K Number
K061456

Validate with FDA (Live)

Date Cleared
2006-06-21

(27 days)

Product Code
Regulation Number
870.1130
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The A&D Medical UM-101 digital blood pressure monitor is intended for use by medical professionals for measuring the systolic and diastolic blood pressure and pulse rate.

Indications For Use:
Measure blood pressure (systolic and diastolic) and pulse rate.

Device Description

The A&D Medical UM-101 digital blood pressure monitor is intended for use by medical professionals for measuring the systolic and diastolic blood pressure and pulse rate. UM-101 uses an inflated cuff which is wrapped around the upper arm. The cuff is inflated manually by the medical professionals. The systolic and diastolic blood pressures are determined by auscultatory method. The deflation rate is regulated by a mechanical valve controlled by the medical professionals. There is a quick release valve so that the pressure of the cuff can be completely released at any time during the measurement. During the deflation, the systolic and diastolic can be marked by the medical professionals through a "MARK" button. UM-101 has an internal counter so it remembers the number of measurements. The information can be retrieved by holding the "MARK" button.

AI/ML Overview

The provided text is a 510(k) summary for the A&D Medical UM-101 Digital Blood Pressure Monitor. It outlines the device's technical characteristics and the standards it meets. However, it does not contain the detailed study information regarding acceptance criteria, reported device performance, sample sizes, expert ground truth establishment, adjudication methods, MRMC studies, or standalone performance that you requested.

The document primarily focuses on demonstrating substantial equivalence to a predicate device by listing the standards the device meets. These standards effectively serve as the acceptance criteria for this particular device submission rather than specific performance metrics from a clinical study.

Here's a breakdown of the information that can be extracted or inferred, and a clear statement about what is missing:

1. A table of acceptance criteria and the reported device performance

Acceptance Criteria (Standards Met)Reported Device Performance
ANSI/AAMI SP-10 standard(Not explicitly detailed in terms of specific performance metrics like accuracy or bias from a clinical study)
European Directive 93/42 EEC for Medical Products(Not explicitly detailed)
EN60601 General Safety Provisions(Not explicitly detailed)
EN60601-2-30 Particular Requirements for the Safety of BP Monitor(Not explicitly detailed)
EN60601-1-2 and EN55011 Electromagnetic Compatibility(Not explicitly detailed)

Missing Information: The document states the device meets these standards, but it does not provide the specific performance metrics (e.g., mean difference, standard deviation of differences for blood pressure readings) that would typically be reported from a clinical validation study against a reference standard to demonstrate compliance with these standards. For devices like blood pressure monitors, ANSI/AAMI SP-10 would have specific accuracy requirements that the device would need to meet.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

Missing Information: The document does not provide any information about a specific clinical test set, sample size, or data provenance. The assessment appears to be based on compliance with general and specific medical device standards rather than a detailed clinical validation study described in the summary.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

Missing Information: This information is not provided as no clinical test set or ground truth establishment process is described beyond adherence to general standards. For a blood pressure monitor, the "ground truth" during testing typically involves comparison against measurements taken by trained observers using a reference method (e.g., auscultation with a manometer) rather than expert consensus on images.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Missing Information: This information is not provided as no clinical test set or adjudication process is described.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not Applicable/Missing Information: This is a blood pressure monitor, not an AI-assisted diagnostic imaging device for human readers. Therefore, an MRMC study is not relevant in this context, and no such study is mentioned.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Applicable (in a different sense): The device itself is a "standalone" blood pressure monitor that provides a reading. It's not an algorithm in the sense of AI performing a diagnostic task. The performance would be its inherent accuracy in measuring blood pressure. However, the document does not provide standalone accuracy metrics from a clinical study against a reference standard. It only states compliance with standards like ANSI/AAMI SP-10, which imply acceptable standalone performance.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

Missing Information (beyond inference): The document does not explicitly state the type of ground truth used for compliance with the listed standards. For blood pressure monitors, the "ground truth" in standard validation protocols (like AAMI SP-10) typically involves simultaneous auscultatory measurements by trained observers or other reference methods.

