K Number
K061675

Validate with FDA (Live)

Manufacturer
Date Cleared
2006-08-02

(49 days)

Product Code
Regulation Number
870.1025
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Vision V x C series 4300 central station monitoring system is intended to be used to provide, using a wireless LAN for communication, centralized surveillance and documentation of patient vital sign data and arrhythmia/ST monitoring for a variable number of Escort Bedside Monitors and a variable number of UHF telemetry transmitters in the hospital environment. It is intended for use by healthcare practitioners trained in the use of the equipment only.

The ST algorithm has been tested for accuracy of the ST segment measurement data. The significance of the ST segment changes must be determined by a physician.

Device Description

The Vision VxC Central Station is designed specifically to provide centralized display for up to 16 patients, storage and recording (or printing) of patient vital sign and waveform data that are being monitored at the bedside by Invivo monitors and telemetry devices.

The Vision VxC Central Station can provide alarm detection and reporting for all vital sign parameters available to the central station for patient alarm surveillance. This alarm surveillance includes alarms reported by the bedside monitors and repeated to the central station as well as primary alarm surveillance for the patient worn WMTS telemetry transmitter device where there is no alarm notification capability on the transmitter worn by the patient.

Arrhythmia monitoring and ST segment detection capability is available as an option. The arrhythmia feature is equipped with password protection to prevent unauthorized users from turning off arrhythmia when a patient is being monitored. When a patient is monitored by arrhythmia the system will provide continuous monitoring of lifethreatening alarms.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the Vision VxC Central Station, based on the provided 510(k) summary:

1. Table of Acceptance Criteria and Reported Device Performance

ParameterAcceptance Criteria (Requirement)Reported Device Performance (Results)
Central Station
Display TypeFlatpanel 19" LCDPass
Central ProcessorIntel PentiumPass
User InterfaceTouchscreen and/or mousePass
Operating SystemMicrosoft Windows XP EmbeddedPass
Number of Patients Monitored1 to 16 (up to 16 designated for telemetry)Pass
Parameters MonitoredECG, Resp, IBP (SYS, DIA, MEAN), NIBP (SYS, DIAS, MEAN), SpO2, ETCO2, TempPass
Max Parameters Monitored20Pass
TrendingTabular for 16 patients, all parameters; up to 72 hours at 1, 2, 3, 4, 5, 15, 30, 60, 120, and 180 min. intervalsPass
Alarm History Storage1000 events within 72 hours per patient; 20 sec per eventPass
DocumentationThermal array recorder and/or laser printerPass
AC Main90-130/180-260 VAC, 47-63 Hz selectable, 6 amps@115VPass
Power Supply235 WPass
Operating Temperature10 to 40° CPass
Storage Temperature-40 to 75° CPass
Relative Humidity5 to 95 %Pass
StandardsUL 60950Pass
Telemetry Receiver Platform
Alarm History Storage1000 events within 72 hours per patient; 20 sec per eventPass
DocumentationThermal array recorder and/or laser printerPass
AC Main115/230 VAC, 60/50 Hz selectable, 4 amps@115VPass
Power Supply235 WPass
Input Voltage100-120 VAC / 200 - 240 VAC, selectable, 50/60 HzPass
Input Current6 Amp max @ 115V (20A Max inrush cold start), 3 Amp max @ 230 V (10A Max inrush cold start)Pass
Telemetry BandFCC WMTS (608-614 MHz)Pass
Operating Temperature10 to 40° CPass
Relative Humidity10 to 90 %Pass
StandardsUL 60950, FCC Part 15 (Spread Spectrum)Pass
Arrhythmia Analysis Option
Number of Arrhythmia Channels1 to 16 (dual vector)Pass
Types of Detected EventsAsystole, VFIB, VTACH, Couplet, High and Low Heart Rate, High Abnormal Count, Bigeminy, Trigeminy, V.RUN, V.Rhythm, Multi-Focal, R-ON-T, PausePass
Type of AlgorithmHeuristic algorithm using template matching and feature extractionPass
QRS Detection SensitivityAHA ≈99.88%, MIT ≈99.93%Pass
QRS Detection Positive PredictivityAHA ≈99.89%, MIT ≈99.85%Pass
PVC Detection SensitivityAHA ≈94.07%, MIT ≈95.44%Pass
PVC Detection Positive PredictivityAHA ≈97.72%, MIT ≈96.60%Pass
PVC Detection False Positive RateAHA ≈0.22%, MIT ≈0.23%Pass
Alarm Displaysarrhythmia alarm must displayPass
Alarm Tonesarrhythmia alarm must soundPass
Alarm Recordingsarrhythmia alarm must generate an alarm recording if configuredPass
Alarm Event Historyarrhythmia alarm must generate event history recordPass
Maximum Patient Load16 patientPass
StandardAAMI/ANSI EC 57: 1998Compliant
ST Analysis Option
Number of ST Channels1 to 16, but no more than arrhythmia channelsPass
Alarmshigh and low for both vectors; can be recorded and/or stored as eventsPass
StandardAAMI/ANSI EC 57: 1998Compliant

