K Number
K062520

Validate with FDA (Live)

Date Cleared
2006-10-11

(44 days)

Product Code
Regulation Number
880.6250
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Powdered vinyl patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

Device Description

powdered vinyl patient examination gloves, Clear(non-colored) that meets all of the requirements of ASTM standard D 5250-004

AI/ML Overview

Here's an analysis of the provided information, focusing on acceptance criteria and supporting studies for the device:

Device: Powdered Vinyl Patient Examination Gloves, Clear (Non-colored)
510(k) Number: K062520


1. Table of Acceptance Criteria and Reported Device Performance

CharacteristicAcceptance Criteria (Standard)Reported Device Performance
DimensionASTM standard D 5250-00e4Meets
Physical PropertiesASTM standard D 5250-00e4Meets
Freedom from Pinholes21 CFR 800.20Meets
Powder AmountASTM standard D 5250-00e4Meets
BiocompatibilityPrimary Skin IrritationPasses (<10mg/dm2)
Dermal SensitizationPasses (Not a Dermal sensitization)

2. Sample Size Used for the Test Set and Data Provenance

The document does not explicitly state the sample sizes used for the tests to demonstrate compliance with the ASTM standards or 21 CFR 800.20. The provenance of the data (e.g., country of origin, retrospective or prospective) is also not specified.


3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

This section is not applicable. The device is a medical glove, where "ground truth" typically refers to objective measurements and adherence to engineering standards rather than expert consensus on medical images or diagnoses. The "ground truth" here is the pass/fail result against established material properties and biocompatibility tests.


4. Adjudication Method for the Test Set

This section is not applicable. Adjudication methods (like 2+1, 3+1) are typically used in studies involving subjective interpretation, such as image analysis for disease detection. For device testing against objective standards, the results are typically binary (pass/fail) based on direct measurement.


5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No MRMC comparative effectiveness study was mentioned or performed. This type of study is relevant for diagnostic devices where human readers interpret results, and the AI's impact on reader performance is evaluated. For a medical glove, this is not applicable.


6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

This section is not applicable. The device is a physical product (a medical glove), not an algorithm or AI system. Therefore, standalone algorithm performance is not relevant. The performance evaluation focuses on the inherent physical and chemical properties of the glove itself.


7. Type of Ground Truth Used

The "ground truth" for this device's performance evaluation is based on objective measurements against established national and international standards and regulations. Specifically:

  • Physical and Dimension Properties: ASTM standard D 5250-00e4
  • Freedom from Pinholes: 21 CFR 800.20
  • Powder Amount: ASTM standard D 5250-00e4
  • Biocompatibility: Primary Skin Irritation in rabbits (ISO10993-10) and Dermal sensitization in the guinea pig (ISO10993-10).

8. Sample Size for the Training Set

This section is not applicable as the device is not an AI/ML algorithm or a diagnostic tool that requires a training set. The manufacturing process of gloves relies on established engineering and materials science principles, not machine learning.


9. How the Ground Truth for the Training Set Was Established

This section is not applicable, as there is no "training set" for this device.

{0}------------------------------------------------

Summary

OCT 1 1 2006

"This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92."

"The assigned 510(k) number is: KOO 252 Qapplicant leave blank)

Premarket Notification [510(k)] Summary

[{a)(1)]. The summary contains on the first page, preferably on your letterhead paper, the submitter's name, address, phone and fax numbers, name of contact person, and date the summary was prepared :

Submitter's name :JIANGSU JAYSUN GLOVE CO., LTD
Submitter's address :NO. 199 JIANLING AVENUE, SUQIAN ECONOMIC DEVELOPMENTZONE, SUQIAN CITY, JIANGSU PROVINCE, 223800, CHINA
Phone number :(86) - 527-4568809
Fax number :(86) - 527-4568806
Name of contact person:Ms. Sea XU
Date the summary was prepared:Aug 20 2006

[(a)(2)]. The name of the device, including the trade or proprietary name if applicable, the common or usual name, and the classification name, if known

Device Name:Powdered Vinyl Patient Examination Gloves,Clear (Non-colored)
Proprietary/Trade name:Powdered Vinyl Patient Examination GlovesOther clients private labeling
Common Name:Patient examination glove
Classification Name:Patient examination glove
Device Classification:1
Regulation Number:21 CFR 880.6250
Panel:General Hospital (80)
Product Code:LYZ

((a)(3)). An identification of the legally marketed device to which your firm is claiming substantial equivalence .

