K Number
K062635

Validate with FDA (Live)

Device Name
MODEL 2200
Manufacturer
Date Cleared
2006-12-19

(104 days)

Product Code
Regulation Number
890.5500
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The CLARIMEDIX device is indicated for use to emit energy in the infrared spectrum to provide topical heating for the purpose of elevating tissue temperature; for the temporary relief of minor muscle and joint pain, arthritis and muscle spasm; relieving stiffness; promoting relaxation of muscle tissue; and to temporarily increase blood circulation where applied.

Device Description

The system is a self-contained unit with four treatment areas.

  • . Popliteal treatment zone
  • Calf treatment zone .
  • Foot treatment zone .
  • . Heel treatment zone
AI/ML Overview

The provided 510(k) summary for the CLARIMEDIX Model 2200 System does not contain the information requested to describe acceptance criteria and a study proving the device meets those criteria.

This document is a 510(k) premarket notification, which establishes substantial equivalence to predicate devices, rather than demonstrating clinical efficacy through performance studies against pre-defined acceptance criteria. The summary focuses on comparing the new device's intended use and technological characteristics to existing devices to show it is "substantially equivalent."

Therefore, I cannot populate the table or provide answers to the specific questions regarding acceptance criteria, study details, sample sizes, ground truth, or MRMC studies, as this information is not present within the provided text.

The closest relevant information is contained in the "Comparison with Predicate Devices" section:

  • Intended Use: The device is indicated for use for the treatment of chronic pain by emitting energy in the IR spectrum for the temporary relief of minor muscle and joint pain, arthritis and muscle spasm; relieving stiffness; promoting relaxation of muscle tissue; and to temporarily increase local blood circulation where applied.
  • Technological Characteristics: Key characteristics like temperature at the skin surface (Less than 95° F), energy source (Multidiodes), and safety features (Remains cool to touch) are listed for the subject device.

However, these are descriptive characteristics and intended uses, not quantified acceptance criteria with reported performance from a study designed to validate those criteria.

In summary, the provided document does not include the requested information about acceptance criteria and a study that proves the device meets those criteria.

{0}------------------------------------------------

K062635

:

510(k) Summary Prepared August 5, 2006

Submitted by:CLARIMEDIX, Inc.1035 Pearl Street, Suite 400Boulder, Colorado 80302
Contact Person:John DunningPresidentTelephone:FAX:e-mail:(303) 905 6163(303) 625 1030jdunning@clarimedix.comDEC 1 9 2006
Product Name:Model 2200 System
Common Name:Infrared lamp
Classification:ILY; Class II; CFR 21 890.5500

Predicate Devices:

Device NameManufacturerK Number
Warp 10Quantum DevicesK032229
Spectropad(a.k.a.RestorativeProducts, Inc.)SMI (A.k.a. AnodyneTherapy System)K931261

{1}------------------------------------------------

Description of Device:

The device is indicated for use for the treatment of chronic pain by emitting energy in the IR spectrum for the temporary relief of minor muscle and joint pain, arthritis and muscle spasm; relieving stiffiess; promoting relaxation of muscle tissue; and to temporarily increase local blood circulation where applied

The system is a self-contained unit with four treatment areas.

  • . Popliteal treatment zone
  • Calf treatment zone .
  • Foot treatment zone .
  • . Heel treatment zone

Intended Use:

The device is indicated for use for the treatment of chronic pain by emitting energy in the IR spectrum for the temporary relief of minor muscle and joint pain, arthritis and muscle spasm; relieving stiffness; promoting relaxation of muscle tissue; and to temporarily increase local blood circulation where applied

Comparison with Predicate Devices:

The submission device and the predicate devices have substantially equivalent intended use and technological specifications as shown in the tables below. The device is compliant with the standards for electrical safety and EMC, IEC 60601-1 IEC 60601-1-2 respectively.

