K Number
K062165

Validate with FDA (Live)

Date Cleared
2007-01-05

(161 days)

Product Code
Regulation Number
862.3150
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

For in vitro diagnostic use only. VITROS Chemistry Products BARB Reagent is used on VITROS 5,1 FS Chemistry Systems for the semi-quantitative or qualitative determination of barbiturates (BARB) in human urine using a cutoff of either 200 ng/mL or 300 ng/mL. Measurements obtained with the VITROS BARB method are used in the diagnosis and treatment of barbiturates use or overdose.

The VITROS Chemistry Products BARB assay is intended for use by professional laboratory personnel. It provides only a preliminary test result. A more specific alternative chemical method must be used to confirm a result obtained with this assay. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method. Clinical consideration and professional judgment should be applied to any drug-of-abuse test result, particularly when evaluating a preliminary positive result.

For in vitro diagnostic use only. VITROS Chemistry Products Calibrator Kit 26 is used to calibrate VITROS 5,1 FS Chemistry Systems for the qualitative or semi-quantitative measurement of drugs of abuse.

For in vitro diagnostic use only. VITROS Chemistry Products FS Calibrator 1 is used in conjunction with VITROS Chemistry Products Calibrator Kits to calibrate VITROS 5,1 FS Chemistry Systems.

For in vitro diagnostic use only. VITROS Chemistry Products DAT Performance Verifiers are assayed controls used to monitor performance of urine drugs of abuse screening assays on VITROS 5,1 FS Chemistry Systems.

Device Description

The VITROS BARB assay is a homogeneous enzyme immunoassay that is performed using the VITROS Chemistry Products BARB Reagent with the VITROS Chemistry Products Calibrator Kit 26, VITROS Chemistry Products FS Calibrator 1, and VITROS Chemistry Products FS Diluent Pack 4 (DAT Diluent / DAT Diluent 2) on VITROS 5,1 FS Chemistry Systems.

The VITROS BARB Reagent is a dual chambered package containing ready-to-use liquid reagents that are used to detect barbiturates in urine. Sample, calibrators, and controls are automatically treated with surfactant (DAT Diluent 2) prior to addition of reagents. Treated sample is added to Reagent 1 containing antibody reactive to secobarbital, glucose-6-phosphate and nicotinamide adenine dinucleotide (NAD*), followed by Reagent 2 containing secobarbital labeled with the enzyme glucose-6phosphate dehydrogenase (G6P-DH). The assay is based on competition between barbiturates in the treated urine sample and secobarbital labeled with the enzyme glucose-6phosphate dehydrogenase (G6P-DH) for antibody binding sites. Enzyme activity decreases upon binding to the antibody, therefore the concentration of barbiturates in the urine sample is directly proportional to measured enzyme activity. Active enzyme converts oxidized nicotinamide adenine dinucleotide (NAD') to NADH, resulting in an absorbance change that is measured spectrophotometrically at 340 nm.

The VITROS Chemistry Products Calibrator Kit 26 is prepared from human urine to which drugs of abuse, metabolites of drugs of abuse, organic salt, surfactants, and preservative have been added. VITROS Chemistry Products FS Calibrator 1 is prepared from sodium chloride and processed water. These standards are used to calibrate VITROS 5,1 FS Chemistry Systems for the qualitative and semiquantitative measurement of barbiturates (BARB).

The VITROS Chemistry Products DAT Performance Verifiers I, II, III, IV, and V are quality control materials prepared from a human urine pool to which analytes, surfactant, and preservative have been added. These are assayed controls used to monitor performance of the VITROS BARB assay on VITROS 5,1 FS Chemistry Systems.

The VITROS Chemistry Products FS Diluent Pack 4 (DAT Diluent / DAT Diluent 2 is a common reagent that is used by multiple assays on the VITROS 5,1 FS Chemistry System. This is a dual chambered package containing two ready-to-use liquid diluents. DAT Diluent is prepared from human urine to which organic salt, surfactants, and preservative have been added. DAT Diluent 2 is prepared from processed water to which surfactant and preservative have been added.

The VITROS 5,1 FS Chemistry System is a clinical chemistry instrument that provides automated use of the VITROS Chemistry Products MicroTip® and MicroSlides® range of products. The VITROS 5,1 FS System was cleared for market by 510(k) premarket notification (K031924).

AI/ML Overview

The provided text describes a 510(k) premarket notification for the VITROS Chemistry Products BARB Reagent and associated calibrators and verifiers. This is a submission for an in-vitro diagnostic device, which typically involves demonstrating substantial equivalence to a predicate device rather than conducting a new clinical efficacy study with acceptance criteria often seen in device approvals.

