDL, Hip joint metal/metal semi-constrained, with a cemented acetabularcomponent, prosthesis (21 CFR §888.3320)KRO, Knee joint femorotibial metal/polymer constrained cementedprosthesis (21 CFR §888.3510)KWA, Hip joint metal/metal semi-constrained, with an uncementedacetabular component, prosthesis (21 CFR §888.3330)KWY, Hip joint femoral (hemi-hip) metal/polymer cemented oruncemented prosthesis (21 CFR §888.3390)KWZ, Hip joint metal/polymer constrained cemented or uncementedprosthesis (21 CFR §888.3310)LPH, Hip joint metal/polymer/metal semi-constrained porous-coateduncemented prosthesis (21 CFR §888.3358)LZO, Hip joint metal/ceramic/polymer semi-constrained cemented ornonporous uncemented prosthesis (21 CFR §888.3353)LZY, Hip joint (hemi-hip) acetabular metal cemented prosthesis (21 CFR§888.3370)MEH, Hip joint metal/ceramic/polymer semi-constrained cemented ornonporous uncemented prosthesis (21 CFR §888.3353) |
{1}------------------------------------------------
Biomet Manufacturing Corp. Compress® Segmental Femoral Replacement System (Short Spindle and Anchor Plug) Page 2 of 2
Legally Marketed Devices To Which Substantial Equivalence Is Claimed:
Compress® Segmental Femoral Replacement System - K043547 (Biomet Manufacturing Corp.)
Device Description: The Compress® Seqmental Femoral Replacement System (Short Spindle and Anchor Plug) utilizes a spring-enhanced and stress-sharing design; to generate stress on the host bone which helps prevent bone atrophy and promotes bone growth. It is intended to be used in conjunction with a knee or hip implant. The device consists of the three main components, the Anchor Plug, Spindle, and the Taper Adapter.
Intended Use:
Indications for Use:
-
- Correction of revision of unsuccessful osteotomy, arthrodesis, or previous joint replacement.
-
- Tumor resections.
-
- Revision of previously failed total joint arthroplasty.
-
- Trauma.
The Compress® Seqmental Femoral Replacement System (Short Spindle and Anchor Pluq) components are intended for uncemented use only.
Summary of Technologies: The Compress® Segmental Femoral Replacement System (Short Spindle and Anchor Pluq) utilizes the same spring-enhanced, stress-sharing design as the predicate Compress® device to generate stress on the host bone which helps prevent bone atrophy and promotes bone growth. The technological characteristics of the Compress® Segmental Replacement System (Short Spindle and Anchor Plug) is identical to the predicate device. All the dimensional specifics of these components are similar to those of the predicate device. The Compress® Segmental Femoral Replacement System (Short Spindle and Anchor Plug) is used in conjunction with the same hip or knee components as the predicate device. Finally, the Compress® Segmental Femoral Replacement System (Short Spindle and Anchor Plug) is made from the same materials conforming to the same standards as the predicate device.
Non-Clinical Testing: Successful mechanical testing on Compress® Segmental Femoral Replacement System (Short Spindle and Anchor Plug) indicated that the device is functional within its intended use and substantially equivalent to the predicate device in performance characteristics.
Clinical Testing: None provided as a basis for substantial equivalence.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features the department's emblem, which consists of a stylized caduceus-like symbol with three curved lines representing the human form. The emblem is positioned to the right of the text, which reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement around the emblem.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
APR 2 3 2009
Biomet Manufacturing Corp. % Jing Xie, Ph.D. Manager, Clinical Affairs P. O. Box 587 Warsaw, Indiana 46581-0587
Re: K062998
Trade/Device Name: Compress Segmental Femoral Replacement System (Short Spindle and Anchor Plug) Regulation Number: 21 CFR 888.3330 Regulation Name: Hip joint metal/metal semi-constrained, with an uncemented acetabular component, prosthesis Regulatory Class: III Product Code: KWA, JDL, KRO, KWZ, JDI, LZO, MEH, LPH, KWY Dated: January 12, 2007 Received: January 12, 2007
Dear Dr. Xie:
This letter corrects our substantially equivalent letter of January 19, 2007.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other
{3}------------------------------------------------
Page 2 - Jing Xie, Ph.D.
Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please contact the CDRH/Office of Surveillance and Biometrics/Division of Postmarket Surveillance at 240-276-3464. For more information regarding the reporting of adverse events, please go to http://www.fda.gov/cdrh/mdr/.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Mark N. Melkerson
Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Indications for Use
10(k) Number (if known): _____________________________________________________________________________________________________________________________________________________
bevice Name:
Compress® Segmental Femoral Replacement System (Short Spindle and Anchor Plug)
Indications for Use:
Indications:
-
- Correction of revision of unsuccessful osteotomy, arthrodesis, or previous joint replacement.
-
- Tumor resections.
-
- Revision of previously failed total joint arthroplasty.
-
- Trauma.
The Compress® Segmental Femoral Replacement System (Short Spindle and Anchor Plug) components are intended for uncemented use only.
Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use NO (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Valeria Bueno
(Division S Division of Ceneral, Destorative, and Neuro... Collection Cester
510(k) Number: K062998.
N/A