K Number
K062262

Validate with FDA (Live)

Date Cleared
2006-11-14

(102 days)

Product Code
Regulation Number
886.5925
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

CIBA Vision® (sifilcon A) spherical soft contact lenses are indicated for the optical correction of refractive ametropia (myopia and hyperopia) in phakic or aphakic persons with non-diseased eyes who have 1.50 diopters or less of astigmatism.

CIBA Vision® Toric (sifilcon A) soft contact lenses are indicated for the optical correction of refractive ametropia (myopia and hyperopia) in phakic or aphakic persons with non-diseased eyes who have 10.00 diopters (D) or less of astigmatism.

CIBA Vision® Multifocal (sifilcon A) soft contact lenses are indicated for the optical correction of refractive ametropia (myopia and hyperopia) and/or presbyopia in phakic or aphakic persons with non-diseased eyes who may require a reading addition of +4.00 diopters (D) or less and who have 1.50 diopters or less of astigmatism.

CIBA Vision® Multifocal Toric (sifilcon A) soft contact lenses are indicated for the optical correction of refractive ametropia (myopia, hyperopia, astigmatism and presbyopia) in phakic or aphakic persons with non-diseased eyes who may require a reading addition of +4.00 diopters (D) or less and have 4.00 diopters (D) or less of astigmatism.

The lenses may be prescribed for daily wear with removal for cleaning and disinfection (chemical, not heat) prior to reinsertion, as recommended by the eve care professional.

Device Description

The lens material is 32% water and 68% sifilcon A a fluoro-silicone containing hydrogel which is surface treated.

Sifilcon B lens designs include spherical, toric, multifocal and multifocal toric in the following parameter ranges:
Diameter Range: 12.0 to 15.0 mm
Base Curve Range: 7.0 to 9.2 mm
Power Range: -20.00D to +20.00D
Center Thickness: 0.070 mm for -3.00D spherical (varies with power)

Lenses contain the color additive Phthalocyanine Green (PCN Green), a light green handling tint, which makes them easier to see when handling.

Lenses have the following properties:
Refractive index: 1.43 (hydrated)
Light transmittance: approximately 92 %
Water content : 32% by weight in normal saline
Oxygen permeability: 82 x 10 -11 [(cm2 /sec)(ml O2 /ml=mmHg)] measured at 35°C (intrinsic Dk-Polarographic method)

Lenses are supplied sterile in sealed blister-packs containing isotonic phosphate buffered saline solution.

AI/ML Overview

The provided document describes the safety and substantial equivalence of CIBA Vision's Sifilcon A Soft Contact Lenses for Daily Wear (K062262) to a previously cleared predicate device (Lotrafilcon B).

Here's a breakdown of the requested information:

1. Table of Acceptance Criteria and Reported Device Performance:

The document doesn't explicitly list "acceptance criteria" in a quantitative table for clinical performance. Instead, it describes demonstrating "similar overall performance" to the predicate in key areas. For non-clinical, it uses terms like "consistent with or better than" and "substantially equivalent."

AspectAcceptance Criteria (Implied)Reported Device Performance (Sifilcon A)
Non-Clinical
Lens Physical/Material PropertiesConsistent with or better than industry marketed lenses; substantially equivalent to predicate lens.Consistent with or better than industry marketed lenses; substantially equivalent to the predicate lens.
Compatibility with Lens Care ProductsCompatible with commonly available lens care products.The lens material is compatible with commonly available lens care products.
BiocompatibilityNon-toxic and biocompatible; substantially equivalent to predicate device and non-toxic/non-irritating.Substantially equivalent to the predicate device and are non-toxic and non-irritating.
Clinical
Overall PerformanceSimilar performance to predicate in vision, health, comfort, and fit.Demonstrated similar overall performance in the clinically relevant areas of vision, health, comfort and fit as compared to the control lens.

2. Sample Size Used for the Test Set and Data Provenance:

  • Test Set Sample Size: The document mentions a "daily wear clinical study" but does not specify the sample size for this clinical evaluation.
  • Data Provenance: The clinical study was conducted in accordance with current Good Clinical Practices (GCP) and published regulations (21 CFR Parts 50, 56, 312, and 812). It compares the sifilcon A lens to an "FDA approved and commercially available contact lens." The country of origin for the data is not explicitly stated, but given FDA submission, it's highly likely to be within the USA or a region adhering to similar regulatory standards for clinical trials. The study is a prospective clinical investigation.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications:

  • Number of Experts & Qualifications: The document does not specify the number of experts or their qualifications used to establish ground truth in the clinical study. It implies that "eye care professionals" would recommend the lenses for daily wear, but this is about prescription, not ground truth establishment for the study itself. Clinical studies for contact lenses typically involve ophthalmologists or optometrists assessing patient outcomes, but specific details are absent.

