K Number
K061002

Validate with FDA (Live)

Manufacturer
Date Cleared
2006-11-15

(218 days)

Product Code
Regulation Number
888.3060
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Maxima™ Anterior Cervical Plate System is intended for anterior fixation to the cervical spine. The specific clinical indications include:

degenerative disc disease (DDD) (defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), spinal stenosis, deformities or curvatures (i.e., scoliosis, kyphosis, and/or lordosis), tumor, pseudoarthrosis, and failed previous fusion.

Device Description

Maxima™ Anterior Cervical Plate System consists of a variety of shapes and sizes of Main Plates, screw, sub-plate, rivets and the associated instruments. The sub-plate is pre-assembled to the main plate and designed to prevent screws from backing out using the elastic behavior during the screw insertion. The rivets are also pre-assembled to the main plate and designed to assemble the sub-plate to the main plate firmly. Each component is subjected to a color anodizing process to differentiate the screw type and diameter and to make the surgical process easy. Screws can be inserted in the center hole of Main Plate - S type. Main Plate - S type are offered in no hole, one hole, two hole, three hole and ranges from 24mm to 108mm. Screws are available in lengths from 10mm to 20mm in 1mm increments. The screws have either a 3,5mm or 4.0mm diameter. Variable screw and Semi-fixed screw are provided. Variable screw provides more rotational and translational degree of freedom than Semi-Fixed screws.

The Maxima™ Anterior Cervical Plate System components are supplied non-sterile, are single use and are fabricated from titanium alloy (Ti-6AI-4V ELI) that conforms to ASTM F 136.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (from predicate device K994239, K030327)Reported Device Performance (Maxima™ Anterior Cervical Plate System)
Mechanical performance according to ASTM F1717-01Demonstrated substantial equivalence to ZEPHIR™ Anterior Cervical Plate System in mechanical testing per ASTM F1717-01.

2. Sample Size Used for the Test Set and Data Provenance

The document does not specify a separate "test set" in the context of clinical data for performance evaluation. The study relies on mechanical testing according to a standard. Therefore, there's no information about:

  • Sample size for the test set: Not applicable in the context of this submission.
  • Data provenance: Not applicable. The data provenance is mechanical testing in a lab setting.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

Not applicable. The "ground truth" for this device's performance is established by adherence to a recognized mechanical testing standard (ASTM F1717-01), not human expert consensus on clinical cases.

4. Adjudication Method for the Test Set

Not applicable, as the evaluation method is based on a mechanical testing standard, not clinical adjudication.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

No. This type of study is for comparing performance with and without an AI system, which is not relevant for this medical device submission.

6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) was Done

Not applicable. This device is a physical implant, not an algorithm.

7. The Type of Ground Truth Used

The ground truth used for demonstrating device performance is based on mechanical testing standards (ASTM F1717-01). The objective of the study was to show that the new device's mechanical performance is equivalent to that of existing, legally marketed predicate devices.

8. The Sample Size for the Training Set

Not applicable. This device is a physical implant; there is no "training set" in the context of machine learning or AI. The development process would involve engineering design and iterative physical testing, not data-driven training.

9. How the Ground Truth for the Training Set was Established

Not applicable, as there is no "training set" for this type of device.

{0}------------------------------------------------

K061002 510(k) SUMMARY

NOW 15 2006

The following 510(k) summary is being submitted as required by 21 CFR 807.92(a):

Submission Information:
Contact:Betty LimRegulatory Affairs
Sponsor:U&i CorporationYongHyun-Dong 529-1, EuijungbuKyonggi-Do, Korea 480-050Phone: 82-31-852-0102(ext. 302)Fax: 82-31-852-0107E-mail: willerry@lycos.co.kr
Date Prepared:Mar. 14, 2006
Device Identification
Trade Name:Maxima™ Anterior Cervical Plate System
Classification Name:Spinal Intervertebral Body Fixation Orthosis (KWQ)

Substantially Equivalent Predicate Legally Marketed Devices

The subject Maxima™ Anterior Cervical Plate System is substantially equivalent in intended use, materials, anatomical site to ZEPHIR™ Anterior Cervical Plate System (K994239, K030327)

