K Number
K062128

Validate with FDA (Live)

Manufacturer
Date Cleared
2006-08-21

(26 days)

Product Code
Regulation Number
862.1215
Age Range
All
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Dimension Vista™ Creatine Kinase MB Isoenzyme (CKMB) Flex® reagent cartridge is a device intended to measurement the activity of the creatine kinase MB isoenzyme in plasma and serum. Measurements with isoenzyme are used in the diagnosis and treatment of myocardial infarction and muscle diseases such as progressive, Duchenne-type muscular dystrophy.

The HAIC assay used on the Dimension Vista™ integrated system is an in-vitro diagnostic assay for the quantitative determination of percent hemoglobin A1c(HbA1c) in anticoagulated whole blood. Measurements of percent hemoglobin A1c are effective in monitoring long term glucose control in individuals with diabetes mellitus.

The AMPH Flex® reagent cartridge used on the Dimension Vista™ integrated system provides reagents for an in-vitro diagnostic test intended for the qualitative and semiquantitative determination of amphetamines in human urine using a cutoff of either 300, 500, or 1000 ng/mL. Measurements obtained with the AMPH method are used in the diagnosis and treatment of amphetamines use or overdose. The AMPH method provides only a preliminary analytical test result. A more specific alternate chemical method must be used in order to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method. Other chemical confirmation methods are available. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are used.

The BARB Flex® reagent cartridge used on the Dimension Vista™ integrated system provides reagents for an in-vitro diagnostic test intended for the qualitative and semiquantitative determination of barbiturates in human urine. Measurements obtained with the BARB method are used in the diagnosis and treatment of barbiturates use or overdose. The BARB method provides only a preliminary analytical test result. A more specific alternate chemical method must be used in order to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method. Other chemical confirmation methods are available. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are used.

The BENZ Flex® reagent cartridge used on the Dimension Vista™ integrated system provides reagents for an in-vitro diagnostic test intended for the qualitative and semiquantitative determination of benzodiazepines in human urine. Measurements obtained with the BENZ method are used in the diagnosis and treatment of benzodiazepines use or overdose. The BENZ method provides only a preliminary analytical test result. A more specific alternate chemical method must be used in order to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method. Other chemical confirmation methods are available. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are used.

The COC Flex® reagent cartridge used on the Dimension Vista™ integrated system provides reagents for an in-vitro diagnostic test intended for the qualitative and semiquantitative determination of benzoylecgonine (cocaine metabolite) in human urine using a cutoff of 150 or 300 ng/mL. Measurements obtained with the COC method are used in the diagnosis and treatment of cocaine use or overdose. The COC method provides only a preliminary analytical test result. A more specific alternate chemical method must be used in order to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method. Other chemical confirmation methods are available. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are used.

The EXTC Flex® reagent cartridge used on the Dimension Vista™ integrated system provides reagents for an in-vitro diagnostic test intended for the qualitative and semiquantitative determination of methylenedioxymethamine (MDMA), and closely related drugs in human urine using a cutoff of either 300 or 500 ng/mL. Measurements obtained with the EXTC method are used in the diagnosis and treatment of ecstasy use or overdose. The EXTC method provides only a preliminary analytical test result. A more specific alternate chemical method must be used in order to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method. Other chemical confirmation methods are available. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are used.

The METH Flex® reagent cartridge used on the Dimension Vista™ integrated system provides reagents for an in-vitro diagnostic test intended for the qualitative and semiquantitative determination of methadone in human urine. Measurements obtained with the METH method are used to detect methadone use or overdose and to determine compliance with methadone maintenance treatment. The METH method provides only a preliminary analytical test result. A more specific alternate chemical method must be used in order to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method. Other chemical confirmation methods are available. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are used.

The OPI Flex® reagent cartridge (300 ng/mL cutoff) used on the Dimension Vista™ integrated system provides reagents for an in-vitro diagnostic test intended for the qualitative and semi-quantitative determination of opiates in human urine. Measurements obtained with the OPI method are used in the diagnosis and treatment of opiates use or overdose. The OPI method provides only a preliminary analytical test result. A more specific alternate chemical method must be used in order to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method. Other chemical confirmation methods are available. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are used.

