(57 days)
The Aequalis Reversed Shoulder Prosthesis is indicated for patients with a functional deltoid muscle as a total shoulder replacement for the relief of pain and significant disability following arthropathy associated to massive and non repairable rotator cuff-tear. This device is also indicated for the prosthetic revisions with massive and non repairable rotator cuff-tear. Only the humeral components are for cemented use. The glenoid implant is anchored to the bone with 4 screws and is for non-cemented fixation.
When during the primary surgery the glenoid bone stock appears to be insufficient to bear the reversed glenoid components or when glenoid bone fracture occurs during the surgical procedures, the hemi-prosthesis adaptor and the union screw can be adapted to the humeral components in order to transform the Aequalis Reversed prosthesis into a non reversed hemi-prosthesis.
When, in case of revision of a Aequalis Reversed prosthesis, the glenoid bone stock appears to be insufficient to implant a base plate and a sphere of Aequalis Reversed range again, the use of the hemi-prosthesis adaptor and the union screw allows for the transformation of the Aequalis Reversed prosthesis into a non reversed hemi-prosthesis in order to avoid the revision of the humeral components.
The Aequalis Reversed Shoulder Prosthesis is a semi-constrained system composed of a humeral and a glenoid parts.
The present device modification submission consists in some additions and changes for metaphyseal screw, glenoid baseplate, glenoid sphere and insert.
Metaphyseal inserts are made of ultra high molecular weight polyethylene (UHMWPE). The base of the glenoid implant is manufactured from Titanium alloy. The sphere is manufactured from Cobalt-Chromium alloy and the screw is manufactured from Titanium alloy.
The hydroxylapatite coating conforms to the ASTM standard F 1185. The coating is performed by BioCoat, Inc. according to their Master File MAF-339.
This is a 510(k) premarket notification for a medical device (Aequalis Reversed Shoulder Prosthesis), not a study report detailing acceptance criteria and performance data. Therefore, the requested information (acceptance criteria, device performance, sample sizes, expert qualifications, etc.) is not available in the provided document.
510(k) submissions typically demonstrate substantial equivalence to a legally marketed predicate device rather than presenting new clinical study data with specific acceptance criteria and performance metrics for the novel device itself.
Here's what can be inferred from the document regarding the device and its regulatory status:
- Device Name: Aequalis Reversed Shoulder Prosthesis
- Intended Use: Restore shoulder joint function and relieve pain for patients with functional deltoid muscle, suffering from arthropathy associated with massive and non-repairable rotator cuff-tear, and for prosthetic revisions with massive and non-repairable rotator cuff-tear. It can also be adapted into a non-reversed hemi-prosthesis if glenoid bone stock is insufficient during primary surgery or revision.
- Predicate Devices: Aequalis Reversed Shoulder Prosthesis (K030941, K041873, K050316), Aequalis Shoulder System (K952928, K012212, K041339, K060209), Delta Shoulder (K021478). This indicates the device is considered substantially equivalent to these devices.
- Regulatory Clearance: The FDA has issued a substantial equivalence determination (K061439) for this device on July 20, 2006, allowing it to be marketed.
The document does NOT contain:
- A table of acceptance criteria and reported device performance.
- Sample sizes used for test sets or data provenance for a study proving acceptance criteria.
- Number or qualifications of experts used to establish ground truth for a test set.
- Adjudication method for a test set.
- Information about a multi-reader multi-case (MRMC) comparative effectiveness study or related effect sizes.
- Information about a standalone (algorithm only) performance study.
- Type of ground truth used (e.g., pathology, outcomes data).
- Sample size for a training set.
- How the ground truth for a training set was established.
This document is a regulatory approval letter based on substantial equivalence, not a clinical trial report.
