K Number
K062027

Validate with FDA (Live)

Date Cleared
2007-01-19

(185 days)

Product Code
Regulation Number
870.1130
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The A&D Medical UA-789 digital blood pressure monitor is intended for used by adults with 12 years older to measure the systolic and diastolic blood pressure and pulse rate.
Indications For Use: Measure blood pressure (systolic and diastolic) and pulse rate.

Device Description

The A&D Medical UA-789 digital blood pressure monitor is intended for used by adults with 12 years older to measure the systolic and diastolic blood pressure and pulse rate. UA-789 uses an inflatable cuff which is wrapped around the patient's upper arm. The cuff is inflated automatically by an internal pump in the device. The systolic and diastolic blood pressures are determined by oscillometric method. The deflation rate is controlled by the internal electronic-controlled exhaust valve. There is a quick exhaust mechanism so that the pressure of the cuff can be completely released. The arm distribution range is from 9.4" to 23.6" (24 cm to 60 cm).

AI/ML Overview

Here's an analysis of the provided text regarding the A&D Medical UA-789 Digital Blood Pressure Monitor, incorporating the requested information:

Understanding the Device and its Validation

The A&D Medical UA-789 is a non-invasive digital blood pressure monitor that uses the oscillometric method to measure systolic and diastolic blood pressure and pulse rate. The validation of this device primarily relies on its adherence to established blood pressure monitor standards.


1. Table of Acceptance Criteria and Reported Device Performance

Acceptance CriteriaReported Device Performance
Accuracy and Effectiveness Standard: ANSI/AAMI SP-10 standardMet (specifically for extra-large cuffs used in UA-789)
European Medical Products Directive: European Directive 93/42 EECMet
General Safety Provisions: EN60601Met
Safety of BP Monitor (Particular Requirements): EN60601-2-30Met
Electromagnetic Compatibility: EN60601-1-2 and EN55011Met

2. Sample Size Used for the Test Set and the Data Provenance

The provided document does not explicitly state the sample size used for the test set or the data provenance (e.g., country of origin, retrospective/prospective). It only mentions that the accuracy and effectiveness of the extra-large cuffs have been "validated through the ANSI/AAMI SP-10 standard." The ANSI/AAMI SP-10 standard itself specifies requirements for clinical validation, including minimum sample sizes (typically a diverse group of subjects). However, the specific study details for the UA-789 are not present in this summary.


3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

The document does not specify the number of experts used or their qualifications for establishing ground truth for the clinical validation. The ANSI/AAMI SP-10 standard requires comparison against a reference method (e.g., auscultation by trained observers), but the details of who performed this for the UA-789 study are not provided in this summary.


4. Adjudication Method for the Test Set

The document does not specify the adjudication method used for the test set.


5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

This question is not applicable to the A&D Medical UA-789 Digital Blood Pressure Monitor. This device is a standalone measurement device for blood pressure and pulse rate, not an AI-assisted diagnostic tool that would involve multi-reader multi-case studies or assessment of human reader improvement.


6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

Yes, the testing described for the A&D Medical UA-789 is inherently a standalone performance evaluation. The device itself is designed to measure blood pressure and pulse rate without human intervention beyond proper application and operation. The validation against standards like ANSI/AAMI SP-10 assesses the accuracy of the device's measurements directly.


7. The Type of Ground Truth Used

Based on the reference to the ANSI/AAMI SP-10 standard, the ground truth would typically be established by a reference method, most commonly auscultation performed by trained observers (e.g., physicians or nurses) using a mercury sphygmomanometer or validated aneroid device, with readings taken according to a standardized protocol. The standard defines the acceptable difference between the device's readings and these reference measurements.


8. The Sample Size for the Training Set

The document does not provide information regarding a "training set" sample size. For traditional medical devices like blood pressure monitors, the concept of a distinct "training set" and "test set" in the machine learning sense is generally not directly applicable. The device's algorithm would have been developed and refined using internal methods and prior engineering knowledge, and then its performance validated against a clinical test set according to a standard.