2. Sample Size Used for the Test Set and Data Provenance

The document states, "This device was validated using patient simulators under simulated use conditions." This indicates that the test set did not consist of real patient data. The provenance for this simulated data is not specified (e.g., country of origin). The testing would be prospective in the sense that the device was evaluated against controlled simulated conditions, but not retrospective using historical clinical patient data.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

Given that "patient simulators" were used, there were likely no human experts required to establish "ground truth" in the traditional sense of clinical diagnosis. The "ground truth" would have been the pre-programmed and known states of the patient simulator, which were then compared against the device's measurements and detections. The qualifications of the individuals who programmed and operated the simulators are not specified.

4. Adjudication Method for the Test Set

No adjudication method is mentioned for the test set, as the evaluation was against known simulator outputs rather than human interpretation that would require adjudication.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

No, an MRMC comparative effectiveness study was not done. The performance data focuses on the device's technical specifications and algorithm accuracy (for arrhythmia and ST analysis) against predefined standards and simulated conditions, not on comparing human reader performance with or without AI assistance.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

Yes, a standalone performance evaluation was done, particularly for the arrhythmia and ST analysis algorithms. The document explicitly lists performance metrics for QRS and PVC detection (Sensitivity, Positive Predictivity, False Positive Rate) based on the AHA and MIT databases. This demonstrates the algorithm's performance independent of human interaction.

7. The Type of Ground Truth Used

For the arrhythmia analysis, the ground truth was based on established benchmarks from the AHA (American Heart Association) and MIT (Massachusetts Institute of Technology) databases. These databases contain annotated ECG recordings where events (like QRS complexes and PVCs) have been meticulously identified and marked, often through expert review and consensus. For other general device functions (display, processing, etc.), the ground truth would be conformance to the stated technical specifications and environmental standards.

8. The Sample Size for the Training Set

The document does not explicitly state the sample size for the training set used for the arrhythmia and ST analysis algorithms. It mentions the "current arrhythmia detection and ST segment analysis algorithm library was replaced with the Mortara Instrument Incorporated product." It further states that "This same library is used in the Welch Allyn Acuity Central Station, which was cleared to market via 510(k) K022453." While the training set data is not provided, the implication is that the Mortara algorithm was trained on a significant dataset to achieve the reported performance on standard benchmarks.

9. How the Ground Truth for the Training Set Was Established

The document does not directly describe how the ground truth for the training set of the Mortara algorithms was established. However, given that these algorithms are evaluated against AHA and MIT databases for their "ground truth" performance metrics (as seen in the "Arrhythmia Analysis Option" table), it can be inferred that the training process would have also relied on highly curated and expert-annotated ECG datasets similar to, or including, segments of these recognized benchmark databases. The "heuristic algorithm using template matching and feature extraction" also points to an approach that would benefit from large, labeled datasets for developing and refining performance.

{0}------------------------------------------------

510(k) SUMMARY OF SAFETY AND EFFECTIVENESS

SUBMITTER INFORMATION

AUG 0 2 2006

A.Company Name:Invivo Corporation
B.Company Address:12601 Research ParkwayOrlando, FL 32826
C.Company Phone:(407) 275-3220
Company Fax:(407) 206-9658
D.Contact Person:Rusty KellyQuality Control ManagerInvivo Corporation
E.Date Summary Prepared:June 12, 2006

DEVICE IDENTIFICATION

A.Generic Device Name:Central Station
B.Trade/Proprietary Name:Vision VxC Central Station
C.Classification:Class II
D.Product Code:DSI "detector and alarm, arrhythmia"

DEVICE DESCRIPTION

The Vision VxC Central Station is designed specifically to provide centralized display for up to 16 patients, storage and recording (or printing) of patient vital sign and waveform data that are being monitored at the bedside by Invivo monitors and telemetry devices.

