K Number
K061112

Validate with FDA (Live)

Manufacturer
Date Cleared
2006-05-22

(31 days)

Product Code
Regulation Number
870.2700
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The PRO2® Pulse Reflectance Oximeter System is indicated for the continuous, non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate. The PRO2 is intended to monitor arterial saturation on the back and forehead locations in pediatric and adult populations. The PRO2 is for use in hospital, hospital-type facilities, and intra hospital transport environment.

Device Description

The Pro 9 Pulse Reflectance Oximeter System consists of a reusable sensor that emits and detect red and infrared light, a flexible disposable sensor holder to attach the sensor to the skin, a connecting cable, and a monitor incorporating control, processing, and display units.

The Pro2® Monitor contains an internal battery to power the unit when AC power is not available. The monitor displays the percentage of oxygen saturation in the blood, pulse rate, signal quality, and alarms.

The Proo Sensor geometry includes light sources that emit light in three different wavelengths, and detection areas defined by two photodetector rings arranged concentrically with the light sources in the center. The rings constitute an annular shape. which allow a multi-path acquisition of signals from a larger tissue area. The measurement of SpO2 is dependent upon specific wavelengths of light detected by its sensor.

The Pro2 device has a disposable sensor holder; Model # AHL-200 for adults and pediatric use. The Pro Sensor Holder provides optical isolation for external light and is attached to the skin via adhesive that is incorporated as part of the Sensor Holder.

AI/ML Overview

The provided documentation for the ConMed PRO2 Pulse Reflectance Oximeter System (K061112) is a 510(k) summary focused on demonstrating substantial equivalence to a predicate device. While it states that "The Pro2® Pulse Reflectance Oximeter System performance was tested with clinical data and the results met the acceptable criteria," it does not provide the specific acceptance criteria or detailed results of the study.

Therefore, I cannot fulfill all parts of your request with the information provided.

Here's what I can extract and what is missing:


1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria:

  • Not explicitly stated in the provided document. The document only mentions that the "results met the acceptable criteria" without detailing what those criteria were.

Reported Device Performance:

  • Not explicitly quantified in the provided document. The document states that the system's performance "was tested with clinical data and the results met the acceptable criteria," but no numerical performance metrics (e.g., accuracy, bias, precision) are given.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Test Set: Not specified.
  • Data Provenance: "Clinical data" is mentioned, implying human subjects, but the country of origin and whether the study was retrospective or prospective are not specified.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

  • Not Applicable / Not Provided. The document is for an oximeter, where "ground truth" for SpO2 is typically established by co-oximetry (blood gas analysis), not expert review of images or other data. The information about how ground truth was established for the clinical data is not included.

4. Adjudication Method for the Test Set

  • Not Applicable / Not Provided. This concept is more relevant for studies where human review or interpretation is being assessed. For an oximeter, ground truth is typically established by a gold standard measurement technique.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done

  • No. An MRMC study is typically for evaluating observer performance in interpreting medical images or other diagnostic tests. This is not relevant for an oximeter, which provides direct physiological measurements.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

  • Yes, implicitly. The device is an oximeter, which directly measures and displays physiological data (SpO2 and pulse rate). Its performance would be evaluated as a standalone device against a reference standard, not as an algorithm assisting human interpretation. However, the specific details of this standalone performance evaluation are not provided.

7. The Type of Ground Truth Used

  • Implicitly, a reference oximetry method (e.g., co-oximetry). While not explicitly stated, the gold standard for SpO2 measurement in clinical studies typically involves invasive blood sampling and co-oximetry. The document states "functional oxygen saturation of arterial hemoglobin (SpO2)," which aligns with this type of ground truth.

8. The Sample Size for the Training Set

  • Not Applicable / Not Provided. Oximeters are generally based on established physiological principles of light absorption by oxygenated and deoxygenated hemoglobin. While calibration might occur, the concept of a "training set" like in machine learning models is not typically applied in the same way to traditional medical devices like oximeters. The document does not mention any machine learning or AI components that would require a distinct training set.

9. How the Ground Truth for the Training Set Was Established

  • Not Applicable / Not Provided. (Refer to point 8)

Summary of Missing Information:

The provided 510(k) summary is very high-level regarding the clinical performance study. Crucial details such as:

  • Specific quantitative acceptance criteria (e.g., A_rms value, bias, precision).
  • Actual numerical performance results (e.g., reported accuracy, bias, or A_rms from the study).
  • Sample size of subjects for the clinical study.
  • Details of the clinical study design (e.g., patient population, induced hypoxia protocol, comparison method).
  • How ground truth (reference SpO2) was specifically established in the clinical trials.

These details are typically found in the full 510(k) submission, not usually in the publicly available summary.