Therefore, the information traditionally requested for acceptance criteria and a study proving a device meets these criteria (such as sample sizes for test and training sets, expert qualifications, adjudication methods, MRMC studies, standalone performance, and ground truth types) is not directly applicable or available in this type of submission.

Instead, the document focuses on demonstrating substantial equivalence to existing legally marketed predicate devices. The "study" described is a comparison study against a predicate device.

Here's a breakdown of what can be extracted or inferred from the provided text, framed within your request:


1. Table of Acceptance Criteria and Reported Device Performance

Formal, pre-defined acceptance criteria (e.g., target specificity/sensitivity thresholds) for performance against a gold standard are not explicitly stated as they would be for a novel device efficacy claim. Instead, the "acceptance" hinges on demonstrating "good agreement" and "similar performance characteristics" to the predicate device.

CharacteristicAcceptance Criteria (Implied by Substantial Equivalence Goal)Reported Device Performance (VITROS BARB vs. Predicate)
Agreement with Predicate"Good agreement" with the predicate immunoassay method for barbiturate detection in urine."The results demonstrated good agreement between the two immunoassay methods." (VITROS BARB assay and Syva® EMIT II Plus Barbiturate assay)
Intended UseSame as predicate: in vitro diagnostic use for qualitative/semi-quantitative analysis of barbiturates in human urine."The VITROS BARB assay and VITROS DAT Performance Verifiers have the following similarities to the predicate devices: same intended use..."
Cutoff ValuesSame as predicate: 200 or 300 ng/mL."...the same cutoff values of 200 or 300 ng/mL..."
Reagent FormatSimilar to predicate: liquid, ready to use."...are liquid, ready to use reagents..."
Performance CharacteristicsSimilar to predicate."...have similar performance characteristics..."
Instrumentation TypeUsed on an automated clinical chemistry analyzer."...are used on an automated clinical chemistry analyzer..."
Calibration TraceabilityCalibrated against the same drug (secobarbital)."...and calibrated against the same drug, secobarbital."
Other SimilaritiesHomogeneous enzyme immunoassay principle, human urine specimen type, sheep polyclonal antibody source reactive to secobarbital, human urine calibrator and control matrix.Explicitly stated as similar characteristics between the new and predicate device (Table 1, Device Similarities).

2. Sample size used for the test set and the data provenance

  • Test Set Sample Size: Not explicitly stated. The document mentions "patient samples" were used to demonstrate equivalence. The exact number of patient samples used in the comparison study is not provided.
  • Data Provenance: Not specified, but likely from a clinical laboratory or labs in the US, given the submission to the FDA. The term "patient samples" suggests a retrospective or prospective collection, but this detail is also not specified.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

  • Ground Truth Establishment: For in vitro diagnostic assays for drugs of abuse, the "ground truth" is typically established by a highly sensitive and specific reference method, not by expert consensus (e.g., pathologists or radiologists). The document explicitly states: "A more specific alternative chemical method must be used to confirm a result obtained with this assay. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method."
  • Expert Qualifications: Not applicable in the context of establishing ground truth for this type of assay.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

  • Not applicable. Adjudication methods like 2+1 or 3+1 are typically used in imaging studies where subjective interpretation is common. For an in vitro diagnostic assay with a quantitative gold standard (GC/MS), this process would not be used.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • Not applicable. This is not an imaging or AI-assisted diagnostic device that involves human readers interpreting results. It is an automated laboratory assay.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

  • The VITROS BARB assay is an automated homogeneous enzyme immunoassay performed on the VITROS 5,1 FS Chemistry System. Therefore, its performance is inherently standalone (algorithm/instrument only) as it provides a result without direct human interpretation of the reaction signals, though human laboratory personnel operate the system and interpret the final reported numerical or qualitative result. This study implicitly demonstrates standalone performance relative to the predicate device.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

  • The implicit "ground truth" for confirmation is Gas Chromatography/Mass Spectrometry (GC/MS), which is stated as the preferred confirmatory method for positive results. The calibration traceability is also tied to secobarbital confirmed by GC/MS. This is a highly accurate analytical chemistry method.

8. The sample size for the training set

  • Not applicable. For this type of immunoassay, there isn't a "training set" in the machine learning sense. The assay is developed based on chemical and biological principles (antibody-antigen reactions, enzyme kinetics) rather than being trained on a dataset. Reference standards and controls are used for calibration and quality control, but this is distinct from a machine learning training set.

