K Number
K060458

Validate with FDA (Live)

Date Cleared
2006-10-12

(232 days)

Product Code
Regulation Number
876.1500
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Manzi Mach 1 Instrument Cleaner Processor System is indicated for use with the High Level Disinfectant MS10 concentrate (MEC 0.49% PAA, minimum contact temperature of 120°F for a contact time of 15 minutes) for cleaning and high level disinfecting flexible bronchoscopes used in health care settings by health care workers.

Device Description

The Manzi Mach 1 Instrument Cleaner-Processor System consists of a Manzi Mach 1 Instrument Cleaner-Processor, a proprietary Manzi germicide, MS10; and a proprietary Manzi Detergent, MD10.

The Manzi Mach 1 Instrument Cleaner-Processor is a self-contained stand-alone system of hardware and software designed to clean and provide high level disinfection of bronchoscopes using the MD10 detergent, the MS10 germicide, and a patented push-pull agitation system. The push-pull agitation system effectively scrubs the interior and exterior surfaces of the bronchoscope without the use of special connectors. The scope is placed in a processing chamber where it is exposed to a push-pull agitation cleaning cycle followed by two hot water rinses, a push-pull agitation disinfection cycle that provides high level disinfection (spore log reduction of > 10 ° microorganisms with no CFUs) of the device, and a final rinse with an ozonated sanitized water rinse.

The hardware for the Manzi Mach 1 Instrument Cleaner-Processor consists of a stainless steel processing chamber, a push-pull agitation pump, an ozonator, and a variety of components that are mounted in a movable covered frame. The cleaner-processor system utilizes accessories such as disposable water filters, reusable bronchoscope trays, and printer paper.

The Manzi Mach 1 Instrument Cleaner-Processor is designed to: ( ) be used in accordance with the reprocessing instructions provided in the operator's manual of the instruments being processed, and (2) facilitate the health care facility's compliance with reprocessing guidelines published by SGNA, APIC, AORN, ASGE, CDC, and other professional organizations.

MD10 is a low foaming enzyme chemical detergent packaged in single use containers for attachment to the Manzi Cleaner. MD10 is intended to be used with the Manzi Instrument Cleaner-Processor.

MS10 is a peracetic acid based liquid chemical germicide. MS10 is intended to be used with the Manzi Instrument Cleaner-Processor.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the Manzi Mach 1 Instrument Cleaner-Processor, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance:

The document outlines three main sets of qualification testing: FDA Guidance, EU Guidance, and Langford IC Systems (LIC) Requirements.

FDA Guidance (August 1993: "Guidance on Premarket Notification [510(k)] Submissions for Automated Endoscope Washer / Disinfectors, and Disinfectors Intended for Use in Health Care Facilities")

RequirementAcceptance Criteria (Implied)Reported Device Performance
Process Parameter TestsConformity to specified operational parametersPassed
Simulated Use TestsEffectiveness in a simulated environmentPassed
Cleaning EfficacyEffective removal of contaminantsPassed
Disinfection EfficacyEffective high-level disinfectionPassed
Rinsing EfficacyEffective removal of cleaning and disinfection agentsPassed
Other TestsNot explicitly detailed, but part of overall performancePassed
Combined ProcessEffectiveness of the entire cleaning and disinfection processPassed
In – Use TestsEffectiveness in real-world or near real-world conditionsPassed
Software DocumentationConformance to software documentation requirementsPassed
Toxicological Evaluation of ResiduesNo harmful residues remaining after processingPassed

EU Guidance (Draft prEN ISO 15883-1: 2003 and Draft prEN ISO 15883-4: 2001)

RequirementAcceptance Criteria (Implied)Reported Device Performance
Cleaning Efficacy - Scope Ninhydrin (prEN 15883-4 Annex B.1.1)Effective removal of protein from scopesPassed
Cleaning Efficacy - Scope Horse serum (prEN 15883-4 Annex B.1.1)Effective removal of organic load from scopesPassed
Cleaning Efficacy - Surrogate Ninhydrin (prEN 15883-4 Annex B.1.1)Effective removal of protein from surrogatesPassed
Cleaning Efficacy - Surrogate Horse serum (prEN 15883-4 Annex B.1.1)Effective removal of organic load from surrogatesPassed
Disinfection Efficacy - Scope Sheep blood (prEN 15883-4 Annex D)Effective high-level disinfection of scopes with organic loadPassed
Disinfection Efficacy - Surrogate Sheep blood (prEN 15883-4 Annex D)Effective high-level disinfection of surrogates with organic loadPassed

