K Number
K061656

Validate with FDA (Live)

Device Name
LASER-D68
Date Cleared
2006-12-15

(189 days)

Product Code
Regulation Number
878.4810
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Laser-D68 System provides visible red and infrared therapy to provide topical heating for the temporary:

  • Increase in local blood circulation
  • Relief of minor muscles and joint aches, pains and stiffness .
  • Relaxation of muscles .
  • . Relief of muscle spasms
  • . Relief of minor pain and stiffness associated with arthritis
Device Description

The Laser-D68 is a non-invasive, low energy infrared therapeutic medical laser that is intended to perform laser therapy in healthcare centers, physical therapy laboratories, and family doctor practices. The Laser D68 is supplied with three different probes that are used to generate, laser radiation when connected to the D68 driver unit. Probe #1 is a 660nm, Indium Gallium Aluminum Phoshpide (InGaAIP) laser diode (visible red light) with an adjustable treatment power output from 1 milliwatts (mW) to 50 milliwatts (mW) (in 1mW increments). Probe #2 is an 830nm, Gallium Aluminum Arsenide (GaAIAs) Iaser diode (infrared light) with an adjustable treatment power output from 20 milliwatts to 400 milliwatts (in 5mW increments). Probe #3 is a 904nm, Gallium Arsenide (GaAs) laser diode (infrared light) with adjustable treatment power output of either SmW to 90mW (in 5mW increments) (45W laser power source) or 5 mW to 150mW (in 5 mW increment) (75W modulated laser power source). Each of the laser probes are fully interchangeable with the D68 driver.

AI/ML Overview

The provided 510(k) summary for the Laser-D68 device (K061656) does not contain information about acceptance criteria or a study that proves the device meets specific performance criteria related to its intended therapeutic effects.

The document focuses on demonstrating substantial equivalence to predicate devices (Eltech s.r.l K-Laser (K050070) and Advanced Medical Technologies Maestro MDTTL Laser System (K053473)) in terms of intended use, technological characteristics, and safety testing.

Here's a breakdown of the requested information based on the provided text:


1. Table of Acceptance Criteria and Reported Device Performance

Acceptance CriteriaReported Device Performance
Therapeutic Efficacy: - Increase in local blood circulation - Relief of minor muscles and joint aches, pains and stiffness - Relaxation of muscles - Relief of muscle spasms - Relief of minor pain and stiffness associated with arthritisNot explicitly stated. The document asserts the device's intended use is to provide these effects but does not present specific performance metrics, targets, or results from a clinical study to prove these claims against defined acceptance criteria.
Electrical SafetyThe Laser-D68 included electrical safety testing and meets the applicable requirements of CFR 1040. The device has been certified to the requirements of the Medical Devices Directive 93/42/EEC, and is CE Marked.
Laser SafetyThe Laser-D68 included laser safety testing and meets the applicable requirements of CFR 1040. The device has been certified to the requirements of the Medical Devices Directive 93/42/EEC, and is CE Marked.
Electromagnetic Compatibility (EMC)The Laser-D68 included electromagnetic compatibility testing and meets the applicable requirements of CFR 1040. The device has been certified to the requirements of the Medical Devices Directive 93/42/EEC, and is CE Marked.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document does not describe any clinical test set or data related to the therapeutic efficacy of the device. The "testing" mentioned ([1] section VII) refers to electrical safety, laser safety, and electromagnetic compatibility testing, which are engineering and compliance tests, not clinical performance studies involving a test set of human subjects or cases. Therefore, there's no information on sample size, data provenance, or study design for therapeutic effectiveness.


3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

Not applicable. As no clinical test set for therapeutic efficacy is described, there's no mention of experts establishing ground truth for such a set.


4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable. No clinical test set for therapeutic efficacy is described.


5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. The Laser-D68 is a therapeutic medical laser, not an AI-assisted diagnostic device that would involve human readers or MRMC studies.


6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. The Laser-D68 is a therapeutic medical laser, not a diagnostic algorithm. The "standalone" performance here might refer to the device's physical output (e.g., power, wavelength), which is described in section I, but not in the context of a "standalone study" akin to AI performance.


7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

For the mentioned safety and compatibility tests, the "ground truth" would be compliance with established regulatory standards (e.g., CFR 1040, Medical Devices Directive 93/42/EEC). For therapeutic efficacy, no ground truth or clinical data is provided. The submission relies on "substantial equivalence" to predicate devices, implying that the known efficacy of the predicates is sufficient.


8. The sample size for the training set

Not applicable. This device is not an AI/machine learning system that requires a "training set."


9. How the ground truth for the training set was established

Not applicable. This device is not an AI/machine learning system that requires a "training set."


Summary Explanation:

This 510(k) submission for the Laser-D68 relies on demonstrating "substantial equivalence" to already legally marketed predicate devices, rather than presenting a de novo clinical study proving the therapeutic efficacy of the Laser-D68 itself against specific acceptance criteria.

