K Number
K060558

Validate with FDA (Live)

Date Cleared
2006-05-11

(70 days)

Product Code
Regulation Number
866.1645
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The BD Phoenix™ Automated Microbiology System is intended for the rapid identification and in vitro antimicrobial susceptibility testing of isolates from pure culture of most aerobic and facultative anaerobic Gram-negative and Gram-positive bacteria of human origin.

The BD Phoenix™ Automated Microbiology System is intended for in vitro quantitative determination of antimicrobial susceptibility by minimal inhibitory concentration (MIC) of most Gram-negative aerobic and facultative anaerobic bacteria isolates from pure culture for Enterobacteriaceae and Non-Enteriaceae and most Gram-positive bacteria isolates from pure culture belonging to the genera Staphylococcus, Enterococcus, and Streptococcus.

This premarket notification is for additional Gram-negative organism groups and Cephalothin (1-64 ug/mL) on the BD Phoenix Automated Microbiology System.

Device Description

The BD Phoenix Automated Microbiology System (Phoenix System) is an automated system for the rapid identification (ID) and antimicrobial susceptibility testing (AST) of clinically relevant bacterial isolates. The system includes the following components:

  • . BD Phoenix instrument and software.
  • t BD Phoenix panels containing biochemicals for organism ID testing and antimicrobial agents for AST determinations.
  • . BD Phoenix ID Broth used for performing ID tests and preparing AST Broth inoculum.
  • . BD Phoenix AST Broth used for performing AST tests only.
  • BD Phoenix AST Indicator solution added to the AST Broth to aid in bacterial growth ● determination.
    The Phoenix AST method is a broth based microdilution test. The Phoenix System utilizes a redox indicator for the detection of organism growth in the presence of an antimicrobial agent. Measurements of changes to the indicator as well as bacterial turbidity are used in the determination of bacterial growth. Each AST panel configuration contains several antimicrobial agents with a wide range of two-fold doubling dilution concentrations.

The Phoenix panel is a sealed and self-inoculating molded polystyrene tray with 136 micro-wells containing dried reagents. Organisms for susceptibility testing must be a pure culture and preliminarily identified as a Gram-negative or Gram-positive isolate. Phoenix panels are inoculated with a specified organism density and placed into the instrument.

The instrument houses the panels where they are continuously incubated at a nominal temperature of 35°C. The instrument takes readings of the panels every 20 minutes. The readings are interpreted to give an identification of the isolate, minimum inhibitory concentration (MIC) values and category interpretations, S, I, or R (sensitive, intermediate, or resistant).

AI/ML Overview

The BD Phoenix™ Automated Microbiology System for use with cephalothin (1-64 µg/mL) was studied to demonstrate its substantial equivalence to the CLSI reference broth microdilution method for antimicrobial susceptibility testing.

Here's an analysis of the acceptance criteria and the study that proves the device meets those criteria:

Acceptance Criteria and Reported Device Performance

The core acceptance criteria for the BD Phoenix™ Automated Microbiology System when compared to the CLSI reference broth microdilution method are Essential Agreement (EA) and Category Agreement (CA).

The document states:
"The performance of the Phoenix System was assessed by calculating Essential Agreement (EA) and Category Agreement (CA) to expected/reference results for all isolates tested."

Essential Agreement (EA) occurs when the BD Phoenix™ Automated Microbiology System agrees exactly or within ± one two-fold dilution to the reference result.
Category Agreement (CA) occurs when the BD Phoenix™ Automated Microbiology System agrees with the reference method with respect to the FDA categorical interpretive criteria (susceptible, intermediate, and resistant).

Unfortunately, "Table 1" which is supposed to summarize the performance, is illegible in the provided text. Therefore, the specific numerical values for EA and CA cannot be extracted. However, the conclusion states: "The data collected from the substantial equivalence studies demonstrate that testing on the BD Phoenix™ Automated Microbiology System with this antimicrobial agent is substantially equivalent as outlined in the FDA draft guidance document, 'Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA,' February 5, 2003." This implies that the reported performance (which would have been in the illegible Table 1) met the acceptance thresholds specified in that FDA guidance.