4. Adjudication Method for the Test Set:

  • Adjudication Method: The document does not describe any specific adjudication method (e.g., 2+1, 3+1) for the clinical study results.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size:

  • MRMC Study: This is not applicable as the device is a medical product (contact lens), not an AI or imaging diagnostic tool that would typically involve multiple readers interpreting cases. The clinical study was a comparison of device performance between two contact lens types.
  • Effect Size: No effect size in terms of human reader improvement with/without AI assistance is mentioned or relevant.

6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done:

  • Standalone Performance: This is not applicable as the device is a contact lens, not a computational algorithm. The evaluation focuses on the performance of the physical lens as used by human wearers.

7. The Type of Ground Truth Used:

  • Type of Ground Truth: For the clinical testing, the "ground truth" would be established by direct clinical observations and measurements of patient outcomes (e.g., vision, lens fit, comfort, ocular health assessments) performed by qualified medical professionals (e.g., ophthalmologists, optometrists). For non-clinical testing, it would be based on validated laboratory measurements and biocompatibility assessments.

8. The Sample Size for the Training Set:

  • Training Set Sample Size: The concept of a "training set" is not applicable to this type of device (contact lens). There is no machine learning model being trained. The "training" for the device would be the manufacturing process and quality control.

9. How the Ground Truth for the Training Set Was Established:

  • Ground Truth for Training Set: As there is no training set in the context of machine learning, this question is not applicable. The "ground truth" for the device's design and manufacturing would be established through material specifications, engineering design principles, and quality control standards.

{0}------------------------------------------------

K062262

Image /page/0/Picture/24 description: The image shows the logo for CIBA Vision, a Novartis company. The logo is black and white and features the words "CIBA Vision" in a bold, sans-serif font. Below the company name is the text "A Novartis Company" in a smaller font. The logo is simple and modern, and it is likely used on the company's products and marketing materials.

CIBA Vision® Corporation 11460 Johns Creek Parkway Duluth, Georgia USA 30097

3-Aug-2006 Page 1 of 4 V01

Sifilcon A Soft Contact Lenses for Daily Wear 510(k) Summary of Safety and Substantial Equivalence

510(k) Summary

1. Submitter Information:

NOV 1 4 2006

Company:CIBA Vision Corporation11460 Johns Creek ParkwayDuluth, Georgia USA 30097
Contact Person:Alicia M. Plesnarski, RACDirector Global Regulatory Aff

lobal Regulatory Affairs alicia.plesnarski@cibavision.com

Telephone:678-415-3924
FAX:678-415-3454
Date Prepared:3 August 2006

Device Name: 2.

  • Common Name: Soft Contact Lens ●
  • CIBA Vision® (sifilcon A) Trade/Proprietary Name: .
  • Classification Name: Daily Wear Soft Contact Lens ●
  • Class II [21 CFR 886.5925 (b) (1)] Device Classification: .

Predicate Device: 3.

CIBA Vision's (lotrafilcon B) soft contact lenses. Both lotrafilcon B (predicate) and sifilcon A are in FDA Group 1 (low water, nonionic polymer). CIBA Vision obtained FDA 510(k) clearance for (lotrafilcon B) lenses for daily wear on March 12, 2004 (K033919).

4. Description of Device:

The lens material is 32% water and 68% sifilcon A a fluoro-silicone containing hydrogel which is surface treated.

Sifilcon B lens designs include spherical, toric, multifocal and multifocal toric in the following parameter ranges:

Diameter Range:12.0 to 15.0 mm
Base Curve Range:7.0 to 9.2 mm
Power Range:-20.00D to +20.00D
Center Thickness:0.070 mm for -3.00D spherical (varies with power)

Lenses contain the color additive Phthalocyanine Green (PCN Green), a light green handling tint, which makes them easier to see when handling.

