K Number
K062884

Validate with FDA (Live)

Date Cleared
2006-11-09

(44 days)

Product Code
Regulation Number
880.6250
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Powder free vinyl patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

Device Description

powder free vinyl patient examination gloves, Clear(non-colored) that meets all of the requirements of ASTM standard D 5250-00e4

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study information derived from the provided text:

1. Table of Acceptance Criteria and Reported Device Performance:

The device described is a medical device, specifically Powder-Free Vinyl Patient Examination Gloves, Clear (Non-Colored). The acceptance criteria for such a device primarily revolve around its physical properties, safety, and performance standards.

CharacteristicAcceptance Criteria (Standard)Reported Device Performance
DimensionASTM standard D 5250-00e4Meets
Physical PropertiesASTM standard D 5250-00e4Meets
Freedom from Pinholes21 CFR 800.20Meets
Powder ResidualASTM standard D 5250-00e4 and D6124-01<2mg/glove
BiocompatibilityPrimary Skin Irritation in rabbits (ISO10993-10)Passes (Not a Primary Skin Irritation)
Dermal sensitization in guinea pigs (ISO10993-10)Passes (Not a Dermal sensitization)

2. Sample Size for Test Set and Data Provenance:

  • Sample Size for Test Set: The document does not explicitly state the sample size used for the performance tests. It refers to meeting the standards, implying that the number of samples tested was sufficient to demonstrate compliance with ASTM standards and FDA regulations for medical gloves.
  • Data Provenance: The document does not specify the country of origin where the tests were conducted. The study is retrospective in nature, as it presents data and conclusions after the tests were performed to demonstrate compliance with pre-existing standards.

3. Number of Experts and Qualifications for Ground Truth:

This section is not applicable to this type of device and study. The "ground truth" for medical gloves is established by objective, standardized laboratory tests against recognized ASTM standards and FDA regulations, not by expert consensus or interpretation of specialized images or data.

4. Adjudication Method:

This section is not applicable for this type of device. Adjudication methods (like 2+1, 3+1) are used in studies where human interpretation is involved, such as reading medical images. For physical and chemical properties of a device like gloves, the results are definitive based on standardized tests.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

This section is not applicable. MRMC studies are typically used in imaging or diagnostic device evaluations to assess how a device affects human reader performance. This study focuses on the inherent physical and biocompatibility properties of the gloves themselves, not on improving human performance in a diagnostic task.

6. Standalone (Algorithm Only Without Human-in-the-Loop) Performance:

This section is not applicable. The device is a physical product (gloves) and does not involve any algorithms or artificial intelligence. Therefore, there is no "standalone" algorithmic performance to be evaluated.

7. Type of Ground Truth Used:

The ground truth used for this device's evaluation is primarily objective, standardized laboratory test results against established national and international standards (ASTM D5250-00e4, ASTM D6124-01, 21 CFR 800.20, and ISO10993-10). These standards define precise methodologies and acceptable ranges for characteristics like dimensions, physical properties, pinholes, powder residue, and biocompatibility.

8. Sample Size for Training Set:

This section is not applicable. The device is a physical product and does not involve any machine learning algorithms that would require a "training set."

9. How Ground Truth for Training Set Was Established:

This section is not applicable as there is no training set for this device.

{0}------------------------------------------------

Summary

NOV - 9 2006

"This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92."

"The assigned 510(k) number is: 1062-094" (applicant leave blank)

Premarket Notification [510(k)] Summary

[(a)(1)]. The summary contains on the first page, preferably on your letterhead paper, the submitter's name, address, phone and fax numbers, name of contact person, and date the summary was prepared :

Submitter's name :SUQIAN GREEN GLOVE CO., LTD
Submitter's address :NO.1 XIANGSHAN ROAD, EAST AREA OF SUYUDEVELOPMENT ZONE, SUQIAN CITY, JIANGSUPROVINCE, 223800, P.R China
Phone number :(86) 0527-4754999
Fax number :(86) 0527-4754998
Name of contact person:Mr. Zheng Gang
Date the summary was prepared:Sep.22.2006

((a)(2)]. The name of the device, including the trade or proprietary name if applicable, the common or usual name, and the classification name, if known

Device Name:Powder Free Vinyl Patient Examination Gloves, Clear(non-colored)
Proprietary/Trade name:Powder Free Vinyl Patient Examination GlovesOther clients private labeling
Common Name:Patient examination glove
Classification Name:Patient examination glove
Device Classification:I
Regulation Number:21 CFR 880.6250
Panel:General Hospital (80)
Product Code:LYZ

[(a)(3)]. An identification of the legally marketed device to which your firm is claiming substantial equivalence .