Table of Acceptance Criteria and Reported Device Performance:

Acceptance CriteriaReported Device Performance (from Table 1, assumed to meet criteria based on conclusion)
Essential Agreement (EA)Specific value illegible in provided text. Assumed to meet FDA guidance.
Category Agreement (CA)Specific value illegible in provided text. Assumed to meet FDA guidance.
Intra-site Reproducibility> 90%
Inter-site Reproducibility> 95%

Sample Size and Data Provenance

Test Set Sample Size: The document mentions "clinical, stock and challenge isolates were tested." No specific total number of isolates for the test set is provided in the accessible text. However, the table of performance (Table 1) has a column that appears to be "Number of Isolates" but the values are also illegible.

Data Provenance: The study involved multiple geographically diverse sites across the United States. The data is prospective, testing the Phoenix System against the CLSI reference method.

Number of Experts and Qualifications for Ground Truth for Test Set

This section is not applicable in the context of this device. The ground truth for antimicrobial susceptibility testing (AST) is not established by human experts in the way, for example, a radiologist establishes ground truth for an image. Instead, the ground truth is established by a standardized, well-defined laboratory method: the CLSI reference broth microdilution method.

Adjudication Method for Test Set

This section is not applicable. As the ground truth is established by a standardized laboratory method (CLSI reference broth microdilution), there is no human adjudication process involved for determining the "correct" ground truth result.

Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No, an MRMC comparative effectiveness study was not done. This type of study is typically performed for diagnostic devices where human interpretation plays a role (e.g., radiology AI). The BD Phoenix system is fully automated and does not involve human readers interpreting results in the same way. The comparison is between the automated system and a reference lab method.

Standalone Performance

Yes, a standalone performance study was done. The entire "Clinical Studies" section describes the performance of the Phoenix System by itself ("algorithm only," in this context meaning the automated system without human intervention influencing the result) compared to the CLSI reference method. The Essential Agreement (EA) and Category Agreement (CA) values discussed are metrics of this standalone performance.

Type of Ground Truth Used

The type of ground truth used is the CLSI reference broth microdilution method. This is a highly standardized and accepted laboratory method for determining antimicrobial minimal inhibitory concentrations (MICs) and categorical interpretations.

Sample Size for the Training Set

The document does not explicitly mention a "training set" or its sample size. This is common for this type of medical device at the time of its submission (2006). The BD Phoenix system uses pre-established algorithms based on known biochemical reactions and growth characteristics, rather than a machine learning model that would typically have a distinct training phase with a labeled dataset. The development likely involved extensive internal testing and optimization rather than a formal "training set" as understood in modern AI/ML development.

How the Ground Truth for the Training Set Was Established

As there is no distinct "training set" identified, this question is not applicable in the same way it would be for a typical AI/ML device. The underlying principles and mechanisms of the BD Phoenix system's interpretation are based on established microbiology principles and reference methods (like CLSI) that guide the determination of appropriate breakpoints and interpretive criteria.

{0}------------------------------------------------

510(K) SUMMARY

SUBMITTED BY:Becton, Dickinson and Company7 Loveton CircleSparks, MD 21152Phone: (410) 316 - 4278Fax: 410-316-4499MAY 11 2006
CONTACT NAME:Monica E. GiguereRegulatory Affairs Specialist
DATE PREPARED:May 2, 2006
DEVICE TRADE NAME:BD Phoenix™ Automated Microbiology System -Cephalothin – 1-64 µg/mL
DEVICE COMMON NAME:Antimicrobial susceptibility test system-short incubation
DEVICE CLASSIFICATION:Fully Automated Short-Term Incubation Cycle AntimicrobialSusceptibility Device, 21 CFR 866.1645
PREDICATE DEVICES:VITEK® System (PMA No. N50510) and BD Phoenix™Automated Microbiology System with Gatifloxacin (K020321,May 23, 2002), Ofloxacin (K020323, April 14, 2002),Levofloxacin (K020322, March 27, 2002), Cephalothin(K031943, August 6, 2003).
INTENDED USE:The BD Phoenix™ Automated Microbiology System isintended for the rapid identification and in vitro antimicrobialsusceptibility testing of isolates from pure culture of mostaerobic and facultative anaerobic Gram-negative and Gram-positive bacteria of human origin

DEVICE DESCRIPTION:

The BD Phoenix Automated Microbiology System (Phoenix System) is an automated system for the rapid identification (ID) and antimicrobial susceptibility testing (AST) of clinically relevant bacterial isolates. The system includes the following components:

  • . BD Phoenix instrument and software.
  • t BD Phoenix panels containing biochemicals for organism ID testing and antimicrobial agents for AST determinations.
  • . BD Phoenix ID Broth used for performing ID tests and preparing AST Broth inoculum.
  • . BD Phoenix AST Broth used for performing AST tests only.