{1}------------------------------------------------

ROHOTI.A Novariis CompanyCIBA Vision® Corporation11460 Johns Creek ParkwayDuluth, Georgia USA 300973-Aug-2006Page 2 of 4v01
Sifilcon A Soft Contact Lenses for Daily Wear510(k) Summary of Safety and Substantial Equivalence

Lenses have the following properties:

Refractive index:1.43 (hydrated)
---------------------------------------
  • approximately 92 % Light transmittance:
  • . 32% by weight in normal saline Water content : .
    • 82 x 10 -11 Oxygen permeability
      • [(cm2 /sec)(ml O2 /ml=mmHg)]

measured at 35°C (intrinsic Dk-Polarographic method)

Lenses are supplied sterile in sealed blister-packs containing isotonic phosphate buffered saline solution. The compatibility and package integrity of the blister-pack packaging system has been demonstrated and successfully used for other CIBA Vision marketed lens products, and packaged lenses are effectively steam sterilized in a validated autoclave. Blister-pack containers are labeled with the lens parameters, lot number and product expiration date. The expiration date has been established through stability studies that have assessed the chemical stability of he lens and package integrity (ability to Shelf-life studies are ongoing to establish and extend the labeled maintain sterility). expiration date.

5. Indications for Use:

.

CIBA Vision® (sifilcon A) spherical soft contact lenses are indicated for the optical correction of refractive ametropia (myopia and hyperopia) in phakic or aphakic persons with non-diseased eyes who have 1.50 diopters or less of astigmatism.

CIBA Vision® Toric (sifilcon A) soft contact lenses are indicated for the optical correction of refractive ametropia (myopia and hyperopia) in phakic persons with nondiseased eyes who have 10.00 diopters (D) or less of astigmatism.

CIBA Vision® Multifocal (sifilcon A) soft contact lenses are indicated for the optical correction of refractive ametropia (myopia and/or presbyopia in phakic or aphakic persons with non-diseased eyes who may require a reading addition of +4.00 diopters (D) or less and who have 1.50 diopters or less of astigmatism.

CIBA Vision® Multifocal Toric (sifilcon A) soft contact lenses are indicated for the optical correction of refractive ametropia (myopia, hyperopia, astigmatism and presbyopia) in phakic or aphakic persons with non-diseased eyes who may require a reading addition of +4.00 diopters (D) or less and have 10.00 diopters (D) or less of astigmatism.

The lenses may be prescribed for daily wear with removal for cleaning and disinfection (chemical, not heat) prior to reinsertion, as recommended by the eye care professional.

6. Description of Safety and Substantial Equivalence:

A series of non-clinical tests and a clinical study were performed to demonstrate the substantial equivalence of the device to the predicate device. All testing was conducted in

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for CIBA Vision, a Novartis company. The logo is in black and white, with the word "CIBA" in large, bold letters above the word "Vision." The word "Vision" has a stylized "i" with a circle in the center. Below the word "Vision" is the text "A Novartis Company" in smaller letters.

CIBA Vision® Corporation 11460 Johns Creek Parkway Duluth, Georgia USA 30097

3-Aug-2006 Page 3 of 4 V01

Sifilcon A Soft Contact Lenses for Daily Wear 510(k) Summary of Safety and Substantial Equivalence

accordance with the May 1994 FDA guideline titled Premarket Notification 510(k) Guidance Document for Class II Contact Lenses and in conformance to applicable device regulations. Results verify that sifilcon Alenses have material characteristics comparable to or better than other currently marketed soft contact lenses, are compatible with commonly available lens care products and are non-toxic and biocompatible. Clinically, the lens performed satisfactorily in a daily wear investigation. Results from all tests demonstrate the substantial equivalence to a previously FDA approved predicate (control) lenses.

Non-clinical Testing:

A series of non-clinical testing was performed to verify equivalence of the device to the predicate device. Non-clinical biocompatibility testing was conducted in accordance with the GLP regulation (21 CFR Part 58).

The results of all non-clinical testing on (sifilcon A) contact lens demonstrate:

  • Lens physical and material properties of the device are consistent with or better than . industry marketed lenses, and substantially equivalent to the predicate lens.
  • The lens material is compatible with commonly available lens care products. .
  • The lens material and extracts of the device are substantially equivalent to the . predicate device and are non toxic and non-irritating.

Clinical Testing:

The sifilcon A contact lens was investigated in daily wear clinical study. The three-month clinical evaluation was conducted in accordance with current Good Clinical Practices and published regulations (21 CFR Parts 50, 56, 312, and 812). The study assessed the clinical performance of the lenses as compared to an FDA approved and commercially available contact lens.

