K Number
K060686

Validate with FDA (Live)

Date Cleared
2006-06-09

(86 days)

Product Code
Regulation Number
870.1130
Age Range
All
Reference & Predicate Devices
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Microlife Automatic Blood Pressure Monitor, Model BP 3 AC1-1 P C i s a d evice intended to measure the systolic and diastolic blood pressure and pulse rate of an adult individual by using a non-invasive MAM (Microlife Average Mode) technique in which an inflatable cuff is wrapped around the upper arm.

The device detects the appearance of irregular heartbeat during measurement, and gives a warning signal with the reading once the irregular heartbeat is detected.

Device Description

Microlife Automatic Blood Pressure Monitor, Model BP3AC1-1 PC, is designed to measure the systolic and diastolic blood pressure and pulse rate of an individual by using a non-invasive MAM (Microlife Average Mode) technique in which an inflatable cuff is wrapped around the Upper arm. Our method to define systolic and diastolic pressure is similar to the auscultatory method but uses an electronic capacitive pressure sensor rather than a stethoscope and mercury manometer. The sensor converts tiny alterations in cuff pressure to electrical signals, by analyzing those signals to define the systolic and diastolic blood pressure and calculating pulse rate, which is a well - known technique in the market called the "oscillometric method".

The device detects the appearance of irregular heartbeat during measurement, and the arrhythmia symbol " device can be used in connection with your personal computer (PC) running the Microlife Blood Pressure Analyzer (BPA) software. The memory data can be transferred to the PC by connecting the monitor via cable with the PC.

AI/ML Overview

Here's an analysis of the provided text regarding the acceptance criteria and study for the Microlife Automatic Blood Pressure Monitor, Model BP3AC1-1 PC:

Disclaimer: This device submission is from 2006. Modern medical device regulations and testing standards, especially for AI/ML-based devices, are significantly more rigorous than what is presented here. This analysis is based solely on the provided text, which describes a relatively straightforward, non-AI medical device by today's standards. The "AI" mentioned in the questions is not explicitly covered in the document as it pertains to advanced machine learning, but rather refers to the "software algorithm" for oscillometric measurements.

Based on the provided 510(k) summary for the Microlife Automatic Blood Pressure Monitor, Model BP3AC1-1 PC:

1. Table of Acceptance Criteria and Reported Device Performance

The submission does not explicitly state specific numerical acceptance criteria for blood pressure accuracy (e.g., mean difference and standard deviation compared to a reference standard). Instead, it states that the device's clinical performance "will remain unchanged" from its predicate device. The primary acceptance criteria appear to be compliance with relevant consensus standards and internal reliability tests.

Acceptance Criteria CategorySpecific Criteria (Implicit/Explicit)Reported Device Performance
Clinical PerformanceNon-inferiority to predicate device (BP3BT0-AP) for systolic & diastolic blood pressure and pulse rate measurement accuracy."Clinical performance of the modified device will remain unchanged."
Mechanical/ElectricalCompliance with FDA November 1993 Draft "Reviewer Guidance for Premarket Notification Submissions", DCRND, outlining Electrical, Mechanical and Environmental Performance Requirements."The Microlife Automatic Blood Pressure Monitor, Model BP3AC1-1 PC, meets all relevant requirements of the aforementioned tests." This implies acceptance criteria were met.
ReliabilitySuccessful completion of Storage, Operation, Vibration, Drop, and Life tests."None of the tests demonstrated any design characteristics that violated the requirements of the Reviewer Guidance or resulted in any safety hazards." Implies compliance.
EMC (Electromagnetic Compatibility)Compliance with relevant EMC standards."EMC Test" performed; implied compliance with acceptance criteria.
Software (PC-link)Functionality of BPA version 3.1.6 and data transfer."PC-link software BPA version 3.1.6 test report" performed; implied successful operation.
Voluntary StandardsCompliance with ANSI/AAMI Voluntary Standard, SP10-2002."as well as... the ANSI/AAMI Voluntary Standard, SP10-2002 testing results" - implies compliance.
Irregular Heartbeat DetectionDetection and warning signal given with reading.Stated as a function, and that it is present in both predicate and new device, implying it performs as intended. No specific accuracy metrics provided for this feature.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: The document explicitly states: "Clinical performance of the modified device will remain unchanged; therefore another clinical test for the modified device, BP3AC1-1PC is not required." This means no new clinical test set (or patient sample) was used for this modified device. The submission relies entirely on the clinical performance of the predicate device (Microlife Automatic Blood Pressure Monitor, Model BP3BT0-AP).
  • Data Provenance: Not applicable for a new clinical test, as none was performed. For the predicate device, the provenance is not detailed in this submission.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

Not applicable. Since no new clinical tests were conducted for the BP3AC1-1 PC device, there was no new "ground truth" to establish for a test set. The submission relies on the established performance of the predicate device.