{1}------------------------------------------------

  • BD Phoenix AST Indicator solution added to the AST Broth to aid in bacterial growth ● determination.
    The Phoenix AST method is a broth based microdilution test. The Phoenix System utilizes a redox indicator for the detection of organism growth in the presence of an antimicrobial agent. Measurements of changes to the indicator as well as bacterial turbidity are used in the determination of bacterial growth. Each AST panel configuration contains several antimicrobial agents with a wide range of two-fold doubling dilution concentrations.

The Phoenix panel is a sealed and self-inoculating molded polystyrene tray with 136 micro-wells containing dried reagents. Organisms for susceptibility testing must be a pure culture and preliminarily identified as a Gram-negative or Gram-positive isolate. Phoenix panels are inoculated with a specified organism density and placed into the instrument.

The instrument houses the panels where they are continuously incubated at a nominal temperature of 35°C. The instrument takes readings of the panels every 20 minutes. The readings are interpreted to give an identification of the isolate, minimum inhibitory concentration (MIC) values and category interpretations, S, I, or R (sensitive, intermediate, or resistant).

DEVICE COMPARISON:

The BD Phoenix™ Automated Microbiology System demonstrated substantially equivalent performance when compared with the CLSI reference broth microdilution method. This premarket notification provides data for additional gram-negative organism groups with cephalothin and the BD PhoenixTM Automated Microbiology System.

SUMMARY OF SUBSTANTIAL EQUIVALENCE TESTING:

The BD Phoenix™ Automated Microbiology System has demonstrated substantially equivalent performance when compared to the CLSI reference broth microdilution method (AST panels prepared according to NCCLS M7). The system has been evaluated as defined in the FDA Draft guidance document, "Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA," February 5, 2003.

Site Reproducibility

Intra- and inter-site reproducibility of this antimicrobial agent in the BD Phoenix System was evaluated at three sites using a panel chosen isolates. Each site tested the isolates in triplicate on three different days using one lot of Phoenix panels containing the antimicrobial agents and associated reagents.

The results of the study demonstrate for each antimicrobial agent there was an overall intra-site reproducibility of greater than 90% and an overall inter-site reproducibility greater than 95% for the isolates tested.

{2}------------------------------------------------

Clinical Studies

Clinical, stock and challenge isolates were tested across multiple geographically diverse sites across the United States to demonstrate the performance of the Phoenix antimicrobial susceptibility test with the Phoenix panel formats containing antimicrobial agents. Phoenix System results for Challenge set isolates were compared to the expected results. Phoenix System results for clinical isolates were compared to the results obtained from the CLSI reference broth microdilution method.

The performance of the Phoenix System was assessed by calculating Essential Agreement (EA) and Category Agreement (CA) to expected/reference results for all isolates tested. Essential Agreement (EA) occurs when the BD Phoenix™ Automated Microbiology System agrees exactly or within ± one two-fold dilution to the reference result. Category Agreement (CA) occurs when the BD Phoenix™ Automated Microbiology System agrees with the reference method with respect to the FDA categorical interpretive criteria (susceptible, intermediate, and resistant).

Table 1 summarizes the performance for the isolates tested in this study.

Performance of BD Phoenix System by Drug Table 1:

CARDE STATE THE COLLECT OF COLLECT OF CLASS OF COLLECT OF CLASS OF COLLECT COLLEGION OF COLLECT COLLEGION OF COLLECTION OF COLLECTION OF COLLECTION OF COLLECTION OF COLLECTIOBIR C RIFFERENCE THE FELL I E EAntimicrobiaA MICH IS IN CONSUL OF BLIN BER EN LE W1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1oncentrationANN A CANADA CONSULACIONAL COLLECT OF THE COLLECT OF THE COLLECT OF THE CONTRACT OF THE CONTRACT OF THE CONTRACT OF THE CONTRACT OF THE CONTRACT OF THE CONTRACT OF THE CONTRA1 1 . 8 M 1 8100 100 M M M 8 80 M 1 80 000 1 80 0 211 20 - 211 21 - 2 1 1 - 2 1 - 2 1 - 2 1 - 2 1 - 2 1 - 2 1 - 2 1 - 2 1 - 2 1 - 2 1 - 2 1 - 2 1 - 2 1 - 2 1 - 2 1 - 2 1 - 21178.101940.100,000Company of Children & Religion on. 0 .
I am the states and any as a18.1ephalothin.ug/mAna *but in fort and------------------------------------------------------------------------------------------------------------------------------------------------------------------------------0000U ba0007.0- Par Philager Statist of Children Walls Andrew Mar

