K Number
K061590

Validate with FDA (Live)

Date Cleared
2006-06-21

(13 days)

Product Code
Regulation Number
870.2300
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The SMDIE Device Interfacing System is indicated for data transfer between standalone physiological monitors that collect noninvasive blood pressure, temperature, and blood oxygenation level through external cuffs and thermometers to external clinical information systems. The SMDIE is also indicated for patient weight transfer between Scale-Tronix 5002 class physician office scales. The SMDIE Device Interfacing System is not intended for continuous vital sign monitoring purposes.
Prescription Use

Device Description

The SMDIE Device Interfacing System Software is intended for use with standalone physiological monitor and Scale-Tronix 5002 scales to transfer data from the physiological monitor and scale to external devices via Health Level Seven data exchange protocols for vitals results and patient information.
Serial interfaces between physiological monitors and the SMDIE allow patient weight, non-invasive blood pressure, blood oxygenation level, and temperature to be communicated to the Device Connectivity & User Interface Component.

AI/ML Overview

The provided text describes a 510(k) premarket submission for the "SMDIE Device Interfacing System." This system is described as software intended to transfer data from physiological monitors and scales to external devices using Health Level Seven (HL7) protocols. The submission's focus is on establishing substantial equivalence to predicate devices, rather than presenting a standalone study with acceptance criteria and device performance metrics in the typical sense for a diagnostic or therapeutic device.

Therefore, many of the requested elements (like sample sizes for test/training sets, number/qualifications of experts, adjudication methods, MRMC studies, standalone performance metrics, and specific ground truth types) are not explicitly addressed because the regulatory pathway chosen (510(k) for a device interfacing system) primarily relies on demonstrating equivalence through functional and software testing rather than clinical performance trials with patient data.

Acceptance Criteria and Reported Device Performance

The submission states that "Software Testing was performed on the SMDIE Device Interfacing System and was determined to be acceptable." However, it does not explicitly define specific quantitative acceptance criteria or report detailed performance metrics (e.g., accuracy, sensitivity, specificity) in the way a diagnostic device would.

Based on the information provided, the "acceptance criteria" are implied to be the successful execution of software tests demonstrating the device's ability to reliably transfer data as intended, and that these tests found the device to be acceptable, leading to a determination of substantial equivalence.

Implied Acceptance Criteria and Reported Performance:

Acceptance Criteria (Implied from Submission)Reported Device Performance (Summary)
Accurate data transfer via HL7 protocols"Software Testing was performed... and was determined to be acceptable."
Compatibility with specified physiological monitors and scalesImplied by successful software testing and substantial equivalence claim.
No new questions regarding safety and effectiveness compared to predicate devices"The minor differences... do not raise any additional questions regarding safety and effectiveness."
Functionality as a physiological device data retriever and translatorThe device is described as performing this function, and testing deemed it acceptable.

Study Details (Based on available information in the 510(k) Summary)

Given the nature of this 510(k) submission for an interfacing system, a traditional clinical study with patient cohorts, "test sets," "training sets," and "ground truth" derived from expert consensus or pathology, as one might see for a diagnostic AI, is not described. Instead, the "study" referenced is primarily software testing and a comparison to predicate devices for substantial equivalence.

