K Number
K060822

Validate with FDA (Live)

Date Cleared
2006-05-23

(57 days)

Product Code
Regulation Number
886.1570
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Keeler Vamage Plus Indirect Ophthalmoscope is intended to examine the Cornea, aqueous, lens, vitreous and retina of the eye.

Device Description

The Vantage Plus is a head mounted Binocular Indirect ophthalmoscope designed to be used by trained personnel for illuminating and viewing parts of the eye such as the Cornea and retina when used in conjunction with an ophthalmic lens. The instrument has two main elements i) The illumination element ii) the viewing element. The illumination element consists of a 6Volt, 5Watt Xenon filament lamp, a condensing lens, IR filtration, User filters for diagnostic purposes, variable stop sizes for increasing/decreasing light aperture sizes, a projection lens and an illumination mirror used by the user to direct the light. The binocular Viewing element consists of low powered Viewing Lenses (+2 dioptre) and directional mirrors which are adjustable to obtain optimal views of the eye. The unit has adjustable illumination levels and is powered by the same power sources to that used on the Vantage (Re: 510(k) K942104).

To improve the ease of use of the instrument, the Vantage Plus combines the Aperture selector control and the Viewing mirror adjustments to achieve optimal instrument setting with minimum fuss. This feature is not present on the Vantage (Re: 510(k) K942104).

AI/ML Overview

Here's an analysis of the provided text regarding the acceptance criteria and study for the Keeler Vantage Plus Binocular Indirect Ophthalmoscope, structured according to your requested information:

1. Table of Acceptance Criteria and Reported Device Performance

The submission primarily focuses on demonstrating substantial equivalence to a predicate device (Vantage Binocular Indirect Ophthalmoscope) rather than presenting a study with specific acceptance criteria that the device meets to establish a new performance benchmark. Instead, it compares the technological characteristics and safety features of the new device against the predicate.

Here's a table based on the provided "Comparison of Indirect Ophthalmoscope Technological Characteristics," where the "Acceptance Criteria" are implied by the predicate device's characteristics and the new device's performance aligns with or is considered equivalent to those.

CriteriaAcceptance Criteria (Implied by Predicate)Reported Device Performance (Vantage Plus)
TypeBinocular (Headband Mounted)Binocular (Headband Mounted)
Illumination6V, 10 Watt Halogen Filament Bulb (Predicate)6V, 5 Watt Xenon Filament Bulb (Different bulb type, but considered equivalent/improved for intended use)
Safety Filter IR BlockerSelectablePermanent (Considered an improvement for safety)
Light Output (60mm patch size)954 Lux (Average from selection of bulbs)913 Lux (Comparable and within an acceptable range for ophthalmoscopic examination)
Blue FilterSelectableSelectable
Red Free FilterSelectableSelectable
DiffuserSelectableSelectable
Light aperture sizesLarge (60mm), Medium (45mm), Small (22mm)Large (60mm) +/-5%, Medium (45mm) +/-5%, Small (22mm) +/-5% (Within acceptable tolerance of predicate)
Inter Pupillary Distance Adj.52mm to 76mm52mm to 76mm (No change from predicate)
Lens Power Viewing Optics+2 Dioptre+2 Dioptre
AttachmentsHi-Mag, Teaching MirrorHi-Mag, Teaching Mirror
Power SourcesWall Pack, Smart Pack, Wireless battery Pack (Compliant with BS EN 60601-1 and 60601-1-2)Wall Pack, Smart Pack, Wireless Battery Pack (Same as predicate, and stated to be compliant with BS EN 60601-1 and 60601-1-2 with no adverse reports for these products)
Quality ApprovalsISO 9001:2000 and BS EN 13485:2003 compliant; full risk analysis, risk management, design verification/validation conducted to meet/surpass product specifications and relevant standards.Keeler is an ISO 9001:2000 and BS EN 13485:2003 compliant company, follows the stated design process including risk analysis and validation to meet/surpass specifications and comply with relevant standards.
Risk ManagementCompleted as per BS EN ISO 14971:2001Completed as per BS EN ISO 14971:2001
Light HazardWithin limits specified for Optical radiation Hazard (Section 6.3) of BS EN ISO 15004:1998Tested in accordance with BS EN ISO 15004:1998 and within specified limits (See separate report under Technical Documentation).
Electrical SafetyPower sources compliant with BS EN 60601-1 and 60601-1-2 with no adverse reports.Power sources are the same as the predicate, compliant with BS EN 60601-1 and 60601-1-2, and Keeler has not received any adverse reports on these products.
Explosion HazardLow probability; measures taken using non-explosive materials, meeting electrical safety requirements, detailed user instructions.Low probability; measures taken using non-explosive materials, meeting electrical safety requirements, detailed user instructions.
Fire HazardLow probability; self-extinguishing materials (HB or VO rated), materials in vicinity of illumination (5W Xenon lamp) designed for high-temperature environments.Low probability; self-extinguishing materials (HB or VO rated), materials in vicinity of illumination (5W Xenon lamp) designed for high-temperature environments.
Surface TemperatureCompliance with BS EN ISO 60601-1 for temperature of external surfaces and controls; caution instruction "Do not remove bulb whilst hot."Complies with BS EN ISO 60601-1 for temperature of external surfaces and controls; caution instruction "Do not remove bulb whilst hot."