Langford IC Systems (LIC) Requirements

LIC Cleaning RequirementAcceptance CriteriaReported Device Performance
Cleaning Efficacy: Reduction of protein loading of scopes contaminated with a Protein Laden Soil to Remaining Protein levels (Ref: AAMI TIR30: 2003, A Compendium of Processes, Materials, Test Methods, and Acceptance Criteria for Cleaning Reusable Medical Devices).Remaining Protein levels of < 6.4 µg/cm²Remaining Protein levels of < 4.0 µg/cm²
High Level Disinfection Efficacy: Reduction of microbial loading of scopes with no colony forming units (CFUs)Reduction of > 10^6 microbial loading; with no CFUs≥ 6 spore log reduction; with no CFUs
LIC Microbiological Efficacy Tests
Simulated Use Test (15 min.) - ISO 15883-4 Surrogates, Sheep's blood soil (Test Organisms: Bacillus subtilis, Mycobacterium terrae, Candida albicans, Enterococcus faecium, Staphylococcus aureus)> 6 spore log reduction; no Colony Forming Units (CFU)> 6 spore log reduction; no CFU
Simulated Use Test (15 min.) - Olympus Bronchoscopes, Sheep's blood soil (Test Organism: Bacillus subtilis)> 6 spore log reduction; no CFU> 6 spore log reduction; no CFU
Simulated Use Test - Ozonated Water System (Test Organisms: Candida albicans, Staphylococcus aureus, Bacillus subtilis)> 6 spore log reduction; no CFU> 6 spore log reduction; no CFU
LIC Sanitized Ozonated Water System Requirement
Final Rinse System Efficacy: Reduction of microbial loading of scopes with no colony forming units (CFUs)Reduction of ≥ 10^6 microbial loading; with no colony forming units (CFUs)≥ 6 spore log reduction; with no CFUs

2. Sample Size Used for the Test Set and the Data Provenance:

The document does not explicitly state the numerical sample size for the test set (number of scopes, surrogates, or runs). Instead, it refers to broad categories of tests such as "Simulated Use Tests," "In-Use Tests," and testing on "Olympus Bronchoscopes" and "Surrogates".

The data provenance is not specified in terms of country of origin. The study appears to be retrospective in the sense that the test results are reported after the completion of the tests rather than being part of an ongoing prospective clinical trial.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts:

The document does not provide information on the number of experts used or their qualifications for establishing ground truth. The ground truth for effectiveness (cleaning, disinfection, organism reduction) is established by laboratory-based microbiological and chemical testing, not expert consensus in the diagnostic sense.

4. Adjudication Method for the Test Set:

This is not applicable to this type of device evaluation. Adjudication methods like 2+1 or 3+1 are typically used for interpreting diagnostic outputs where human readers might disagree. For device performance in cleaning and disinfection, the results are quantitative (e.g., protein levels, log reduction of microorganisms) and therefore do not require such adjudication.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

This is not applicable. The Manzi Mach 1 is an automated instrument cleaner-processor, not an AI-powered diagnostic or assistive tool for human readers. Therefore, an MRMC study and the concept of human reader improvement with AI are not relevant.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done:

Yes, the studies describe the standalone performance of the Manzi Mach 1 Instrument Cleaner-Processor combined with its proprietary chemicals (MS10 germicide and MD10 detergent). The device operates automatically to clean and disinfect the instruments, without human intervention in the core cleaning/disinfection process itself. The "Results" section in the tables reflects the performance of the automated system.

7. The Type of Ground Truth Used:

The ground truth used in these studies is primarily laboratory-based microbiological and chemical testing against established standards and validated methods. Examples include:

  • Microbiological assays: To determine log reduction of specific test organisms (e.g., Bacillus subtilis, Mycobacterium terrae, Candida albicans) to assess disinfection efficacy.
  • Chemical assays: To measure remaining protein levels (e.g., Ninhydrin test) to assess cleaning efficacy.
  • Biological indicator assessment: Counting Colony Forming Units (CFUs) to confirm complete elimination or reduction of microorganisms.
  • Adherence to recognized standards: FDA Guidance, EU ISO standards (prEN ISO 15883-1 and 15883-4), and AAMI TIR30: 2003.