9. How the Ground Truth for the Training Set Was Established

As noted in point 8, the concept of a "training set" with established ground truth as it applies to machine learning algorithms is likely not directly relevant here. The oscillometric algorithm for blood pressure measurement is based on established physiological principles and engineering. Its initial development and refinement would be based on comprehensive internal testing and optimization, rather than a distinct "training set" that requires external ground truth establishment in the same way an AI model does. The clinical validation then serves as the final proof of accuracy.

{0}------------------------------------------------

Attachment (D) 510(k) Summary

1. DATE PREPARED

July 14, 2006 (original). September 1, 2006 (revised).

JAN 1 9 2007

2. SPONSOR INFORMATION

A&D Engineering, Inc. Mr. Jerry Wang 1555 McCandless Drive, Milpitas, CA 95035 Tel: 408-518-5113 Fax: 408-635-2313 Email: jwang@aandmedical.com

3. DEVICE NAME

Proprietary Name:A&D Medical UA-789 Digital Blood Pressure Monitor
Common/Usual Name:Blood Pressure Monitor
Classification name:Non-invasive blood pressure measurement System21 CFR 870-1130, Class II, 74DXN.

DEVICE DESCRIPTION AND INTENDED USE 4.

The A&D Medical UA-789 digital blood pressure monitor is intended for used by adults with 12 years older to measure the systolic and diastolic blood pressure and pulse rate.

5. PREDICATE DEVICE

A&D LifeSource model UA-787 digital blood pressure monitor with FDA 510(k) K012472.

6. TECHNOLOGICAL CHARACTERISTICS

UA-789 uses an inflatable cuff which is wrapped around the patient's upper arm. The cuff is inflated automatically by an internal pump in the device. The systolic and

{1}------------------------------------------------

diastolic blood pressures are determined by oscillometric method. The deflation rate is controlled by the internal electronic-controlled exhaust valve. There is a quick exhaust mechanism so that the pressure of the cuff can be completely released. The arm distribution range is from 9.4" to 23.6" (24 cm to 60 cm).

The accuracy and effectiveness of the extra large cuffs used in UA-789 have been validated through the ANSI/AAMI SP-10 standard.

7. DEVICE TESTING

A&D Medical UA-789 digital blood pressure monitor meets the following standards:

  • ANSI/AAMI SP-10 standard .
  • . European Directive 93/42 EEC for Medical Products
  • EN60601 General Safety Provisions .
  • EN60601-2-30 Particular Requirements for the Safety of BP Monitor .
  • EN60601-1-2 and EN55011 Electromagnetic Compatibility .

{2}------------------------------------------------

Image /page/2/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three legs. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

A&D Engineering, Inc. c/o Mr. Jerry Wang Director of Engineering and QA 1555 McCandless Dr. Milpitas, CA 95035

JAN 1 9 2007

Re: K062027

Trade Name: A&D Medical LifeSource UA-789 Digital Blood Pressure Monitor Regulation Number: 21 CFR 870.1130 Regulation Name: Non-invasive blood pressure measurement system Regulatory Class: Class II (two) Product Code: DXN Dated: January 3, 2007 Received: January 4, 2007

Dear Mr. Wang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 - Mr. Wang

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Bram D Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

KO62027

Attachment B

Indications for Use

510(k) Number (if known):

Device Name: A&D Medical LifeSource UA-789 Digital Blood Pressure Monitor

Indications For Use:

Measure blood pressure (systolic and diastolic) and pulse rate.

Prescription Use (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use × (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Syn Off
Division of Cardiovascular Devices
510(k) Number K062027

Page 1 of ____________________________________________________________________________________________________________________________________________________________________

§ 870.1130 Noninvasive blood pressure measurement system.

(a)
Identification. A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.(b)
Classification. Class II (performance standards).