  1. Sample size used for the test set and the data provenance:

    • The document mentions "Software Testing was performed," but does not specify the sample size of data points, transactions, or devices used in this testing.
    • Data Provenance: Not specified. It's likely simulated or controlled test data used for software validation, rather than patient data from a specific country or retrospective/prospective collection.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable/Not specified. For a device interfacing system focused on data transfer, "ground truth" would typically refer to the correctness and integrity of the data transfer process itself. This is usually validated through technical verification and validation (V&V) procedures by software engineers and quality assurance professionals, rather than clinicians establishing "ground truth" on medical findings.
  3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not applicable/Not specified. Adjudication methods are typically used in studies involving human interpretation (e.g., radiologists reviewing images) to resolve discrepancies in expert opinions for diagnostic ground truth. This is not relevant to software testing of a data interfacing system.
  4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No MRMC study was done. This type of study assesses the impact of an AI algorithm on human reader performance, typically in diagnostic tasks. The SMDIE is a data transfer system, not an AI-powered diagnostic tool intended to assist human readers.
  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Yes, in a sense. The "Software Testing" can be considered a standalone performance evaluation of the algorithm's ability to transfer data correctly. However, a specific "standalone performance" metric (like sensitivity/specificity for a diagnostic algorithm) is not provided, only a general statement of "acceptability."
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • Not explicitly stated in terms of expert consensus, pathology, or outcomes data, as these are typically for diagnostic/prognostic devices. For this data interfacing system, the "ground truth" would refer to the expected correct data values and their successful transfer according to HL7 standards. This would be established by technical specifications and expected system behavior, verified through testing.
  7. The sample size for the training set:

    • Not applicable/Not specified. The SMDIE system is not described as an AI/ML device that requires a "training set" in the context of machine learning model development. Its functionality is based on predefined protocols and logic for data transfer.
  8. How the ground truth for the training set was established:

    • Not applicable. As there is no described training set for an AI/ML model, the concept of establishing ground truth for it does not apply here.

{0}------------------------------------------------

JUN 2 1 2006 Siemens Medical Solutions Health Services Corporation Traditional 510(k) Premarket Submission SMDIE Device Interfacing System

SECTION 5 - 510(k) SUMMARY

Submission Correspondent:Address:Emergo Group, Inc.2454 McMullen Booth RoadSuite 427Clearwater, FL 33759
Phone:Fax:E Mail:(727) 797-4727(727) 797-4757igordon@emergogroup.com
Contact:Mr. Ian GordonSenior Vice President
Submission Sponsor:Siemens Medical Solutions Health Services Corporation51 Valley Stream ParkwayMalvern, PA 19355
Date Prepared:May 19, 2006
Trade Name:SMDIE Device Interfacing System
Common Name:Physiological device data retriever and translator
Classification:Monitor, Physiological, Patient (Without ArrhythmiaDetection Or Alarms)
Regulation #: 870.2300, Class IIProduct Code: MWI

Description: The SMDIE Device Interfacing System Software is intended for use with standalone physiological monitor and Scale-Tronix 5002 scales to transfer data from the physiological monitor and scale to external devices via Health Level Seven data exchange protocols for vitals results and patient information.1

" Health Level Seven is one of several Ancrican National Standards Institute (ANSI) -accredited Standards Developing Organizations (SDOs) operating in the healtheare arena. Most SDOs produce standards (sometimes of protocols) for a particular healtheare domain such as pharmacy, medical devices, imaging or insurance (claims processing) transactions. Health Level Seven's domain is clinical and administrative data." Source: http://www.hl7.org.

{1}------------------------------------------------

Serial interfaces between physiological monitors and the SMDIE allow patient weight, non-invasive blood pressure, blood oxygenation level, and temperature to be communicated to the Device Connectivity & User Interface Component.

061590

List of Classification names and numbers of externally connected devices are as follows:

  • Cardiac Monitor 870.2300 .
  • Clinical Electronic Thermometer 880.2910 .
  • . Oximeter - 870.2700
  • Non-Invasive Blood Pressure Measurement System -- 870.1130 .
  • Breathing Frequency Monitor 868.2375 ●

The SMDIE Device Interfacing System is indicated for data Intended Use: transfer between standalone physiological monitors that collect noninvasive blood pressure, temperature, and blood oxygenation level through external cuffs and thermometers to external clinical information systems. The SMDIE is also indicated for patient weight transfer between Scale-Tronix 5002 class physician office scales. The SMDIE Device Interfacing System is not intended for continuous vital sign monitoring purposes.

For Prescription Use Only.

Predicate Devices:

The SMDIE Device Interfacing System substantially equivalent to the following predicate devices.