The PCP Flex® reagent cartridge used on the Dimension Vista™ integrated system provides reagents for an in-vitro diagnostic test intended for the qualitative and semiquantitative determination of phencyclidine in human urine. Measurements obtained with the PCP method are used in the diagnosis and treatment of phencyclidine use or overdose. The PCP method provides only a preliminary analytical test result. A more specific alternate chemical method must be used in order to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method. Other chemical confirmation methods are available. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are used.

The THC Flex® reagent cartridge used on the Dimension Vista™ integrated system provides reagents for an in-vitro diagnostic test intended for the qualitative and semiquantitative determination of cannabinoids in human urine. Measurements obtained with the THC method are used in the diagnosis and treatment of cannabinoids use or overdose. The THC method provides only a preliminary analytical test result. A more specific alternate chemical method must be used in order to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method. Other chemical confirmation methods are available. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are used.

Device Description

Dade Behring Dimension Vista™ Flex® reagent cartridges and the HA1C kit are prepackaged Dade Behring products that are specifically designed to be used on the Dade Behring Dimension Vista™ integrated system, a floor model, fully automated, microprocessor-controlled, integrated instrument system. The Dimension Vista™ system was previously cleared with seven associated test methods (K051087). This Special 510(k) is submitted for a packaging modification to in-vitro diagnostic devices that have been cleared under the 510(k) process for use on Dimension® clinical chemistry systems. The packaging change is to allow use on the Dimension Vista™ system. The reagents contained in the Dimension Vista™ Flex® reagent cartridges are the same as those contained in the Flex® reagent cartridges manufactured for the Dimension® clinical chemistry systems, another family of Dade Behring analyzers. The calibrator included in the Dimension Vista™ HAIC kit is the same product (unchanged) as that used in the Dimension® HA1C Kit. The packaging modification, does not affect the intended use of the devices. nor does it alter the fundamental scientific technology of the devices.

AI/ML Overview

Please find the requested information regarding the acceptance criteria and study details below.

1. Table of Acceptance Criteria and the Reported Device Performance

The provided document describes a 510(k) submission for new reagent cartridges and a kit for the Dimension Vista™ system, which are essentially packaging modifications of existing, previously cleared devices for use on a new analyzer system. The core scientific technology and reagents remain the same.

The document mainly focuses on demonstrating substantial equivalence to predicate devices rather than establishing new performance criteria. Therefore, the "acceptance criteria" are implied to be performance comparable to the predicate devices, and the "reported device performance" is framed in terms of achieving this equivalence.

Acceptance Criteria (Implied)Reported Device Performance
Reagents are identical to predicate devices."The reagents contained in the Dimension Vista™ Flex® reagent cartridges are the same as those contained in the Flex® reagent cartridges manufactured for the Dimension® clinical chemistry systems..."
Calibrator for HA1C is identical to predicate device."The calibrator included in the Dimension Vista™ HA1C kit is the same product (unchanged) as that used in the Dimension® HA1C Kit."
Packaging modification does not affect intended use."The packaging modification, does not affect the intended use of the devices."
Packaging modification does not alter fundamental scientific technology."nor does it alter the fundamental scientific technology of the devices."
Performance is substantially equivalent to predicate devices."Comparative testing described in the protocol included in this submission demonstrates substantially equivalent performance."
"The Flex® reagent cartridges/kit... are substantially equivalent in all aspects... for same intended use and indications for use.""Comparative testing also demonstrates substantially equivalent performance."

2. Sample Size Used for the Test Set and the Data Provenance

The document states, "Comparative testing described in the protocol included in this submission demonstrates substantially equivalent performance." However, it does not specify the sample size used for the test set or the data provenance (e.g., country of origin, retrospective or prospective) for these comparative tests.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

This information is not provided in the document. As this is an in-vitro diagnostic device for quantitative and qualitative measurements, the "ground truth" would likely be established through reference methods (e.g., GC/MS for drug screens, or an established method for CKMB and HbA1c) rather than expert consensus on images or clinical cases.

