K Number
K062135

Validate with FDA (Live)

Date Cleared
2006-11-07

(104 days)

Product Code
Regulation Number
892.5700
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ART Shielded Breast Applicator is intended for use with Nucletron or Varian remote afterloading HDR equipment for non-invasive brachytherapy of the breast in areas of intact skin.

Device Description

The Advanced Radiation Therapy (ART) Shielded Breast Applicator is used for non-invasive HDR treatment of the breast. The applicator is designed to be positioned either external to a mammography style paddle which immobilizes the breast or directly against the breast with intact skin. Once the target has been immobilized and imaged the applicator is placed such that it is aligned with the target. The exterior shielding of the applicator will reduce dose to the surrounding tissue. The applicator is available in 5 sizes - 4, 5, 6, 7 and 8 cm ID.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the ART Shielded Breast Applicator, based on the provided text:

Acceptance Criteria and Device Performance

Acceptance CriteriaReported Device Performance
Not specified.Not specified.

Explanation: The provided text explicitly states: "The Advanced Radiation Therapy Shielded Breast Applicator is a passive device. Performance testing was not required." This indicates that the device's substantial equivalence was based on its similarity to predicate devices and not on specific performance metrics or a formal study with acceptance criteria.


Detailed Study Information (Where Applicable)

Given the statement "Performance testing was not required," much of the following information is not present in the provided document.

  1. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective):

    • Not applicable, as no performance testing was required or conducted.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience):

    • Not applicable, as no performance testing was required or conducted.
  3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • Not applicable, as no performance testing was required or conducted.
  4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. This device is a passive medical device, not an AI-powered diagnostic tool.
  5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • Not applicable. This device is a passive medical device, not an algorithm.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • Not applicable, as no performance testing was required or conducted.
  7. The sample size for the training set:

    • Not applicable, as no algorithm or AI model was developed or trained for this passive device.
  8. How the ground truth for the training set was established:

    • Not applicable, as no algorithm or AI model was developed or trained for this passive device.

Summary of Study Conducted:

The premarket notification for the ART Shielded Breast Applicator did not involve a formal performance study with acceptance criteria. Instead, the basis for clearance was a Statement of Equivalency demonstrating substantial equivalence to three legally marketed predicate devices:

  • Advanced Radiation Therapy - ART Breast Brachytherapy Applicator - K060299
  • Mick Nuclear - HDR-IORT Shielded Applicator - K052351
  • Varian Medical Systems - Shielded Applicator Set - K033371

The document states that the ART Shielded Breast Applicator provides "the same or similar functions" and that its "intended use, statement of indications, and technological and performance characteristics support the concept of substantial equivalence" to these predicates. As a passive device, performance testing was deemed unnecessary by the manufacturer and, implicitly, by the FDA for the 510(k) clearance process.

{0}------------------------------------------------

Advanced Radiation Therapy

K062135

July 25, 2006

ART Shielded Breast Applicator Premarket Notification 510k Summary of Safety and Effectiveness

ART Shielded Breast Applicator

1. Sponsor Name

NOV - 7 2006

Advanced Radiation Therapy, LLC 9 Linnell Circle, Billerica, MA 01821 Telephone: 978-663-7300 Fax 978-663-7322 Contact Individual: Raymond J. Bricault Jr., Chief Operating Officer

2. Device Name

ART Shielded Breast Applicator

3. Identification of Predicate or Legally Marketed Device

Advanced Radiation Therapy - ART Breast Brachytherapy Applicator -K060299 Mick Nuclear - HDR-IORT Shielded Applicator - K052351 Varian Medical Systems - Shielded Applicator Set - K033371

ﻬﺎ Device Description

The Advanced Radiation Therapy (ART) Shielded Breast Applicator is used for non-invasive HDR treatment of the breast. The applicator is designed to be positioned either external to a mammography style paddle which immobilizes the breast or directly against the breast with intact skin. Once the target has been immobilized and imaged the applicator is placed such that it is aligned with the target. The exterior shielding of the applicator will reduce dose to the surrounding tissue. The applicator is available in 5 sizes - 4, 5, 6, 7 and 8 cm ID.

5. Intended Use

The ART Shielded Breast Brachytherapy Applicator is intended for use with Nucletron or Varian remote afterloading systems for non-invasive brachytherapy in areas of intact skin.

6. Comparison of Technological Characteristics

The predicate devices and the ART Shielded Applicator are used to provide radiation to tissue via a remote afterloader device; all are intended as accessories to remote afterloaders.

7 Performance Testing

The Advanced Radiation Therapy Shielded Breast Applicator is a passive device. Performance testing was not required.

{1}------------------------------------------------

Advanced Radiation Therapy July 25, 2006 ART Shielded Breast Applicator Premarket Notification

Statement of Equivalency

The ART Shielded Breast Applicator is substantially equivalent to the predicates, which provide the same or similar functions. The intended use, statement of indications, and technological and performance characteristics of the ART Shielded Breast Applicator support the concept of substantial equivalence.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows a logo for the Department of Health & Human Services. The logo features a stylized eagle with three lines representing its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES" is arranged in a circular fashion around the eagle.

Food and Drug Administration 9200 Corporate Blvd. Rockville MD 20850

Mr. Ravmond J. Bricault Jr. Chief Operating Officer Advanced Radiation Therapy, LLC 9 Linnell Circle BILLERICA MA 01821-3902

NOV - 7 2006

Re: K062135

Trade/Device Name: ART Shielded Breast Applicator Regulation Number: 21 CFR 892.5700 Regulation Name: Remote controlled radionuclide applicator system Regulatory Class: II Product Code: JAQ Dated: October 6, 2006 Received: October 10, 2006

Dear Mr. Bricault:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Image /page/2/Picture/10 description: The image is a black and white circular logo. The logo contains the text "1906-2006" at the top, followed by the letters "PA" in a bold, stylized font. Below the letters, the word "Centennial" is written in a cursive font. There are four stars at the bottom of the logo.

Protecting and Promoting Public Health

{3}------------------------------------------------

Page 2 -

Please be advised that FDA's issuance of a substantial equivalence determination does not mean r reast be advised that I Driv issually in that your device complies with other requirements of the Act that I DA has made a regulations administered by other Federal agencies. You must of any I cacal surates und regarants, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set CITY at 807), adomig (21 OF N P 21 CFR Part 820); and if applicable, the electronic forum in all quality systems (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) This letter will anow you'ls cognitudios of substantial equivalence of your device to a legally premarket notication: "The PDF intems
marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific advise to your of the following numbers, based on the regulation number at the top of this letter:

21 CFR 876.xxx(Gastroenterology/Renal/Urology240-276-0115
21 CFR 884.xxx(Obstetrics/Gynecology)240-276-0115
21 CFR 894.xxx(Radiology)240-276-0120
Other240-276-0100

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under (21GT K Fall 607.97): "100 Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150

or at its Internet address http://www.fda.gov/cdrl/industry/support/index.html.

Sincerely yours,

Nancy C. Hogdon

Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

July 25, 2006 Advanced Radiation Therapy ART Shielded Breast Applicator Premarket Notification Indications for Use

510(k) Number (if known):

K062135

Device Name: ART Shielded Breast Applicator

Indications for Use:

The ART Shielded Breast Applicator is intended for use with Nucletron or Varian remote afterloading HDR equipment for non-invasive brachytherapy of the breast in areas of intact skin.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

David A. Leagrom

(Division Sign-Off) Division of Reproductive, Abdominal. and Radiological Devices

510(k) Number K062135

N/A