The Vision VxC Central Station can provide alarm detection and reporting for all vital sign parameters available to the central station for patient alarm surveillance. This alarm surveillance includes alarms reported by the bedside monitors and repeated to the central station as well as primary alarm surveillance for the patient worn WMTS telemetry transmitter device where there is no alarm notification capability on the transmitter worn by the patient.

Arrhythmia monitoring and ST segment detection capability is available as an option. The arrhythmia feature is equipped with password protection to prevent unauthorized users from turning off arrhythmia when a patient is being monitored. When a patient is monitored by arrhythmia the system will provide continuous monitoring of lifethreatening alarms.

{1}------------------------------------------------

SUBSTANTIAL EQUIVALENCE

The Vision VxC Central Station is of comparable type and is substantially equivalent to the following predicate devices:

Predicate DeviceManufacturer510(k) No.Date Cleared
Escort-Link Central StationMonitor Model 20500Medical Data ElectronicsK98210411/25/1998
Acuity Central MonitoringSystemWelch Allyn Protocol,IncorporatedK02245308/21/2002

INTENDED USE

The Vision VxC Series 4300 Central Station Monitor is intended to be used to provide. using a wireless LAN for communication, centralized surveillance and documentation of patient vital sign data and arrhythmia/ST monitoring for a variable number of Escort Bedside Monitors and a variable number of UHF telemetry transmitters in the hospital environment. It is intended for use by healthcare practitioners trained in the use of the equipment only.

The ST algorithm has been tested for accuracy of the ST segment measurement data. The significance of the ST segment changes must be determined by a physician

COMPARISON TO PREDICATE DEVICE:

The current version of the Vision Central Station was cleared to market under 510/k) K982104 under the name "Escort-Link Central Station Monitor Model 20500". The Vision Central Station has been modified as follows:

  • The current arrhythmia detection and ST segment analysis algorithm library was . replaced with the Mortara Instrument Incorporated product. This same library is used in the Welch Allyn Acuity Central Station, which was cleared to market via 510(k) K022453.
  • The current Vision VxC platform was updated with new hardware and electronics to . prevent product obsolescence.
  • The current operating system was updated to Windows XP Embedded. .
  • The Link Auxiliary Base has previously been replaced with a Telemetry Receiver . Platform which incorporates a WMTS receiver in a more compact enclosure (reference 510(k) K012336).

The modified Vision Central Station will be marketed as the Vision VxC Central Station.

TECHNOLOGICAL CHARACTERISTICS

A comparison of the technological characteristics of the Vision Vx C Central Station and the predicate devices has been performed. The results of this comparison demonstrate that the Vision VxC Central Station is equivalent to the marketed predicate devices in technological characteristics.

{2}------------------------------------------------

ENVIRONMENTAL AND NON-CLINICAL TESTING:

Applicable environmental and non-clinical testing was performed per EN IEC 60950. AANSVAAMI EC 57; 1998, EN 55022 and EN 55024. The Vision VxC Central Station passed all tests.

PERFORMANCE DATA

The performance data included in this submission to compare equivalency of the Vision VxC Central Station with the Escort Link Central Station and the Acuity Central Monitoring System confirms that the performance requirements for accuracy and precision were met and indicates substantial equivalence to the predicate devices. Equivalent performance in meeting user requirements was also determined.

Summary of Performance Testing:

This device was validated using patient simulators under simulated use conditions. The functional requirements and user needs were verified to have been met.

PARAMETER:REQUIREMENTRESULTS
Display Type:Flatpanel 19" LCDPass
Central Processor:Intel PentiumPass
User Interface:Touchscreen and/or mousePass
Operating System:Microsoft Windows XP EmbeddedPass
Number of Patients Monitored:I to 16 (up to 16 designated for telemetry)Pass
Parameters Monitored:ECG, Resp, IBP (SYS, DIA, MEAN), NIBP (SYS,DIAS, MEAN), SpO2, ETCO2, TempPass
Max Parameters Monitored :20Pass
Trending:Tabular for 16 patients, all parameters; up to 72hours at 1, 2, 3, 4, 5, 15, 30, 60, 120, and 180 min.intervalsPass
Alarm History Storage:1000 events within 72 hours per patient; 20 sec pereventPass
Documentation:Thermal array recorder and/or laser printerPass
AC Main:90-130/180-260 VAC, 47-63 Hz selectable, 6amps@115VPass
Power Supply:235 WPass
Operating Temperature:10 to 40° CPass
Storage Temperature:-40 to 75° CPass
Relative Humidity:5 to 95 %Pass
StandardsUL 60950Pass