M2367A Device Link System by Agilent Technologies, Inc. 510(k) #: K012094 Decision: Substantially Equivalent (SE)

Unity Network ID by General Electric Medical Systems Information Technology 510(k) #: K021454 Decision: Substantially Equivalent (SE)

DataCaptor by Capsule Technologie 510(k) #: K032142 Decision: Substantially Equivalent (SE)

Integriti Patient Monitor by Alliance Instruments 510(k) #: K980688 Decision: Substantially Equivalent (SE)

Centra View Device Link System by ICU Data Systems, Inc. 510(k) #: K033283 Decision: Substantially Equivalent (SE)

{2}------------------------------------------------

K061590
P 3/3

Siemens Medical Solutions Health Services Corporation Traditional 510(k) Premarket Submission SMDIE Device Interfacing System

Safety and Effectiveness and Summary and Conclusion Regarding Substantial Equivalence:

By definition, a device is substantially equivalent when the device has the same intended use and the same technological characteristics as the predicate device, or has the same intended use and different technological characteristics, but it can be demonstrated that the device is as safe and effective as the predicate device and the new device does not raise different questions regarding safety and effectiveness as compared to the predicate device.

The minor differences between the SMDIE Device Interfacing System and the predicate devices cited do not raise any additional questions regarding safety and effectiveness. Software Testing was performed on the SMDIE Device Interfacing System and was determined to be acceptable.

The device, as designed, is as safe and effective as the predicate device, and the device is determined to be substantially equivalent to the referenced predicate devices.

{3}------------------------------------------------

Image /page/3/Picture/2 description: The image is a black and white logo for the Department of Health and Human Services. The logo features the department's emblem, which is a stylized representation of a human figure. The figure is composed of three abstract shapes that resemble a person's head, torso, and legs. The text "DEPARTMENT OF HEALTH AND HUMAN SERVICES - USA" is arranged in a circular pattern around the emblem.

JUN 2 1 2006

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Siemens Medical Solutions Health Services Corporation. c/o Mr. Daniel W. Lehtonen Responsible Third party Official Intertek Testing Services NA, Inc. 2307 East Aurora Rd. Unit B7 Twinsburg, OH 44087

Re: K061590

Trade Name: SMDIE Device Interfacing System Regulation Number: 21 CFR 870.2300 Regulation Name: Cardiac Monitor (including cardiotachometer and rate alarm). Regulatory Class: II (two) Product Code: MWI Dated: June 7, 2006 Received: June 8, 2006

Dear Mr. Daniel W. Lethtonen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2 Mr. Daniel W. Lethtonen

Please be advised that FDA's issuance of a substantial equivalence determination does not mean I lease be devilode that I Drimination that your device complies with other requirements of the Act that I Dr Hab Intact a and regulations administered by other Federal agencies. You must or any I odolar statutes and equirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set Of It Far 8077, laceming (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) I mis loter will anow you to ought inding of substantial equivalence of your device to a legally prematic is notificate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire of Compliance at (240) 276- 0120 . Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small other general international and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Bffmmimor for

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

INDICATIONS FOR USE

510(k) Number (if known):

Device Name: SMDIE Device Interfacing System

Indications for Use:

The SMDIE Device Interfacing System is indicated for data transfer between standalone physiological monitors that collect noninvasive blood pressure, temperature, and blood oxygenation level through external cuffs and thermometers to external clinical information systems. The SMDIE is also indicated for patient weight transfer between Scale-Tronix 5002 class physician office scales. The SMDIE Device Interfacing System is not intended for continuous vital sign monitoring purposes.

Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D)

AND/OR Over-The-Counter Use_ (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE -CONTINUE ON ANOTHER PAGE IF NEEDED)

B.Bennumton

Page 1 of 1

§ 870.2300 Cardiac monitor (including cardiotachometer and rate alarm).

(a)
Identification. A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.(b)
Classification. Class II (performance standards).