K Number
K062097

Validate with FDA (Live)

Date Cleared
2006-10-13

(81 days)

Product Code
Regulation Number
884.1720
Age Range
All
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Pajunk insufflation cannula acc. Veress and the modular insufflation cannula acc. Veress is employed in minimal invasive surgery. The cannula is designed for the initial puncture with subsequent gas insuffation for laparoscopic operations. It is used to establish a pneumoperitoneum.

Device Description

The insufflation cannulae according Veress were developed especially for the safe and efficient r rio moumation oannalae and insufflation cannulae are equipped with maintenance free cocks and a female LuerLock connector. Before application, a sharp outer cannula (sterile packed for modular Veress, reusable for Standard Veress) is mounted onto the reusable inner cannula and fastened by means of a LuerLock connector. Therefore within the modular system , a new, absolutely sharp disposable outer cannula is used for every application. In comparison with totally disposable Veress-cannulae, with the same advantages, the modular system offers a large potential to reduce cost.

AI/ML Overview

This is a premarket notification for a Veress and Modular Veress Insufflation Cannula. This type of submission typically focuses on demonstrating substantial equivalence to a predicate device rather than presenting extensive de novo clinical study data with specific acceptance criteria and detailed device performance metrics. Therefore, many of the requested data points (e.g., sample sizes for test and training sets, number of experts for ground truth, MRMC study details, acceptance criteria table) are not applicable or explicitly provided in this document.

Here's the information that can be extracted or inferred based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

This type of information is generally not included in 510(k) submissions for devices like the Veress cannula, which are often cleared based on substantial equivalence to predicate devices rather than meeting specific performance criteria in a new clinical study. The device's performance is assumed to be equivalent to the predicate devices.

2. Sample Size Used for the Test Set and Data Provenance

A formal "test set" in the context of a prospective clinical study with a defined sample size to prove performance against specific acceptance criteria is not described. The submission relies on substantial equivalence and "clinical evaluation and summarizing literature." Therefore, specific sample sizes for a 'test set' derived from a new study are not provided. The data provenance is described as "clinical evaluation and summarizing literature," suggesting a review of existing data rather than a new prospective study.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

Not applicable. The submission does not describe a new clinical study with a ground truth established by experts.

4. Adjudication Method for the Test Set

Not applicable. The submission does not describe a new clinical study with an adjudication process.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of How Much Human Readers Improve with AI vs without AI Assistance

Not applicable. This device is a surgical instrument (insufflation cannula), not an AI-powered diagnostic or assistive tool. Therefore, an MRMC study and AI-related metrics are irrelevant.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done

Not applicable. This device is a surgical instrument, not an algorithm.

7. The Type of Ground Truth Used

The "ground truth" for the device's safety and effectiveness is primarily based on substantial equivalence to existing legally marketed predicate devices, which have a history of safe and effective use. Additionally, the submission mentions "clinical evaluation and summarizing literature," suggesting that the medical community's existing knowledge and published findings on similar devices serve as a form of "ground truth" regarding the general principles of Veress cannula use.

8. The Sample Size for the Training Set

Not applicable. This is not an AI/ML device that requires a training set.

9. How the Ground Truth for the Training Set Was Established

Not applicable. This is not an AI/ML device that requires a training set with established ground truth.


Summary of Device Rationale from the Submission:

The submission for the Veress and Modular Veress Insufflation Cannula focuses on demonstrating substantial equivalence to two predicate devices:

  1. Cannula with sharp obturator & Veress Needle by Karl Storz (K800668)
  2. Veress Cannula et al. by Richard Wolf Medical (K041321)

The rationale provided is that the proposed devices are "as safe and effective as and therefore substantial equivalent to the predicate devices" in terms of intended use, indication, and technical characterization. The document explicitly states: "The comparison between the predicate devices and the proposed devices... demonstrates that the proposed devices are safe and effective, as well as substantially equivalent to the predicate devices." It also mentions "The cannula system acc. Veress has been used for years now. The clinical evaluation and summarizing literature, which is part of this submission (10.0), makes this aspects evident." This implies that prior clinical experience and published data on this type of device support its safety and effectiveness without requiring a new, formal clinical trial with specific performance criteria.