The testing mentioned (electrical safety, laser safety, and EMC) confirms that the device functions safely and as designed according to regulatory standards (CFR 1040, MDD 93/42/EEC), but these are not studies evaluating its direct therapeutic performance (e.g., how much it increases blood circulation or reduces pain). The assumption is that because it is technologically similar and has the same intended use as the predicate devices, it will achieve the same therapeutic outcomes. No performance metrics or clinical study results for the stated therapeutic indications are provided in this document.

{0}------------------------------------------------

KO6/656

510(k) Summary (per 21 CFR 807.92)

DEC 1 5 2006

I. Applicant Laser Pain Clinics 300 N Highland, Suite 105 WNJ Medical Office Building Sherman, Texas 75092

Contact Person:Holly Wright, Manager and PresidentLaserMD Management GP, LLC (General Partner)c/o LaserMD Management GP, LLC903.868.1350 telephone206.203.2001 Facsimilee-mail: holly.wright@laserpainclinics.com
Date Prepared:November 6, 2006

II. Device Name

Proprietary Name:Laser-D68
Common/ Usual Name:Infrared Lamp
Classification Name:Lamp, infrared, therapeutic heating
Regulation Number:890.5500
Product Codes:ILY
Classification:2
Classification Panel:Physical Medicine

III. Predicate Device

Laser Pain Clinics is substantially equivalent to the Eltech s.r.l K-Laser (K050070) and the Advanced Medical Technologies Maestro MDTTL Laser System (K053473). The devices are therapeutic medical lasers designed to deliver light energy to the target tissue.

IV. Intended Use of the Device

Laser-D68 System provides visible red and infrared therapy to provide topical heating for the temporary:

  • Increase in local blood circulation �
  • Relief of minor muscles and joint aches, pains and stiffness .
  • Relaxation of muscles .
  • . Relief of muscle spasms
  • . Relief of minor pain and stiffness associated with arthritis

V. Description of the Device

The Laser-D68 is a non-invasive, low energy infrared therapeutic medical laser that is intended to perform laser therapy in healthcare centers, physical therapy laboratories, and family doctor practices.

Page 5-1

{1}------------------------------------------------

The Laser D68 is supplied with three different probes that are used to generate, laser radiation when connected to the D68 driver unit. Probe #1 is a 660nm, Indium Gallium Aluminum Phoshpide (InGaAIP) laser diode (visible red light) with an adjustable treatment power output from 1 milliwatts (mW) to 50 milliwatts (mW) (in 1mW increments). Probe #2 is an 830nm, Gallium Aluminum Arsenide (GaAIAs) Iaser diode (infrared light) with an adjustable treatment power output from 20 milliwatts to 400 milliwatts (in 5mW increments). Probe #3 is a 904nm, Gallium Arsenide (GaAs) laser diode (infrared light) with adjustable treatment power output of either SmW to 90mW (in 5mW increments) (45W laser power source) or 5 mW to 150mW (in 5 mW increment) (75W modulated laser power source).

Each of the laser probes are fully interchangeable with the D68 driver.

Summary of the Technical Characteristics for the Laser-D68 to the VI. referenced predicate devices.

The Laser-D68 and the aforementioned predicate devices are infrared lamps as defined in 21 CFR 890.5500. These devices use infrared diodes to emit invisible photonic energy to tissue. The intended use is identical for all devices.

VII. Testing

The testing of the Laser-D68 included electrical safety testing, laser safety testing and electromagnetic compatibility testing.

Safety & Effectiveness VIII.

There are no substantial differences between the Laser-D68 defined in this 510(k) submission and the predicate devices. They are similar to the technologies that are currently used in other similar medical devices.

The Laser-D68 included electrical safety testing, laser safety testing and electromagnetic compatibility testing.

The Laser-D68 meets the applicable requirements of CFR 1040.

The Laser-D68, as with the predicate devices, has been certified to the requirements of the Medical Devices Directive 93/42/EEC, and is CE Marked

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like symbol with three curved lines forming its body and wings. The logo is surrounded by text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Laser Pain Clinics % Ms. Holly Wright Manager and President 300 N. Highland, Suite 105 Sherman, Texas 75092

DEC 1 5 2006

Re: K061656

Trade/Device Name: Laser D-68 Regulation Number: 21 CFR 890.5500 Regulation Name: Infrared lamp Regulatory Class: II Product Code: ILY Dated: November 6, 2006 Received: November 7, 2006

Dear Ms. Wright:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

Page 2 - Ms. Holly Wright

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Mark N. Melkerson Director

Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

LASER
PAIN CLINICS

4. Indication for Use Statement

510(k) Number (if known): K061656

Device Name: Laser D-68

Indications for Use:

Laser-D68 System provides visible red and infrared therapy to provide topical heating for the temporary:

  • Increase in local blood circulation ●
  • Relief of minor muscles and joint aches, pains and stiffness .
  • Relaxation of muscles
  • Relief of muscle spasms
  • Relief of minor pain and stiffness associated with arthritis

Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off)

Division of General Restorative,

and Neurological Devices

Page 1 of
-----------
510(k) NumberK061656
K061656 MNM

§ 878.4810 Laser surgical instrument for use in general and plastic surgery and in dermatology.

(a)
Identification. (1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
(b)
Classification. (1) Class II.(2) Class I for special laser gas mixtures used as a lasing medium for this class of lasers. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 878.9.