(21 days)
The Indomitable Magnetic Resonance Imaging System is indicated for use in producing images of multiple planes in the head and body. These images correspond to the distribution of hydrogen nuclei exhibiting nuclear magnetic resonance (NMR) and depend for their contrast upon NMR parameters [hydrogen nuclei concentration and flow velocity, T1 (spinlattice relaxation time) and T2 (spin-spin relaxation time)]. As a result of the acquisition and processing of the NMR data, these images display the internal structure of the head and body, and when interpreted by a trained physician, can vield diagnostically useful information.
FONAR Corporation has modified the existing control limits for the Indomitable scanner gradient subsystem to allow the maximum gradient current to increase by 50 percent, with a corresponding increase in the maximum slew rate for the gradient waveforms. All other subsystems remain essentially unchanged from their previously approved (K002490) state.
This modified gradient subsystem follows the same basic design principles as our previously approved gradient systems. It functions on the same basic operating and physical principles as the predicate device, the Indomitable gradient system (K002490). It is constructed with the same manufacturing materials, and is fabricated in the same fashion to the previously approved gradients.
The primary differences of this modification to the gradient subsystem are the software limits for maximum gradient current value and the increased slew rate that accompanies the changed current value. It is, in summary, a significant equivalent of FONAR's currently approved predicate device.
The provided text describes a special 510(k) submission for a device modification, specifically an increase in the operational limits for the gradient subsystem of an MRI scanner. The goal of the study was to demonstrate that this modification does not adversely affect patient safety and that the device remains substantially equivalent to its predicate device. This is primarily a safety validation rather than a diagnostic performance study in the typical sense of AI-powered diagnostic devices.
Here's an analysis of the acceptance criteria and the study that proves the device meets them:
Acceptance Criteria and Reported Device Performance
The acceptance criteria for this device modification are primarily focused on maintaining patient safety and demonstrating substantial equivalence to the predicate device, particularly regarding the absence of Peripheral Nerve Stimulation (PNS) at the new operational limits. The "reported device performance" here refers to the outcomes of the safety testing.
| Acceptance Criteria | Reported Device Performance |
|---|---|
| Operational Mode (Safety): Continue to operate in the NORMAL mode without exceeding safety limits for the public. | Testing shown that an increase of the maximum current from 200 Amps to 300 Amps would still allow the scanner to be operated in the NORMAL mode, since no PNS was detected in the test subjects. The maximum dB/dt is 20.27 mT/m with a corresponding maximum slew rate of 33.34 mT/m/ms. |
| Peripheral Nerve Stimulation (PNS): No PNS detected in test subjects at increased gradient current/slew rate. | No volunteer experienced any sensations associated with PNS during the clinical trial. |
| Maintenance of MRI Effectiveness Parameters | MRI effectiveness parameters such as spatial resolution, geometric distortion, specification volume, image uniformity, and slice spacing are unchanged by the modifications. |
| Substantial Equivalence | The modified gradient subsystem is found to be substantially equivalent to the previously approved gradient subsystem of the predicate device (K002490) under the current conditions for intended use, with no adverse effects to patient health or safety. |
| Compliance with Standards | All testing was conducted in accordance with current FDA guidance documents and international standards (specifically ISO standard IEC 60601-2-33 for clinical testing). |
| Specific Safety Parameters (Non-clinical Testing) | Max dB/dt (pulsing X, Y, and Z gradients) meets the new specified value of 20.27 mT/m. Other parameters like static field strength, acoustic noise, SAR, emergency conditions, and biocompatibility are either unaffected or comply with standards. |
Study Details
This submission focuses on a safety validation for a hardware/software modification, not a diagnostic AI algorithm. Therefore, many standard AI/diagnostic study metrics (like specific test set sizes for diagnostic accuracy, expert qualifications for ground truth in diagnostic tasks, MRMC studies, or training set details) are not directly applicable or reported in the same way.