8. The Sample Size for the Training Set:

The document does not mention a "training set" in the context of machine learning or AI. This is a traditional medical device pre-market notification (510(k)) for a physical cleaning and disinfection system. Therefore, the concept of a training set as used in AI development is not applicable here. The device itself is "trained" during its engineering and design phase, and validated through the qualification testing described.

9. How the Ground Truth for the Training Set Was Established:

As there is no "training set" in the AI sense, this question is not applicable. The device's performance parameters were likely established through iterative design, engineering, and testing phases to meet the specified performance metrics, but this is not analogous to establishing ground truth for an AI training dataset.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the letters "L/C" in a stylized font. The "/" is represented by a lightning bolt. The letters are black and white and have a textured appearance.

MANZI MACH 1 - INSTRUMENT CLEANER - PROCESSOR Pre-Market Notification 510(k) # K060458 Section II Tab 10 510(k) Summary

Applicant's Name, Address, Telephone, Fax, Contact Person

Langford IC Systems, Inc. 310 S. Williams Blvd. Suite 270 Tucson, Arizona 85711 Tel: 520-745-6201 Fax: 520-745-6286 E-mail: terrylangford(@)licsystems.com

Contact Person

Terry Langford President and CEO, Langford IC Systems, Inc. Tel: 520-745-6201 Fax: 520-745-6286 E-mail: terrylangford@licsystems.com

1.0 CLASSIFICATION, COMMON OR USUAL NAME, DEVICE NAME

  • Classification Name: Pending Class II a) Common / Usual Name: Endoscope and accessories Device Classification: 21 CFR § 876.1500, -Endoscope and accessories Proprietary Name: Manzi Mach 1 Instrument Cleaner- Processor System with MS10 High Level Disinfector
  • b) Classification Name: Pending - Class II Common / Usual Name: Endoscope and accessories Device Classification: 21 CFR & 876.1500. Proprietary Name: Manzi MS10

2.0 PREDICATE DEVICE

  • a) System 83 Plus™ Washer-Disinfector, K983017 Manzi Cleaner System, K043314 (Washer)
  • b) Steris 20 Sterilant, K875280

3.0 INDICATIONS FOR USE

The Manzi Mach 1 Instrument Cleaner Processor System is indicated for use with the High Level Disinfectant MS10 concentrate (MEC 0.49% PAA, minimum contact temperature of 120% for a contact time of 15 minutes) for cleaning and high level disinfecting flexible bronchoscopes used in health care settings by health care workers.

4.0 DESCRIPTION OF THE DEVICE

The Manzi Mach 1 Instrument Cleaner-Processor System consists of a Manzi Mach 1 Instrument Cleaner-Processor, a proprietary Manzi germicide, MS10; and a proprietary Manzi Detergent, MD10.

The Manzi Mach 1 Instrument Cleaner-Processor is a self-contained stand-alone system of hardware and software designed to clean and provide high level disinfection of bronchoscopes using the MD10 detergent, the MS10 germicide, and a patented push-pull agitation system. The push-pull agitation system effectively scrubs the interior and exterior surfaces of the bronchoscope without the use of special connectors. The scope is placed in a processing chamber where it is exposed to a push-pull

{1}------------------------------------------------

Image: LIC logo with lightning bolt
LANGFORD IC SYSTEMS, INC.

MANZI MACH 1 - INSTRUMENT CLEANER - PROCESSOR Pre-Market Notification 510(k) # K060458 Section II Tab 10 510(k) Summary

agitation cleaning cycle followed by two hot water rinses, a push-pull agitation disinfection cycle that provides high level disinfection (spore log reduction of > 10 ° microorganisms with no CFUs) of the device, and a final rinse with an ozonated sanitized water rinse.

The hardware for the Manzi Mach 1 Instrument Cleaner-Processor consists of a stainless steel processing chamber, a push-pull agitation pump, an ozonator, and a variety of components that are mounted in a movable covered frame. The cleaner-processor system utilizes accessories such as disposable water filters, reusable bronchoscope trays, and printer paper.