{0}------------------------------------------------
JUL 20 2006
1061439 (pg 1 of 2) Implants Chirurgicaux
Summary of Safety and Effectiveness information Special 510(k) Premarket Notification - Aegualis Reversed Shoulder Prosthesis
Regulatory authority: Safe Medical Devices Act of 1990, 21 CRF 807.92
-
Device name AEOUALIS Reversed Shoulder Prosthesis Trade name: Total-Shoulder System and Hemi-Shoulder System Common name: Classification name: Shoulder joint metal/polymer semi-constrained cemented prosthesis
-
Submitter Tornier Rue Doven Gosse 38330 Saint Ismier - France
3) Company contact
Tornier Mrs Mireille Lémery Regulatory affairs Manager 161, rue Lavoisier - Montbonnot 38334 Saint Ismier Cedex - France Tel: 00 33 4 76 61 38 98 Fax: 00 33 4 76 61 35 33 e-mail : mireille lemery@tornier.fr
-
Classification Device class: Class II Classification panel: Orthopedic Product code: KWS
-
Equivalent / Predicate device
Aequalis Reversed Shoulder Prosthesis, TORNIER SA, K030941, K041873, K050316 Aequalis Shoulder System, TORNIER SA, K952928, K012212, K041339, K060209 Delta Shoulder, DePuy Inc, K021478
6) Device description
The Aequalis Reversed Shoulder Prosthesis is intended to relieve pain and significant disability following massive and non repairable cuff-tear associated to arthropathy and following massive cuff-tear arthropathy. In this case, the rotator muscles of the shoulder (supraspinatus, teres minor and long head of the biceps) are no more useful for mobility, and only the deltoid (for abduction and external rotation) and the subscapularis (for internal rotation) are functional.
Therefore, the usual goal of such surgery is to restore the shoulder joint to facilitate its working condition and to reduce or eliminate pain. The Aequalis Reversed Shoulder Prosthesis is intended to accomplish these
Image /page/0/Picture/14 description: The image shows a logo with a stylized letter 'T' inside a hexagon. The 'T' is bold and has a unique design, with the top part of the 'T' appearing to be connected to the vertical stem. The logo is simple and monochromatic, with a clean and modern aesthetic. The hexagon provides a defined border for the central 'T' symbol.
Tél. : 33 (0)4 76 61 35 00 TORNIER S.A.S. 161, rue Lavoisier - Montbonnot Fax : 33 (0)4 76 61 35 33 38334 SAINT-ISMIER CEDEX FRANCE
S.A.S. au capital de 288 000 € SIRET : 070 501 275 000 13 R.C.S. : 070 501 275 CODE APE : 331 B
Page 1/ page 2
{1}------------------------------------------------
K06)439 (pg 2 of 2)
goals. Its reversed design allows to medialize the center of rotation of the shoulder, lengthening the deltoid muscle lever arm.
The Aequalis Reversed Shoulder Prosthesis is a semi-constrained system composed of a humeral and a glenoid parts.
The present device modification submission consists in some additions and changes for metaphyseal screw, glenoid baseplate, glenoid sphere and insert.
7) Materials
Metaphyseal inserts are made of ultra high molecular weight polyethylene (UHMWPE). The base of the glenoid implant is manufactured from Titanium alloy. The sphere is manufactured from Cobalt-Chromium alloy and the screw is manufactured from Titanium alloy.
The hydroxylapatite coating conforms to the ASTM standard F 1185. The coating is performed by BioCoat, Inc. according to their Master File MAF-339.
8) Indications
The Aequalis Reversed Shoulder Prosthesis is indicated for patients with a functional deltoid muscle as a total shoulder replacement for the relief of pain and significant disability following arthropathy associated to massive and non repairable rotator cuff-tear. This device is also indicated for the prosthetic revisions with massive and non repairable rotator cuff-tear. Only the humeral components are for cemented use. The glenoid implant is anchored to the bone with 4 screws and is for non-cemented fixation.
When during the primary surgery the glenoid bone stock appears to be insufficient to bear the reversed glenoid components or when glenoid bone fracture occurs during the surgical procedures, the hemiprosthesis adaptor and the union screw can be adapted to the humeral components in order to transform the Aequalis Reversed prosthesis into a non reversed hemi-prosthesis.