2. Sample Size Used for the Test Set and the Data Provenance

The document does not describe a clinical "test set" in the traditional sense for evaluating diagnostic accuracy or a specific performance metric against a ground truth. Instead, the submission details a technical comparison and safety assessment against a predicate device.

  • Test Set/Sample Size: Not explicitly applicable in the context of a clinical performance study. The "measurements" for characteristics like light output and aperture sizes were taken from a "selection of bulbs" and from "440mm from the front of the instrument" on "a fully functional prototype." Specific numbers for how many prototypes were tested for these physical characteristics are not provided.
  • Data Provenance: The testing appears to have been conducted internally by Keeler Ltd., located in Windsor, Berkshire, United Kingdom. The data is retrospective in the sense that it's a comparison to an already marketed device, but the testing on the "Vantage Plus" prototype would be prospective for the new device.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

This information is not provided because the submission does not involve a clinical study requiring expert-established ground truth for a diagnostic outcome. The "ground truth" for the technical characteristics is established by the measurements and adherence to international standards.

4. Adjudication Method for the Test Set

Not applicable, as there was no clinical test set requiring adjudication of diagnostic outcomes.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

No MRMC study was performed or mentioned. This device is a manual ophthalmoscope, not an AI-powered diagnostic device.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done

Not applicable. The Vantage Plus Binocular Indirect Ophthalmoscope is a medical instrument used by a human operator, not an algorithm.

7. The Type of Ground Truth Used

The "ground truth" here is based on:

  • Physical measurements and engineering specifications: For characteristics like light output, aperture size, lens power, etc.
  • International safety and quality standards: BS EN ISO 15004:1998 (Light Hazard), BS EN ISO 14971:2001 (Risk Management), BS EN 60601-1 and 60601-1-2 (Electrical Safety).
  • Performance of the legally marketed predicate device (Keeler Vantage): The predicate's characteristics serve as the baseline for substantial equivalence.

8. The Sample Size for the Training Set

Not applicable. This is not an AI/machine learning device requiring a training set.

9. How the Ground Truth for the Training Set Was Established

Not applicable, as there is no training set mentioned for this device.

{0}------------------------------------------------

Special 510K Summary

Keeler Instruments Inc

Submitter Contact Information:

Company Name/Address:

456, Parkway, Broomall, PA 19008MAY 23 2006
Company Phone No:(610) 353 4350
Company Fax No:(610) 353 7814
Contact Person:Mr. Eugene VanArsdale (Marketing Director).
Manufacturing Location:Keeler Ltd Clewer Hill Road, Windsor, Berkshire. SL4 4AA. United Kingdom.
Company Phone No:(44) 1753 857177
Company Fax No:(44) 1753 830247
Contact Person:Mr. Neil Atkins (Development Manager).
Device Information:
Device Trade Name:Vantage Plus Binocular Indirect Ophthalmoscope
Common Name:Ophthalmoscope
Class:II
Classification Panel:86
Product Code:HLI (AC) and HLJ (DC)
Regulation Number:886.1570
Substantial Equivalence:
Predicate Device:Vantage Binocular Indirect Ophthalmoscope.
510(k) No:K942104
Manufacturer:Keeler Ltd (FDA Registration No 1000391004) (Address as above).
Distributor:Keeler Instruments Inc (FDA Registration No 2519784) (Address as above).
Device Description:The Vantage Plus is a head mounted Binocular Indirect ophthalmoscope designed to be used by trained personnel for illuminating and viewing parts of the eye such as the Cornea and retina when used in conjunction with an ophthalmic lens. The instrument has two main elements i) The illumination element ii) the viewing element. The illumination element consists of a 6Volt, 5Watt Xenon filament lamp, a condensing lens, IR filtration, User filters for diagnostic purposes, variable stop sizes for increasing/decreasing light aperture sizes, a projection lens and an illumination mirror used by the user to direct the light. The binocular Viewing element consists of low powered Viewing Lenses (+2 dioptre) and

The binocular Viewing element consists of low powered Viewing Lenses (+2 dioptre) and directional mirrors which are adjustable to obtain optimal views of the eye. The unit has adjustable illumination levels and is powered by the same power sources to that used on the Vantage (Re: 510(k) K942104).