The Manzi Mach 1 Instrument Cleaner-Processor is designed to: ( ) be used in accordance with the reprocessing instructions provided in the operator's manual of the instruments being processed, and (2) facilitate the health care facility's compliance with reprocessing guidelines published by SGNA, APIC, AORN, ASGE, CDC, and other professional organizations.

MD10 is a low foaming enzyme chemical detergent packaged in single use containers for attachment to the Manzi Cleaner. MD10 is intended to be used with the Manzi Instrument Cleaner-Processor.

MS10 is a peracetic acid based liquid chemical germicide. MS10 is intended to be used with the Manzi Instrument Cleaner-Processor.

5.0 SUMMARY OF NONCLINICAL TESTS for the MANZI MACH 1 INSTRUMENT CLEANER-PROCESSOR

5.1 Qualification Testing - FDA Guidance

The Manzi Mach 1 Instrument Cleaner-Processor System was tested and found to conform with the requirements of the "Guidance on Premarket Notification [510(k)] Submissions for Automated Endoscope Washer / Disinfectors, and Disinfectors Intended for Use in Health Care Facilities", dated August, 1993". The table below identifies the qualifications performed and the results obtained:

RequirementResults
G.Performance Data
1 Process Parameter TestsPassed
2 Simulated Use TestsPassed
2.c. Effectiveness Tests
2.c.(1) Cleaning EfficacyPassed
2.c.(2) Disinfection EfficacyPassed
2.c.(3) Rinsing EfficacyPassed
2.c.(4) Other TestsPassed
2.c.(5) Combined ProcessPassed
3. In – Use TestsPassed
H.Software DocumentationPassed
I.Toxicological Evaluation of ResiduesPassed

5.2 Qualification Testing - EU Guidance

10/12/06

{2}------------------------------------------------

The Manzi Mach 1 Instrument Cleaner-Processor System was also tested and found to conform with the requirements of the Draft prEN ISO 15883-1: 2003, Washer-disinfectors - Part 1: General Requirements, Definitions and Tests and Draft prEN ISO 15883-4: 2001, Washer-disinfectors - Part 4: Requirements and Tests for Washer-Disinfectors Employing Chemical Disinfection for Thermo-Labile Endoscopes as identified in the Table below.

prEN ISO 15883-1: 2003 RequirementResults
6.10Annex BAnnex ECleaning Efficacy - Scope NinhydrinHorse serum- prEN 15883-4 Annex B.1.1Cleaning Efficacy - Surrogate NinhydrinHorse serum- prEN 15883-4 Annex B.1.1Surrogate - prEN 15883-4PassedPassed
6.11Annex DDisinfection Efficacy - ScopeSheep blood - prEN 15883-4 Annex DDisinfection Efficacy - SurrogateSheep blood - prEN 15883-4 Annex DPassedPassed

5.3 Qualification Testing - Langford IC Systems (LIC) Requirements

The Manzi Mach 1 Instrument Cleaner-Processor System was also tested and found to conform to the LIC requirements identified in the table below.

LIC Cleaning RequirementResults
Cleaning Efficacy: Reduction of protein loading of scopescontaminated with a Protein Laden Soil to Remaining Proteinlevels of < 6.4 µg/cm² (Ref: AAMI TIR30: 2003, ACompendium of Processes, Materials, Test Methods, andAcceptance Criteria for Cleaning Reusable Medical Devices.).PassedRemaining Proteinlevels of < 4.0 µg/cm²
LIC High Level Disinfection RequirementResults
High Level Disinfection Efficacy: Reduction of > $10^6$ microbialloading of scopes with no colony forming units (CFUs)Passed≥ 6 spore log reductionwith no CFUs
LIC Microbiological Efficacy Tests
Test MethodTest OrganismsResults
Simulated Use Test (15min.)ISO 15883-4 SurrogatesSheep's blood soilBacillus subtilisMycobacterium terraeCandida albicansEnterococcus faeciumStyphlococcus aureus> 6 spore log reduction;no Colony FormingUnits (CFU)
Simulated Use Test (15min.)Olympus BronchoscopesSheep's blood soilBacillus subtilis> 6 spore log reduction;no CFU
Simulated Use TestOzonated Water SystemCandida albicans, Staphylococcusaureus, Bacillus subtilis> 6 spore log reduction;no CFU

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Langford IC Systems. The logo consists of the text "LANCFORD IC SYST" in a stylized font. Above the text is a graphic element that appears to be a stylized checkmark or abstract design. The text and graphic are in black and white.

