K Number
K062052

Validate with FDA (Live)

Device Name
LMT NOMAG IC 3.0
Date Cleared
2006-07-31

(11 days)

Product Code
Regulation Number
880.5410
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The LMT nomag IC 3.0 MR-incubator is an infant incubator system for temporary use during MR-imaging in clinical environments provided with a suitable trolley for intra hospital transport.

The LMT nomag IC 3.0 MR-incubator provides a controlled environment of warmth and humidity for premature babies and sick infants up to a body weight of 4.5 kg (10 lbs) and a maximum body length of 55 cm (21.7 inches).

The LMT nomag IC 3.0 is suitable for use with MR scanners of field strengths up to 3.0T.

Device Description

The LMT Nomag IC 3.0 is an incubator for preterm and term-born infants who are scheduled for Magnetic Resonance (MR) Imaging and are dependant on warming therapy. It provides a microclimate of air temperature and humidity suiting the infants needs. The incubator has a trolley for intra clinical transport providing power.

The incubator runs in air control mode and has closed loop control for both air temperature and humidity. Humidification works hygienically via vaporizing of distilled water.

AI/ML Overview

The acceptance criteria for the LMT nomag IC 3.0 neonatal transport incubator and the study that proves its performance are described below based on the provided text.

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance CriteriaReported Device Performance
1. Basic Incubator PerformanceThe basic incubator performance of the LMT nomag IC 3.0 is defined by the IEC particular standard. The design changes over the nomag IC 1.5. do not affect these performance figures.
(Implicit) 2. Compatibility with MR Scanners (No Adverse Effects on Incubator)During testing, the incubator has shown no adverse effects in its behavior due to the MR scanner.
(Implicit) 3. Compatibility with MR Scanners (No Significant Artifacts in Imaging)Imaging has not shown significant artifacts caused by the incubator.
(Implicit) 4. Compatibility with MR Scanners (No Influence on Other Scanner Functions)Other functions of the scanner were also not influenced by the incubator.
5. Compatibility with MR Scanners up to 3T Field StrengthThe compatibility to MR scanners with 3T field strength has been proven.
6. Compliance with Harmonized StandardsThe LMT Nomag IC 3.0 incubator is compliant with international harmonized standards for incubators by the IEC, specifically IEC 60601-2-20: 1996 and IEC 60601-1 (Ed.2): 1995.

2. Sample Size Used for the Test Set and Data Provenance

The text does not explicitly state a specific sample size for the test set in terms of individual incubators or the number of tests performed. However, it mentions:

  • "extensive bench testing"
  • "testing under MR influence"
  • "compatibility with MR scanners of different manufacturers (Siemens, Philips, GE 1.5T and 3T) have been tested according to an established protocol using phantoms."

The data provenance is prospective bench testing conducted in, or on behalf of, LMT Lammers Medical Technology GmbH in Germany, based on the applicant's address.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts

The text does not provide information regarding the number of experts used or their qualifications for establishing ground truth. The evaluation appears to be based on objective measurements and established protocols for MR compatibility and incubator performance.

4. Adjudication Method for the Test Set

The text does not describe an adjudication method. The testing appears to rely on direct measurement and observation of incubator function and MR image quality against established protocols and standards ("extensive bench testing," "established protocol using phantoms").

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

A Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not performed, nor is it applicable to this device. The LMT nomag IC 3.0 is a medical device (neonatal transport incubator) and not an AI-powered diagnostic or assistive tool for human readers.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

A "standalone" performance study in the context of an algorithm or AI was not done, nor is it applicable. This device is a physical medical incubator, not a software algorithm. The "standalone" performance here refers to the device's inherent function, which was evaluated through bench testing.

7. The Type of Ground Truth Used

The ground truth used for evaluating the LMT nomag IC 3.0's performance was based on:

  • Compliance with International Standards: Specifically, IEC 60601-2-20: 1996 and IEC 60601-1 (Ed.2): 1995.
  • Established Test Protocols: "established protocol using phantoms" for MR compatibility.
  • Comparison to a Legally Marketed Predicate Device: The LMT nomag IC 3.0 was evaluated against the legally marketed LMT nomag IC 1.5 device to demonstrate substantial equivalence, particularly for factors not affected by design changes and for image evaluation performed "to the same standards."

8. The Sample Size for the Training Set

The concept of a "training set" is not applicable to this device as it is a physical medical device and not a machine learning or AI algorithm.

9. How the Ground Truth for the Training Set was Established

As the concept of a training set is not applicable, the method of establishing its ground truth is also not relevant for this device.

{0}------------------------------------------------

Sec. 5 510(k) Summary LMT nomag IC 3.0

Image /page/0/Picture/1 description: The image shows the text "K062952 (p. 1 of 3)" written in black ink at the top. Below this text is a logo consisting of five black dots followed by the letters "LMT" in a bold, sans-serif font. The letters are also black, and the overall image has a high contrast.