Conclusions Drawn from Substantial Equivalence Studies

The data collected from the substantial equivalence studies demonstrate that testing on the BD Phoenix™ Automated Microbiology System with this antimicrobial agent is substantially equivalent as outlined in the FDA draft guidance document, "Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA," February 5, 2003. Technological characteristics of this system are substantially equivalent to those used in the VITEK® system, which received approval by the FDA under PMA number N50510 and BD Phoenix™ Automated Microbiology System with Gatifloxacin (K020321, May 23, 2002), Ofloxacin (K020323, April 14, 2002), Levofloxacin (K020322, March 27, 2002),and Cephalothin (K031943, August 6, 2003).

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle-like symbol with flowing lines, representing the department's mission. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the symbol. The logo is in black and white.

Public Health Service

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

MAY 11 2006

Ms. Monica E. Giguere Regulatory Affairs Specialist BD Diagnostics Systems Becton, Dickinson and Company 7 Loveton Circle Sparks, MD 21152

Re: K060558 Trade/Device Name: BD Phoenix TM Automated Microbiology System Cephalothin (1-64 ug/mL) - Gram-Negative ID/AST or AST Regulation Number: 21 CFR 866.1645 Regulation Name: Fully Automated Short-Term Incubation Cycle Antimicrobial Susceptibility Devices Regulatory Class: Class II Product Code: LON Dated: May 2, 2006 Received: May 4, 2006

Dear Ms. Giguere:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).

{4}------------------------------------------------

Page 2 -

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240)276-0484. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html

Sincerely yours,

Sally a story

Sally A. Hojvat, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Page 1 of 1

510(k) Number: K060558

Device Name: BD Phoenix™ Automated Microbiology System for use with the antimicrobial agent Cephalothin (1-64 ug/mL) - Gram-negative ID/AST or AST only Phoenix panels.

Indications for Use:

The BD Phoenix™ Automated Microbiology System is intended for in vitro quantitative determination of antimicrobial susceptibility by minimal inhibitory concentration (MIC) of most Gram-negative aerobic and facultative anaerobic bacteria isolates from pure culture for Enterobacteriaceae and Non-Enteriaceae and most Gram-positive bacteria isolates from pure culture belonging to the genera Staphylococcus, Enterococcus, and Streptococcus.

This premarket notification is for additional Gram-negative organism groups and Cephalothin (1-64 ug/mL) on the BD Phoenix Automated Microbiology System.

Cephalothin has been shown to be active in vitro against most strains of microorganisms listed below. Because the drug insert for cephalothin (CF) was unavailable, the organisms that were included are those organisms listed in other sources such as "The Sanford Guide to Antimicrobial Therapy" for 1st generation cephalosporins and/or those organisms listed in the drug inserts for those drugs for which cephalothin is considered the predictor of activity per the CLSI M100-S15, Table 2C.

Active In Vitro and in Clinical Infections Against:

Escherichia coli Klebsiella pneumoniae Proteus mirabilis

Prescription Use (Per 21 CFR 801.109) Over-the-Counter Use

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Fuddlu lu. Poole
Division Sign Off

Division Sign-Off

Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) K060558

BD Diagnostic Systems Becton, Dickinson and Company

Page 6

§ 866.1645 Fully automated short-term incubation cycle antimicrobial susceptibility system.

(a)
Identification. A fully automated short-term incubation cycle antimicrobial susceptibility system is a device that incorporates concentrations of antimicrobial agents into a system for the purpose of determining in vitro susceptibility of bacterial pathogens isolated from clinical specimens. Test results obtained from short-term (less than 16 hours) incubation are used to determine the antimicrobial agent of choice to treat bacterial diseases.(b)
Classification. Class II (special controls). The special control for this device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA.”