9. How the ground truth for the training set was established

  • Not applicable, as there is no "training set" in the machine learning context. The calibrators and controls used are prepared standards with known concentrations, traceable to secobarbital with GC/MS confirmation (as mentioned for calibration traceability).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for Ortho-Clinical Diagnostics, which is a Johnson & Johnson company. The logo features a stylized symbol on the left, followed by the company name in a clean, sans-serif font. Below the company name, the text "a Johnson-Johnson company" is written in a smaller, cursive font.

JAN - 5 2007

100 Indigo Creek Drive Rochester, New York 14626-5101

5.0 510(k) Summary

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

The assigned 510(k) number is: كت كت

Ortho-Clinical Diagnostics, Inc. 1. Submitter name, 100 Indigo Creek Drive address, Rochester, New York 14626-5101 contact (585) 453-4041 Contact Person: Marlene Hanna

    1. Preparation July 27, 2006 Date

3. Device name Trade or Proprietary Names:

VITROS Chemistry Products BARB Reagent VITROS Chemistry Products Calibrator Kit 26 VITROS Chemistry Products FS Calibrator 1 VITROS Chemistry Products DAT Performance Verifiers I, II, III, IV, and V

Common Names: Barbiturates assay and controls

Classification Names:

  1. Barbiturates test system (21 CFR 862.3150) Class II

  2. Clinical toxicology calibrators (21 CFR 862.3200) Class II

  3. Clinical toxicology control material (21 CFR 862.3280) DAT Performance Verifiers I, II ,III, IV, andV) are assayed controls, so they meet the reserved criteria under Section 510(l) of the Food, Drug, and Cosmetic Act.

Continued on next page

{1}------------------------------------------------

    1. Predicate The VITROS Chemistry Products BARB assay is substantially equivalent to the Devices Syva® EMIT II Plus Barbiturate assay. The VITROS Chemistry Products DAT Performance Verifiers are substantially equivalent to the Bio-Rad Liquicheck™ UrineToxicology Controls.
    1. Device The VITROS BARB assay is a homogeneous enzyme immunoassay that is description performed using the VITROS Chemistry Products BARB Reagent with the VITROS Chemistry Products Calibrator Kit 26, VITROS Chemistry Products FS Calibrator 1, and VITROS Chemistry Products FS Diluent Pack 4 (DAT Diluent / DAT Diluent 2) on VITROS 5,1 FS Chemistry Systems.

The VITROS BARB Reagent is a dual chambered package containing ready-to-use liquid reagents that are used to detect barbiturates in urine. Sample, calibrators, and controls are automatically treated with surfactant (DAT Diluent 2) prior to addition of reagents. Treated sample is added to Reagent 1 containing antibody reactive to secobarbital, glucose-6-phosphate and nicotinamide adenine dinucleotide (NAD*), followed by Reagent 2 containing secobarbital labeled with the enzyme glucose-6phosphate dehydrogenase (G6P-DH). The assay is based on competition between barbiturates in the treated urine sample and secobarbital labeled with the enzyme glucose-6-phosphate dehydrogenase (G6P-DH) for antibody binding sites. Enzyme activity decreases upon binding to the antibody, therefore the concentration of barbiturates in the urine sample is directly proportional to measured enzyme activity. Active enzyme converts oxidized nicotinamide adenine dinucleotide (NAD') to NADH, resulting in an absorbance change that is measured spectrophotometrically at 340 nm.

The VITROS Chemistry Products Calibrator Kit 26 is prepared from human urine to which drugs of abuse, metabolites of drugs of abuse, organic salt, surfactants, and preservative have been added. VITROS Chemistry Products FS Calibrator 1 is prepared from sodium chloride and processed water. These standards are used to calibrate VITROS 5,1 FS Chemistry Systems for the qualitative and semiquantitative measurement of barbiturates (BARB).

The VITROS Chemistry Products DAT Performance Verifiers I, II, III, IV, and V are quality control materials prepared from a human urine pool to which analytes, surfactant, and preservative have been added. These are assayed controls used to monitor performance of the VITROS BARB assay on VITROS 5,1 FS Chemistry Systems.