-
Sample size used for the test set and the data provenance:
- Sample Size: This is a clinical trial involving "volunteers," implying a relatively small number of healthy individuals rather than a large patient cohort for diagnostic validation. The exact number of volunteers is not specified in the provided text, only that "All volunteers completed the test."
- Data Provenance: Prospective. The study was a "voluntary clinical trial" specifically conducted for this submission. The origin of the volunteers (e.g., country) is not specified but is implicitly within the context of FDA regulations in the US.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Ground Truth Establishment: For this safety study, the "ground truth" is the absence of PNS. This was established directly by the test subjects themselves reporting any "sensations associated with PNS." There isn't a need for multiple external experts to establish a "ground truth" in the diagnostic sense, as the criterion is subjective patient experience combined with adherence to international safety standards.
- Expert Qualifications: The study was implicitly supervised by medical professionals (as required for a clinical trial), but no specific number or qualifications of experts involved in establishing ground truth (i.e., PNS assessment beyond patient self-reporting) are detailed.
-
Adjudication method (e.g., 2+1, 3+1, none) for the test set:
- Adjudication Method: Not applicable in the context of this safety study. The primary "adjudication" for PNS was the direct report from the "volunteers" themselves according to the IEC 60601-2-33 standard. There is no mention of independent reviewers or a consensus process for assessing PNS beyond this.
-
If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- MRMC Study: No. This is not an AI-assisted diagnostic device; it's a modification to an MRI scanner's gradient system. Therefore, an MRMC study and effects on human reader performance are not applicable.
-
If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
- Standalone Performance: Not applicable. This device is an MRI scanner modification, not a standalone algorithm. The "algorithm" here refers to the system's operating software, which sets hardware limits. The performance evaluated is the physical system's safety, not an algorithm's diagnostic output.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- Ground Truth Type: Patient self-reporting of physiological sensation (absence of PNS) per the guidelines of international safety standards (IEC 60601-2-33). This is complemented by non-clinical technical measurements to confirm adherence to new physical operating parameters.
-
The sample size for the training set:
- Training Set Sample Size: Not applicable. This is not a machine learning or AI algorithm in the context of a "training set" for diagnostic performance.
-
How the ground truth for the training set was established:
- Training Set Ground Truth: Not applicable, as there is no training set mentioned for this type of device modification.
{0}------------------------------------------------
7/5/2006
Fonar Corporation 110 Marcus Drive Melville, New York 11747-4292
JUL 2 8 2006
June 30, 2006
Special 510(k) Summary - Device Modification
Submitter Information:
| Company | FONAR CorporationRegistration Number 2432211110 Marcus Drive, Melville, New York 11747-4292 |
|---|---|
| Contact: | Luciano Bonanni Executive Vice PresidentPhone: (631) 694-2929 Fax: (631) 753-5150E-mail: lbonanni@fonar.com |
Device Designation:
| Trade Name: | Indomitable Magnetic Resonance Imaging Scanner Accessory |
|---|---|
| Common Name: | Magnetic Resonance Imaging Scanner (MRI Scanner) |
| Classification: | System, Nuclear Magnetic Resonance Imaging (NMR/MRI) |
| Product Code: | LNH |
| Class: | 2 |
| Tier: | 2 |
| C.F.R. Section | 892.1000 |
| Classification Panel: | Radiology |
Reason for Submission: Change in operational limits for gradient subsystem.
FONAR Corporation has modified the existing control limits for the Indomitable scanner gradient subsystem to allow the maximum gradient current to increase by 50 percent, with a corresponding increase in the maximum slew rate for the gradient waveforms. All other subsystems remain essentially unchanged from their previously approved (K002490) state.
Device Modifications
This modified gradient subsystem follows the same basic design principles as our previously approved gradient systems. It functions on the same basic operating and physical principles as the predicate device, the Indomitable gradient system (K002490). It is constructed with the same manufacturing materials, and is fabricated in the same fashion to the previously approved gradients.
