(18 days)
RadiForce GS320 is intended to be used in displaying and viewing digital images for diagnosis of X-ray or MRI, etc. by trained medical practitioners. The device is not specified for digital mammography system.
RadiForce GS320 is a 54cm (21.3") Monochrome LCD display for medical image viewing. GS310 displays high-definition medical imaging.
The provided document is a 510(k) summary for a medical display monitor (EIZO NANAO CORPORATION Monochrome LCD Monitor, RadiForce GS320), not a device that utilizes AI or provides diagnostic interpretations. Therefore, many of the requested fields regarding acceptance criteria, study details, expert involvement, and ground truth are not applicable to this type of device and are not present in the submission.
The 510(k) process for a monitor like this focuses on demonstrating substantial equivalence to a predicate device through technical specifications and intended use, rather than clinical performance studies that directly assess AI accuracy or human reader improvement.
Here's the information that can be extracted or noted as not applicable based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
For medical display monitors, "acceptance criteria" and "reported device performance" are typically defined by technical specifications and adherence to standards, rather than clinical accuracy metrics. The comparison is primarily against a predicate device's technical specifications.
| Characteristic | Acceptance Criteria (Predicate Device K060845 - RadiForce G310) | Reported Device Performance (RadiForce GS320) |
|---|---|---|
| Panel Size and Type | 53 cm (20.8") TFT Monochrome LCD panel | 54 cm (21.3") TFT Monochrome LCD panel |
| Pixel Pitch | 0.207 x 0.207mm | 0.2115 x 0.2115mm |
| Ratio of Sub-pixel Opening Areas | 4: 4: 4 | Same as predicate |
| Grayscale Tones | 1,024 from a pallet of 8,161 | Same as predicate |
| Viewing Angles | H: 170°, V: 170° | Same as predicate |
| Scanning Frequency (H, V) | 31-100kHz, 48-71.5Hz (VGA Text: 69-71Hz) Frame synchronous mode: 59-61Hz | Same as predicate |
| Native Resolutions | 1536 x 2048 | Same as predicate |
| Brightness | 700 cd/m² (Typical) | Same as predicate |
| Contrast Ratio | 900: 1 (typical) | 850: 1 (typical) |
| DOT Clock | 165 MHz | Same as predicate |
| Response Time | 50 ms (typical) | Same as predicate |
| Active Display Size (H x V) | 318 x 424 mm | 324 x 433 mm |
| Viewable Image Size | 529 mm (20.8") (diagonal) | 541mm (21.3") (diagonal) |
| Luminance Calibration | Built-in swing calibration sensor provided | Same as predicate |
| Input Signals | DVI Standard 1.0 | Same as predicate |
| Input Terminals | DVU-D 24 pin | Same as predicate |
| USB Ports / Standard | 1 upstream, 2 downstream | 1 upstream, 2 downstream/Standard Rev.2.0 |
| Power | AC100-120V, 200-240V, 50/60Hz | Same as predicate |
| Power Management | DVI-DMPM | Same as predicate |
| Dimensions (W x H x D) | With Stand: 368 x 515.5 mm – 597.5 x 209 mm Without Stand: 368 x 486 x 90 mm | With Stand: 376 x 522.5 mm – 604.5x 208.5 mm Without Stand: 376 x 500 x 92 mm |
| Certifications & Standards | TUV/GM, CE Medical Device Directive, CB (EN60601-1), cTUVus (UL2601-1, CSA C22.2 No. 601-1), VCCI-B, FCC-B, Canadian ICES-003-A, CCC | Same as predicate |
The key acceptance criterion here is demonstrating "substantial equivalence" to the predicate device (RadiForce G310, K060845), meaning the new device has similar technological characteristics and intended use, and does not raise new questions of safety or effectiveness. The reported performance of the GS320 is compared directly against the specifications of the GS310. The submission states: "Comparison table of the principal characteristics of 2 devices in Attachment 1 shows that new and predicate devices are substantially equivalent in the areas of technical characteristics, general functions." The modifications noted are a slightly larger panel size and a lower contrast ratio, with the submission implicitly arguing these do not negate substantial equivalence for the intended use.
2. Sample size used for the test set and the data provenance
N/A. This is a medical display monitor, not an AI or diagnostic device that would typically undergo testing with a "test set" of patient data. The evaluation is based on engineering specifications and comparison to a predicate device.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
N/A. No ground truth or expert evaluation of diagnostic performance is mentioned or required for this type of device submission.
4. Adjudication method for the test set
N/A. No test set for diagnostic performance is mentioned.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
N/A. This is not an AI-assisted device.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
N/A. This is a medical display monitor, not an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
N/A. Ground truth is not relevant for a medical display monitor's 510(k) submission, which focuses on technical specifications and substantial equivalence.
8. The sample size for the training set
N/A. This device does not involve a training set as it's a hardware monitor, not a machine learning algorithm.
9. How the ground truth for the training set was established
N/A. Not applicable for a hardware device.
{0}------------------------------------------------
510(k) Summary as required by 807.92
AUG - 7 2006
- Company Identification EIZO NANAO CORPORATION 153 Shimokashiwano-cho, Hakusan-shi, Ishikawa-ken, 924-8566, Japan Tel: +81-76-274-2468 Fax: +81-76-274-2484
-
- Official Correspondent Hiroaki Hashimoto (Mr.) Manager of Engineering Management Section
-
- Date of Submission June 28, 2006
-
- Device Trade name Monochrome LCD Monitor, RadiForce GS320
-
- Common/Usual Name Image display system, medical image workstation, image monitor/display, and others
6. Classification Number
Medical displays classified in Class II per 21 CFR 892.2050.