Continued on next page

{2}------------------------------------------------

6. Deviceintended usesThe VITROS Chemistry Products FS Diluent Pack 4 (DAT Diluent / DAT Diluent 2 is a common reagent that is used by multiple assays on the VITROS 5,1 FS Chemistry System. This is a dual chambered package containing two ready-to-use liquid diluents. DAT Diluent is prepared from human urine to which organic salt, surfactants, and preservative have been added. DAT Diluent 2 is prepared from processed water to which surfactant and preservative have been added.
The VITROS 5,1 FS Chemistry System is a clinical chemistry instrument that provides automated use of the VITROS Chemistry Products MicroTip® and MicroSlides® range of products. The VITROS 5,1 FS System was cleared for market by 510(k) premarket notification (K031924).
VITROS Chemistry Products BARB Reagent: For in vitro diagnostic use only.VITROS Chemistry Products BARB Reagent is used on VITROS 5,1 FS Chemistry Systems for the semi-quantitative or qualitative determination of barbiturates (BARB) in human urine using a cutoff of either 200 ng/mL or 300 ng/mL.Measurements obtained with the VITROS BARB method are used in the diagnosis and treatment of barbiturates use or overdose.
The VITROS Chemistry Products BARB assay is intended for use by professional laboratory personnel. It provides only a preliminary test result. A more specific alternative chemical method must be used to confirm a result obtained with this assay. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method. Clinical consideration and professional judgment should be applied to any drug-of-abuse test result, particularly when evaluating a preliminary positive result.
VITROS Chemistry Products Calibrator Kit 26: For in vitro diagnostic use only.VITROS Chemistry Products Calibrator Kit 26 is used to calibrate VITROS 5,1 FS Chemistry Systems for the qualitative or semi-quantitative measurement of drugs of abuse.
VITROS Chemistry Products FS Calibrator 1: For in vitro diagnostic use only.VITROS Chemistry Products FS Calibrator 1 is used in conjunction with VITROS Chemistry Products Calibrator Kits to calibrate VITROS 5,1 FS Chemistry Systems
VITROS Chemistry Products DAT Performance Verifiers I, II, III, IV, and V:For in vitro diagnostic use only. VITROS Chemistry Products DAT Performance

Verifiers are assayed controls used to monitor performance of urine drugs of abuse screening assays on VITROS 5,1 FS Chemistry Systems.

The VITROS Chemistry Products BARB assay and VITROS Chemistry Products 7. Comparison to predicate DAT Performance Verifiers are substantially equivalent to the Syva® EMITOII Plus Barbiturate assay (K993987) and Bio-Rad Liquicheck™ Urine Toxicology Controls devices: (K022707) predicate devices which were cleared by the FDA for IVD use.

Continued on next page

{3}------------------------------------------------

The performance of the VITROS BARB assay on the VITROS 5,1 FS Chemistry System was compared to the Syva® EMIT II Plus Barbiturate assay on the SYVA 30R System. The results demonstrated good agreement between the two immunoassay methods.

The VITROS BARB assay and VITROS DAT Performance Verifiers have the following similarities to the predicate devices: same intended use, the same cutoff values of 200 or 300 ng/mL, are liquid, ready to use reagents, have similar performance characteristics, are used on an automated clinical chemistry analyzer and calibrated against the same drug, secobarbital.

Table 1
DAT controls and the Syva EMIT II Plus Barbiturate assay and Bio-Rad Urine Toxicology
controls.
Device Similarities
Device CharacteristicDescription
Indications for UseFor in vitro diagnostic use only. The assays are intended for use in the qualitativeand semi-quantitative analysis of barbiturates in human urine. The controls areassayed controls used to monitor the performance of Chemistry Systems.
Test PrincipleHomogeneous enzyme immunoassay
Cut-Off values200 or 300 ng/mL
Specimen TypeHuman Urine
Reagent FormatLiquid ready to use
Antibody sourceSheep polyclonal antibodies reactive to secobarbital
Calibration traceabilityTraceable to secobarbital with confirmation by GC/MS.
Calibrator matrixHuman urine
Control matrixHuman urine
Device Differences
Device CharacteristicVITROS BARB assay(New device)Emit® II Plus Barbiturates assay(Predicate device)
Calibrator levels6 levelsQualitative: Three levelsSemi-quantitative: Five levels
Calibrator formatFrozen Liquid ready to useRefrigerated liquid ready to use
InstrumentationTo be used on VITROS 5,1 FSChemistry SystemsMultiple automated clinical chemistrysystems
Controls: DrugsReportedCocaine metabolites(benzoylecgonine),benzodiazepines (lormetazepam),methadone, amphetamines (d-methamphetamine), opiates(morphine), cannabinoids (11-nor-delta-THC-9-COOH),phencyclidine and barbiturates(secobarbital).Methamphetamine, secobarbital,lormetazepam, tetrahydrocannabinol(THC), benzoylecgonine, ethanol, lysergicacid diethylamide (LSD), methadone,methaqualone, morphine, (Free),phencyclid ine, propoxyphene,nortriptyline and addition of creatinine,pH, specific gravity.
Controls: Number ofLevelsFiveTwo