{0}------------------------------------------------

510(k) Summary of Safety and Effectiveness

K061112

PRO2® Pulse Reflectance Oximeter System

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92.

MAY 2 2 2006

Submitter

ConMed Corporation 525 French Road Utica, NY 13502

Contact Person

lra Duesler Director, Requlatory Affairs ConMed Corporation 525 French Road Utica, NY 13502

Phone: 315-624-3072 Fax: 315-624-3225 e-mail: ira_duesler@mail.conmed.com

Date Prepared:

April 18, 2006

Name of Device

Pro2 Pulse Reflectance Oximeter System

Classification Names

Oximeter

Device Classification

Regulatory Class:Class II
Product Code:74 DQA
Classification Panel:Cardiovascular Device Panel
Regulation Number:21 CFR 870.2700

Predicate Devices

ConMed Pro2 Pulse Reflectance K032831 ConMed Corporation Oximeter System

{1}------------------------------------------------

Description of Device

The Pro 9 Pulse Reflectance Oximeter System consists of a reusable sensor that emits and detect red and infrared light, a flexible disposable sensor holder to attach the sensor to the skin, a connecting cable, and a monitor incorporating control, processing, and display units.

The Pro2® Monitor contains an internal battery to power the unit when AC power is not available. The monitor displays the percentage of oxygen saturation in the blood, pulse rate, signal quality, and alarms.

The Proo Sensor geometry includes light sources that emit light in three different wavelengths, and detection areas defined by two photodetector rings arranged concentrically with the light sources in the center. The rings constitute an annular shape. which allow a multi-path acquisition of signals from a larger tissue area. The measurement of SpO2 is dependent upon specific wavelengths of light detected by its sensor.

The Pro2 device has a disposable sensor holder; Model # AHL-200 for adults and pediatric use. The Pro Sensor Holder provides optical isolation for external light and is attached to the skin via adhesive that is incorporated as part of the Sensor Holder.

Indications For Use

The Pro2 Pulse Reflectance Oximeter System is indicated for the continuous, non-invasive monitoring Pro> Pulse Reflectance Oximeter System of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate. The Pro2 is intended to monitor arterial saturation on the back and forehead locations in pediatric and adult populations. The Pro is for use in hospital, hospital-type facilities, and intra hospital transport environment.

Nonclinical Performance

The Pro2 Pulse Reflectance Oximeter System was tested and passed all required electrical and biocompatibility testing.

Clinical Performance

The Pro2® Pulse Reflectance Oximeter System performance was tested with clinical data and the results met the acceptable criteria.

Conclusion

The Pro2 Pulse Reflectance Oximeter System is substantially equivalent to the following 510(k) cleared devices:

ConMed Pro2® Pulse Reflectance Oximeter System (K032831)

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the Department of Health and Human Services (HHS) in the United States. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH AND HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol that resembles a human figure embracing a bird, representing the department's mission to protect and promote the health and well-being of Americans.

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MAY 2 2 2006

Mr. Ira Duesler Director, Regulatory Affairs ConMed Corporation 525 French Road Utica, New York 13502

Re: K061112

Trade/Device Name: ConMed™ PRO2® Pulse Reflectance Oximeter System Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: II Product Code: DQA Dated: April 19, 2006 Received: April 25, 2006

Dear Mr. Duesler:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 - Mr. Duesler

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. or the For ar) - ruth all the Act's requirements, including, but not limited to: registration 1 od intoc comply Trail and and (21 CFR Part 801); good manufacturing practice and ibuing (= = = = = forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) This letter wifi anow Jourse FDA finding of substantial equivalence of your device to a premiured notification - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 11 - 11 - 11 - 11 - 11 - 11 - 11 - 11 your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), If you desire specific act Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the may obtain of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Chiu Lin, Ph.D.

Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

{4}------------------------------------------------

Indications For Use

510(k) Number (if known) _ K
Device Name:ConMed PRO2 Reflectance Oximeter System

Indications for use:

The PRO2® Pulse Reflectance Oximeter System is indicated for the continuous, กดก-The PRO2 - Fulse Reliectanice Oximocor Oytenh is natural hemoglobin (SpO2) and pulse invasive monitoring of functional oxygon on the back or forehead locations in the frealitions in rate. The PRO2 is intended to monitor aronton outch one in hospital-type facilities and intra-hospital transport environment.

Prescription Use X (per 21 CFR 801.109)

and/or

Over-the-counter Use _________________________________________________________________________________________________________________________________________________________

PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED

Concurrence of CDRH, Office of Device Evaluation (ODE)

Martindale

Anesthesiology, General Hospital.

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

§ 870.2700 Oximeter.

(a)
Identification. An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.(b)
Classification. Class II (performance standards).