K Number
K061748

Validate with FDA (Live)

Date Cleared
2006-07-17

(26 days)

Product Code
Regulation Number
888.3030
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The proposed Mini Fragment Plates are intended for essentially non load bearing stabilization and fixation of small bone fragments in fresh fractures, revision procedures, joint fusion, and reconstructions of small bones of the hand, foot, wrist, ankle, humerus, scapula, finger, toe, pelvis, and craniomax, illofacial skeleton.

Device Description

The proposed Mini Fragment Plates are a set of orthopedic plates and fasteners supplied in a sterilization tray together with several reusable and disposable tools. The set of orthopedic plates and fasteners provided in the sterilization tray consists of the following implantable devices: Fragment Plates that are provided in the following design configurations: Straight, Y, Right Hockey Stick, and Left Hockey Stick. Pegs; Screws; and K-Wires. Included in the sterilization tray are the following re-useable instruments: Peg drivers, Other standard surgical tools. In addition, the following are non-reusable instruments included in the sterilization tray: F.A.S.T. Guide Technology™ Drill Guide, Drill bits.

AI/ML Overview

The provided text is for a 510(k) premarket notification for "Mini Fragment Plates", a type of orthopedic implant. It focuses on demonstrating substantial equivalence to a predicate device rather than presenting a performance study with acceptance criteria in the context of device performance metrics like sensitivity, specificity, or reader studies.

Therefore, the requested information, such as acceptance criteria, reported device performance, sample sizes for test and training sets, expert qualifications, adjudication methods, MRMC studies, standalone performance, and ground truth types and establishment, are not present in the provided document.

Instead, the document details:

  • Device Name: Mini Fragment Plates
  • Manufacturer: Hand Innovations, LLC
  • Predicate Device: Fragment Plate System (K041081)
  • Classification: Class II
  • Indications for Use: Stabilization and fixation of small bone fragments in fresh fractures, revision procedures, joint fusion, and reconstructions of small bones of the hand, foot, wrist, ankle, humerus, scapula, finger, toe, pelvis, and craniomaxillofacial skeleton.
  • Device Description: Orthopedic plates and fasteners (Straight, Y, Right Hockey Stick, Left Hockey Stick), pegs, screws, K-wires, reusable instruments (peg drivers, standard surgical tools), and non-reusable instruments (F.A.S.T. Guide Technology™ Drill Guide, drill bits).
  • Biocompatibility: Not required as the material (Ti-6Al-4V ELI per ASTM F 136-00) is the same as the previously approved predicate device and has a long history of safe use.
  • Summary of Substantial Equivalence: The proposed Mini Fragment Plates have the same indications for use, intended use, basic shape/design, operating principle, and materials as the predicate device. The changes do not alter the fundamental scientific technology.

This document is a regulatory submission focused on confirming that the new device is "substantially equivalent" to an already approved device, meaning it doesn't raise new questions of safety and effectiveness. It does not involve performance testing against specific clinical or system-level acceptance criteria in the way a novel diagnostic or AI-driven device would.

{0}------------------------------------------------

K061748 1 of 2 00

JUL 1 7 2006

510(k) Summary of Safety and Effectiveness

Name, AddressandEstablishmentRegistrationNo.The address and registration number of the manufacturer is as follows:
Hand Innovations, LLC8905 SW 87 Avenue, Suite 220Miami, FL 33176-2227
Establishment Registration No.: 9042874
Tel.: (305) 270-6899
Fax: (305) 412-8060
GeneralProvisionsThe name of the device is:
Proprietary NameCommon or Usual Name
Mini Fragment PlatesPlate Fixation Bone
Name ofPredicateDevicesThe device is substantially equivalent to:• Fragment Plate System (510(k) # K041081 - July 1, 2003) - Hand Innovations,LLC.
ClassificationClass II.
PerformanceStandardsPerformance standards have not been established by the FDA under section 514 ofthe Food, Drug and Cosmetic Act.
Indications forUseThe proposed Mini Fragment Plates are intended for essentially non load bearingstabilization and fixation of small bone fragments in fresh fractures, revisionprocedures, joint fusion, and reconstructions of small bones of the hand, foot, wrist,ankle, humerus, scapula, finger, toe, pelvis, and craniomax, illofacial skeleton.

:

:

{1}------------------------------------------------

K061748 (pg 2 of 2)

DeviceDescriptionThe proposed Mini Fragment Plates are a set of orthopedic plates and fastenerssupplied in a sterilization tray together with several reusable and disposable tools.The set of orthopedic plates and fasteners provided in the sterilization tray consists ofthe following implantable devices: Fragment Plates that are provided in the following design configurations: Straight, Y, Right Hockey Stick, and Left Hockey Stick. Pegs; Screws; and K-Wires.
Included in the sterilization tray are the following re-useable instruments: Peg drivers, Other standard surgical tools.
In addition, the following are non-reusable instruments included in the sterilizationtray: F.A.S.T. Guide Technology™ Drill Guide, Drill bits.
Biocompati-bilityThe Proposed Mini Fragment Plates do not require biocompatibility testing becausethe material is the same as the previously approved Fragment Plate System. Thematerial is Ti-6Al-4V ELI per ASTM F 136-00 and also has a long history of safeuse in the Orthopedic industry.
Summary ofSubstantialEquivalenceThe proposed Mini Fragment Plates has the following similarities to the previouslycleared Fragment Plate System : (K041081). Have the same indications for use and intended use Have the same basic shape/design Use the same operating principle Utilize the same materials
In summary, the changes incorporated in the proposed Mini Fragment Plates do notalter the fundamental scientific technology of the predicate device and therefore, inour opinion, is substantially equivalent to the predicate Fragment Plate System.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the seal of the Department of Health and Human Services (HHS) of the United States. The seal features a stylized eagle with three lines representing its wings and body. The text "DEPARTMENT OF HEALTH AND HUMAN SERVICES. USA" is arranged in a circular pattern around the eagle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Hand Innovations, LLC % DePuv Orthopaedics, Inc. Ms. Natalie S. Heck Manager of Regulatory Affairs 700 Orthopaedic Drive PO Box 988 Warsaw, Indiana 46581

JUL 1 7 2006

Re: K061748

Trade/Device Name: Mini Fragment Plates Regulation Number: 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulation Class: II Product Code: KTT Dated: June 19, 2006 Received: June 21, 2006

Dear Ms Heck:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality

{3}------------------------------------------------

Page 2 – Ms. Natalie S. Heck

systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120 . Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Barbara Buchner

Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

K061748

Indications for Use

510(k) Number (if known): Device Name: Mini Fragment Plates

Indications for Use:

The Fragment Plate System is intended for essentially non load bearing stabilization and fixation of small bone fragments in fresh fractures, revision procedures, joint fusion, and reconstructions of small bones of the hand, foot, wrist, ankle, humerus, scapula, finger, toe, pelvis, and craniomax, illofacial skeleton.

Over-The-Counter Use Prescription Use X AND/OR (21 CFR 807 Subpart C) (Part 21 CFR 801 Subpart D) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Larbere Buck MD for Myra
Division Sign Off

(Posted Nove Storas i and Neurological Devices

Page of of

510(k) Number K061746

Mini Fragment Plate System Special 510(k)

0000026

§ 888.3030 Single/multiple component metallic bone fixation appliances and accessories.

(a)
Identification. Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.(b)
Classification. Class II.