K Number
K060790

Validate with FDA (Live)

Manufacturer
Date Cleared
2006-05-25

(63 days)

Product Code
Regulation Number
868.5450
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Breas HA01 is a humidifier intended to humidify gases delivered to a patient with a Breas CPAP device or a bilevel ventilator constructed to use an internal mask leak and no active exhalation valve.

Device Description

The HA01 is a humidifier that only can be used together with Breas breathing therapy equipment. such as iSleep or Viyo. It uses pass-over humidification to humidify the air delivered to the patient. It consists of the following major components: 1 Water container - a sealed container with a water capacity of 400 ml. It has one inlet port where the air from the ventilator enters the humidifier and one 22 mm standard conical male outlet port where various non-invasive patient interfaces, such as nasal masks may be connected. It has a detachable lid that is removed when filling water in the container. The water container is attached to other Breathing therapy equipment, such as iSleep or Vivo when in use. Dimensions: W × H × D148 × 155 × 98 mm Weight; 260 g (empty container) 2 Electronic heating unit that via a heating plate heats the water container. Power. 12-30 V DC. max 40 W. is supplied from the Breas breathing therapy unit that the humidifier is connected to. The output from the HA01 unit is controlled by software included in the attached Breas unit. This software, which is a modularized part of the connected unit, controls temperature and it includes alarm supervision.

AI/ML Overview

The provided text describes a 510(k) premarket notification for a medical device, the Breas HA01 humidifier. It outlines the device's technical specifications and the testing performed to demonstrate its substantial equivalence to a predicate device. However, it does not contain the kind of detailed information typically found in a clinical study report regarding acceptance criteria, device performance, ground truth establishment, or sample sizes related to AI/algorithm performance.

Here's a breakdown of why it's not possible to extract the requested information from this document:

  • This is a 510(k) summary, not a clinical study report. 510(k) submissions focus on demonstrating substantial equivalence to a legally marketed predicate device, primarily through non-clinical testing and comparison of technological characteristics and intended use. They are generally not focused on clinical efficacy studies or detailed performance metrics against specific acceptance criteria in the same way a PMA or de novo submission might require.
  • The device is a medical device (humidifier), not an AI/algorithm-driven device. The document refers to "software included in the attached Breas unit" that "controls temperature and it includes alarm supervision," but this is supervisory software for a physical medical device, not an AI or algorithm that would require a test set, ground truth established by experts, or MRMC studies.
  • Clinical studies were explicitly stated as NOT required. The document states, "Clinical studies were not required to support a substantial equivalence determination." This confirms that no such clinical performance study, as typically described in the request, was conducted or submitted for this specific notification.

Therefore, I cannot provide the requested information because it is not present in the provided text. The questions are geared toward AI/algorithm performance studies, which were not performed or reported for this device in this submission.

{0}------------------------------------------------

510(k) Premarket Notification

Breas Medical AB

Kobo79c

510(k) SUMMARY

MAY 2 5 2006

Submitter

Breas Medical AB

Företagsvagen 1

Karl-Johan Holm

SE 435 33 Molnlycke

Sweden

Contact Person

Summary Date

Quality Assurance and Regulatory Affairs Manager

Heated Humidifier or humidifier, respiratory gas,

(direct patient interface) (21 CFR 868.5450)

Phone: +46 31 868830

Fax: +46 31 868810

March 06, 2006

HA01 Humidifier

HA01

BTT

Name of Device

Common Name

Classification Name

Product Code

Predicate Device

Breas HA50 (K002454)

{1}------------------------------------------------

Device Description:

The HA01 is a humidifier that only can be used together with Breas breathing therapy equipment. such as iSleep or Viyo. It uses pass-over humidification to humidify the air delivered to the patient.

It consists of the following major components:

1 Water container - a sealed container with a water capacity of 400 ml. It has one inlet port where the air from the ventilator enters the humidifier and one 22 mm standard conical male outlet port where various non-invasive patient interfaces, such as nasal masks may be connected. It has a detachable lid that is removed when filling water in the container. The water container is attached to other Breathing therapy equipment, such as iSleep or Vivo when in use. Dimensions: W × H × D148 × 155 × 98 mm Weight; 260 g (empty container)

2 Electronic heating unit that via a heating plate heats the water container. Power. 12-30 V DC. max 40 W. is supplied from the Breas breathing therapy unit that the humidifier is connected to. The output from the HA01 unit is controlled by software included in the attached Breas unit. This software, which is a modularized part of the connected unit, controls temperature and it includes alarm supervision.

