(127 days)
The Lapdoc navigation system is a computerized navigational system intended to provide assistance in both the preoperative planning and the intra-operative surgical phases of dental implantation surgery. The system provides precise navigational guidance of surgical instruments such as a dental bur, with regard to the pre-operative planning in the dental implantation procedure.
The Lapdoc System is a computer based assistance system for dental implantologists. It supports surgeons during planning and execution phase of dental implant procedures by providing highly precise quiding assistance during the implantation process within the human jaw. The systems supports the surgical implantation process by providing visual guidance for position and orienting of a dental bur relative to a planned implant position.
The Lapdoc system is designed to support clinicians by decreasing complication rates of surgical interventions in the field of dental implantology. Using the system will allow precise orientation of the surgeon within the human jaw both in the medical image model of the jaw during the planning phase as well as in real anatomy during the execution phase. This is achieved by objective measuring the spatial positions of surgical instruments and the patients anatomy. Based on three dimensional medical image data, the surgeon plans an intervention by virtually placing models of implants within the graphical depiction (of the medical image set) of the jaw. The surgeon views and interprets the medical images which are graphically augmented with the computer models of the implants. It is further possible to manually mark the outlines of sensitive anatomical structures (i.e. nerves) within the medical images. To perfectly understand the geometric dimensions of the patients anatomy and the later intervention, a three dimensional model is created from isodense surfaces in the CT images and displayed on the computer screen. Lapdoc uses the geometric information of the planned implants within the medical images to generate highly precise information to guide the surgeon's instrument directly to the intended implant position and orientation.
An optical measurement system consisting of an infrared-based stereo camera system and a set of two rigid structures (also known as rigid bodies or trackers) with attached reflectors is used to measure the spatial positions of the patient and the instruments. Therefore, the trackers are reproducible attached to the patient (using a dental splint) and to the surgical bur (using a cone nut connector). The stereo camera system comprises an internal position measurement processor and is connected to the Lapdoc computer. It senses the positions of the instruments and the patients tracker and communicates it to the Lapdoc computer. The computer calculates in real time the deviation between the actual and the desired (from the planning phase) transformation of the positions of both trackers. The actual bur position and the planned implant position and orientation are graphically displayed to the surgeon using a mini-display is placed within the vicinity of the surgeon and near the patient's mouth. The surgeon minimizes this deviation during the drill procedure by moving the bur in a way, that position and orientation of both the dental bur and planned implant are congruent. The system supports the surgeon by actieving a highly precise surgical outcome according to the planned scenario.
Lapdoc system will graphically and audibly signal to the surgeon, if the plan is fulfilled and the surgical bur has reached its designated position and orientation. It will further warn the surgeon by reaching a minimal proximity value close to sensitive anatomical structures (i.e. nerves) and therefore avoids injury of such structures. Lapdoc is specifically designed to minimize the interaction between the user and the system. The application of the system follows simple ergonomic rules. The usage and working principle is easy to understand by the surgeon. The clinical workflow is not altered using Lapdoc.
To describe the patients anatomy and the medical image model in one common coordinate system, it is necessary to perform an image registration process. This registration process comprises a navigation bow of known geometry and an algorithm to detect and measure the position of the navigation bow within the medical image data set. Lapdoc correlates the positions of the navigation bow within the image data with its measured spatial position. Since the navigation bow is attached reproducible to the patients jaw by click-snaping it onto the patient's teeth, positions on or within the patients anatomy can also be described within the medical image data set using the registration function.
This document describes the Lapdoc system, a computer-based assistance system for dental implantologists that supports surgeons during the planning and execution phases of dental implant procedures by providing highly precise guiding assistance.
Safety & Performance Studies:
1. Acceptance Criteria and Device Performance:
The primary acceptance criterion for the Lapdoc system is its navigational accuracy.
| Acceptance Criteria | Reported Device Performance |
|---|---|
| +/- 1.0 mm accuracy | Achieved +/- 1.0 mm accuracy for overall system accuracy. |
This acceptance criterion was established based on:
- Review of state-of-the-art technology (Appendix documents No. 8 to No. 12).
- Opinions by experts in the field (Appendix documents No. 6 to No. 8).
- Accuracy specified for FDA cleared predicate devices (Appendix documents No. 13 to No. 16).
