K Number
K060351

Validate with FDA (Live)

Device Name
MEDTOX OXYCODONE
Date Cleared
2006-05-12

(88 days)

Product Code
Regulation Number
862.3650
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The MEDTOX® OXYCODONE Test System uses immunochromatographic test strips for the rapid, qualitative detection of oxycodone in human urine. It is intended for prescription point-of-care use including physician office laboratories and central laboratory settings. It is also intended for workplace settings, criminal justice or forensic settings, and drug rehabilitation centers. MEDTOX® OXYCODONE is not for over-the-counter sale. The test detects oxycodone at concentrations 100 ng/mL and above. THE MEDTOX® OXYCODONE PROVIDES ONLY A PRELIMINARY ANALYTICAL TEST RESULT. A MORE SPECIFIC ALTERNATE CHEMICAL METHOD MUST BE USED IN ORDER TO OBTAIN A CONFIRMED ANALYTICAL RESULT. GAS CHROMATOGRAPHY/MASS SPECTROMETRY (GC/MS) IS THE PREFERRED CONFIRMATORY METHOD. CLINICAL CONSIDERATION AND PROFESSIONAL JUDGMENT SHOULD BE APPLIED TO ANY DRUG OF ABUSE TEST RESULT.

Device Description

Each test strip contains antibody colloidal gold, a drug conjugate and a control line. A mouse monoclonal antibody specific to oxycodone is mixed with colloidal gold and applied to the sample well pad of the strip. Drug is conjugated to protein and immobilized at the test line. Strips have an anti-mouse immunoglobulin antibody immobilized at the control line. The anti-mouse antibody binds the mouse antibody coated on the colloid gold, When urine is applied to the sample well of the device, the dried antibody-colloidal gold on the sample pad dissolves and the urine wicks up the white test strip carrying the red antibody-colloidal gold with it.

AI/ML Overview

MEDTOX® OXYCODONE Test System: Acceptance Criteria and Performance Study Summary

This document describes the acceptance criteria and performance study results for the MEDTOX® OXYCODONE Test System, an immunochromatographic test strip for the rapid, qualitative detection of oxycodone in human urine. The information is extracted from the 510(k) summary K060351.

1. Table of Acceptance Criteria and Reported Device Performance

Performance CharacteristicAcceptance Criteria (Implied)Reported Device Performance
Precision/Reproducibility
75% below cutoff (25 ng/mL)Consistent negative results100% negative results
50% below cutoff (50 ng/mL)Majority negative results93% negative results
At cutoff (100 ng/mL)Majority positive results (or clear distinction at cutoff)93% positive results
25% above cutoff (125 ng/mL)Consistent positive results98% positive results
50% above cutoff (150 ng/mL)Consistent positive results100% positive results
Point-of-Care (POC) PrecisionSimilar performance to in-house professional study"Similar to the professional study" for most levels; consistent at 0 and 150 ng/mL
Analytical Specificity(Cross-Reactivity)Oxycodone: Positive at 100 ng/mL (100% reactivity) Oxymorphone: Significant cross-reactivity expected (compared to predicate) Other opiates/compounds: Low or no cross-reactivityOxycodone: Positive at 100 ng/mL (100% reactivity) Oxymorphone: Positive at 200 ng/mL (50% reactivity) Codeine: Positive at 5,000 ng/mL (2%) Ethylmorphine: Positive at 5,000 ng/mL (2%) Dihydrocodeine: Positive at 10,000 ng/mL (1%) Hydrocodone: Positive at 75,000 ng/mL (<1%) Hydromorphone: Positive at 50,000 ng/mL (<1%) Morphine: Positive at 50,000 ng/mL (<1%) Naloxone: Positive at 50,000 ng/mL (<1%) Norcodeine: Positive at 100,000 ng/mL (<1%) Many others: Negative at 50,000-100,000 ng/mL (<1%)
pH InterferenceNo interference with results from pH values 4.0-9.0All pH levels (4.0-9.0) gave negative results at 25 ng/mL and positive results at 150 ng/mL
Specific Gravity InterferenceNo interference with results from specific gravity 1.003-1.035All specific gravity levels (1.003-1.035) gave negative results at 25 ng/mL and positive results at 150 ng/mL
Common Drugs InterferenceNo interference with results from common OTC/prescription drugsNone of the tested common drugs (e.g., Acetaminophen, Ibuprofen, Caffeine) affected expected results at 100 µg/mL
Method Comparison (vs. GC/MS)High agreement with GC/MS for both positive and negative samples, particularly around the cutoff.Agreement among positives: 96% Agreement among negatives: 97%
Comparison to Predicate DeviceEquivalent Indications for Use, ability to detect true negative and true positive samples, detect primary metabolite (oxymorphone), and limited cross-reactivity with other opiates.Equivalent Indications for Use. Ability to detect true negative samples (100%). Ability to detect true positive samples (+50% above cut-off) (98%). Cross-reactivity to oxymorphone (50% vs. 103% for predicate). Cross-reactivity to other opiate compounds (2% or less vs. <1% for predicate).