Class I* powdered vinyl patient examination gloves, Clear(non-colored) that meets all of the requirements of ASTM standard D 5250-0064.

Predicate device : FUGUAN (Brand) Powdered Vinyl Patient Examination Gloves, Shijiazhuang Fuguan Plastic Products Co., Ltd.. K032907.

[(a)(4)] A description of the device

Device Description : powdered vinyl patient examination gloves, Clear(non-colored) that meets all of the requirements of ASTM standard D 5250-004

{1}------------------------------------------------

K062520

1(a)(5)] The summary describes the intended use of the device

Device Intended Use: powdered vinyl patient examination glove, Clear(non-colored) is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

[(a)(6)| A summary of the technological characteristics of new device compared to the predicate device.

The powdered vinyl patient examination gloves, Clear (non-colored) non sterile are summarized with the following technological characteristics compared to ASTM or equivalent standard.

CharacteristicsStandardDevice performance
DimensionASTM standard D 5250-00e4Meets
Physical PropertiesASTM standard D 5250-00e4Meets
Freedom from pinholes21 CFR 800.20Meets
Powder AmountASTM standard D 5250-00e4Meets
BiocompatabilityPrimary Skin Irritation in rabbits<10mg/dm2
Passes
Dermal sensitization in the guinea pigNot a Primary Skin Irritation
PassesNot a Dermal sensitization

[(b)(1)| A brief discussion of the nonclinical submitted, reference, or relied on in the premarket notification submission for a determination of substantial equivalence .

Powdered vinyl patient examination gloves . Clear(non-colored) meet requirements per ASTM D5250-00-4, per ASTM D6124-01, per 21 CFR 800.20 and ISO10993-10.

[(b)(2)| A brief discussion of the clinical submitted, reference, or relied on in the premarket notification submission for a determination of substantial equivalence .

Clinical data is not needed for gloves or for most devices cleared by the 510(k) process.

[(b)(3)| The conclusions drawn from the nonclinical and clinical tests that demonstrate that the device is as safe, as effective, and performed as well or better than the legally marketed device identified in (a)(3).

It can be concluded that the Powdered Vinyl Patient Examination Gloves, Clear(non-colored) meet the ASTM standard or equivalent standard and FDA requirements for waterleak test on pinhole AQL, meet labeling claims .

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the logo is an abstract symbol that resembles a stylized caduceus or a human figure embracing a globe.

Public Health Service

OCT 1 1 2006

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Jiangsu Jaysun Glove Company Limited C/O Mr. Chu Xiaoan Beijing Easy-Link Company Room 1606, Building 1, Jianxiang Yuan. No. 209 Bei Si Huan Zhong Road, Haidian District Beijing, P. R. CHINA 100083

Re: K062520

Trade/Device Name: Powdered Vinyl Patient Examination Gloves, Clear (Non-colored) Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: I Product Code: LYZ Dated: September 26, 2006 Received: September 26, 2006

Dear Mr. Xiaoan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 - Mr. Xiaoan

Please be advised that FDA's issuance of a substantial equivalence defermination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Clue

Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

INDICATIONS FOR USE

Applicant: JIANGSU JAYSUN GLOVE CO. LTD

510(k) Number (if known): * KO62520

Device Name:__Powdered Vinyl Patient Examination Gloves, Clear(Non-colored)

Indications For Use:

Powdered vinyl patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

Prescription Use (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use × (21 CFR 801 Subpart C)

  • !

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Stale H. Murphy KD 10/11/04

n Sign-Off) sion of Anesthesiology, General Hospital, Control Control, Dental Devices

LabelsValues
Number4062520

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.