Clinical evaluation of the sifilcon Alens demonstrated similar overall performance in the clinically relevant areas of vision, health, comfort and fit as compared to the control lens when used under daily wear conditions.

{3}------------------------------------------------

CIBA VisionA Novartis CompanyCIBA Vision® Corporation11460 Johns Creek ParkwayDuluth, Georgia USA 300973-Aug-2006Page 4 of 4v01
---------------------------------------------------------------------------------------------------------------------------------------------------------

Sifilcon A Soft Contact Lenses for Daily Wear

510(k) Summary of Safety and Substantial Equivalence

Substantial Equivalence:

The CIBA Vision (sifilcon A) contact lens is substantially equivalent to the predicate lens and similar to other daily wear soft contact lenses in terms of water content (32% water) and ionic characteristics (FDA Group I: low water, nonionic), clinical performance, and indications for use. In addition, the lenses may be disinfected using a chemical, not heat, disinfection regimen.

Any differences which may exist between the sifilcon A soft contact lens and other Group I soft hydrophilic contact lenses do not adversely affect the safety and effectiveness of the device.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a stylized eagle with three lines representing its body and wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the eagle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

NOV I 4 2006

Ciba Vision Corporation c/o Alicia M. Plesnarski, RAC 11460 Johns Creek Parkway Duluth, GA 30097

Re: K062262

Trade/Device Name: Ciba Vision® (sifilcon A) Soft Contact Lenses for Daily Wear Regulation Number: 21 CFR 886.5925 Regulation Name: Soft (hydrophilic) contact lenses Regulatory Class: Class II Product Code: LPL Dated: October 19, 2006 Received: October 20, 2006

Dear Ms. Plesnarski:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{5}------------------------------------------------

Page 2 – Alicia M. Plesnarski, RAC

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

M.B. Eychlemi S.M.D

Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

Part II. Indications for Use Statement, v01, 3-Aug-2006, Page 1 of 1 Premarket Notification 510(k): (sifilcon A) Soft Contact Lens for Daily Wear

PART II. INDICATIONS FOR USE STATEMENT

KO62262 510(k) Number:

CIBA Vision® (sifilcon A) Soft Contact Lenses Device Name:

Indications For Use:

CIBA Vision® (sifilcon A) spherical soft contact lenses are indicated for the optical correction of refractive ametropia (myopia and hyperopia) in phakic or aphakic persons with non-diseased eyes who have 1.50 diopters or less of astigmatism.

CIBA Vision® Toric (sifilcon A) soft contact lenses are indicated for the optical correction of refractive ametropia (myopia and hyperopia) in phakic or aphakic persons with non-diseased eyes who have 10.00 diopters (D) or less of astigmatism.

CIBA Vision® Multifocal (sifilcon A) soft contact lenses are indicated for the optical correction of refractive ametropia (myopia and hyperopia) and/or presbyopia in phakic or aphakic persons with non-diseased eyes who may require a reading addition of +4.00 diopters (D) or less and who have 1.50 diopters or less of astigmatism.

CIBA Vision® Multifocal Toric (sifilcon A) soft contact lenses are indicated for the optical correction of refractive ametropia (myopia, hyperopia, astigmatism and presbyopia) in phakic or aphakic persons with non-diseased eyes who may require a reading addition of +4.00 diopters (D) or less and have 4.00 diopters (D) or less of astigmatism.

The lenses may be prescribed for daily wear with removal for cleaning and disinfection (chemical, not heat) prior to reinsertion, as recommended by the eve care professional.

Concurrence of CDRH, Office of Device Evaluation (ODE)
Division Sign-Off)
Division of Ophthalmic Ear,
Nose and Throat Devises

510(k) NumberK062262
------------------------
Prescription Use:X or Over the Counter Use
----------------------------------------------------------------------------------------------------------------------------------

§ 886.5925 Soft (hydrophilic) contact lens.

(a)
Identification. A soft (hydrophilic) contact lens is a device intended to be worn directly against the cornea and adjacent limbal and scleral areas of the eye to correct vision conditions or act as a therapeutic bandage. The device is made of various polymer materials the main polymer molecules of which absorb or attract a certain volume (percentage) of water.(b)
Classification. (1) Class II if the device is intended for daily wear only.(2) Class III if the device is intended for extended wear.
(c)
Date PMA or notice of completion of a PDP is required. As of May 28, 1976, an approval under section 515 of the act is required before a device described in paragraph (b)(2) of this section may be commercially distributed. See § 886.3.