2. Sample Size Used for the Test Set and Data Provenance

  • Precision/Reproducibility (Professional Study):
    • Sample Size: Drug-free urines spiked to 6 concentrations (0, 25, 50, 75, 100, 125, 150 ng/mL). Tested in triplicate on 6 different occasions, resulting in a total of 18 tests per concentration level. (The table shows "Number Tested" as 1, which appears to be a typo and likely refers to the number of panels or a mislabeling, given the description of triplicate testing on 6 occasions).
    • Data Provenance: Retrospective, conducted in-house by MEDTOX employees. Origin: likely USA (Burlington, NC is applicant's address).
  • Point-of-Care (POC) Precision:
    • Sample Size: 6 concentration levels (0, 25, 50, 100, 125, 150 ng/mL). Each operator tested 5 replicates per level. With 9 operators across 3 sites, this translates to 45 tests per concentration level per site (15/15/15 listed in the table is likely for the number of positive/negative results per site, not the total tested). Assuming 5 replicates, this would be 5 replicates * 3 sites * 3 operators = 45 tests per concentration. The table shows 15/15/15 under "Number Tested" for each site, implying 15 tests per concentration per site. So, 15 tests * 6 concentrations * 3 sites = 270 total tests.
    • Data Provenance: Prospective, conducted by 9 POC operators at 3 different sites. Origin: not explicitly stated beyond "POC sites," likely USA.
  • Analytical Specificity (Cross-Reactivity and Non-Reactive Compounds):
    • Sample Size: Each compound evaluated in triplicate. Total number of compounds not explicitly stated, but includes 19 cross-reactive compounds listed in the table and 11 non-cross-reactive endogenous compounds listed. An "extensive list" of unrelated compounds (prescription/OTC) was also evaluated.
    • Data Provenance: Retrospective, conducted by in-house operators. Origin: likely USA.
  • Interference (pH, Specific Gravity, Common Drugs):
    • Sample Size: Each condition (pH level, specific gravity level, common drug) evaluated in triplicate. pH: (6 levels * 3 replicates) * 2 concentrations = 36 tests. Specific Gravity: (8 levels * 3 replicates) * 2 concentrations = 48 tests. Common Drugs: Numerous drugs (14 listed) * 3 replicates * 2 concentrations (25% and 150% of cutoff) = 84 tests.
    • Data Provenance: Retrospective, conducted by in-house operators. Origin: likely USA.
  • Method Comparison (vs. GC/MS):
    • Sample Size: Total of (0+2+2+6+37) + (103+5+4+1+1) = 68 positive samples (by candidate device/GC/MS positive) + 114 negative samples (by candidate device/GC/MS negative) = 182 clinical urine samples.
    • Data Provenance: Retrospective clinical urine samples obtained from MEDTOX Laboratories. Screened using DRI Oxycodone 100 ng/mL cutoff Assay (predicate device) and confirmed by GC/MS for positive samples. Origin: likely USA.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

  • Precision/Reproducibility: 3 MEDTOX employees who did not develop the test. Qualifications are not specified beyond being "MEDTOX employees." This was an internal study.
  • Point-of-Care (POC) Studies: 9 POC operators at 3 different sites. Qualifications are specified as individuals with "a minimum of a high school education who also satisfy specific training and certification guidelines" (demonstration of test, use of QA samples, proper technique, basic understanding of results, sample preparation for confirmation, review of training program, 80% on written exam).
  • Analytical Specificity and Interference Studies: In-house operators. Qualifications are not specified.
  • Method Comparison:
    • Ground Truth Establishment: GC/MS (Gas Chromatography/Mass Spectrometry) was used as the ground truth. This is a highly accurate and generally accepted confirmatory method for drug testing.
    • Experts for GC/MS: The document doesn't explicitly state the number or qualifications of experts operating the GC/MS. However, GC/MS results are considered an objective and analytical gold standard, typically operated by trained analytical chemists or lab technicians specializing in toxicology.

4. Adjudication Method for the Test Set

  • Precision/Reproducibility, Analytical Specificity, Interference: No explicit adjudication method described beyond raw results generated by the operators.
  • Method Comparison: No explicit adjudication method for the comparison between the candidate device and GC/MS. The comparison is direct against the GC/MS results, which serve as the reference standard. The results are presented as counts of agreement/disagreement with GC/MS.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size

No MRMC comparative effectiveness study was explicitly performed comparing human readers with AI vs. without AI assistance. This device is a rapid test strip intended for visual interpretation by trained operators, not an AI-assisted diagnostic tool.