The primary differences of this modification to the gradient subsystem are the software limits for maximum gradient current value and the increased slew rate that accompanies the changed current value. It is, in summary, a significant equivalent of FONAR's currently approved predicate device.
{1}------------------------------------------------
7/5/2006
Description of Device
Gradient Subsystem
The gradients provide the ability of the system to accurately determine spatial location within the imaging volume. They are also used to help establish control of other imaging factors such as slice thickness and field-of-view. The gradients used on the Indomitable magnet, which are unchanged in this submission have the same overall diameter, and are still comprised of the same three quadrupolar (linear) gradients installed on all Fonar scanners.
One pair is placed in each of the orthogonal imaging planes (X, Y, and Z), developing a typical readout gradient of 1.2 mT/m. Note that while the typical gradient remains at 1.2 mT/m for this submission, the value of maximum gradient change has increased from a value of 12 mT/m to a new maximum of 20.27 mT/m. The gradient linearity remains unchanged at better than 2% over a 28 cm DSV with a stability of + 0.01%.
Gradient Current Amplifiers
The gradient current amplifier power required does not affect the actual gradient strength (which is still specified at a typical value of 1.2 mT/m with a typical amplifier power dissipation of 12 kW). The amplifiers for the Indomitable are still designed for accurate control of the current output to the gradient coils with very low distortion. and are rated for a maximum current output of 300 Amps. The gradient amplifiers are unchanged from previous submissions.
Gradient System Control
The operating characteristics for the hardware of the gradient system described above are controlled by the Indomitable's Sympulse scanner operating software. The software is programmed to operate the scanner system in a safe and efficient manner, and is substantially equivalent to the current operating software in the Indomitable system as approved in K002490. The software has always included "safety limits", which were established through laboratory testing and clinical imaging.
Prior to this submission, the maximum current value allowed by software for the gradient current amplifiers was 200 Amps. This setting effectively limited the maximum gradient amplitude (dB/dt) to 12 mT/m for the system. This rate of change produced a maximum slew rate of 19.73 mT/m/ms and allowed the Indomitable scanner to operate in the NORMAL mode. Testing performed for this submission has shown that an increase of the maximum current from the previous maximum of 200 Amps to the new maximum value of 300 Amps would still allow the scanner to be operated in the NORMAL mode, since no PNS was detected in the test subjects.
Therefore, the maximum gradient current for the gradient amplifiers will now be established by the Sympulse operating software at a value of 300 Amps. This will translate to a maximum dB/dt of 20.27 mT/m with a corresponding maximum slew rate of 33.34 mT/m/ms. The scanner will continue to operate in the NORMAL mode.
{2}------------------------------------------------
Magnet and Gradient Subsystem Comparisons
The tables of information presented below compare and summarize the common specifications for the Indomitable magnets and gradient subsystems as configured in the original submission (K002490) and in this revision. These tables show that the specifications compare favorably and therefore are substantially equivalent.