8. The sample size for the training set

Not Applicable/Missing Information: This device is a traditional blood pressure monitor, not an AI/ML-based algorithm that requires a "training set." Therefore, this concept is not relevant, and no such information is provided.

9. How the ground truth for the training set was established

Not Applicable/Missing Information: As mentioned above, this is not an AI/ML device, so a training set and its associated ground truth establishment are not applicable.

In summary: The provided 510(k) summary focuses on demonstrating compliance with recognized standards (ANSI/AAMI SP-10, European Directives, EN standards for safety and EMC) to establish substantial equivalence. It does not provide the specific detailed clinical validation study results (like sample sizes, expert qualifications, exact performance metrics) that would typically be contained within a more comprehensive clinical data section of a 510(k) submission, even though such studies are generally performed to meet these standards.

{0}------------------------------------------------

K06145.6

Attachment (D) 510(k) Summary

DATE PREPARED 1.

May 23, 2006

SPONSOR INFORMATION 2.

A&D Engineering, Inc. Mr. Jerry Wang 1555 McCandless Drive, Milpitas, CA 95035 Tel: 408-518-5113 Fax: 408-635-2313 Email: jwang@andmedical.com

DEVICE NAME 3.

Proprietary Name:A&D Medical UM-101 Digital Blood Pressure Monitor
Common/Usual Name:Blood Pressure Monitor
Classification name:Non-invasive blood pressure measurement System21 CFR 870-1130, Class II, 74DXN.

DEVICE DESCRIPTION AND INTENDED USE 4.

The A&D Medical UM-101 digital blood pressure monitor is intended for use by medical professionals for measuring the systolic and diastolic blood pressure and pulse rate.

PREDICATE DEVCIE 5.

A&D LifeSource model UA-704 digital blood pressure monitor with FDA 510(k) K032499.

TECHNOLOGICAL CHARACTERISTICS 6.

UM-101 uses an inflated cuff which is wrapped around the upper arm. The cuff is inflated manually by the medical professionals. The systolic and diastolic blood pressures are determined by auscultatory method. The deflation rate is regulated by a

{1}------------------------------------------------

mechanical valve controlled by the medical professionals. There is a quick release valve so that the pressure of the cuff can be completely released at any time during the measurement. During the deflation, the systolic and diastolic can be marked by the medical professionals through a "MARK" button. UM-101 has an internal counter so it remembers the number of measurements. The information can be retrieved by holding the "MARK" button.

7. DEVICE TESTING

A&D Medical UM-101 digital blood pressure monitor meets the following standards:

  • . ANSI/AAMI SP-10 standard
  • European Directive 93/42 EEC for Medical Products .
  • . EN60601 General Safety Provisions
  • . EN60601-2-30 Particular Requirements for the Safety of BP Monitor
  • EN60601-1-2 and EN55011 Electromagnetic Compatibility .

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the seal of the Department of Health and Human Services (HHS). The seal features an eagle with its wings spread, surrounded by the words "DEPARTMENT OF HEALTH AND HUMAN SERVICES, USA" in a circular arrangement. The eagle is stylized with three distinct lines forming its body and wings. The overall design is simple and monochromatic.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JUN 2 1 2006

A&D Engineering, Inc. c/o Jerry Wang Director of Engineering & QA 1555 McCandless Drive Milpitas, CA 959035

Re: K061456

Trade Name: UM-101 Digital Blood Pressure Monitor Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: II (two) Product Code: DXN Dated: May 24, 2006 Received: May 25, 2006

Dear Mr. Wang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act

{3}------------------------------------------------

Page 2 -- Mr. Jerry Wang

or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Bhimiman for

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use

K061456 510(k) Number (if known):

Device Name: A&D Medical UM-101 Digital Blood Pressure Monitor

Indications For Use:

Measure blood pressure (systolic and diastolic) and pulse rate.

Prescription Use (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use × (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Blymmema

Sign-Off) Division of Cardloyascular Devices 510(k) Number

Page 1 of __ 1 __

§ 870.1130 Noninvasive blood pressure measurement system.

(a)
Identification. A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.(b)
Classification. Class II (performance standards).