K Number
K062872

Validate with FDA (Live)

Manufacturer
Date Cleared
2007-03-13

(168 days)

Product Code
Regulation Number
888.3030
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Synthes (USA) Pediatric LCP Hip Plate System is intended for fixation of fractures and osteotomies of the proximal femur in children, adolescents, and small statured adults.

Specific indications include:

  • intertrochanteric derotation and varus osteotomies (100° and . 110° plates)
  • femoral neck and petrochanteric fractures (120° plates) .
  • intertrochanteric valgus osteotomies (150° plates)
Device Description

The Synthes (USA) Pediatric LCP Hip Plate System is available in 3.5 and 5.0 mm versions which include 100, 110, 120, and 150 degree angles. The plates are designed with three conical locking screw holes located at the head in conjunction with 3 or 4 combination locked-compression screw holes at the shaft. The fixed angle construct is created by means of three standard locking screws inserted through the head of the plate. The shaft of the plate accepts locking or cortex screws, depending on the nature of the fracture or the quality of bone.

In addition, the plates are manufactured from 316L stainless steel and available both sterile and non-sterile.

AI/ML Overview

I am sorry, but the provided text does not contain information about acceptance criteria or a study that proves a device meets such criteria. The document is a 510(k) summary and corresponding FDA letter for the "Synthes (USA) Pediatric LCP Hip Plate System," which outlines the device description, intended use, and substantial equivalence to a predicate device. It does not include details about device performance metrics, study designs, sample sizes, expert qualifications, or ground truth establishment.

Therefore, I cannot generate the requested table and information.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the word "SYNTHES" in bold, black letters, with a registered trademark symbol to the right of the word. To the left of the word is a circular logo with a stylized image inside. The logo appears to be a representation of a person or figure. A thin line is present underneath the word "SYNTHES".

K062872

MAR 1 2007

Page 1 of of __ of ___________________________________________________________________________________________________________________________________________________________

3.0 510(k) Summary

Synthes (USA) Sponsor: 1301 Goshen Parkway West Chester, PA 19380 (610) 719-5000

Synthes (USA) Pediatric LCP Hip Plate System Device Name:

21 CFR 888.3030: Single/multiple component metallic bone Classification: fixation appliances and accessories

Synthes Angled Blade Plates Predicate Devices:

The Synthes (USA) Pediatric LCP Hip Plate System is available in Device Description: 3.5 and 5.0 mm versions which include 100, 110, 120, and 150 degree angles. The plates are designed with three conical locking screw holes located at the head in conjunction with 3 or 4 combination locked-compression screw holes at the shaft. The fixed angle construct is created by means of three standard locking screws inserted through the head of the plate. The shaft of the plate accepts locking or cortex screws, depending on the nature of the fracture or the quality of bone.

In addition, the plates are manufactured from 316L stainless steel and available both sterile and non-sterile.

The Synthes (USA) Pediatric LCP Hip Plate System is intended for Intended Use: fixation of fractures and osteotomies of the proximal femur in children, adolescents, and small statured adults. Specific intertrochanteric derotation and varus indications include: neck and osteotomies (100° and 110° plates), femoral petrochanteric fractures (120° plates), and intertrochanteric valgus osteotomies (150° plates).

Information presented supports substantial equivalence. Substantial Equivalence:

{1}------------------------------------------------

Image /page/1/Picture/12 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus, which is a symbol often associated with healthcare. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the caduceus.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Synthes (USA) % Ms. Deborah Jackson, RAC Regulatory Affairs Specialist 1301 Goshen Pkwy West Chester, PA 19380

MAR 1 3 2007

Re: K062872

Trade/Device Name: Synthes (USA) Pediatric LCP Hip Plate System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: II Product Code: JDS Dated: March 02, 2007 Received: March 05, 2007

Dear Ms. Jackson:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{2}------------------------------------------------

Page 2 - Ms. Deborah Jackson, RAC

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely vours.

Barbara Brehm

Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the word "SYNTHES" in bold, black letters. To the left of the word is a circular logo with a stylized image inside. A registered trademark symbol is to the right of the word.

Page 1 of 1

2.0 Indications for Use

510(k) Number (if known):

Device Name:

Indications for Use:

KO62872

Synthes (USA) Pediatric LCP Hip Plate System

The Synthes (USA) Pediatric LCP Hip Plate System is intended for fixation of fractures and osteotomies of the proximal femur in children, adolescents, and small statured adults.

Specific indications include:

  • intertrochanteric derotation and varus osteotomies (100° and . 110° plates)
  • femoral neck and petrochanteric fractures (120° plates) .
  • intertrochanteric valgus osteotomies (150° plates)

X Prescription Use (Per 21 CFR 801.109)

AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Barbave Brichum

Division of General, Restorative, and Neurological Devices

510(k) Number K062872