4. Adjudication Method (e.g., 2+1, 3+1, none) for the Test Set

This information is not provided in the document. Given the nature of these in-vitro diagnostic tests, adjudication methods like 2+1 or 3+1, which are common in image interpretation or clinical expert review, are not applicable. Instead, the performance would be compared against a reference method or predicate device.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done, What was the Effect Size of How Much Human Readers Improve with AI vs without AI Assistance

No MRMC comparative effectiveness study was done, nor is it applicable to this submission. The device involves in-vitro diagnostic reagent cartridges for automated analyzer systems, not an AI-assisted interpretation or diagnostic tool for human readers.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

Yes, this describes a standalone performance evaluation in the context of an in-vitro diagnostic device. The Dimension Vista™ system is described as a "floor model, fully automated, microprocessor-controlled, integrated instrument system." The reagents are designed for use on this automated system. The comparative testing would confirm the analytical performance of the reagents on the new system, without human interpretation of the raw data beyond standard laboratory procedures.

7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

For the drug screens (Amphetamine, Barbiturates, Benzodiazepines, Cocaine, Ecstasy, Methadone, Opiates, Phencyclidine, Cannabinoids), the "Indications For Use" sections explicitly state: "A more specific alternate chemical method must be used in order to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method." This indicates that GC/MS is the ground truth or gold standard against which the screening tests are compared.

For Creatine Kinase MB Isoenzyme (CKMB) and Hemoglobin A1c (HA1C), while not explicitly stated as GC/MS, the "ground truth" would be established by comparison to accepted reference methods or the performance of the predicate devices themselves, as the submission aims to demonstrate substantial equivalence to these existing methods.

8. The Sample Size for the Training Set

The document is a 510(k) submission for modifications to existing products for use on a new instrument system. It focuses on comparative testing to establish substantial equivalence. It does not describe a "training set" in the context of machine learning or algorithm development. The reagents and their fundamental scientific technology are stated to be unchanged from previously cleared devices.

9. How the Ground Truth for the Training Set Was Established

As noted in point 8, there is no mention of a "training set" in the context of algorithm development or machine learning in this document. The device's underlying technology is based on established biochemical assays, not a learnable algorithm requiring a training set.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows a series of handwritten alphanumeric characters. The characters appear to be "K062128". The characters are written in a dark ink, and the background is white. The handwriting is somewhat stylized, with some characters having loops or curves.

510(k) Summary of Safety and Effectiveness Information

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

Submitter's Name:Lorraine H PiestrakDade Behring Inc.P.O. Box 6101Newark, DE 19714-6101AUG 2 1 2006
Date of Preparation:July 25, 2006

Name of Products:

Dimension Vista™ Creatine Kinase MB Isoenzyme (CKMB) Flex® reagent cartridge Dimension Vista™ Hemoglobin A1c Kit (HA1C) Dimension Vista™ Urine Amphetamine Screen (AMPH) Flex® reagent cartridge Dimension Vista™ Urine Barbiturates Screen (BARB) Flex® reagent cartridge Dimension Vista™ Urine Benzodiazepines Screen (BENZ) Flex® reagent cartridge Dimension Vista™ Urine Cocaine metabolite Screen (COC) Flex® reagent cartridge Dimension Vista™ Urine Ecstasy Screen (EXTC) Flex® reagent cartridge Dimension Vista™ Urine Methadone Screen (METH) Flex® reagent cartridge Dimension Vista™ Urine Opiates Screen (OPI) Flex® reagent cartridge Dimension Vista™ Urine Phencyclidine Screen (PCP) Flex® reagent cartridge Dimension Vista™ Urine Cannabinoids Screen (THC) Flex® reagent cartridge

FDA Classification Name:

Classification Name:Common/Usual Name:
862.1215 Differential Rate Kinetic Method, Cpk or IsoenzymesCreatine kinase isoenzyme test system
864.7470 Assay, Glycosylated HemoglobinGlycosylated hemoglobin test system
862.3100 Enzyme Immunoassay, AmphetamineAmphetamine test system
862.3150 Enzyme Immunoassay, BarbiturateBarbiturate test system
862.3170 Enzyme Immunoassay, BenzodiazepineBenzodiazepine test system
862.3250 Enzyme Immunoassay, Cocaine and MetabolitesCocaine and metabolite test system
862.3100 Enzyme Immunoassay, AmphetamineEcstasy test system
862.3620 Enzyme Immunoassay, MethadoneMethadone test system
862.3650 Enzyme Immunoassay, OpiatesOpiate test system
862.3100 Enzyme Immunoassay, PhencyclidinePhencyclidine test system
862.3870 Enzyme Immunoassay, CannabinoidsCannabinoid test system

{1}------------------------------------------------

Predicate Device:

The following table describes the predicate devices, device classification, regulation and product code associated with this pre-market notification:

New ProductPredicatePredicate510(k) #DeviceclassRegulationProductCode
Dimension Vista™CKMB Flex®reagent cartridgeDimension® Flex®CKMB reagent cartridgeK943024II862.1215JHS
Dimension Vista™HA1C KitDimension®HA1C KitK011852II864.7470LCP
Dimension Vista™AMPH Flex®reagent cartridgeDimension® Flex®AMPH reagent cartridgeK040133II862.3100DKZ
Dimension Vista™BARB Flex® reagentcartridgeDimension® Flex®BARB reagent cartridgeK000459II862.3150DIS
Dimension Vista™BENZ Flex® reagentcartridgeDimension® Flex®BENZ reagent cartridgeK000458II862.3170JXM
Dimension Vista™COC Flex® reagentcartridgeDimension® Flex®COC reagent cartridgeK033713II862.3250DIO
Dimension Vista™EXTC Flex® reagentcartridgeDimension® Flex®EXTC reagent cartridgeK053337II862.3100DKZ
Dimension Vista™METH Flex®reagent cartridgeDimension® Flex®METH reagent cartridgeK000466II862.3620DJR
Dimension Vista™OPI Flex® reagentcartridgeDimension® Flex®OPI reagent cartridgeK003209II862.3650DJG
Dimension Vista™PCP Flex® reagentcartridgeDimension® Flex®PCP reagent cartridgeK000462II862.3100LCM
Dimension Vista™THC Flex® reagentcartridgeDimension® Flex®THC reagent cartridgeK000461II862.3870LDJ

Device Description:

Dade Behring Dimension Vista™ Flex® reagent cartridges and the HA1C kit are prepackaged Dude Delining Sincers on that are specifically designed to be used on the Dade m-viro diagnosio took model (asted system, a floor model, fully automated, microprocessor-Dontrolled, integrated instrument system. The Dimension Vista™ system was previously cleared

{2}------------------------------------------------

with seven associated test methods (K051087). This Special 510(k) is submitted for a packaging modification to in-vitro diagnostic devices that have been cleared under the 510(k) process for use on Dimension® clinical chemistry systems. The packaging change is to allow use on the Dimension Vista™ system.

The reagents contained in the Dimension Vista™ Flex® reagent cartridges are the same as those contained in the Flex® reagent cartridges manufactured for the Dimension® clinical chemistry systems, another family of Dade Behring analyzers. The calibrator included in the Dimension Vista™ HAIC kit is the same product (unchanged) as that used in the Dimension® HA1C Kit. The packaging modification, does not affect the intended use of the devices. nor does it alter the fundamental scientific technology of the devices.

Intended Use:

Creatine Kinase MB Isoenzyme

The CKMB method is an in vitro diagnostic test for the quantitative measurement of creatine kinase MB Isoenzyme in human serum and plasma on the Dimension Vista™ System.

Hemoglobin A1c Kit

The HA1C assay used on the Dimension Vista™ System is an in vitro diagnostic assay for the quantitative determination of percent hemoglobin A1c (HbA1c) in anticoagulated whole blood. Measurements of percent hemoglobin Alc are effective in monitoring long term glucose control in individuals with diabetes mellitus.