When, in case of revision of a Aequalis Reversed prosthesis, the glenoid bone stock appears to be insufficient to implant a base plate and a sphere of Aequalis Reversed range again, the use of the hemiprosthesis adaptor and the union screw allows for the transformation of the Aequalis Reversed prosthesis into a non reversed hemi-prosthesis in order to avoid the revision of the humeral components.
Image /page/1/Picture/12 description: The image shows a logo with a stylized letter 'T' inside a hexagon. The 'T' is bold and has a unique design, with a smaller, geometric shape integrated into its upper portion. The logo is black and white, with the 'T' standing out against the white background of the hexagon.
TORNIER S.A.S. 161, rue Lavoisier - Montbonnot 38334 SAINT-ISMIER CEDEX FRANCE
Tél. : 33 (0)4 76 61 35 00 Fax : 33 (0)4 76 61 35 33
S.A.S. au capital de 288 000 € SIRET : 070 501 275 000 13 R.C.S. : 070 501 275 CODE APE : 331 B
Page 2/ page 2
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized depiction of an eagle or bird-like figure with three wing-like shapes. The text "DEPARTMENT OF HEALTH AND HUMAN SERVICES, USA" is arranged in a circular pattern around the bird-like figure. The logo is black and white.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUL 20 2006
Tornier S.A.S % Mireille Lemery Regulatory Affairs Manager 161, Rue Lavoisier - Montbonnot 38334 Saint-Ismier Cedex France
Re: K061439
Trade/Device Name: Aequalis Reversed Shoulder Prosthesis Regulation Number: 21 CFR 888.3660 Regulation Name: Shoulder joint metal/polymer semi-constrained cemented prosthesis Regulatory Class: Class II Product Code: KWS Dated: June 30, 2006 Received: July 3, 2006
Dear Ms. Lemery:
We have reviewed your Section 510(k) premarket notification of intent to market the device we nave rowed your be received the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use stated in the encreate) is ttgeenactment date of the Medical Device Amendments, or to conimer to rial to rial to rise and in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The r ou may, aterers, interes of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device is elaborned toor as courols. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set
{3}------------------------------------------------
Page 2 - Ms. Mireille Lemery
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at 240-276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Barbara Invelmo
Mark N. Melkerson Director Division of Generative, Restorative And Neurological Devices Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Indications for Use
Kob1431
510(k) Number (if known):
Device Name: Aequalis Reversed Shoulder Prosthesis
Indications For Use:
The Aequalis Reversed Shoulder Prosthesis is indicated for patients with a functional deltoid muscle as a total shoulder replacement for the relief of pain and significant disability following arthropathy associated to massive and non repairable rotator cuff-tear. This device is also indicated for the prosthetic revisions with massive and non repairable rotator cuff-tear. Only the humeral components are for cemented use. The glenoid implant is anchored to the bone with 4 screws and is for non-cemented fixation.
When during the primary surgery the glenoid bone stock appears to be when darmy the primary over glenoid components or when glenoid bone fracture occurs during the surgical procedures, the hemi-prosthesis adaptor and the union screw can be adapted to the humeral components in order to the union the Aequalis Reversed prosthesis into a non reversed hemiprosthesis.
When, in case of revision of a Aequalis Reversed prosthesis, the glenoid bone which, in base of roinsufficient to implant a base plate and a sphere of Aequalis Reversed range again, the use of the hemi-prosthesis adaptor and the union screw allows for the transformation of the Aequalis Reversed prosthesis into a non reversed hemi-prosthesis in order to avoid the revision of the humeral components.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off)
Division of General, Restorative,
and Neurological Devices
| Page 1 of 1 | |
|---|---|
| -- | ------------- |
| Tornier | 510(k) Number | Section 4 |
|---|---|---|
| K961439 |
N/A