(28 days)
This device is indicated for patients with severe knee pain and disability due to:
• Rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis.
• Collagen disorders, and/or avascular necrosis of the femoral condyle.
· Post-traumatic loss of joint configuration, particularly when there is patellofemoral erosion, dysfunction or prior patellectomy.
· Moderate valgus, varus, or flexion deformities. · The salvage of previously failed surgical attempts or for a knee in which satisfactory stability in flexion cannot be obtained at the time of surgery.
Specific uses with LPS-Flex femorals: · Provides increased flexion capability for patients who have both the flexibility and desire to increase their flexion range. · The LPS-Flex femoral, when used with the LPS-Flex articular surfaces, is designed for use with both cruciate ligaments excised and when load bearing ROM is expected to be less than or equal to 155 degrees.
The LPS-Flex Fixed Bearing Knee is a semiconstrained, condylar system for use without the cruciate ligaments when additional stability is required to prevent anterior subluxation of the femur relative to the tibia in flexion.
This document is a 510(k) premarket notification for a medical device modification, specifically for the NexGen® Complete Knee Solution Legacy® Posterior Stabilized (LPS)-Flex Fixed Bearing Tivanium® Ti-6Al-4V Alloy Femoral Components. It seeks to demonstrate substantial equivalence to a predicate device.
Here's an analysis of the provided information concerning acceptance criteria and supporting studies:
-
A table of acceptance criteria and the reported device performance
This document does not explicitly list quantitative acceptance criteria in a table format in the way one might expect for a software or diagnostic device. Instead, the "acceptance criteria" are implicitly met by demonstrating substantial equivalence through non-clinical (laboratory) performance testing. The main criterion is that the modified device performs comparably to the predicate device and meets safety and effectiveness standards for its intended use.
Acceptance Criteria (Implicit) Reported Device Performance Device is safe and effective. The results of non-clinical (laboratory) performance testing demonstrate that the device is safe and effective. Device is substantially equivalent to the predicate device (NexGen® Complete Knee Solution Legacy® Posterior Stabilized (LPS)-Flex Fixed Bearing Femoral and Articular Surface Components, K991581). The results of non-clinical (laboratory) performance testing demonstrate that the device is substantially equivalent to the predicate device. The modifications (change in material and surface hardening processing) do not change the intended use or the fundamental scientific technology. Device is packaged using the same materials and processes. -
Sample sizes used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
The document states: "The results of non-clinical (laboratory) performance testing demonstrate that the device is safe and effective and substantially equivalent to the predicate device."
- Sample Size for Test Set: The document does not specify the sample size for the non-clinical (laboratory) performance testing.
- Data Provenance: The data provenance is "non-clinical (laboratory) performance testing." There's no information about the country of origin or whether it was retrospective or prospective, as these terms typically apply to clinical data.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
This is not applicable. The study conducted was non-clinical (laboratory) performance testing of a mechanical device (femoral components for a knee replacement). There was no "ground truth" to be established by human experts in the context of interpreting medical images or data. Performance was likely evaluated against engineering specifications and comparison to the predicate device's characteristics.
-
Adjudication method (e.g. 2+1, 3+1, none) for the test set
This is not applicable. Adjudication methods like 2+1 or 3+1 are used for establishing ground truth in studies involving human interpretation or clinical outcomes, typically for diagnostic devices or AI algorithms. This submission pertains to a mechanical orthopedic implant, where "adjudication" in this sense is not relevant.
-
If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
This is not applicable. An MRMC study is relevant for evaluating the impact of an AI algorithm on human reader performance, particularly in diagnostic imaging. The device in question is a knee implant component, not a diagnostic AI tool.
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
This refers to a mechanical device, not an algorithm. Therefore, "standalone algorithm only" testing is not applicable. The "standalone performance" was the non-clinical laboratory testing of the physical components.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc)
This is not explicitly stated in terms of a "ground truth" as it would be for a diagnostic device. For a mechanical implant, "ground truth" generally relates to adherence to design specifications, material properties, mechanical strength, wear characteristics, and biocompatibility, as confirmed through laboratory testing. The implicit ground truth would be that the modified device's performance characteristics (e.g., strength, durability, wear resistance) are equivalent or superior to the predicate device, and meet a predefined set of engineering and biocompatibility standards.
-
The sample size for the training set
This is not applicable. This device is a mechanical implant, not an AI/ML algorithm that requires a "training set."
-
How the ground truth for the training set was established
This is not applicable, as there is no training set for a mechanical implant.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the word "zimmer" with a stylized "Z" in a circle to the left of the word. The "Z" is black and bolded, and the circle is white with a black outline. The word "zimmer" is in a bold, sans-serif font. The image is in black and white.