{2}------------------------------------------------

Predicate DevicesSubject Device
Quantum Warp 10K032229SMI SpectropadK931261(a.k.a. Anodyne TherapySystem)CLARIMEDIX Model 2200
FDAclassification21 CFR 890.5500 InfraredLamp (ILY)a) Identification. An infraredlamp is a device intended formedical purposes that emitsenergy at infrared frequencies(approximately 700 nanometersto 50,000 nanometers) toprovide topical heatingb) classification Class II21 CFR 890.5500 InfraredLamp (ILY)a) Identification. An infraredlamp is a device intended formedical purposes that emitsenergy at infrared frequencies(approximately 700nanometers to 50,000nanometers) to provide topicalheatingb) classification Class II21 CFR 890.5500 InfraredLamp (ILY)a) The CLARIMEDIX deviceis indicated for use to emitenergy in the infraredspectrum to provide topicalheating for the purpose ofelevating tissue temperature;for the temporary relief ofminor muscle and joint pain,arthritis and muscle spasm;relieving stiffness; promotingrelaxation of muscle tissue;and to temporarily increaseblood circulation whereapplied.b) classification Class II
Intended UseIndicated for use for thetreatment of chronic pain byemitting energy in the Near-IRspectrum for the temporaryrelief of minor muscle and jointpain, arthritis and musclespasm; relieving stiffness;promoting relaxation of muscletissue; and to temporarilyincrease local blood circulationwhere appliedIntended for relief of minormuscle and joint pain andimprovement of superficialcirculation 1indicated for use for thetreatment of chronic pain byemitting energy in the IRspectrum for the temporaryrelief of minor muscle andjoint pain, arthritis and musclespasm; relieving stiffness;promoting relaxation of muscletissue; and to temporarilyincrease local blood circulationwhere applied
IntendedUsersPhysical medicine health careprofessionalsPhysical medicine health careprofessionalsPhysical medicine health careprofessionals
Intended Siteof UseMedical health care facilitiesMedical health care facilitiesMedical health care facilities

Table 1 Comparison of Indications for Use

{3}------------------------------------------------

Predicate DevicesPredicate DevicesSubject Device
Quantum Warp 10K032229SMI SpectropadK931261(a.k.a. Anodyne TherapySystem)CLARIMEDIX Model 2200
Method ofadministrationTopicalTopicalTopical
Treatment areas10cm2Leg, foot, arm, neckLower leg, foot, adjustablelocation
Number ofzonesNot specified in productlabeling4-8 therapy padsFour zones
Temperature atthe skin surfaceNot specified in labelingNot specified in labelingLess than 95° F
Energy sourceMultidiodes dispersed overtreatment zones60 Infrared diodes per arrayMultidiodes
WaveformConstantNot specified in labelingModulating
Power supply8 each 1.5 AA batteriesprovides 50 dosesNot specified in labeling24V
Safety featuresRemains cool to the touchNot specified in labelingRemains cool to touch
BiocompatibilityNot specified in labelingNot specified in labeling;Biocompatible draperecommended
ConfigurationHandheld4-8 pads applied4 zones applied
Length andwidth5.55in x 2.65 inNot specified in labeling25.1 in. x 10.0 in.
PortableYesYesYes
Electrical safetyNot specified in labelingNot specified in labelingCompliant with IEC 60601-1
FirmwarecontrolYesYesYes

·

:

Table 2 Comparison of Technological Characteristics

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" arranged in a circular pattern around the symbol. The logo is black and white.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

CLARIMEDEX, Inc. % Regulatory Technology Services, LLC Mr. Mark Job 1394 25TH Street, Northwest Buffalo, Minnesota 55313

DEC 1 9 2006

Re: K062635

Trade/Device Name: Model 2200 Regulation Number: 21 CFR 890.5500 Regulation Name: Infrared lamp Regulatory Class: II Product Code: ILY Dated: December 6, 2006 Received: December 7, 2006

Dear Mr. Job:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{5}------------------------------------------------

Page 2 - Mr. Mark Job

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

For Pablo

Mark N. Melkerson

DEP D.R.SECTOR

Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

510(k) Number (if known):

Device Name: Model 2200

Indications For Use:

The CLARIMEDIX device is indicated for use to emit energy in the infrared spectrum to provide topical heating for the purpose of elevating tissue temperature; for the temporary relief of minor muscle and joint pain, arthritis and muscle spasm; relieving stiffness; promoting relaxation of muscle tissue; and to temporarily increase blood circulation where applied.

X Prescription Use (Per 21CFR 801)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence Of CDRH, Office Of Device Evaluation (ODE)

(Division Sign-Off)
Division of General, Restorative,
and Neurological Devices

(K) Number 2062635

§ 890.5500 Infrared lamp.

(a)
Identification. An infrared lamp is a device intended for medical purposes that emits energy at infrared frequencies (approximately 700 nanometers to 50,000 nanometers) to provide topical heating.(b)
Classification. Class II (special controls). The device, when it is an infrared therapeutic heating lamp, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 890.9.