{1}------------------------------------------------

Device Description

Maxima™ Anterior Cervical Plate System consists of a variety of shapes and sizes of Main Plates, screw, sub-plate, rivets and the associated instruments. The sub-plate is pre-assembled to the main plate and designed to prevent screws from backing out using the elastic behavior during the screw insertion. The rivets are also pre-assembled to the main plate and designed to assemble the sub-plate to the main plate firmly. Each component is subjected to a color anodizing process to differentiate the screw type and diameter and to make the surgical process easy. Screws can be inserted in the center hole of Main Plate - S type. Main Plate - S type are offered in no hole, one hole, two hole, three hole and ranges from 24mm to 108mm. Screws are available in lengths from 10mm to 20mm in 1mm increments. The screws have either a 3,5mm or 4.0mm diameter. Variable screw and Semi-fixed screw are provided. Variable screw provides more rotational and translational degree of freedom than Semi-Fixed screws.

The Maxima™ Anterior Cervical Plate System components are supplied non-sterile, are single use and are fabricated from titanium alloy (Ti-6AI-4V ELI) that conforms to ASTM F 136.

Indications for Use:

The Maxima™ Anterior Cervical Plate System is intended for anterior fixation to the cervical spine. The specific clinical indications include:

degenerative disc disease (DDD) (defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), spinal stenosis, deformities or curvatures (i.e., scoliosis, kyphosis, and/or lordosis), tumor, pseudoarthrosis, and failed previous fusion.

Statement of Technological Comparison:

Mechanical testing was carried out according to ASTM F1717-01 to validate the Maxima" Anterior Cervical Plate System. The testing demonstrated substantially equivalent to the ZEPHIR™ Anterior Cervical Plate System (K994239, K030327) in terms of intended use, materials, and anatomical sites.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes entwined around it, on the right side. To the left of the caduceus is a circular inscription that reads "DEPARTMENT OF HEALTH AND HUMAN SERVICES - USA" in capital letters.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

U&I Corporation c/o Mr. D. K. Rah 529-1 YongHyun-Dong Euijungbu, Kyonggi-Do, Korea 480-050

NOV 15 2006

Re: K061002

Trade Name: Maxima™ Anterior Cervical Plate System Regulation Number: 21 CFR 888.3060 Regulation Name: Spinal Intervertebral Body Fixation Orthosis Regulatory Class: II Product Code: K WQ Dated: October 23, 2006 Received: October 25, 2006

Dear Mr. Rah:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, l'isting of devices, good manufacturing practice, labeling, and prohibitions against mishranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, TDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the seconic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 100-1050.

{3}------------------------------------------------

Page 2 - Mr. D. K. Rah

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. Young ust comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the storonic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 100-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permitte your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation one it led, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may ohtain other general information on your responsibilities under the Act from the Division of Sonall Manufacturers, International and Consumer Assistance at its toll-free number (800) 633-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Mark N. Melkerson

Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

INDICATIONS for USE STATEMENT

510(k) Number (if known): K061002

Device Name: Maxima™ Anterior Cervical Plate System.

Indications for Use: The Maxima™ Anterior Cervical Plate System is intended for anterior fixation to the cervical spine. The specific clinical indications include:

degenerative disc disease (DDD) (defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), spinal stenosis, deformities or curvatures (i.e., scoliosis, and/or lordosis), tumor, pseudoarthrosis, and failed previous fusion.

Prescription Use X (Per 21 CFR 801 Subpart D)

OR

Over-the-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Of Division of General, Restorative and Neurological Devices

510(k) Number K061062

§ 888.3060 Spinal intervertebral body fixation orthosis.

(a)
Identification. A spinal intervertebral body fixation orthosis is a device intended to be implanted made of titanium. It consists of various vertebral plates that are punched into each of a series of vertebral bodies. An eye-type screw is inserted in a hole in the center of each of the plates. A braided cable is threaded through each eye-type screw. The cable is tightened with a tension device and it is fastened or crimped at each eye-type screw. The device is used to apply force to a series of vertebrae to correct “sway back,” scoliosis (lateral curvature of the spine), or other conditions.(b)
Classification. Class II.