Urine Amphetamine/Methamphetamine Screen

The AMPH method is an in vitro diagnostic test for the qualitative and semi-quantitative determination of amphetamines in human urine using a cutoff of either 300. 500. or 1000 ne/mL on the Dimension Vista™ System. Measurements obtained with the AMPH method are used in the diagnosis and treatment of amphetamines use or overdose.

Urine Barbiturates Screen

The BARB method is an in vitro diagnostic test for the qualitative and semi-quantitative determination of barbiturates in human urine on the Dimension Vista™ System. Measurements obtained with the BARB method are used in the diagnosis and treatment of barbiturates use or overdose.

Urine Benzodiazepines Screen

The BENZ method is an in vitro diagnostic test for the qualitative and semi-quantitative determination of benzodiazepines in human urine on the Dimension Vista™ System. Measurements obtained with the BENZ method are used in the diagnosis and treatment of benzodiazepines use or overdose.

{3}------------------------------------------------

Urine Cocaine metabolite Screen

The Cocaine method is an in vitro diagnostic test for the qualitative and semi-quantitative I he COC Methou is an in Thiro Chaghostic construction in human urine using a cutoff of 150 or delemmation of belizeyneegommo (60cm) System. Measurements obtained with the COC method are used in the diagnosis and treatment of cocaine use or overdose.

Urine Ecstasy Screen

The EXTC method is an in vitro diagnostic test for the qualitative and semi-quantitative I IIE EATC Inchiou is all in vir o diaglies betamine (MDMA) and closely related drugs in human urine using a cutoff of 300 or 500 ng/mL on the Dimension Vista™ System. Measurements urine using a cutorn of 500 or over in the diagnosis and treatment of ecstasy use or overdose.

Urine Methadone Screen

The METH method is an in vitro diagnostic test for the qualitative and semi-quantitative I he MILTTT method is all in Viro diagnoine on the Dimension Vista™ System. Measurements determination of method are used to detect methadone use or overdose, and to determine compliance with methadone maintenance treatment.

Urine Opiates Screen

The OPI method is an in vitro diagnostic test for the qualitative and semi-quantitative I he Oi i niculou is an in 'nive alagarine on the Dimension Vista™ System. Measurements decemmation of oplates in maintained in the diagnosis and treatment of opiates use or overdose.

Urine Phencyclidine Screen

The PCP method is an in vitro diagnostic test for the qualitative and semi-quantitative I IC I CT nichlod is an in viro canghouse worke on the Dimension Vista™ System. Measurements determination of phenoyondation in names and treatment of phencyclidine use or overdose.

Urine Cannabinoids Screen

The THC method is an in vitro diagnostic test for the qualitative and semi-quantitative I he 111c mediod is an in 71/7 o' diagnosurine on the Dimension Vista™ System. Measurements determination or cannaction the diagnosis and treatment of cannabinoids use or overdose.

{4}------------------------------------------------

Comparison to Predicate Device:

Compullion to 1 realert of Flex® reagent cartridges/kit and the predicate Dimension® Flex® Dour the Difficilision Visia - Pieres reagents in flexible plastic cartridges. A comparison of the reagult cartruges in contain propactials as reagues as as a cartridges is provided in the following table:

FeatureDimension Vista™Flex® reagent cartridgeDimension® AnalyzerFlex® reagent cartridge
ReagentsPrepackaged, 12-well plastic, DadeBehring Flex® reagent cartridgesPrepackaged, 6 & 8 well plastic, DadeBehring Flex® reagent cartridges
Intended Usein vitro diagnostic usein vitro diagnostic use
Indications for UseSame as Dimension® analyzerAs described in 510(k)s for eachpreviously cleared method.
Final concentrationof sample/reagentratio in test milieuSame as Dimension® analyzerAs described in 510(k)s for eachpreviously cleared method
Tablet Sizes7/32"7/32" & 9/32"
Total tests containedin each Flex®cartridgeApproximately three times morethan contained in Dimension®Flex® reagent cartridgesAs described in 510(k)s for eachpreviously cleared method.
Calibration30 to 90 days(determined for each method)30 to 90 daysAs described in 510(k)s for eachpreviously cleared method.
HA1C calibratorSame product as Dimension®HA1C calibratorAs described in K011852

Comments on Substantial Equivalence:

The Dade Behring Dimension Vista™ Flex® reagent cartridges and the Dimension® Flex® The Dado Bolling Dimensioned similarly for the same purpose. Both contain prepackaged reagents for in-vitro diagnostic tests that are processed on microprocessor-controlled, integrated instrument systems to analyze a variety of analytes in human specimens.