{4}------------------------------------------------

  • The data presented in the premarket notification provide a reasonable assurance that 8. Conclusions the VITROS Chemistry Products BARB Reagent, VITROS Chemistry Products Calibrator Kit 26, VITROS Chemistry Products FS Calibrator 1, and the VITROS Chemistry Products DAT Performance Verifiers are safe and effective for the stated intended uses and are substantially equivalent to the cleared predicate devices. Equivalence to the predicates was demonstrated using a commercially available assay along with patient samples.

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized eagle-like symbol with three curved lines representing wings or feathers.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

JAN 0 5 2007

Marlene Hanna Ortho-Clinical Diagnostics, Inc. 100 Indigo Creek Drive Rochester, NY 14626-5101

K062165 Re:

Trade/Device Name: VITROS Chemistry Products BARB Reagent VITROS Chemistry Products Calibrator Kit 26 VITROS Chemistry Products FS Calibrator 1 VITROS Chemistry Products DAT Performance Verifiers I, II, III, IV, and V Regulation Number: 21 CFR 862.3150 Regulation Name: Barbiturate test system Regulatory Class: Class II Product Code: DIS, DKB, DIF

Dated: December 20, 2006

Received: December 21, 2006

Dear Ms. Hanna:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements conceming your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21

{6}------------------------------------------------

Page 2 -

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240) 276-0490. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address at http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Jean M. Cooper, M.S., D.V.M.

Yéan M. Cooper, M.S., D.V.M. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{7}------------------------------------------------

Indications for Use

510(k) Number (if known):

x062165

known):

Device Name: VITROS Chemistry Products BARB Reagent

For in vitro diagnostic use only. VITROS Chemistry Products BARB Indications for Reagent is used on VITROS 5,1 FS Chemistry Systems for the semi-Use: quantitative or qualitative determination of barbiturates (BARB) in human urine using a cutoff of either 200 ng/mL or 300 ng/mL. Measurements obtained with the VITROS BARB method are used in the diagnosis and treatment of barbiturates use or overdose.

The VITROS Chemistry Products BARB assay is intended for use by professional laboratory personnel. It provides only a preliminary test result. A more specific alternative chemical method must be used to confirm a result obtained with this assay. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method. Clinical consideration and professional judgment should be applied to any drug-of-abuse test result, particularly when evaluating a preliminary positive result.

Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Division Sign-Off

Office of In Vitro Diagnostic Device Evaluation and Safety

SNG(K) K062165

Page 1 of 2

{8}------------------------------------------------

Indications for Use

Page 2 of 2

510(k) Number (if known):K062165
Device Name:VITROS Chemistry Products Calibrator Kit 26VITROS Chemistry Products FS Calibrator 1VITROS Chemistry Products DAT Performance Verifiers I, II, III, IV, and V.
Indications for Use:For in vitro diagnostic use only. VITROS Chemistry ProductsCalibrator Kit 26 is used to calibrate VITROS 5,1 FS ChemistrySystems for the qualitative or semi-quantitative measurement of drugsof abuse.For in vitro diagnostic use only. VITROS Chemistry Products FSCalibrator 1 is used in conjunction with VITROS Chemistry ProductsCalibrator Kits to calibrate VITROS 5,1 FS Chemistry Systems.For in vitro diagnostic use only. VITROS Chemistry Products DATPerformance Verifiers are assayed controls used to monitorperformance of urine drugs of abuse screening assays on VITROS 5,1FS Chemistry Systems.
Prescription UseX
Over-The-Counter Use
AND/OR

(Part 21 CFR 801 Subpart D)

(21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Division Sign-Off

Office of In Vitro Diagnostic Device Evaluation and Safety

062/65

§ 862.3150 Barbiturate test system.

(a)
Identification. A barbiturate test system is a device intended to measure barbiturates, a class of hypnotic and sedative drugs, in serum, urine, and gastric contents. Measurements obtained by this device are used in the diagnosis and treatment of barbiturate use or overdose and in monitoring levels of barbiturate to ensure appropriate therapy.(b)
Classification. Class II (special controls). A barbiturate test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (e.g., programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).