K Number
K061666

Validate with FDA (Live)

Date Cleared
2006-08-24

(71 days)

Product Code
Regulation Number
876.1500
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Cellvizio® (-GI, -LUNG) with Confocal Miniprobe™ (Coloflex, Gastroflex, Alveoflex) is a confocal laser imaging system with fiber optic probes that is intended to allow confocal laser imaging of the internal microstructure of tissues in the anatomical tract, i.e., gastrointestinal or pulmonary, accessed by the endoscope.

Device Description

The Cellvizio® (-GI, -LUNG) with Confocal Miniprobe™ (Coloflex, Gastroflex, Alveoflex) is a confocal microscope with a fiber optic probe which allows in vivo visual inspection of tissues with a microscopic resolution during an endoscopic procedure. The F-600 system has been designed to allow real-time observations of tissues. The device is based on a common laser scanning technology adapted for imaging through a bundle of optical fibers and is thus composed of several elements: a Laser Scanning Unit, proprietary software running on a remote computer, a flat panel display and Miniaturized Fiber Optic Probes. The F-600 system can be used with any legally marketed endoscope with a working channel of 2.8 mm or greater.

AI/ML Overview

The provided text describes the Mauna Kea Technologies F-600 System, a special 510(k) submission for the Cellvizio® (-GI, -LUNG) with Confocal Miniprobe™ (Coloflex, Gastroflex, Alveoflex). However, it does not contain details about acceptance criteria or a study proving the device meets specific performance criteria.

The document is a Premarket Notification 510(k) Summary, which primarily focuses on:

  • Submitter information
  • Device name and classification
  • Predicate devices
  • Device description
  • Intended use
  • Comparison to predicate devices (specifically comparing it to their previously marketed F-400 system, which itself was based on predicate devices)
  • FDA's substantial equivalence determination letter.

The purpose of a 510(k) submission is to demonstrate that the new device is "substantially equivalent" to a legally marketed predicate device. This often involves comparing technological characteristics and intended use, rather than conducting new performance studies that establish acceptance criteria and device performance in the same way a PMA (Premarket Approval) might require.

Therefore, the requested information regarding acceptance criteria and a study proving the device meets those criteria cannot be extracted from this document. The FDA's letter states: "We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices..." This indicates that the regulatory decision was based on substantial equivalence, not on specific performance data against predefined acceptance criteria for the new device.

{0}------------------------------------------------

Ko61664

Mauna Kea Technologies Special 510(k): F-600 System

Premarket Notification Special 510(k) Summary 10.0

In accordance with the Safe Medical Devices Act of 1990 and 21 CFR 807.92, a Premarket Notification 510(k) Summary for the F-600 system Premarket Notification is provided below

10.1 Submitter Information

Company Name and Address:

Mauna Kea Technologies 9, rue d'Enghien 75010 Paris France

Contact Name:

Fouad Tarabah, PharmD. Director of Quality and Regulatory Affairs Mauna Kea Technologies Telephone: +33 1 70 08 09 61 Fax: +33 1 48 24 12 18 e-mail: fouad@maunakeatech.com

Date Prepared: May 2006

10.2 Name of Device

Proprietary Name: Cellvizio® (-GI, -LUNG) with Confocal Miniprobe™ (Coloflex, Gastroflex, Alveoflex)

Classification Name: Endoscope and/or Accessories

10.3 Predicate Device(s) Information

Pentax Confocal Laser System, K042740

Pentax EC-3870CILK Confocal Video Colonoscope, K042741

superDimension Bronchus, K042438

10.4 Device Description

The Cellvizio® (-GI, -LUNG) with Confocal Miniprobe™ (Coloflex, Gastroflex, Alveoflex) is a confocal microscope with a fiber optic probe which allows in vivo visual inspection of tissues with a microscopic resolution during an endoscopic procedure. The F-600 system has been designed to allow real-

{1}------------------------------------------------

61666

Mauna Kea Technologies Special 510(k): F-600 System

time observations of tissues. The device is based on a common laser scanning technology adapted for imaging through a bundle of optical fibers and is thus composed of several elements: a Laser Scanning Unit, proprietary software running on a remote computer, a flat panel display and Miniaturized Fiber Optic Probes. The F-600 system can be used with any legally marketed endoscope with a working channel of 2.8 mm or greater.

10.5 Intended Use

The Cellvizio® (-GI, -LUNG) with Confocal Miniprobe™ (Coloflex, Gastroflex, Alveoflex) is a confocal faser system that is intended to allow confocal laser imaging of the internal microstructure of tissues in the anatomical tract, i.e., gastrointestinal or respiratory, accessed by the endoscope.

10.6 Comparison to Predicate Device(s)

The comparison to the predicate devices was based on a review of the F-400 system information included in the company's 510(k) Premarket Notification K051585 and information concerning the predicate devices that was available in the public domain. The F-400 system is the Mauna Kea Technologie's legally marketed device (K051585) which serves as the unmodified device for the Cellvizio® (-GI, -LUNG) with Confocal Miniprobe™ (Coloflex, Gastroflex, Alveoflex) special 510(k) submission.

{2}------------------------------------------------

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

AUG 2 4 2006

Mauna Kea Technologies % Fouad Tarabah Director of Quality and Regulatory Affairs 9, Rue D'Enghien Paris, France 75010

Re: K061666

Trade/Device Name: Cellvizio® (-GI, -LUNG) with Confocal Miniprobe™ (Coloflex, Gastroflex, Alveoflex) Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: GCJ Dated: June 20, 2006 Received: June 25, 2006

Dear Fouad Tarabah:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product

{3}------------------------------------------------

Page 2 -- Fouad Tarabah

radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely vours.

Vabare Preelup

Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Mauna Kea Technologies Special 510(k): F-600 System

ODE Indications Statement

Special 510(k) Number (if known): K061666

Device Name: Cellvizio® (-GI, -LUNG) with Confocal Miniprobe™ (Coloflex, Gastroflex, Alveoflex)

Indications for Use:

The Cellvizio® (-GI, -LUNG) with Confocal Miniprobe™ (Coloflex, Gastroflex, Alveoflex) is a confocal laser imaging system with fiber optic probes that is intended to allow confocal laser imaging of the internal microstructure of tissues in the anatomical tract, i.e., gastrointestinal or pulmonary, accessed by the endoscope.

Prescription Use: ____________________________________________________________________________________________________________________________________________________________

AND/OR

Over-the-Counter Use:

(Part 21 CFR 801 Subpart D)

(21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Tarbar Buchen for wwn
(Division Sign-Off)

Division of General, Restorative, and Neurological Devices

510(k) Number K061666

F-600 System Special 510(k) July 17, 2006

Confidential Page 10 of 63

§ 876.1500 Endoscope and accessories.

(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.