-
- Predicate Device
| Manufacturer | : | EIZO NANAO CORPORATION |
|---|---|---|
| Device Name | : | Monochrome LCD Monitor |
| Model Name | : | RadiForce G310 |
| 510(k) No. | : | K060845 |
| Items | GS310 | GS320 |
| 510(k) Number | K060845 | Not known |
| Panel Size and Type | 53 cm (20.8") TFT MonochromeLCD panel | 54 cm (21.3") TFT MonochromeLCD panel |
| Cabinet Color | Black | It is same as the following. |
| Pixel Pitch | 0.207 x 0.207mm | 0.2115 x 0.2115mm |
| Ratio of Sub-pixelOpening Areas | 4: 4: 4 | It is same as the following. |
| Grayscale Tones | 1,024 from a pallet of 8,161 | It is same as the following. |
| Viewing Angles | H: 170°, V: 170° | It is same as the following. |
| Scanning Frequency(H, V) | 31-100kHz, 48-71.5Hz(VGA Text: 69-71Hz)Frame synchronous mode: 59-61Hz | It is same as the following. |
| Native Resolutions | 1536 x 2048 | It is same as the following. |
| Brightness | 700 cd/m² (Typical) | It is same as the following. |
| Contrast Ratio | 900: 1 (typical) | 850: 1 (typical) |
| DOT Clock | 165 MHz | It is same as the following. |
| Response Time | 50 ms (typical) | It is same as the following. |
| Active Display Size(H x V) | 318 x 424 mm | 324 x 433 mm |
| Viewable Image Size | 529 mm (20.8") (diagonal) | 541mm (21.3") (diagonal) |
| Luminance Calibration | Built-in swing calibration sensorprovided. | It is same as the following. |
| Input Signals | DVI Standard 1.0 | It is same as the following. |
| Input Terminals | DVU-D 24 pin | It is same as the following. |
| USB Ports / Standard | 1 upstream, 2 downstream | 1 upstream, 2downstream/Standard Rev.2.0 |
| Power | AC100-120V, 200-240V, 50/60Hz | It is same as the following. |
| Power Management | DVI-DMPM | It is same as the following. |
| Dimensions (W x H x D) | With Stand:368 x 515.5 mm– 597.5 x 209 mmWithout Stand:368 x 486 x 90 mm | With Stand:376 x 522.5 mm– 604.5x 208.5 mmWithout Stand:376 x 500 x 92 mm |
| Certifications &Standards | TUV/GM, CE Medical DeviceDirective, CB (EN60601-1),cTUVus (UL2601-1, CSA C22.2 No.601-1), VCCI-B, FCC-B, CanadianICES-003-A, CCC | TUV/GM, CE Medical DeviceDirective, CB (EN60601-1),cTUVus (UL2601-1, CSA C22.2 No.601-1), VCCI-B, FCC-B, CanadianICES-003-A, CCC |
8. Description of Device
RadiForce GS320 is a 54cm (21.3") Monochrome LCD display for medical image viewing. GS310 displays high-definition medical imaging.
-
- Intended Use
RadiForce GS320 is intended to be used in displaying and viewing digital images for diagnosis of Xray or MRI, etc. by trained medical practitioners. The device is not specified for digital mammography system.
- Intended Use
10. Technological Characteristics
GS320 employs smaller grayscale tones than that of GS310. The only modification is that the panel size became big with 54cm(21.3") from 53cm(20.8"). Comparison table of the principal characteristics of 2 devices in Attachment 1 shows that new and predicate devices are substantially equivalent in the areas of technical characteristics, general functions.
{1}------------------------------------------------
・
:
Appendix 1: Comparison Table with Predicate Device
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines representing its wings, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular pattern around the eagle. The logo is black and white.
Food and Drug Administration 9200 Corporate Blvd. Rockville MD 20850
AUG - 7 2006
Mr. Hiraoki Hashimoto Manager of Engineering Management Section EIZO NANAO Corporation 153 Shimokashiwano, Hakusan, Ishikawa 924-8566 JAPAN
Re: K062053
Trade/Device Name: Monochrome LCD Monitor, RadiForce GS320 Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: July 3, 2006 Received: July 20, 2006
Dear Mr. Hashimoto:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Image /page/2/Picture/10 description: The image shows a circular logo with the letters "FDA" in a stylized font at the center. Above the letters, the years "1906-2006" are printed. Below the letters, the word "Centennial" is written in cursive, followed by three stars. The text around the circle reads "U.S. Food and Drug Administration".
Protecting and Promoting Public Health
{3}------------------------------------------------
Page 2 -
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 21 CFR 876.xxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|---|---|---|
| 21 CFR 884.xxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 894.xxx | (Radiology) | 240-276-0120 |
| Other | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150
or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Nancy C. Hogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Indications for Use
510(k) Number (if known): Not known
Device Name : Monochrome LCD Monitor, RadiForce GS320
Indications For Use:
RadiForce GS320 is intended to be used in displaying and viewing digital images for diagnosis of X-ray or MRI, etc. by trained medical practitioners. The device is not specified for digital mammography system.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
David A. Lynn
(Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number
N/A