JUL 3 1 2006

Preparation Date: Feb, 17th 2006

Applicant LMT Lammers Medical Technology GmbH Osterweide 8 23568 Luebeck Germany

Establishment Registration No. 3004083324

Contact Person:Thomas Michael BohnenTel: +49-451-58098-0Fax: +49-451-58098-15Email: bohnen@lammersmedical.com
-----------------------------------------------------------------------------------------------------------------------------

Common Device Name Neonatal Transport Incubator

Device Panel: General Hospital

Product Code FPL (Incubator, Neonatal Transport) Regulation No. 880.5410

Class

Propietary Name LMT nomag IC 3.0 REF: LMT-0100004-A

  • Performance Standard: IEC 60601-2-20: 1996 IEC 60601-1 (Ed.2): 1995 ISO 10993: 1997
    2

  • Legally marketed device LMT nomag IC 1.5 K#033565
    Reason for 510(k): extended indications for use

{1}------------------------------------------------

Sec. 5 510(k) Summary LMT nomag IC 3.0

Image /page/1/Picture/1 description: The image shows the text "K462452 (0.2043)" at the top. Below this text are four black dots in a horizontal row, followed by the letters "LMT" in a bold, sans-serif font. The letters are aligned horizontally and appear to be part of a logo or identifier.

Intended Use of the LMT nomag IC 3.0

The LMT nomag IC 3.0 is an infant incubator system for temporary use during MR-imaging in clinical environments and provided with a suitable trolley for intra hospital transport.

The LMT nomag IC 3.0 MR-incubator provides a controlled environment of warmth and humidity for premature babies and sick infants up to a body weight of 4.5 kg (10 lbs) and a maximum body length of 55 cm (21.7 inches).

The LMT nomag IC 3.0 is suitable for use with MR scanners of field strengths up to 3.0T.

Function of the Device

The LMT Nomag IC 3.0 is an incubator for preterm and term-born infants who are scheduled for Magnetic Resonance (MR) Imaging and are dependant on warming therapy. It provides a microclimate of air temperature and humidity suiting the infants needs. The incubator has a trolley for intra clinical transport providing power.

The incubator runs in air control mode and has closed loop control for both air temperature and humidity. Humidification works hygienically via vaporizing of distilled water.

Design and Specifications

The basic incubator performance of the LMT nomag IC 3.0 is defined by the IEC particular standard and the design changes over the nomag IC 1.5. do not affect these performance figures.

Testing under MR influence

The compatibility to MR scanners with 3T field strength has been proven in extensive bench testing.

lmage evaluation was performed to the same standards as with the nomag IC 1.5.

{2}------------------------------------------------

Sec. 5 510(k) Summary LMT nomaq IC 3.0

Image /page/2/Picture/1 description: The image shows the text "Kψ62φ52 (P.3ΩΔ3)" written in black ink at the top. Below this text is a row of three black dots followed by the letters "LMT" in a bold, sans-serif font. The letters are also black, and the overall image has a simple, clean appearance.

The compatibility with MR scanners of different manufacturers (Siemens, Philips, GE 1.5T and 3T) have been tested according to an established protocol using phantoms.

During this testing the behaviour of the incubator has shown no adverse effects in neither direction. The incubator performance was not influenced in any way by the scanner and the imaging has not shown significant artefacts caused by the incubator. Other functions of the scanner were also not influenced by the incubator.

SE Statement

The LMT Nomag IC 3.0 incubator is substantially equivalent to the legally marketed device. It is compliant with international harmonized standards for incubators by the IEC.

The testing did not rise new risks or questions associated with the use of a MR compatible incubator.

The LMT Nomag IC 3.0 works as safe and effective as the legally marketed device.

V.N.BoC

Thomas Michael Bohnen RA LMT Lammers Medical Technology GmbH, Germany

{3}------------------------------------------------

Image /page/3/Picture/2 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three lines forming its body and wings. The text "DEPARTMENT OF HEALTH AND HUMAN SERVICES, USA" is arranged in a circular pattern around the eagle.

JUL 3 1 2006

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

LMT Lammers Medical Technology GmbH C/O Mr. Jeffrey D. Rongero Responsible Third Party Official Underwriters Laboratories, Incorporated 12 Laboratory Drive Research Triangle, North Carolina 27709

Re: K062052

Trade/Device Name: LMT Nomag IC 3.0 Regulation Number: 21 CFR 880.5410 Regulation Name: Neonatal Transport Incubator Regulatory Class: II Product Code: FPL Dated: June 19, 2006 Received: July 20, 2006

Dear Mr. Rongero:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device. subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice. labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affècting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2 - Mr. Rongero

.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97), You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrlv/industrv/support/index.html.

Sincerely vours,

Chiu Lin, Ph.D.

Chiu Lin. Ph.D. Director Division of Anesthesiology. General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indications for Use

510(k) Number (if known):

______________________________________________________________________________________________________________________________________________________________________________ Device Name:

Indications for Use:

The LMT nomag IC 3.0 MR-incubator is an infant incubator system for temporary use during MR-imaging in clinical environments provided with a suitable trolley for intra hospital transport.

The LMT nomag IC 3.0 MR-incubator provides a controlled environment of warmth and humidity for premature babies and sick infants up to a body weight of 4.5 kg (10 lbs) and a maximum body length of 55 cm (21.7 inches).

The LMT nomag IC 3.0 is suitable for use with MR scanners of field strengths up to 3.0T.

Prescription Use × (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

His C. Murphy & Elizabeth M. Petton (Widow of John)
1789

sthesiblogy, General Hospi

Page 1 of 1

umber Krizdz

§ 880.5410 Neonatal transport incubator.

(a)
Identification. A neonatal transport incubator is a device consisting of a portable rigid boxlike enclosure with insulated walls in which an infant may be kept in a controlled environment while being transported for medical care. The device may include straps to secure the infant, a battery-operated heater, an AC-powered battery charger, a fan to circulate the warmed air, a container for water to add humidity, and provision for a portable oxygen bottle.(b)
Classification. Class II (performance standards).