| MAGNET CONFIGURATIONS | FROM 510(K) K002490 | THIS SUBMISSION |
|---|---|---|
| Magnet Specification | Indomitable Electromagnet | Indomitable Electromagnet |
| Type | Iron-core Electromagnet | Iron-core Electromagnet |
| Field Strength | $0.6 T \pm 5%$ | $0.6 T \pm 5%$ |
| Operational Mode | Normal Operating Mode for both dB/dt and SAR | Normal Operating Mode for both dB/dt and SAR |
| Calculated SAR | 0.76 W/kg | 0.76 W/kg |
| Power Consumption | 98 KVA | 98 KVA |
| Magnet Coil Winding Material | Copper Bar with hollow core cooling | Copper Bar with hollow core cooling |
| Cooling | 30 ton closed loop liquid chiller | 30 ton closed loop liquid chiller |
| Stability (magnet drift) | long-term < 2 ppm/hr - short term < . 3 ppm/min | long-term < 2 ppm/hr - short term < . 3 ppm/min |
| Field Homogeneity | 7 ppm within 30 cm DSV - 1 ppm within 20 cm DSV | 7 ppm within 30 cm DSV - 1 ppm within 20 cm DSV |
| Shimming | passive only | passive only |
| Open Gap™ Configuration | 18" transverse clearance - 8' vertical axis | 18" transverse clearance - 8' vertical axis |
| Fringe Field(5 Gauss Line -from magnet center) | 9.77 ft vertically - 13 ft to either side9.3 ft toward front or rear | 9.77 ft vertically - 13 ft to either side9.3 ft toward front or rear |
| Pole Cap Eddy Current Compensation | steel pole with inlaid laminatesand self-shield gradients | steel pole with inlaid laminatesand self-shield gradients |
| Outer Dimensions (h x 1 x w) | 128" x 142" x 87" | 128" x 142" x 87" |
| Weight (w/ base) | 290,000 lbs. (Approx.) | 290,000 lbs. (Approx.) |
| GRADIENT SUBSYSTEM | FROM 510(K) K002490 | THIS SUBMISSION |
|---|---|---|
| Magnet Specification | Indomitable Electromagnet | Indomitable Electromagnet |
| Gradient Type | three orthogonal - quadrupolar (linear) X, Y, Z | three orthogonal - quadrupolar (linear) X, Y, Z |
| Gradient Strength | typical readout gradient 1.2 mT/m (12 mT/m max) | typical readout gradient 1.2 mT/m (20.27 mT/m max) |
| Typical Rise Time | typical 0.75 msec per 15 A (1.2 mT/m equivalent) | typical 0.75 msec per 15 A (1.2 mT/m equivalent) |
| Maximum Slew Rate | 19.73 mT/m/ms | 33.34 mT/m/ms |
| Stability | $\pm$ 0.01% | $\pm$ 0.01% |
| Heat Dissipation | 4 kW typical; 12 kW max (Air Cooled) | 4 kW typical; 12 kW max (Air Cooled) |
{3}------------------------------------------------
7/5/2006
Indications for Use
The Indomitable Magnetic Resonance Imaging System is indicated for use in producing images of multiple planes in the head and body. These images correspond to the distribution of hydrogen nuclei exhibiting nuclear magnetic resonance (NMR) and depend for their contrast upon NMR parameters [hydrogen nuclei concentration and flow velocity, T1 (spinlattice relaxation time) and T2 (spin-spin relaxation time)]. As a result of the acquisition and processing of the NMR data, these images display the internal structure of the head and body, and when interpreted by a trained physician, can vield diagnostically useful information.
WARNING: This device is limited by U.S. Federal law to investigational use for indications not in the indications statement.
Under the requirements of the law, the non-indicated applications can be used only under an Institutional Review Board approved protocol for a non-significant risk device or an Investigational Device Exemption application approved by the FDA for a significant risk device. The procedures to be followed, under the sponsorship of FONAR Corporation, are determined by the current guidelines established by the FDA, which should provide the IRB with sufficient guidance to determine the level of risk for a MRI device.
WARNING: U.S. Federal law restricts the sale, distribution and use of this device by or on the order of a physician.
Description of Safety and Substantial Equivalence
MRI effectiveness parameters such as spatial resolution, geometric distortion, specification volume, image uniformity, and slice spacing are unchanged by the modifications to the gradient subsystem. A series of clinical and non-clinical tests were performed to demonstrate the safety and effectiveness of the system, and to verify substantial equivalence to the predicate device. All testing was conducted in accordance with current FDA guidance documents and international standards. Results from all testing demonstrate substantial equivalence to the predicate device.
Non-Clinical Testing
The following specific parameters have been identified by the FDA as being pertinent to patient safety and thereby have been highlighted for inclusion in 510(k) submissions. The majority of these concerns are specifically tested by comparison to the ISO and NEMA standards, and where utilized are described individually.