2. Sample Size and Data Provenance for Test Set:
The provided document does not explicitly state the sample size used for a test set or the data provenance (e.g., country of origin, retrospective or prospective) for validating the accuracy. The reported accuracy of +/- 1.0 mm appears to be a general system specification derived from predicate devices and expert opinion rather than a specific performance study on a defined test set.
3. Number and Qualifications of Experts for Ground Truth:
The document mentions that the accuracy level was defined based on "opinions by experts in the field (see Appendix documents No. 6 to No. 8)." However, it does not specify the exact number of experts used or their specific qualifications (e.g., "radiologist with 10 years of experience").
4. Adjudication Method for Test Set:
No information is provided regarding an adjudication method for a test set, as a specific performance study with a test set and ground truth generation is not detailed.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:
No MRMC comparative effectiveness study is reported in the provided text. The document focuses on the system's standalone accuracy in guiding surgical instruments rather than comparing human reader performance with and without AI assistance.
6. Standalone Performance (Algorithm Only):
Yes, a standalone performance assessment was conducted for the device's accuracy. The stated "overall system accuracy" of +/- 1.0 mm represents the standalone accuracy of the Lapdoc system in guiding surgical instruments according to pre-operative planning. The system calculates in real-time the deviation between the actual and desired positions and orientations, and the surgeon minimizes this deviation. The system itself measures and guides, indicating a standalone performance metric.
7. Type of Ground Truth Used:
The ground truth for the +/- 1.0 mm accuracy criterion seems to be a combination of:
- Expert Consensus: "opinions by experts in the field".
- Predicate Device Specifications: "accuracy specified for the FDA cleared predicate devices."
- State-of-the-Art Technology Review: This implies a benchmark derived from existing reliable technologies.
It is not based on pathology or outcomes data for this specific accuracy metric.
8. Sample Size for Training Set:
The document does not provide information about a distinct training set sample size. The Lapdoc system is described as a computer-based assistance system that uses geometric information and optical measurement in real-time, rather than a machine learning model that would typically have a separate training phase with a labeled dataset.
9. How Ground Truth for Training Set Was Established:
As there is no mention of a traditional training set for a machine learning model, the concept of establishing ground truth for a training set in that context does not apply here. The system's functionality relies on established principles of optical tracking and computer-assisted navigation rather than learning from a vast dataset.
{0}------------------------------------------------
Ko61224
Name
510(k) Notification Lapdoc
SECTION E - 510(k) SUMMARY
SEP - 5 2006
Datum
| Section E - 510(k) Summary | ErstelliGeprüft | |||
|---|---|---|---|---|
| Applicant´s Name: | RoboDent GmbHKurfürstendamm 21310719 BerlinTel: +49 (30) 49 500 820Fax: +49 (30) 49 500 822 | GenehmigtFreigegeben | ||
| Contact Person : | Prof. Dr. mult. h.c. Jürgen BierRoboDent GmbHKurfürstendamm 21310719 BerlinTel: +49 (30) 49 500 823Fax: +49 (30) 49 500 822i.bier@robodent.com | |||
| Date prepared : | April 2006 | |||
| Trade Name | Lapdoc | |||
| Common name | Image Guided Surgery System, CAS / Stereotaxic Instrument | |||
| Classification name | Stereotaxic Instrument (per 21 CFR section 880.4560 HAW) | |||
| Regulatory Class: | Class II | |||
| Predicate Devices: | IGI-System (K023424)OrthoPilot (K003347)Vectorvision Cranial / ENT (K023651)Vectorvision2 (K983831) |
Classification: The FDA has classified "stereotaxic Instruments" devices as class II, pursuant to 21 CFR 882.4560 (product code: HAW).
Description of the Device: The Lapdoc System is a computer based assistance system for dental implantologists. It supports surgeons during planning and execution phase of dental implant procedures by providing highly precise quiding assistance during the implantation process within the human jaw. The systems supports the surgical implantation process by providing visual guidance for position and orienting of a dental bur relative to a planned implant position.