{5}------------------------------------------------

The reagents contained in the Dimension Vista™ Flex® reagent cartridges are the same as those The reagents connamed in the Damelible manufactured for the Dimension® clinical chemistry comained in the Frex® reagone cartring analyzers. The calibrator included in the Dimension Vista™ Systems, another railing of the same product (unchanged) as that used in the Dimension® HA1C Kit. The packaging modifications do not affect the intended use of the devices, nor do they alter the fundamental scientific technology of the devices.

Comparative testing described in the protocol included in this submission demonstrates substantially equivalent performance.

Conclusion:

Conclusion:
The Flex® reagent cartridges/kit, containing reagents for testing CKMB, HA1C, AMPH, BARB, THE FIEA® FEAGERE Canning Concenting Concension® Vista™ Integrated system are DENZ, COC, LATC, METTI, OF I, I OF , Les , Lean performance to the Dimension® system Flex® Substantially equivalent in abongs, for same intended use and indications for use. Comparative testing also demonstrates substantially equivalent performance.

Louise H. Prestvik

Lorraine H Piestrak Regulatory Affairs & Compliance Manager July 25, 2006

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized emblem featuring an abstract representation of a human figure.

Public Health Service

Food and Druq Administration 2098 Gaither Road Rockville MD 20850

Ms. Lorraine Piestrak Regulatory Affairs & Compliance Manager Dade Behring, Inc. PO Box 6101, M/S 514 Newark, DE 19714-6101

AUG 2 1 2006

Re: K062128

Trade/Device Name: Dimension Vista™ Creatine Kinase MB Isoenzyme (CKMB) Flex® reagent cartridge Dimension Vista™ Hemoglobin A1c Kit (HA1C) Dimension Vista™ Urine Amphetamine/Methamphetamine Screen (AMPH)Flex® reagent cartridge Dimension Vista™ Urine Barbiturates Screen (BARB) Flex® reagent cartridge Dimension Vista™ Urine Benzodiazepines Screen (BENZ) Flex® reagent cartridge Dimension Vista™ Urine Cocaine metabolite Screen (COC) Flex® reagent cartridge Dimension Vista™ Urine Ecstasy Screen (EXTC) Flex® reagent cartridge Dimension Vista™ Urine Methadone Screen (METH) Flex® reagent cartridge Dimension Vista™ Urine Opiates Screen (OPI) Flex® reagent cartridge Dimension Vista™ Urine Phencyclidine Screen (PCP) Flex® reagent cartridge Dimension Vista™ Urine Cannabinoids Screen (THC) Flex® reagent cartridge Regulation Number: 21 CFR$862.1215 Regulation Name: Creatine phosphokinase/creatine kinase or isoenzymes test system Regulatory Class: Class II Product Code: JHS, LCP, DKZ, DIS, JXM, DIO, DJR, DJG, LCM, LDJ Dated: July 25, 2006 Received: July 26, 2006

Dear Ms. Piestrak:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

{7}------------------------------------------------

Page 2 - Ms. Lorraine Piestrak

,

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240) 276-0450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrl/industry/support/

Sincerely yours,

Alberto
Alberto Gutierrez, Ph.D.

Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{8}------------------------------------------------

510(k) Number (if known): K06212

K062128

Dimension Vista™ Creatine Kinase MB Isoenzyme (CKMB) Flex® Device Name: reagent cartridge

Indications For Use:

The Dimension Vista™ Creatine Kinase MB Isoenzyme (CKMB) Flex® reagent cartridge is a r no Difficial of Your - See the activity of the creatine kinase MB isoenzyme in plasma and de vice intendou to measurement with isoenzyme are used in the diagnosis and treatment Scrum: "Infarction and muscle diseases such as progressive, Duchenne-type muscular dystrophy.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Page 1 of ||

Division Sign-Off

Division Sign-Off

510(k) K062128

{9}------------------------------------------------

< 062128 510(k) Number (if known):

Dimension Vista™ Hemoglobin A1c Kit (HA1C) Device Name:

Indications For Use:

The HAIC assay used on the Dimension Vista™ integrated system is an in-vitro diagnostic assay for the quantitative determination of percent hemoglobin A1c(HbA1c) in anticoagulated whole blood. Measurements of percent hemoglobin A1c are effective in monitoring long term glucose control in individuals with diabetes mellitus.

X Prescription Use (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Page ) of li

Division Sign-Off

KOi 2128

{10}------------------------------------------------

< 062128 510(k) Number (if known):

Device Name: Dimension Vista™ Urine Amphetamine/Methamphetamine Screen (AMPH) Flex® reagent cartridge

Indications For Use:

The AMPH Flex® reagent cartridge used on the Dimension Vista™ integrated system provides reagents for an in-vitro diagnostic test intended for the qualitative and semiquantitative determination of amphetamines in human urine using a cutoff of either 300, 500, or 1000 ng/mL. Measurements obtained with the AMPH method are used in the diagnosis and treatment of amphetamines use or overdose.

The AMPH method provides only a preliminary analytical test result. A more specific alternate chemical method must be used in order to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method. Other chemical confirmation methods are available. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are used.

Prescription Use x (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

C. Taylor

Page 3 of 11

Division Sign/Off

KOC 2128

{11}------------------------------------------------

002128 510(k) Number (if known):

Device Name: Dimension Vista™ Urine Barbiturates Screen (BARB) Flex® reagent cartridge

Indications For Use:

The BARB Flex® reagent cartridge used on the Dimension Vista™ integrated system provides reagents for an in-vitro diagnostic test intended for the qualitative and semiquantitative determination of barbiturates in human urine. Measurements obtained with the BARB method are used in the diagnosis and treatment of barbiturates use or overdose.

The BARB method provides only a preliminary analytical test result. A more specific alternate chemical method must be used in order to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method. Other chemical confirmation methods are available. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are used.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Division Sign-Off

Page of 11

062128

{12}------------------------------------------------

510(k) Number (if known): < Olo 2128

Dimension Vista™ Urine Benzodiazepines Screen (BENZ) Flex® Device Name: reagent cartridge

Indications For Use:

The BENZ Flex® reagent cartridge used on the Dimension Vista™ integrated system provides reagents for an in-vitro diagnostic test intended for the qualitative and semiquantitative determination of benzodiazepines in human urine. Measurements obtained . with the BENZ method are used in the diagnosis and treatment of benzodiazepines use or overdose.

The BENZ method provides only a preliminary analytical test result. A more specific alternate chemical method must be used in order to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method. Other chemical confirmation methods are available. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are used.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

GTC

Page of 11

Jivision Sign-Off

K062128

{13}------------------------------------------------

06200 510(k) Number (if known):

Dimension Vista™ Urine Cocaine metabolite Screen (COC) Flex® Device Name: reagent cartridge

Indications For Use:

The COC Flex® reagent cartridge used on the Dimension Vista™ integrated system provides reagents for an in-vitro diagnostic test intended for the qualitative and semiquantitative determination of benzoylecgonine (cocaine metabolite) in human urine using a cutoff of 150 or 300 ng/mL. Measurements obtained with the COC method are used in the diagnosis and treatment of cocaine use or overdose.

The COC method provides only a preliminary analytical test result. A more specific alternate chemical method must be used in order to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method. Other chemical confirmation methods are available. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are used.