6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

Yes, the precision, analytical specificity, and interference studies effectively represent standalone performance as the device's output (positive/negative line on the strip) is visually interpreted by the operator, but the inherent analytical performance of the strip itself is being evaluated under controlled conditions. The "Method Comparison" study also evaluates the standalone performance of the device against the GC/MS reference.

7. The Type of Ground Truth Used

  • Precision/Reproducibility, Analytical Specificity, Interference: Spiked drug-free urine samples with known concentrations of oxycodone or interfering substances. This is a form of analytical truth (known concentration).
  • Method Comparison: GC/MS (Gas Chromatography/Mass Spectrometry) results. This is considered the gold standard analytical confirmatory method for drug detection.

8. The Sample Size for the Training Set

This submission does not describe a "training set" in the context of machine learning. The device is an immunochromatographic assay, not an AI/ML algorithm. The performance studies described are for validation of the ready-to-use device.

9. How the Ground Truth for the Training Set Was Established

Not applicable, as there is no training set in the context of AI/ML for this device.

{0}------------------------------------------------

K060351

Image /page/0/Picture/1 description: The image shows the logo for MEDTOX Diagnostics, Inc. The logo is in black and features the word "MEDTOX" in large, bold letters, with a registered trademark symbol to the right of the "X". Below "MEDTOX" is the phrase "DIAGNOSTICS, INC." in smaller, sans-serif font.

510(k) Summary

Purpose for Submission:Traditional 510(k) for New Assay
Analyte:Oxycodone
Applicant:MedTox Diagnostics, Inc.1238 Anthony RoadBurlington, NC 27215Telephone: 1-800-334-1116FAX: 1-336-229-4471Establishment Number: 1050155MAY 12 2006
Contact Person:Phillip HartzogAssociate Director, Research and DevelopmentTelephone: 1-336-226-6311 ext 230Cell Phone: 336-953-6576E-mail: phartzog@medtox.com
Date Prepared:February 10, 2006
Device Trade Name:MEDTOX® OXYCODONE
Regulatory Information:862.3650 Enzyme Immunoassay, Opiates
Classification:Class II
Product Code:DJG
Panel:91 (Toxicology)
Predicate Device:DRI Oxycodone Assay, 100 ng/mL Cutoff510(k) Number: K040411

Indications for Use

The MEDTOX® OXYCODONE Test System uses immunochromatographic test strips for the rapid, qualitative detection of oxycodone in human urine. It is intended for prescription point-of-care use including physician office laboratories and central laboratory settings. It is also intended for workplace settings, criminal justice or

{1}------------------------------------------------

forensic settings, and drug rehabilitation centers. MEDTOX® OXYCODONE is not for over-the-counter sale.

Workplace operators that may use this device are defined as individuals with a minimum of a high school education who also satisfy the following training and certification guidelines:

(1)Training should be conducted by a qualified professional and include a demonstration of the MEDTOX® OXYCODONE test system and (2) the use of quality assurance samples for monitoring and confirming the performance of the test system. Trainers should observe and confirm that the operator (3) uses proper technique when running a test sample and quality assurance samples. (4) has a basic understanding of test results, including the potential for false positive and false negative results, (5) knows how to prepare a sample for shipment to the laboratory for confirmation testing, (6) has reviewed the information contained in the MEDTOX® Training and Certification Program (available at www.medtox.com) and (7) that the operator minimally achieves a score of 80% on the written exam provided by MEDTOX®.

Operators achieving a score of 80% will be provided with a certificate of training participation. Quality assurance samples appropriate for training are available from MEDTOX® Diagnostics, Inc. Additionally, MEDTOX® Technical Support will provide access to assistance from individuals who are experienced in the interpretation of drug testing results.

The test detects oxycodone at concentrations 100 ng/mL and above.

THE MEDTOX® OXYCODONE PROVIDES ONLY A PRELIMINARY ANALYTICAL TEST RESULT. A MORE SPECIFIC ALTERNATE CHEMICAL METHOD MUST BE USED IN ORDER TO OBTAIN A CONFIRMED ANALYTICAL RESULT. GAS CHROMATOGRAPHY/MASS SPECTROMETRY (GC/MS) IS THE PREFERRED CONFIRMATORY METHOD. CLINICAL CONSIDERATION AND PROFESSIONAL JUDGMENT SHOULD BE APPLIED TO ANY DRUG OF ABUSE TEST RESULT.