- a. Static Field Strength: 0.6T ± 5%
- b. Peak and A-Weighted acoustic noise: 98.3 DBA ± 0.05
- c. Description of operational modes of the system: Normal Mode Only
- d. Maximum SAR for transmit coil: (Unaffected for this submission)
e. Maximum dB/dt (pulsing X, Y and Z gradients): 20.27 mT/m
f. Potential emergency conditions and shutdown: (Unaffected for this submission)
g. Biocompatibility of materials: (Unaffected for this submission)
{4}------------------------------------------------
Clinical Testing
A voluntary clinical trial was conducted under the testing guidelines described in the ISO standard IEC 60601-2-33, which describes the criteria for experimental studies of human subjects to directly determine limits related to minimizing PNS for any given type of gradient system. All volunteers participated in the study under informed consent. All volunteers completed the test. During the testing process, no volunteer experienced any sensations associated with PNS. This result meets the criteria for continued operation in the normal mode for the Indomitable scanner with the new maximum current limits.
Summary
The results of the testing, combined with the additional material within this submission has demonstrated that the modified gradient subsystem of the Indomitable magnet is substantially equivalent to the previously approved gradient subsystem of the predicate device (K002490) under the current conditions for intended use, with no adverse effects to patient health or safety reported.
{5}------------------------------------------------
Food and Drug Administration 9200 Corporate Blvd. Rockville MD 20850
JUL 2 8 2006
Mr. Luciano Bonanni Executive Vice President Fonar Corporation 110 Marcus Drive MELVILLE NY 11747-4292
Re: K061930
Trade/Device Name: Indomitable Magnetic Resonance Imaging Scanner Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: II Product Code: LNH Dated: July 5, 2006 Received: July 7, 2006
Dear Mr. Bonanni:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Image /page/5/Picture/9 description: The image is a black and white circular logo. The logo commemorates a centennial, with the years 1906-2006 displayed at the top. The letters "FDA" are prominently featured in the center of the logo. The word "Centennial" is written below the letters "FDA".
Protesting and Promoting Public Health
{6}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 21 CFR 876.xxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|---|---|---|
| 21 CFR 884.xxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 894.xxx | (Radiology) | 240-276-0120 |
| Other | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150
or at its Internet address http://www.fda.gov/cdrl/industry/support/index.html.
Sincerely yours,
Nancy C. Hogdon
Nancy C. Brogdon Director, Division of Reproductive. Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
STATEMENT OF INDICATIONS FOR USE
Applicant: FONAR CORPORATION 510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________ Device Name: Indomitable Magnetic Resonance Imaging Scanner
Indications For Use:
The Indomitable Magnetic Resonance Imaging System is indicated for use in producing images of multiple planes in the head and body. These images correspond to the distribution of hydrogen nuclei exhibiting nuclear magnetic resonance (NMR) and depend for their contrast upon NMR parameters [hydrogen nuclei concentration and flow velocity, T1 (spin-lattice relaxation time) and T2 (spin-spin relaxation time)]. As a result of the acquisition and processing of the NMR data, these images display the internal structure of the head and body, and when interpreted by a trained physician, can yield diagnostically useful information.
WARNING: This device is limited by U.S. Federal law to investigational use for indications not in the indications statement.
Under the requirements of the law, the non-indicated applications can be used only under an Institutional Review Board approved protocol for a non-significant risk device or an Investigational Device Exemption application approved by the FDA for a significant risk device. The procedures to be followed, under the sponsorship of FONAR Corporation, are determined by the current guidelines established by the FDA, which should provide the IRB with sufficient guidance to determine the level of risk for a MRI device.
WARNING: U.S. Federal law restricts the sale, distribution and use of this device by or on the order of a physician.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER LINE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Per 21 CFR 801.109)
(Optional Format 1-2-96)
Nancy Broughton
Division Sign-C Division of Reproductive, and Radiological Devic 510(k) Number
N/A