CENTRAL STATION

{3}------------------------------------------------

TELEMETRY RECEIVER PLATFORM

PARAMETER:REQUIREMENTRESULTS
Alarm History Storage:1000 events within 72 hours per patient; 20 sec perPass
event
Documentation:Thermal array recorder and/or laser printerPass
AC Main:115/230 VAC, 60/50 Hz selectable, 4 amps@115VPass
Power Supply:235 WPass
Input Voltage:100-120 VAC / 200 - 240 VAC, selectable, 50/60 HzPass
Input Current:6 Amp max @ 115V (20A Max inrush cold start),Pass
3 Amp max @ 230 V (10A Max inrush cold start)
Telemetry Band:FCC WMTS (608-614 MHz)Pass
Operating Temperature:10 to 40° CPass
Relative Humidity:10 to 90 %Pass
StandardsUL 60950, FCC Part 15 (Spread Spectrum)Pass

ARRHYTHMIA ANALYSIS OPTION

PARAMETER:REQUIREMENTRESULTS
Number of ArrhythmiaChannels:1 to 16 (dual vector)Pass
Types of Detected Events:Asystole, VFIB, VTACH, Couplet, High and LowHeart Rate, High Abnormal Count, Bigeminy,Trigeminy, V.RUN, V.Rhythm, Multi-Focal, R-ON-T, PausePass
Type of Algorithm:Heuristic algorithm using template matching andfeature extractionPass
QRS Detection SensitivityAHA ≈99.88%, MIT ≈99.93%Pass
QRS Detection Positive PredictivityAHA ≈99.89%, MIT ≈99.85%Pass
PVC Detection SensitivityAHA ≈94.07%, MIT ≈95.44%Pass
PVC Detection Positive PredictivityAHA ≈97.72%, MIT ≈96.60%Pass
PVC Detection False Positive RateAHA ≈0.22%, MIT ≈0.23%Pass
Alarm Displaysarrhythmia alarm must displayPass
Alarm Tonesarrhythmia alarm must soundPass
Alarm Recordingsarrhythmia alarm must generate an alarm recording ifconfiguredPass
Alarm Event Historyarrhythmia alarm must generate event history recordPass
Maximum Patient Load16 patientPass
StandardAAMI/ANSI EC 57: 1998Compliant

ST ANALYSIS OPTION

PARAMETER:REQUIREMENTRESULTS
Number of ST Channels:1 to 16, but no more than arrhythmia channelsPass
Alarms:high and low for both vectors; can be recordedand/or stored as eventsPass
StandardAAMI/ANSI EC 57: 1998Compliant

Conclusion:

The verification and validation activities for the modified Vision Central Station confirm that all identified risks have been mitigated and that this device operates as designed and intended. The test results demonstrate the modified Vision Central Station is substantially equivalent to the predicate device cleared to market via 510(k) K982104 and the other predicate devices identified in this submittal.

K061675 +4/4

{4}------------------------------------------------

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

AUG 11 2 2006

Invivo Corporation c/o Ms. Maria Keelan Regulatory Affairs Specialist 12601 Research Parkway Orlando, FL 32826

Re: K061675 Vision VxC Central Station Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia detector and alarm Regulatory Class: Class II (two) Product Code: MHX Dated: July 7. 2006 Received: July 10, 2006

Dear Ms. Keelan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments. or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register.

{5}------------------------------------------------

Page 2 -- Ms. Maria Keelan

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours.

Blymmerman for

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

Indications for Use

510 (k) Number (if known):

Device Name: Model 4300 Vision VxC Central Station

Indications for Use:

The Vision V x C series 4300 central station monitoring system is intended to be used to provide, using a wireless LAN for communication, centralized surveillance and documentation of patient vital sign data and arrhythmia/ST monitoring for a variable number of Escort Bedside Monitors and a variable number of UHF telemetry transmitters in the hospital environment. It is intended for use by healthcare practitioners trained in the use of the equipment only.

The ST algorithm has been tested for accuracy of the ST segment measurement data. The significance of the ST segment changes must be determined by a physician.

Bhumima

lon Sign-Off Division of Cardiovascular Devices 510(k) Number k

Prescription Use

(Part CFR 801 Subpart D)

Over-The Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

AND/OR

Concurrence of CDRH, Office of Device Evaluation (ODE)

§ 870.1025 Arrhythmia detector and alarm (including ST-segment measurement and alarm).

(a)
Identification. The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.(b)
Classification. Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.