To improve the ease of use of the instrument, the Vantage Plus combines the Aperture selector control and the Viewing mirror adjustments to achieve optimal instrument setting with minimum fuss. This feature is not present on the Vantage (Re: 510(k) K942104).

Dia

그 | 강

D

K0608zz

{1}------------------------------------------------

Comparison of Technological Characteristics:

:

Comparison of Indirect Ophthalmoscope Technological Characteristics
CriteriaPredicate DeviceKeeler Vantage(Re:510(k) K942104)Keeler Vantage Plus (see note 4)
TypeBinocular (Headband Mounted)Binocular (Headband Mounted)
Illumination6V, 10 Watt Halogen FilamentBulb6V, 5 Watt Xenon Filament Bulb
Safety Filter IR BlockerSelectablePermanent
Light Output (See note 1)954 Lux913 Lux
Blue FilterSelectableSelectable
Red Free FilterSelectableSelectable
DiffuserSelectableSelectable
Light aperture sizes (See note 2)Large (60mm)Large (60mm) +/-5%
Medium (45mm)Medium (45mm) +/-5%
Small (22mm)Small (22mm)+/-5%
Inter Pupillary Distance Adjustment52mm to 76mm
Lens Power Viewing Optics+2 Dioptre+2 Dioptre
AttachmentsHi-MagTeaching MirrorHi-MagTeaching Mirror
Power Sources (See Note 2)Wall PackSmart PackWireless battery PackWall PackSmart PackWireless Battery Pack
Note 1: Average light output from a selection of bulbs for a patch size of 60mm.Note 2: Measurements taken from 440mm from the front of the instrument.Note 3: All Power supplies comply with relevant parts of BS EN 60601-1 and 60601-1-2.Note 4: All testing has been carried out on a fully functional prototype.
Quality approvals:Keeler is an ISO 9001:2000 and BS EN 13485:2003 compliant company and are obliged to ensurethat our operating and design practices fully comply. As part of the design process, all necessary riskanalysis, risk management, design verification and validation are conducted to ensure that the safetyand effectiveness of product designs either meet or surpass the product specification and comply toothe relevant standards.
Risk Management:Completed as per BS EN ISO 14971:2001
Light Hazard:The Vantage Plus has been tested in accordance with BS EN ISO 15004:1998 and is within the limitsspecified for Optical radiation Hazard (Section 6.3) (See separate report under TechnicalDocumentation).
Electrical:The Vantage Plus is using the same Power Sources as the Vantage detailed in 510(k) K942104. Allpower sources are compliant with BS EN 60601-1 and 60601-1-2 and Keeler has not received anyadverse reports on these products (See separate reports under Technical Documentation).
Explosion:Low probability. All measures have been taken to reduce the risk of explosion by using non-explosive materials, ensuring the product meets all of the relevant electrical safety requirements anddetailed user instructions defining the requirements for safe usage of the instrument.
Fire Hazard:Low Probability. All measures have been taken to use self extinguishing materials which are eitherHB or VO rated. The system is illuminated using a 5watt Xenon lamp and all materials used in thevicinity are specifically designed to operate in high temperature environments safely.
Surface Temperature:Complies with BS EN ISO 60601-1 for temperature of external surfaces and controls. Caution in use instructions indicates 'Do not remove bulb whilst hot'.

{2}------------------------------------------------

Image /page/2/Picture/2 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of an eagle with its wings spread, forming a human-like figure. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.

JUL 3 2008 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Keeler Instruments, Inc. c/o Mr. Eugene VanArsdale Marketing Director 456 Parkway Broomall, PA 19008

Re: K060822

Trade/Device Name: Vantage Plus Binocular Indirect Ophthalmoscope Regulation Number: 21 CFR 886.1570 Regulation Name: Ophthalmoscope Regulatory Class: II Product Code: HLI and HLJ Dated: May 4, 2006 Received: May 8, 2006

Dear Mr. VanArsdale:

This letter corrects our substantially equivalent letter of May 23, 2006.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Malvina B. Edwards, ud

Malvina B. Eydelman, M.I Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Copy

Indications for Use

510(k) Number (if known):

Device Name: Vantage Plus Binocular Indirect Ophthalmoscope

Indications for Use:

i

The Keeler Vamage Plus Indirect Ophthalmoscope is intended to examine the Cornea, aqueous, lens, vitreous and retina of the eye.

Prescription Use (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Oft)
Division of Ophthalmic Ear,
Nose and Throat Devises

510(k) Number K060822

Page 1 of 1

22

§ 886.1570 Ophthalmoscope.

(a)
Identification. An ophthalmoscope is an AC-powered or battery-powered device containing illumination and viewing optics intended to examine the media (cornea, aqueous, lens, and vitreous) and the retina of the eye.(b)
Classification. Class II (special controls). The device, when it is an AC-powered opthalmoscope, a battery-powered opthalmoscope, or a hand-held ophthalmoscope replacement battery, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.