The Lapdoc system is designed to support clinicians by decreasing complication rates of surgical interventions in the field of dental implantology. Using the system will allow precise orientation of the surgeon within the human jaw both in the medical image model of the jaw during the planning phase as well as in real anatomy during the execution phase. This is achieved by objective measuring the spatial positions of surgical instruments and the patients anatomy. Based on three dimensional medical image data, the surgeon plans an intervention by virtually placing models of implants within the graphical depiction (of the medical image set) of the jaw. The surgeon views and interprets the medical images which are graphically augmented with the computer models of the implants. It is further possible to manually mark the outlines of sensitive anatomical structures (i.e. nerves) within
510(k) Summary printed by: Dirk Mucha Print date: 2005-04-04 File printed: W.'SRL. Verwallung IBZWM PDF PROJEKTAKTEN2004 FDA LapDocl2006 FDA files'06-04-03 Summary.doo
complete file Page 15 of 616 Pages
P. 1 of 5
{1}------------------------------------------------
510(k) Notification Lapdoc
SECTION E - 510(k) SUMMARY
the medical images. To perfectly understand the geometric dimensions of the patients anatomy and the later intervention, a three dimensional model is created from isodense surfaces in the CT images and displayed on the computer screen. Lapdoc uses the geometric information of the planned implants within the medical images to generate highly precise information to guide the surgeon's instrument directly to the intended implant position and orientation.
An optical measurement system consisting of an infrared-based stereo camera system and a set of two rigid structures (also known as rigid bodies or trackers) with attached reflectors is used to measure the spatial positions of the patient and the instruments. Therefore, the trackers are reproducible attached to the patient (using a dental splint) and to the surgical bur (using a cone nut connector). The stereo camera system comprises an internal position measurement processor and is connected to the Lapdoc computer. It senses the positions of the instruments and the patients tracker and communicates it to the Lapdoc computer. The computer calculates in real time the deviation between the actual and the desired (from the planning phase) transformation of the positions of both trackers. The actual bur position and the planned implant position and orientation are graphically displayed to the surgeon using a mini-display is placed within the vicinity of the surgeon and near the patient's mouth. The surgeon minimizes this deviation during the drill procedure by moving the bur in a way, that position and orientation of both the dental bur and planned implant are congruent. The system supports the surgeon by actieving a highly precise surgical outcome according to the planned scenario.
Lapdoc system will graphically and audibly signal to the surgeon, if the plan is fulfilled and the surgical bur has reached its designated position and orientation. It will further warn the surgeon by reaching a minimal proximity value close to sensitive anatomical structures (i.e. nerves) and therefore avoids injury of such structures. Lapdoc is specifically designed to minimize the interaction between the user and the system. The application of the system follows simple ergonomic rules. The usage and working principle is easy to understand by the surgeon. The clinical workflow is not altered using Lapdoc.
To describe the patients anatomy and the medical image model in one common coordinate system, it is necessary to perform an image registration process. This registration process comprises a navigation bow of known geometry and an algorithm to detect and measure the position of the navigation bow within the medical image data set. Lapdoc correlates the positions of the navigation bow within the image data with its measured spatial position. Since the navigation bow is attached reproducible to the patients jaw by click-snaping it onto the patient's teeth, positions on or within the patients anatomy can also be described within the medical image data set using the registration function.
510(k) Summary printed by: Dirk Mucha Print date: 2006-04-04 File printed: W.'SRL. Verwallung!BZWM PDF PROJEKTAKTEN2004 FDA LapDocl2006 FDA files\06-04-03 Summary.doc
complete file Page 16 of 616 Pages
/2 of 5
{2}------------------------------------------------
SECTION E - 510(k) SUMMARY
The Lapdoc provides accurate guidance during the process of dental implantation. Surgical instruments are guided according to a pre-operative planning. The surgeon is assisted by avoiding the risk of causing damage to critical anatomical structures. Lapdoc is defined as a clinical assistance system, to support the surgeon during the implantation process with the ability of precise determination of position and orientation of surgical instruments within the patients anatomy. Lapdoc does not replace the human judgment. It is always the surgeons choice to define specific locations and angulations of the implant and to pursue this decision within the clinical treatment phase. Thus. the surgeon will not be released from his/her sole and ultimate clinical responsibility.
Intended use: The Lapdoc navigation system is a computerized navigational system intended to provide assistance in the preoperative planning and the intra-operative surgical phases of dental implantation surgery. The system provides precise navigational guidance of surgical instruments such as a dental bur, according to the preoperative planning in the dental implantation procedure.
