K Number
K062674

Validate with FDA (Live)

Date Cleared
2006-12-01

(84 days)

Product Code
Regulation Number
862.1660
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Mission Trinity B™ Controls are intended to be used as a quality control material for monitoring the performance of pH, pCO2, pO2, Na+, K+, Ca++, Cl-, glucose, lactate, tHb, %O2Hb, %COHb, %metHb & %HHb on Bayer instrumentation. For In Vitro Diagnostic Use

Device Description

Trinity B is a specially formulated, three-level, aqueous material intended for use to monitor all analytes measured by the Bayer line of blood gas, electrolyte, metabolite and CO-Oximetry analyzers. Trinity B controls provide a convenient method of performing periodic QC checks for laboratories selecting to measure liquid QC as a part of their quality assurance program. Trinity B contains clinically relevant quantities of pH, PCO2, PO2, sodium, potassium, chloride, ionized calcium, glucose, lactate and suitable concentrations of dyes to simulate clinically relevant values of hemoglobin, and hemoglobin derivatives: O2Hb, COHb, MetHb and HHB.

AI/ML Overview

Here's the information regarding the acceptance criteria and study for the Mission Trinity B device, extracted and organized as requested:

1. Table of Acceptance Criteria and Reported Device Performance

The provided text does not explicitly state quantitative acceptance criteria or detailed performance data in a table format. It mentions "Tests were conducted to verify specific performance requirements" and that "side by side testing of Mission Trinity B vs. Predicate (OEM) product on intended instruments to o demonstrate equivalence" was performed.

Criterion TypeAcceptance Criteria (Not Explicitly Stated)Reported Device Performance
Stability(Implied: Demonstrated stability over time under specified conditions)"Accelerated (high temperature) stress test to support stability" was conducted. The specific results are not detailed, but the conclusion states "Comparison of technological characteristics, formulation and intended use to predicate devices listed in this summary support the claim of substantial equivalence."
Equivalence(Implied: Performance comparable to the predicate device for all specified analytes on intended instruments)"Side by side testing of Mission Trinity B vs. Predicate (OEM) product on intended instruments to o demonstrate equivalence" was conducted. The conclusion states the device is substantially equivalent, implying this testing confirmed comparable performance. Specific data on individual analytes' equivalence is not provided.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: Not explicitly stated. The document mentions "Side by side testing," suggesting that samples of both the Mission Trinity B and the predicate device were analyzed multiple times on the intended instruments.
  • Data Provenance: Not explicitly stated (e.g., country of origin). The study appears to be a prospective comparison study conducted by the manufacturer, Diamond Diagnostics Inc.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

This information is not applicable as the device is a quality control material intended for instrument calibration/quality verification, not a diagnostic device that produces results requiring human expert interpretation for "ground truth." The "ground truth" for the performance of a quality control material such as this would be its known assayed values, and the performance is measured by how accurately and precisely the instruments measure these known values compared to the predicate control.

4. Adjudication Method for the Test Set

This information is not applicable for the reasons stated in point 3.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This type of study is typically used for diagnostic imaging devices or other interpretation-based systems, not for quality control materials for automated analyzers.

6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

The concept of "standalone performance" of an algorithm is not applicable in this context. Mission Trinity B is a chemical control solution, not an algorithm or a software-based device. Its "performance" is its stability and its ability to serve as a reliable reference material for analytical instruments. The testing involved instrument analysis of the control material, similar to a "standalone" chemical analysis.

7. The Type of Ground Truth Used

For a quality control material, the "ground truth" is its established assayed values for the analytes it contains. These values are determined through rigorous analytical methods (e.g., reference methods, highly precise laboratory techniques) by the manufacturer. The purpose of the study was to demonstrate that instruments measure these values comparably to how they measure the predicate control.

8. The Sample Size for the Training Set

This information is not applicable. Quality control materials like Mission Trinity B do not involve machine learning algorithms that require a "training set." The product is a chemical formulation.

9. How the Ground Truth for the Training Set Was Established

This information is not applicable for the reasons stated in point 8.

{0}------------------------------------------------

K062674

5. 510(k) Summary

Owner's Name & Address:

Diamond Diagnostics Inc 333 Fiske Street Ho!liston MA 01746 FAX: 508-429-0452

Establishment Registration Number: 3003030793

Date of Preparation: Identification of the Device:

DEC - 1 2006

Contact Person: Randolph Quinn Development & Compliance Manager 508-429-0450 (x318)

September 1, 2006

Proprietary/Trade name: Mission Trinity B Common name blood gas controls with electrolytes, glucose and dyes Classification name: controls for blood-gases (assayed and un-assayed) (Classification # 862.1660, Product Code JJY)

Substantial Equivalence Table of Product Part Numbers & Trade Names

Diamond/Mission ProductBayer Equivalent
DD-96001 Trinity B Level 1108860 Rapid Complete Level 1
DD-96002 Trinity B Level 2108868 Rapid Complete Level 2
DD-96003 Trinity B Level 3108869 Rapid Complete Level 3
DD-96123 Trinity B Levels 1,2,3Not available

Description of New Device:

Trinity B is a specially formulated, three-level, aqueous material intended for use to monitor all analytes measured by the Bayer line of blood gas, electrolyte, metabolite and CO-Oximetry analyzers. Trinity B controls provide a convenient method of performing periodic QC checks for laboratories selecting to measure liquid QC as a part of their quality assurance program.