{0}------------------------------------------------

Veress and Modular Veress Insufflation Cannula

Premarket Notification Submission

510(k) Premarket Notification Submission:

Summary of Safety and Effectiveness

Date of Preparation: July 20th, 2006

OCT 1 3 2006

Submitter Information/ production site:

Pajunk GmbH Karl-Hall-Strasse 01 78187 Geisingen Germany Fon: +49(0)7704-9291-0 Fax: +49(0)7704-9291-610 Establishment Registration Number: 9611612

USA Contact:

Pajunk USA German American Trade Center 5126 South Royal Atlanta Drive 30084 Tucker, Georgia USA

Contact:

Christian Quass, Director Regulatory Affairs Fon: +49(0)7704-9291-586 Fax: +49(0)7704-9291-605 E-Mail: christian.quass@pajunk.com

Contact

Stefan Dayagi Fon: +01(0)770-493-9305

E-Mail: stefan.dayagi@pajunk-usa.com

Contract Sterilizer:

SteriPro Lab & EO Facility Dreieichstrasse. 7 64546 Mörfelden Germany Tel +49 6105 23091 or +49 (0) 6105 93470 Fax +49 6105 24760

Device Information:

Trade Names: Veress Cannula and Modular Veress Cannula

Common Name: Veress Insufflation Cannula

Classification Name: (reusable) laparoscope, general & plastic surgery

Classification Reference: 21 CfR 876.1500

Poposed Classification: Regulatory Class: II

Proposed Product Code: GCJ, NLM

Review Panel General & Plastic Surgery

Requlation description Endoscopes and accessories

Predicate Devices:

    1. Cannula with sharp obturator & Veress Needle by Karl Storz K800668
      1. Veress Cannula et al. by Richard Wolf Medical K041321

Image /page/0/Picture/32 description: The image contains the number 021. The numbers are black and are on a white background. The numbers are aligned horizontally.

Image /page/0/Picture/33 description: The image shows the logo for PAJUNK Medizintechnologie. Above the logo is the text K062097 and page 1 of 3. The logo is in black and white and features the company name in a bold, sans-serif font.

{1}------------------------------------------------

Veress and Modular Veress Insufflation Cannula

Premarket Notification Submission

Image /page/1/Picture/2 description: The image contains the logo for PAJUNK MEDIZINTECHNOLOGIE. There is also some handwritten text that says K062097 and Page 243. The PAJUNK logo is in a black box with white text.

Device Description:

The insufflation cannulae according Veress were developed especially for the safe and efficient r rio moumation oannalae and insufflation cannulae are equipped with maintenance free cocks and a female LuerLock connector. Before application, a sharp outer cannula (sterile packed for modular Veress, reusable for Standard Veress) is mounted onto the reusable inner cannula and fastened by means of a LuerLock connector. Therefore within the modular system , a new, absolutely sharp disposable outer cannula is used for every application. In comparison with totally disposable Veress-cannulae, with the same advantages, the modular system offers a large potential to reduce cost.

Predicate Devices:

The predicate devices are the Veress cannula of Karl Storz (K800668) and the Veress cannula of Richard Wolf (K041321). They are found to be substantially equivalent in intended use, indication and technical characterization.

The Pajunk Veress cannula and the Pajunk modular Veress cannula is found to be as safe and effective as and therefore substantial equivalent to the predicate devices.

The detailed discussion of substantial eqivalence can be found in Section 12 of this submission.

Sterilization

The Insufflation cannula according Veress is supplied non-sterile. It has to be cleaned, disinfected and sterilized before each use by the customer facility according to ist own sterilization protocols.

Cleaning, disinfection and sterilization information for devices supplied non-sterile can be found in section 14.0 of this submission.

The modular insufflation system acc. Veress consists of a sterile sharp outer cannula for skin penetration and a reusable or sterile insufflation cannulal blunt obturator acc. Veress for safe and effective surgical intervention.

The contract sterilizer and the sterilizing process at Sterigenics is the same as that one used for all further Pajunk products already cleared for market in the USA. The Veress cannula is not the worst-case-product which indeed is the Sprotte (K911221, K911202, K911260) and the StimuLong Kit (K043130, K033018) within Pajunks sterilization process.

Packaging and Labeling

The non-sterile Veress insufflation cannula is packaged and labeled just like all the other Pajunks FDA-approved products marketed non-sterile, for example Pajunks Trocar and Balloon Systems K012771 or Pajunks Handles and Electrodes K011997, K033249. Samples and process descriptions for labeling and packaging can be found in section 13.0 and 14.0 of this submission.

Biocompatibility status

All materials employed in the manufacturing process that may come in contact with blood, tissue or fluids to be injected have been cleared in Pajunks former 510(k) applications. Furthermore these materials are long term proven materials for the use with medical devices.