MANZI MACH 1 – INSTRUMENT CLEANER - PROCESSOR Pre-Market Notification 510(k) # K060458 Section II Tab 10 510(k) Summary

LIC Sanitized Ozonated Water System RequirementResults
Final Rinse System Efficacy: Reduction of ≥ 106 microbialloading of scopes with no colony forming units (CFUs)Passed≥ 6 spore log reductionwith no CFUs

6.0 SUMMARY OF NONCLINICAL TESTS for the MANZI MS10

5.1 Qualification Testing - FDA Guidance

The Manzi MS10 germicide was tested to and met the requirements of the current edition of "Guidance for Industry and FDA Reviewers, Content and Format of Premarket Notification [510(k)] Submissions for Liquid Chemical Sterilants / High Level Disinfectants'?, doted January 3, 2000. The table below identifies the qualifications performed and the results obtained:

RequirementResults
5.4Potency TestPassed
5.5Simulated Use TestsPassed
5.6In-Use TestsPassed
6.0BiocompatibilityPassed
Microbiological Efficacy Summary
Test MethodTest OrganismsResults
Sporicidal Activity of Sterilants;AOAC Official Method 966.04Bacillus subtilisClostridium sporogenes> 6 spore log reduction;No CFUs
Fungicidal Activity of Sterilants;AOAC Official Method 955.17TrichophytonmentagrophytesMS10 is fungicidal
Use-Dilution Method; AOACOfficial Method 955.14, 955.15,964.02Salmonella choleraesuisStaphylococcus aureusPseudomonas aeruginosaMS10 is bactericidal
Virucidal TestingPoliovirus Type 1MS10 is virucidal
Quantitative Tuberculocidal TestMycobacterium bovisMS10 is tuberculocidal

7.0 OVERALL PERFORMANCE CONCLUSIONS

The studies demonstrate that the Manzi Mach 1 Instrument Cleaner-Processor System is safe and effective for the cleaning and high level disinfection of bronchoscopes within the stated indications for use for the Manzi Mach 1 Instrument Cleaner-Processor, the Manzi MS10 germicide, and the Manzi Detergent, MD10, and establishes substantial equivalence of the Manzi Mach 1 Instrument Cleaner Processor System to the predicate devices identified in Section 2.0 above.

{4}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/4/Picture/1 description: The image shows the seal of the Department of Health & Human Services, USA. The seal features a stylized eagle with three stripes forming its wing, symbolizing service, science, and security. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES, USA" is arranged in a circular pattern around the eagle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

OCT 1 2 2006

Langford IC Systems, Incorporated C/O Jimmie T. Johnston, Ph.D. Vice President of Regulatory and Clinical Affairs Proven Process Medical Devices 141 Washington Street East Walpole, Massachusetts 02032

Re: K060458

Trade/Device Name: Manzi Mach 1 Instrument Cleaner-Processor System Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: II Product Code: FEB Dated: September 19, 2006 Received: September 20, 2006

Dear Dr. Johnston:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{5}------------------------------------------------

Page 2 -- Dr. Johnston

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Chiu Lin, Ph.D.

Chiu Lin. Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

Indications for Use

510(k) Number: K060458

Device.Name: Manzi Mach 1 Instrument Cleaner - Processor System with MS10 High Level Disinfector

Indications for Use:

The Manzi Mach 1 Instrument Cleaner Processor System is indicated for use with the High Level Disinfectant MS10 concentrate (MEC 0.49% PAA, minimum contact temperature of 120°F for a contact time of 15 minutes) for cleaning and high level disinfecting flexible bronchoscopes used in health care settings by health care workers.

Prescription Use (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use × (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Clerz

on Sign-Off) ion of Anesthesiology, General Hospital. con Control, Dental Devices

) Number: K060458

§ 876.1500 Endoscope and accessories.

(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.