Prescription Use X · (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

CAC

Division Sign-Off

Page of II

062128

{14}------------------------------------------------

ﭗ ﮨ k 062128 510(k) Number (if known):

Dimension VistaTM Urine Ecstasy Screen (EXTC) Flex® reagent Device Name: cartridge

Indications For Use:

The EXTC Flex® reagent cartridge used on the Dimension Vista™ integrated system provides reagents for an in-vitro diagnostic test intended for the qualitative and semiquantitative determination of methylenedioxymethamine (MDMA), and closely related drugs in human urine using a cutoff of either 300 or 500 ng/mL. Measurements obtained with the EXTC method are used in the diagnosis and treatment of ecstasy use or overdose.

The EXTC method provides only a preliminary analytical test result. A more specific alternate chemical method must be used in order to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method. Other chemical confirmation methods are available. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are used.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Ote

Jusion Sign-Off

Page 7 of 11

K062128

{15}------------------------------------------------

510(k) Number (if known): KO6 2129

Dimension Vista™ Urine Methadone Screen (METH) Flex® reagent Device Name: cartridge

Indications For Use:

The METH Flex® reagent cartridge used on the Dimension Vista™ integrated system provides reagents for an in-vitro diagnostic test intended for the qualitative and semiquantitative determination of methadone in human urine. Measurements obtained with the METH method are used to detect methadone use or overdose and to determine compliance with methadone maintenance treatment.

The METH method provides only a preliminary analytical test result. A more specific alternate chemical method must be used in order to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method. Other chemical confirmation methods are available. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are used.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

G.C.

Author Signature OR

Page Bof 11

c 2128

{16}------------------------------------------------

KO62128 510(k) Number (if known):

Dimension Vista™ Urine Opiates Screen (OPI) Flex® reagent Device Name: cartridge

Indications For Use:

The OPI Flex® reagent cartridge (300 ng/mL cutoff) used on the Dimension Vista™ integrated system provides reagents for an in-vitro diagnostic test intended for the qualitative and semi-quantitative determination of opiates in human urine. Measurements obtained with the OPI method are used in the diagnosis and treatment of opiates use or overdose.

The OPI method provides only a preliminary analytical test result. A more specific alternate chemical method must be used in order to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method. Other chemical confirmation methods are available. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are used.

Prescription Use (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Vision Sign-Off

Page of of 11

K062128

{17}------------------------------------------------

2062124 510(k) Number (if known):

Dimension Vista™ Urine Phencyclidine Screen (PCP) Flex® reagent Device Name: cartridge

Indications For Use:

The PCP Flex® reagent cartridge used on the Dimension Vista™ integrated system provides reagents for an in-vitro diagnostic test intended for the qualitative and semiquantitative determination of phencyclidine in human urine. Measurements obtained with the PCP method are used in the diagnosis and treatment of phencyclidine use or overdose.

The PCP method provides only a preliminary analytical test result. A more specific alternate chemical method must be used in order to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method. Other chemical confirmation methods are available. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are used.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Approval Sign-Off

Page 10of 11

2062128

{18}------------------------------------------------

510(k) Number (if known):

1062121

Dimension Vista™ Urine Cannabinoids Screen (THC) Flex® reagent Device Name: cartridge

Indications For Use:

The THC Flex® reagent cartridge used on the Dimension Vista™ integrated system provides reagents for an in-vitro diagnostic test intended for the qualitative and semiquantitative determination of cannabinoids in human urine. Measurements obtained with the THC method are used in the diagnosis and treatment of cannabinoids use or overdose.

The THC method provides only a preliminary analytical test result. A more specific alternate chemical method must be used in order to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method. Other chemical confirmation methods are available. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are used.

Prescription Use (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Vision Sign Off

Page Hof 1

KSC 2128

§ 862.1215 Creatine phosphokinase/creatine kinase or isoenzymes test system.

(a)
Identification. A creatine phosphokinase/creatine kinase or isoenzymes test system is a device intended to measure the activity of the enzyme creatine phosphokinase or its isoenzymes (a group of enzymes with similar biological activity) in plasma and serum. Measurements of creatine phosphokinase and its isoenzymes are used in the diagnosis and treatment of myocardial infarction and muscle diseases such as progressive, Duchenne-type muscular dystrophy.(b)
Classification. Class II.