.. . ":

Class I* powder free vinyl patient examination gloves , Clear(non-colored) that meets all of the requirements of ASTM standard D 5250-00e4.

Predicate device : FUGUAN (Brand) Powder-Free Vinyl Patient Examination Gloves, Shijiazhuang Fuguan Plastic Products Co., Ltd.. K032908 .

{(a)(4)] A description of the device

{1}------------------------------------------------

Device Description : powder free vinyl patient examination gloves, Clear(non-colored) that meets all of the requirements of ASTM standard D 5250-00e4

[(a)(5)] The summary describes the intended use of the device

Device Intended Use: powder free vinyl patient examination glove, Clear(non-colored) is a disposable device intended for medical purposes that is worn on the examinor's hand on finger to prevent contamination between patient and examiner.

[(a)(6)] A summary of the technological characteristics of new device compared to the predicate device.

The powder free vinyl patient examination gloves, non sterile are summarized with the following technological characteristics compared to ASTM or equivalent standard.

CharacteristicsStandardDevice performance
DimensionASTM standard D 5250-00e4.Meets
Physical PropertiesASTM standard D 5250-00e4.Meets
Freedom from pinholes21 CFR 800.20Meets
Powder ResidualASTM standard D 5250-00e4Meets
Biocompatabilityand D6124-01Primary Skin Irritation inrabbits<2mg/glovePassesNot a Primary Skin Irritation
Dermal sensitization in theguinea pigPasses
Not a Dermal sensitization

((b)(1)) A brief discussion of the nonclinical submitted, reference, or relied on in the premarket notification submission for a determination of substantial equivalence .

Powder free vinyl patient examination gloves , Clear(non-colored) meet requirements per ASTM D5250-006, per ASTM D6124-01, per 21 CFR 800.20 and ISO10993-10.

((b)(2)) A brief discussion of the clinical submitted, reference, or relied on in the premarket notification submission for a determination of substantial equivalence .

Clinical data is not needed for gloves or for most devices cleared by the 510(k) process.

[(b)(3)] The conclusions drawn from the nonclinical and clinical tests that demonstrate that the device is as safe,as effective, and performed as well or better than the legally marte tate device identified in (a)(3).

It can be concluded that the Powder Free Vinyl Patient Examination Gloves meet the ASTM standard or equivalent standard and FDA requirements for waterleak test on pinhole AQL, meet labeling claims .

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle.

Public Health Service

NOV - 9 2006

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Suqian Green Glove Company Limited C/O Mr. Chu Xiaoan Beijing Easy-Link Company No. 209 Bei Si Huan Zhong Road Rm. 1606, Building 1, Jianxiang Yuan Beijing CHINA 100083

Re: K062884

Trade/Device Name: Powder-Free Vinyl Patient Examination Gloves, Clear (Non-Colored) Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: 1 Product Code: LYZ Dated: September 22, 2006 Received: September 26, 2006

Dear Mr. Xiaoan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 - Mr. Xiaoan

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely vours.

Uuey tholeom for
Chiu Lin, Ph.D.

Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

INDICATIONS FOR USE

Applicant:_SUQIAN_GREEN_GLOVE_CO., LTD

510(k) Number (if known):_* KO62884

Device Name:__Powder Free Vinyl Patient Examination Gloves, Clear(Non-colored)

Indications For Use:

Powder free vinyl patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

Prescription Use (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Shiela L. Murphy lo

f Anesthesiology, General Hospital.
Control, Dental Devices

Number: K06.2884

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.