Summary of Safety and Effectiveness
OCT 1 3 2006
| Submitter: | Zimmer, Inc.P.O. Box 708Warsaw, IN 46581-0708 | |
|---|---|---|
| Contact Person: | Anthony FrancalanciaSenior Associate, Regulatory AffairsTelephone: (574) 372-4570Fax: (574) 372-4605 | |
| Date: | September 14, 2006 | |
| Trade Name: | NexGen® Complete Knee Solution Legacy®Posterior Stabilized (LPS)-Flex Fixed BearingTivanium® Ti-6Al-4V Alloy Femoral Components | |
| Common Name: | LPS-Flex Fixed Bearing Tivanium FemoralComponents | |
| Classification Name: | Knee joint patellofemorotibialpolyethylene/metal/polyethylene semiconstrainedcemented total knee prosthesis | |
| Classification Reference: | 21 CFR § 888.3560 | |
| Predicate Device: | NexGen® Complete Knee Solution Legacy®Posterior Stabilized (LPS)-Flex Fixed BearingFemoral and Articular Surface Components,manufactured by Zimmer, Inc., K991581, clearedJuly 30, 1999 | |
| Device Description: | The LPS-Flex Fixed Bearing Knee is asemiconstrained, condylar system for use withoutthe cruciate ligaments when additional stability isrequired to prevent anterior subluxation of thefemur relative to the tibia in flexion. | |
| Intended Use: | This device is indicated for patients with severeknee pain and disability due to:• Rheumatoid arthritis, osteoarthritis, traumaticarthritis, polyarthritis.• Collagen disorders, and/or avascular necrosis ofthe femoral condyle. |
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the text "Special 510(k): Device Modification". The text is written in a bold, sans-serif font. The words "Special" and "Device Modification" are larger than the number "510(k)". The text is centered on the image.
· Post-traumatic loss of joint configuration,
dysfunction or prior patellectomy.
particularly when there is patellofemoral erosion,
· Moderate valgus, varus, or flexion deformities. · The salvage of previously failed surgical attempts or for a knee in which satisfactory stability in
Image /page/1/Picture/1 description: The image shows the logo for Zimmer Biomet. The logo consists of a large, stylized "Z" enclosed in a circle, followed by the word "zimmer" in lowercase letters. The "Z" is bold and black, and the word "zimmer" is also black and in a sans-serif font.
flexion cannot be obtained at the time of surgery. Specific uses with LPS-Flex femorals: · Provides increased flexion capability for patients who have both the flexibility and desire to increase their flexion range. · The LPS-Flex femoral, when used with the LPS-Flex articular surfaces, is designed for use with both cruciate ligaments excised and when load bearing ROM is expected to be less than or equal to 155 degrees. Comparison to Predicate Device: Except for a change in material and surface hardening processing, the LPS-Flex Fixed Bearing Tivanium Femoral Components are identical to the predicate device's femoral components. The modifications do not change the intended use or the fundamental scientific technology. The device is packaged using the same materials and processes. Performance Data (Non-clinical Non-Clinical Performance and Conclusions: and/or Clinical): The results of non-clinical (laboratory) performance testing demonstrate that the device is safe and effective and substantially equivalent to the predicate device. Clinical Performance and Conclusions: Clinical data and conclusions were not needed for
this device.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like figure with outstretched wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the emblem.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Zimmer, Inc. % Mr. Anthony Francalancia Senior Associate, Regulatory Affairs P.O. Box 708 Warsaw, Indiana 46581-0708
OCT 1 3 2006
Re: K062768
Trade/Device Name: NexGen® Complete Knee Solution Legacy® Posterior Stabilized (LPS)-Flex Fixed Bearing Tivanium® Ti-6AI-4V Alloy Femoral
Components Regulation Number: 21 CFR 888.3560 Regulation Name: Knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis Regulatory Class: Class II Product Code: JWH Dated: September 14, 2006 Received: September 15, 2006
Dear Mr. Francalancia:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2 -- Mr. Anthony Francalancia
This letter will allow you to begin marketing your device as described in your Section 510(k) This letter will and w Jourse organ finding of substantial equivalence of your device to a legally premaince hourieation. The results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you active of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Other golfinational and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
fa Vah
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Indications for Use
510(k) Number (if known):
Device Name:
NexGen® Complete Knee Solution Legacy® Posterior Stabilized (LPS)-Flex Fixed Bearing Tivanium® Ti-6AI-4V Alloy Femoral Components
Indications for Use:
- · This device is indicated for patients with severe knee pain and disability due to:
- Rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis.
- Collagen disorders, and/or avascular necrosis of the femoral connivale.
- Post-traumatic loss of joint configuration, particularly when there is patellofemoral erosion, dysfunction or prior patellectomy
- Moderate valgus, varus, or flexion deformities.
- The salvage of previously failed surgical attempts or for a knee in which satisfactory stability in flexion cannot be obtained at the time of surgery.
- Specific uses with LPS-Flex femorals:
- Provides increased flexion capability for patients who have both the flexibility and desire to increase their flexion range.
- The LPS-Flex femoral, when used with the LPS-Flex articular surfaces, is designed for use with both cruciate ligaments excised and when load bearing ROM is expected to be less than or equal to 155 degrees.
| Prescription Use | X (Part 21 CFR 801 Subpart D) |
|---|---|
| AND/OR | |
| Over-The-Counter Use (21 CFR 807 Subpart C) |
(Please do not write below this line - Continue on another page if needed)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of General, Restoration and Neurological Devices
Page 1 of 1
510(k) Number: k06 2768
N/A