The Breas HA01 may, together with other Breas breathing therapy equipment such as iSleep or Vivo, be used in clinical settings (e.g., hospitals, sleep laboratories, sub-acute care institutions) and home environments. It must always be prescribed by a licensed physician.

Intended Use:

The Breas HA01 is a humidifier intended to humidify gases delivered to a patient with a Breas CPAP device or a bilevel ventilator constructed to use an internal mask leak and no active exhalation valve.

Comparison of Use and Technological Characteristics:

The HA01 can be used in clinical settings (e.g., hospitals, sleep laboratories, sub-acute care institutions) and home environments.

As compared with the cited predicate device, the Breas HA01 has:

Same intended uses

{2}------------------------------------------------

Similar design

Same fundamental scientific technology

The differences that do exist are minimal. The HA01 is intended to be used integrated with other Breas products (e.g. Breas Vivo and iSleep systems). The features are described in the modified device information section 6 and appendix 6 (manuals and sell sheets).

Summary of Performance Testing:

    1. Non-clinical testing was conducted to verify that the Breas HA01 is capable of meeting its stated performance specifications and that all Risk Analysis issues have been appropriately addressed. The device passed all tests.
    1. Comparative testing to predicate device has been performed, both devices were tested according to ISO 8185. This bench testing confirmed that the HA01 is substantial equivalent to its predicate device.
    1. Testing was conducted to demonstrate compliance with applicable requirements in the November 1993 draft "Reviewer Guidance for Premarket Notification Submissions" published by the FDA's Division of Cardiovascular, Respiratory, and Neurological Devices and the July 1995 "Draft Reviewer Guidance for Ventilators". The testing included but was not limited to:
  • Electrical Safety testing per IEC 60601-1 ●
  • Safety and Performance testing per ISO 17510-1 .
  • . Medical use humidifiers testing per ISO 8185
  • . Electromagnetic Compatibility testing (EMC testing)
  • . Mechanical Safety testing
  • Environmental testing .
  • . Functional testing
  • Particle matter testing ◆

The testing was performed as an integral part of the development of the Breas Vivo 30, Breas Vivo 40 and Breas iSleep 20 systems.

The device passed all tests.

    1. Clinical studies were not required to support a substantial equivalence determination.

{3}------------------------------------------------

Conclusions:

The Breas HA 01 meets its stated performance specifications and criteria outlined in the Reviewer Guidance publications referenced above. We conclude that the device is capable of operating safely in their intended environments and will be effective in fulfilling its intended use.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized caduceus symbol, which features a staff with a snake winding around it. The text "DEPARTMENT OF HEALTH AND HUMAN SERVICES - USA" is arranged in a circular pattern around the symbol. The logo is black and white.

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MAY 2 5 2006

Mr. Karl-Johan Holm Quality Assurance & Regulatory Affairs Mananger Breas Medical AB Företagsvagen 1 Molnlycke SWEDEN SE 435 33

Re: K060790

Trade/Device Name: Breas, Model HA01 Regulation Number: 21 CFR 868.5450 Regulation Name: Respiratory Gas Humidifier Regulatory Class: II Product Code: BTT Dated: April 25, 2006 Received: May 1, 2006

Dear Mr. Holm:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{5}------------------------------------------------

Page 2 - Mr. Holm

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits vour device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Signature

Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

Indications for Use

510(k) Number (if known): K060790

Device Name: Breas HA01

Indications for Use:

The Breas HA01 is a humidifier intended to humidify gases delivered to a patient with a Breas CPAP device or a bilevel ventilator constructed to use an internal mask leak and no active exhalation valve.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use = (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Ausylhom

General Hospital i di Andalhesionegy, on Control, Denta

Number K060790

Page 1_of_1

§ 868.5450 Respiratory gas humidifier.

(a)
Identification. A respiratory gas humidifier is a device that is intended to add moisture to, and sometimes to warm, the breathing gases for administration to a patient. Cascade, gas, heated, and prefilled humidifiers are included in this generic type of device.(b)
Classification. Class II (performance standards).