Substantial Equivalence: Lapdoc shares technological and clinical features and an identical Intended use with different FDA-cleared tracking devices. The DenX-IGI (K023424), the Orthopilot (K003347), the VectorVision Cranial / ENT® (K023651) and the VectorVision 2 ® (K983831) system. Together, all this devices share some important features (not exclusively):
- All devices serve as assistance systems by providing visual support during the freehand ﻪ positioning and orientation process of surgical instruments relative to the patients anatomy;
- All devices measure positions of the human anatomy and surgical instruments by means of optical infrared-based measurement systems consisting of stereo camera systems;
- All devices calculate the position and orientation of surgical instruments within the medical image data set by using the position measurement data;
- All devices use sophisticated, yet easily understandable software tools to visualize medical ・ image data in different manners (slices and three dimensional models) and the positions and orientation of surgical instruments;
- All devices reduce the risk of harming or damaging anatomical structures by visualization of ﻴﺮ ﺍﻟﻤﺴﺎﺣﺔ ﺍﻟﻤﺴﺘﻘﻠﺔ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍ such with respect to the actual positions of surgical tools;
- All devices provide navigational accuracy of +/- 1,0 mm (overall system accuracy), which can not be achieved by brut human visual inspection.
The DenxIGP (K023424) system is a system for computer aided navigation of surgical instruments, whose purpose is to position a dental bur on the human jaw to optimally place and drill implant cavities. The system is based on the planning of such implant position within a three dimensional medical image data set (CT) and to pursue this position and angulation during the implantation process using an integrated position measurement device. The Orthopilot® (K003347) is
{3}------------------------------------------------
SECTION E - 510(k) SUMMARY
a computer aided system to navigate surgical instruments and to optimize the positions of cutting templates/guides for total knee replacement surgery and providing intra-operative measurements of bone alignment. The VectorVision Cranial / ENT® (K023651) and the VectorVision 2 °(K983831) are computer-aided systems for open or percutaneous surgery. They are indicated for any medical condition, where a reference to a rigid anatomical structure, such as the skull, a long bone or vertebra can be identified relative to a CT, MR or X-ray based model of anatomy.
Lapdoc and its predicate devices provide precise and accurate guidance of surgical tools and instruments within the patient's anatomy, near to organs or parts of it. Guidance is achieved by measuring positions of both the instruments and the anatomy and objective visualization of proportions and/or coherences of this spatial relationship.
All described devices do not intend to replace the surgeons expert knowledge and to release the clinicians responsibility.
The following navigation systems are used for the same body organs:
| System | organ / body part / clinical application |
|---|---|
| DenX IGI® (K023424) | Navigation in dental implantology |
| Lapdoc | Navigation in dental implantology |
Since the general principle of surgical navigation can be applied to any surgical field the following described navigation systems are used on different body organs:
| System | organ / body part / clinical application |
|---|---|
| DenX IGI® (K023424) | Navigation in dental implantology |
| Orthopilot® (K003347) | Navigation for knee replacement surgery and bone alignment surgery |
| VectorVision 2 ® (K983831) | Navigation in open/percutaneous surgery (skull/long-bones/vertebraes) |
| VectorVision Cranial / ENT® | Navigation in open/percutaneous surgery (skull/long-bones/vertebraes) |
| (K023651) | |
| Lapdoc | Navigation in dental implantology |
The effectiveness of all navigation support systems is based on their accuracy during the actual spatial guiding process for surgical instruments. The underlying principles of measuring positions of parts of the human anatomy and the positions of instruments, the registration of medical image data to the anatomy does not change with respect to the specific field of application remain the same. It can be assumed, that the safety and performance assumptions are not altered, when the principle of navigation is applied to the field of dental navigation. Summarized, the LapDoc-system is substantially equivalent to the declared predicate devices.
P. 4 of 5
{4}------------------------------------------------
SECTION E - 510(k) SUMMARY
Conclusions: Data presented above clearly demonstrate that the Lapdoc-System is substantially equivalent to its predicate devices in its technological features, clinical intended uses, its safety and effectiveness.