4. Adjudication Method for the Test Set

Not applicable, as no new clinical test set was used.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

  • No MRMC study was done. This device is a standalone blood pressure monitor.
  • No "AI" in the modern sense (machine learning) is described. The "software algorithm" mentioned refers to the standard oscillometric method used to derive blood pressure and pulse rate. The submission does not involve human readers interpreting AI outputs or any AI assistance to human tasks.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

Yes, the testing described is effectively a standalone performance assessment for the device itself. The device (algorithm + hardware) directly measures and displays blood pressure and pulse rate without human intervention in the measurement process (other than putting on the cuff and starting the device). The "clinical performance unchanged" statement refers to the device's standalone accuracy against a reference standard (likely auscultation) as established for the predicate device.

7. The Type of Ground Truth Used

For the predicate device's clinical validation (upon which this submission relies), the ground truth for blood pressure measurement would almost certainly have been established using expert auscultation simultaneous with the device's measurement, often with a double-blinded approach (e.g., using a Y-tube and two independent observers, with a third to break ties if necessary, and a mercury manometer as the reference standard). This standard methodology is typical for AAMI and BHS (British Hypertension Society) protocols.

8. The Sample Size for the Training Set

The concept of a "training set" as understood in modern machine learning is not applicable here. This device uses a pre-defined oscillometric algorithm, not a trainable AI model. Therefore, no distinct training set was used.

9. How the Ground Truth for the Training Set Was Established

As above, the concept of a training set is not applicable. The underlying algorithm (oscillometric method) is based on well-established physiological principles and signal processing, not on iterative training with a labeled dataset. The parameters of the oscillometric algorithm would typically be refined through engineering and calibration rather than machine learning training.

{0}------------------------------------------------

510(K) SUMMARY

This summary of 5l0(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.

The assigned 5l0(k) number is: K ص8666068

Submitter's Identification: 1.

JUN - 9 2006

Microlife Intellectual Property GmbH, Switzerland

Max Schmidheiny-Strasse 201 9435 Heerbrugg / Switzerland

Date Summary Prepared: January 11, 2006

2. Name of the Device:

Microlife Automatic Blood Pressure Monitor, Model BP3AC1-1 PC.

3. Information for the 510(k) Cleared Device (Predicate Device):

Microlife Automatic Blood Pressure Monitor, Model BP3BT0-AP, K#041411.

4. Device Description:

Microlife Automatic Blood Pressure Monitor, Model BP3AC1-1 PC, is designed to measure the systolic and diastolic blood pressure and pulse rate of an individual by using a non-invasive MAM (Microlife Average Mode) technique in which an inflatable cuff is wrapped around the Upper arm. Our method to define systolic and diastolic pressure is similar to the auscultatory method but uses an electronic capacitive pressure sensor rather than a stethoscope and mercury manometer. The sensor converts tiny alterations in cuff pressure to electrical signals, by analyzing those signals to define the systolic and diastolic blood pressure and calculating pulse rate, which is a well - known technique in the market called the "oscillometric method".

The device detects the appearance of irregular heartbeat during measurement, and the arrhythmia symbol " device can be used in connection with your personal computer (PC) running the Microlife Blood Pressure Analyzer (BPA) software. The memory data can be transferred to the PC by connecting the monitor via cable with the PC.

Exhibit #1-1

page 1 of 4

{1}------------------------------------------------

5. Intended Use:

The Microlife Automatic Blood Pressure Monitor, Model BP 3AC1-1 PC, is a device intended to measure the systolic and diastolic blood pressure and pulse rate of an adult individual by using a non-invasive MAM (Microlife Average Mode) technique in which an inflatable cuff is wrapped around the upper arm.

The device detects the appearance of irregular heartbeat during measurement, and gives a warning signal with the reading once the irregular heartbeat is detected.

Comparison to the 510(k) Cleared Device (Predicate Device): 6.

Both devices use the well-known oscillometric method within the software algorithm to determine the systolic and diastolic blood pressure and pulse rate. An upper arm cuff is inflated automatically; deflate rate is controlled by a factory set bleed valve and the deflation pressures are transferred via tubing to a sensor in both units. Moreover, both devices have irregular heart beat detection function.

The differences between BP3AC1-1 PC and the predicate device are described as the following:

1.1. Average Mode function (MAM function):

BP3AC1-1 PC contains a switchable average mode in which the device automatically repeats 3 individual measurements cycles. After that, the average of these 3 individual measurements is calculated and also be shown on the display.