{2}------------------------------------------------

Veress and Modular Veress Insufflation Cannula

Premarket Notification Submission

Image /page/2/Picture/2 description: The image contains the logo for PAJUNK MEDIZINTECHNOLOGIE. The logo has the word "PAJUNK" in bold, white letters inside of a black rectangle. Below the rectangle, the word "MEDIZINTECHNOLOGIE" is written in smaller, black letters. There is also some handwritten text to the right of the logo.

Indicated Patient population

The indicated patient population is adult, even though there are several studies for pediatric use.

Technology Characteristics:

The inner and outer cannulas consist of stainless steel. Biocompatibility is proven and documented. The outer cannulas guide cone of the reusable Veress as well as the body of the inner cannula consist of brass while the guide cone/ hub of the disposable cannula consists of inner cannula somala is intended to penetrate the skin. Then the inner cannula is pushed through and spring locked. Now the pneumoperitoneum can be created by insufflation via the inner cannula.

Conclusion:

The comparison between the predicate devices and the proposed devices in section 12 of this submission demonstrates that the proposed devices are safe and effective, as well as substantially equivalent to the predicate devices.

The cannula system acc. Veress has been used for years now. The clinical evaluation and summarizing literature, which is part of this submission 10.0), makes this aspects evident.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the top and bottom. Inside the circle is a stylized symbol featuring three curved lines resembling a bird or a wave-like design.

Food and Drug Administration 9200 Corporate Blvd. Rockville MD 20850

Mr. Christian Quass Director of Regulatory Affairs Pajunk GmbH Medizintechnologies Karl-Hall-Str. 1, 78187 Geisingen GERMANY

OCT 1 3 2006

Re: K062097

Trade/Device Name: Veress Cannula and Modular Veress Cannula Regulation Number: 21 CFR 884.1720 Regulation Name: Gynecologic laparoscope and accessories Regulatory Class: II Product Code: HET Dated: September 25, 2006 Received: September 27, 2006

Dear Mr. Quass:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Image /page/3/Picture/10 description: The image is a circular logo or emblem. The text "1906-2006" is at the top of the circle. The letters "FDA" are in the center of the circle in a bold, stylized font. Below the letters, the word "Centennial" is written in a cursive font. Three stars are at the bottom of the circle.

Protecting and Promoting Public Health.

{4}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic

product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:

21 CFR 876.xxx(Gastroenterology/Renal/Urology)240-276-0115
21 CFR 884.xxx(Obstetrics/Gynecology)240-276-0115
21 CFR 894.xxx(Radiology)240-276-0120
Other240-276-0100

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150

or at its Internet address http://www.fda.gov/cdrl/industry/support/index.html.

Sincerely vours.

Nancy C. Hogdon

Nancy C. Brogdon Director, Division of Reproductive. Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

Page 2 -

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the logo for PAJUNK Medizintechnologie. The word "PAJUNK" is in large, bold, white letters against a black background. Below that, "MEDIZINTECHNOLOGIE" is written in smaller, white letters.

Indications for use

510(k) Number:

Device Name:

Veress cannula and Modular Veress Cannula

Indications for Use:

The Pajunk insufflation cannula acc. Veress and the modular insufflation cannula acc. Veress is employed in minimal invasive surgery. The cannula is designed for the initial puncture with subsequent gas insuffation for laparoscopic operations. It is used to establish a pneumoperitoneum.

Prescription Use X
(Per 21 CFR 801.109)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

David A. Legorre

(Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number

Page 1 of 1

§ 884.1720 Gynecologic laparoscope and accessories.

(a)
Identification. A gynecologic laparoscope is a device used to permit direct viewing of the organs within the peritoneum by a telescopic system introduced through the abdominal wall. It is used to perform diagnostic and surgical procedures on the female genital organs. This generic type of device may include: Trocar and cannula, instruments used through an operating channel, scope preheater, light source and cables, and component parts.(b)
Classification. (1) Class II (performance standards).(2) Class I for gynecologic laparoscope accessories that are not part of a specialized instrument or device delivery system, do not have adapters, connector channels, or do not have portals for electrosurgical, lasers, or other power sources. Such gynecologic laparosope accessory instruments include: the lens cleaning brush, biopsy brush, clip applier (without clips), applicator, cannula (without trocar or valves), ligature carrier/needle holder, clamp/hemostat/grasper, curette, instrument guide, ligature passing and knotting instrument, suture needle (without suture), retractor, mechanical (noninflatable), snare, stylet, forceps, dissector, mechanical (noninflatable), scissors, and suction/irrigation probe. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 884.9.