Trinity B contains clinically relevant quantities of pH, PCO2, PO2, sodium, potassium, chloride, ionized calcium, glucose, lactate and suitable concentrations of dyes to simulate clinically relevant values of hemoglobin, and hemoglobin derivatives: O2Hb, COHb, MetHb and HHB.

Intended Use:

Mission Trinity B blood gas, electrolyte, metabolite, CO-Oximeter control is an assayed quality control material used for monitoring the performance of blood gas, electrolyte, metabolite, and CO-Oximeter (Bayer 270/800/400 series) instrumentation for the analytes listed on the Expected Values Chart.

Comparison of Technological Characteristics with Predicate Device:

CharacteristicsMission Trinity BBayer Rapid Complete
PNDD-96001, 96002, 96003, 96123108860, 108868, 108869
Contents:Any LevelAqueous solution of buffers, electrolytes, glucose, lactate, dyes, equilibrated with CO2, O2 and N2.Contains NO human or animal materials.Aqueous solution of buffers, electrolytes, glucose, lactate, dyes, equilibrated with CO2, O2 and N2.Contains NO human or animal materials.
ContainerGlass ampule (heat sealed, score break)Glass ampule (heat sealed, score break)
ColorRed/purple solutionRed/purple solution
Package30 X 1.8mL (instrument only requires a fraction aof an mL to make a measurement)30 X 2.5mL
Intended UseFor in-vitro diagnostics use for quality control ofpH/Blood Gas analyzers, ISE analyzers, CO-oximeters, and Metabolite analyzers.For in-vitro diagnostics use for quality control ofpH/Blood Gas analyzers, ISE analyzers, CO-oximeters, and Metabolite analyzers.
Storage / ShelfLife2-8°C (36-months) or 18-25°C 12-months2-8°C or 18-25°C (until expiration date)

{1}------------------------------------------------

Assessment of Non-Clinical Performance Data:

Tests were conducted to verify specific performance requirements:

  • Accelerated (high temperature) stress test to support stability #
  • Side by side testing of Mission Trinity B vs. Predicate (OEM) product on intended instruments to o demonstrate equivalence

Assessment of Clinical Performance Data:

NA

Conclusions:

Comparison of technological characteristics, formulation and intended use to predicate devices listed in this summary support the claim of substantial equivalence.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows a black and white logo. The logo features a stylized eagle with three curved lines representing its body and wings. The eagle is facing right. The logo is encircled by text that reads 'DEPARTMENT OF HEALTH & HUMAN SERVICES USA'.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Randolph Quinn Development & Compliance Manager Diamond Diagnostics Inc. 333 Fiske Street Holliston, MA 01746

DEC - 1 2006

K062674 Re: Trade/Device Name: Mission Trinity B Regulation Number: 21 CFR 862.1660 Regulation Name: Quality control material (assayed and unassayed) Regulatory Class: Class I Product Code: JJY Dated: September 1, 2006 Received: September 8, 2006

Dear Mr. Quinn:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to dcvices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adultcration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).

{3}------------------------------------------------

Page 2 -

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240) 276-0484. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Alberto Guti

Alberto Gutierrez, Ph.D. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Page 3 -

cc: HFZ-401 DMC
D.O.

{5}------------------------------------------------

4. Indications for Use Statement

510(k) Number (if known):
K062674

Device Name: Mission Trinity B

Indications For Use:

Mission Timity B™ Controls are intended to be used as a quality control material for monitoring the performance of pH, pCO2, pO2, Na+, K+, Ca++, Cl-, glucose, lactate, tHb, %O2Hb, %COHb, %metHb & %HHb on Bayer instrumentation.

For In Vitro Diagnostic Use

Prescription Use _ X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Carol C Benson

vision Sign-Off

Office of In Vitro Diagnostic Device

K 062674

Diamond Diagnostics 510(k) Submission for Trinity B Controls

§ 862.1660 Quality control material (assayed and unassayed).

(a)
Identification. A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.(b)
Classification. Class I (general controls). Except when intended for use in donor screening tests, quality control materials (assayed and unassayed) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.