Performance standards:
510(k) Notification
Lapdoc
| StandardsNo. | StandardsOrganization | Standards Title | Version | Date |
|---|---|---|---|---|
| 601-1 | IEC | Medical electrical equipment - Part 1: General requirementsfor safety | German versionEN 60601-:1990 +A1:1993 +.A2:1995 | 03/1996 |
| 60601-1-2 | IEC | Equipment-PartMedicalElectrical1-2:GeneralRequirementsSafety -CollateralforStandard:Electromagnetic Compatibility Requirements and Tests. | German versionEN 60601-1-2:2001 | 10/2002 |
| 60601-1-1 | IEC | Medical electrical equipment - Part 1-1: General requirementsfor safety - Collateral standard: Safety requirements formedical electrical systems | German versionEN 60601-1-1:2001 | 08/2002 |
| 60601-1-4 | IEC | Medical electrical equipment Part 1-4: General requirementsfor safety - Collateral standard: Programmable electricalmedical systems | German versionIEC 60601-1-4:1996+A1:1999 | 04/2001 |
| 60601-1/A13 | EN | Medical electrical equipment - Part 1: General requirementsfor safety | German versionEN 60601-1 : 1990/AM 13 :1996 | 10/1996 |
| 15223 | ાકભ | Medical Devices - Symbols to be used with medical devicelabels, labelling and information to be supplied | 2nd edition | 04/2000 |
| 10993-5 | ાટળ | Biological Evaluation of medical devicesPart 5: Tests for cytotoxicity: in-vitro methods | German VersionEN ISO 10993-5:1999 | 11/1999 |
| 10993-10 | ાકેર | Biological Evaluation of medical devicesPart 10: Tests for irritation and delayed-type hypersensitivity | German versionEN ISO 10993-10:2002 | 02/2003 |
| 14791 | ાકર | Medical devices - Application of risk management to medicaldevices | German versionEN ISO14971:2001 | 03/2001 |
| 880 | EN | Graphical symbols for use in the labelling of medical devices | German versionEN 980:2003 | 08/2003 |
| 1041 | EN | Information supplied by the manufacturer with medical devices | German versionEN 1041:1998 | 04/1998 |
| E889 | ASTM | Standard Specification for Stainless Steel for SurgicalInstruments | 02 | 04/2002 |
| A182 | ASTM | Forged or Rolled Alloy and Stainless Steel Pipe Flanges,Forged Fittings, and Valves and Parts for High-TemperatureService | 04 | 10/2004 |
| 8265 | ASTM | Standard Specification for Titanium and Titanium Alloy Strip | 03 | 11/2003 |
| F67 | ASTM | Standard Specification for unalloyed Titanium for surgicalImplant Applications | 00 | 10/2000 |
Accuracy and functional performance standard: As of today, there are no specific FDA Guidance documents regarding the Image-Guided surgery systems for oral and maxillofacial surgery. The level of accuracy, defined to be acceptable in dental implantation navigation, was +/- 1,0 mm. This level of accuracy was defined based on review of state of the art technology (see Appendix documents No. 8 to No. 12), opinions by experts in the field (see Appendix documents No. 6 to No. 8) and the accuracy specified for the FDA cleared predicate devices (see Appendix documents No. 13 to No. 16).
510(k) Summary printed by: Dirk Mucha Print data: 2006-04-04 File printed: W:'SRL. Vervallung'BZMM PDF PROJEKTAKTE:N2004 FDA LapDocl2006 FDA files'06-04-03 Summary.coc
Ps.ofs
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for the Department of Health and Human Services (HHS). The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH AND HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized graphic of three human figures connected by flowing lines, representing the department's mission to protect the health of all Americans and provide essential human services.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
SEP - 5 2006
Robodent GMBH % Mr. Juergen Bier Kurfurstendamm 213 Berlin, Germany 10719
Re: K061224
Trade/Device Name: Lapdoc Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic instrument Regulation Class: Class II Product Code: HAW Dated: August 14, 2006 Received: August 21, 2006
Dear Mr. Bier:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{6}------------------------------------------------
Page 2 - Mr. Juergen Bier
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
fra
Mark N. Mekerson
Mark N. Mekerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
Indications for Use
510(k) Number (if known): K061224
Device Name: Lapdoc
Indications For Use:
The Lapdoc navigation system is a computerized navigational system intended to provide assistance in both the preoperative planning and the intra-operative surgical phases of dental implantation surgery. The system provides precise navigational guidance of surgical instruments such as a dental bur, with regard to the pre-operative planning in the dental implantation procedure.
Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
CHIVIS ign-Off
Division of General, Restorative, and Neurological Devices
510(k) Number.
Page 1 of
complete file Page 13 of 616 Pages
§ 882.4560 Stereotaxic instrument.
(a)
Identification. A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.(b)
Classification. Class II (performance standards).