1.2. PC-link function:

BP3AC1-1 PC can be used in connection with your personal computer (PC) running the Microlife Blood Pressure Analyzer (BPA) software. The memory data can be transferred to the PC by connecting the blood pressure monitor to personal computer via USB cable. All the memory data can be transferred to the connected computer through USB cable and be shown on the computer

Exhibit #1-2

Page 2 of 4

{2}------------------------------------------------

monitor. After transferred to computer, the memory data can then be saved in the personal computer memory.

Discussion of Non-Clinical Tests Performed for Determination of 7. Substantial Equivalence are as follows:

Testing information demonstrating safety and effectiveness of the Microlife Automatic Blood Pressure Monitor, Model BP3AC1-1 PC in the intended environment of use is supported by testing that was conducted in accordance with the FDA November 1993 Draft "Reviewer Guidance for Premarket Notification Submissions", DCRND, which outlines Electrical, Mechanical and Environmental Performance Requirements.

The following tests were conducted accordingly:

  • a. Reliability Test - Storage test
  • Reliability Test Operation test b.
  • Reliability Test Vibration test C.
  • d. Reliability Test - Drop test
  • Reliability Test Life test e.
  • f. EMC Test
  • PC-link software BPA version 3.1.6 test report ದ.

None of the tests demonstrated any design characteristics that violated the requirements of the Reviewer Guidance or resulted in any safety hazards. It is then concluded that the Microlife Automatic Blood Pressure Monitor, Model BP3AC1-1 PC, meets all relevant requirements of the aforementioned tests.

8. Discussion of Clinical Tests Performed:

Clinical performance of the modified device will remain unchanged; therefore anther clinical test for the modified device, BP3AC1-1PC is not required.

9. Conclusions:

We have demonstrated that the Microlife Automatic Blood Pressure Monitor, Model BP3AC1-1 PC, is as safe and effective as the predicate device, the Microlife Automatic Blood Pressure Monitor, Model BP 3BT0-AP based on electrical,

Exhibit #1-3

page 3 of 4

{3}------------------------------------------------

mechanical and environmental testing results as well as the FDA DCRND November 1993 Draft "Reviewer Guidance for Premarket Notification Submissions", and, the ANSI/AAMI Voluntary Standard, SP10-2002 testing results.

page 404

Exhibit #1-4

and and the country of the county of

:

.

:

{4}------------------------------------------------

Image /page/4/Picture/2 description: The image shows the logo for the Department of Health and Human Services (HHS). The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the text "DEPARTMENT OF HEALTH AND HUMAN SERVICES - USA" arranged in a circular fashion around the symbol. The text is in all capital letters and is evenly spaced around the circle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JUN - 9 2006

Microlife Intellectual Property c/o Mr. Tzu-Wei Li Center for Measurement Standards/ Industrial Technology Research Institute Bldg 16, 321 Kuang Fu Rd. Sec. 2 Hsinchu, Taiwan 30042 R.O.C. R.O.C.

Re: K060686

Trade/ Name: Microlife Automatic Blood Pressure Monitor, Mode BP3AC1-PC Regulation Number: 21 CFR 870.1130 Regulation Name: Blood Pressure Measurement System, Non-invasive Regulatory Class: II (two) Product Code: DXN Dated: March 6 and May 17, 2006 Received: March 15 and May 23, 2006

Dear Mr. Li:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{5}------------------------------------------------

Page 2 - Mr. Tzu-Wei Li

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must or any I with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) rms leket notification. The FDA finding of substantial equivalence of your device to a legally premated predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at 240-276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

B. Gimmerman for

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

Indications For Use Form

Page 1 of 1

510(k) Number (if known):_

Device Name: Microlife Automatic Blood Pressure Monitor , Model BP3AC1-1PC Indications for Use:

The Microlife Automatic Blood Pressure Monitor, Model BP 3 AC1-1 P C i s a d evice intended to measure the systolic and diastolic blood pressure and pulse rate of an adult individual by using a non-invasive MAM (Microlife Average Mode) technique in which an inflatable cuff is wrapped around the upper arm.

The device detects the appearance of irregular heartbeat during measurement, and gives a warning signal with the reading once the irregular heartbeat is detected.

Prescription Use (Part 21 CFR 801 Subpart D) AND/ OR

Over-The-Counter Use _ X (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

B.B. Commemor

§ 870.1130 Noninvasive blood pressure measurement system.

(a)
Identification. A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.(b)
Classification. Class II (performance standards).