Test Principle

Each test strip contains antibody colloidal gold, a drug conjugate and a control line. A mouse monoclonal antibody specific to oxycodone is mixed with colloidal gold and applied to the sample well pad of the strip. Drug is conjugated to protein and immobilized at the test line. Strips have an anti-mouse immunoglobulin antibody immobilized at the control line. The anti-mouse antibody binds the mouse antibody coated on the colloid gold,

When urine is applied to the sample well of the device, the dried antibody-colloidal gold on the sample pad dissolves and the urine wicks up the white test strip carrying the red antibody-colloidal gold with it.

{2}------------------------------------------------

Interpretation of Results

Negative: When no drug is present in the urine sample, the red antibody-colloidal gold migrates up the test strip and binds to the drug conjugate immobilized on the membrane. The binding of the antibody-colloidal gold to the drug conjugate generates a line at the test line position on the device.

Positive: When drug is present in the sample the antibody-colloidal gold binds the drug before it migrates up the test strip. However, when the antibody-colloidal gold binds the drug in the urine, the antibody-colloidal gold can not bind to the drug coniugate immobilized on the test strip. When the drug concentration is at or above the cutoff concentration, the majority of the antibody-colloidal gold is bound to the drug in the urine. Therefore, as the drug bound antibody-colloidal gold migrates up the test strip it is unable to bind to the drug conjugate immobilized on the membrane. Therefore no line is generated at the test line position on the device.

Control Line: The test strip has an internal procedural control. A control line forms when the antibody-colloidal gold binds to the anti-mouse immunoglobulin antibody immobilized on the membrane at the control line position on the device. A line must form at the control line position on the device to indicate that there was an adequate volume of sample, the reagents migrated properly, and that the test strip is intact.

Summary of MEDTOX® OXYCODONE Performance Testing

1. Analytical performance:

a. Precision/Reproducibility:

Performance around the cutoff concentration was evaluated by testing drugfree urines spiked with standard solutions. Drug free urine was also tested. Testing was performed in triplicate on 6 different occasions by 3 MEDTOX employees who did not develop the test. Samples were randomized and presented as a blind panel.

Data from this study is presented below. The test consistently gave negative results (100%) for natural urine spiked to 75% below the cutoff (25 ng/mL). and negative results the majority of the time (93%) for 50% below the cutoff (50 ng/mL). The test consistently gave positive results for natural urine spiked above the cut-off (98% 125 ng/mL and 100% for 150 ng/mL), and positive results the majority of the time (93%) at the cutoff. Most of the variability in test results occurred at 25% below the cutoff (75 ng/mL), and here the results were as variable between replicates of the sample as between operators.

A Property and Property of ChildrenA C------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------Oxycodone Cutoff = 100 ng/mL------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Conc.na/mNumber Tested1Positive------
BURICAL------------------------------------------------------------------------------------------------------------------------------------------------------------------------------A C - Children Charles on A B A B A B A B A B A B A B - A B - A B - A - A - A - A - A - A - A - A - A - A - A - A - A - A - A - A - A - A - A - A - A - A - A - A - A - A - A
A COLLECT OF CONSULERS40-44-4-4
-------------------------------------CORPORACIALA LAND BLO------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

{3}------------------------------------------------

------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
.Autober 4000ి విశ
---------------------------------------1400***Acres Annual A- Children:A PROPERTY
446CALL CONA MINE AN1 APt------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ﻟﺴﺎﻋﺔ ﺍﻟﻤﺴﺎﺣﺔ ﺍﻟﻤﺴﺎﺣﺔ ﺍﻟﻤﺴﺎﺣﺔ ﺍﻟﻤﺴﺎﺣﺔ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺎﺣﺔ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺎﺣﺔ ﺍﻟﻤﺴﺘﻮﻯ1
--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------t.-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------4-4-400-40.***--------1

POC Studies:

Precision studies were performed by 9 POC operators at 3 different sites using randomized blind samples. Each operator tested 5 replicates of the 6 levels below. The results obtained from each of the 3 sites (Site1, Site2, Site3) are listed below. Results at 75% below the cutoff and 50% above the cutoff were consistent and identical to the results with MEDTOX employees. Results at the other levels were similar to the professional study.

Oxycodone Cutoff = 100 ng/mL
Number TestedPositiveNegative
Conc. (ng/mL)Site1/ Site2/ Site3Site1/ Site2/ Site3Site1/ Site2/ Site3
015/15/150/0/015/15/15
2515/15/150/0/015/15/15
5015/15/152/0/113/15/14
10015/15/1515/12/120/3/3
12515/15/1515/13/140/2/1
15015/15/1515/15/150/0/0

b. Linearity/assay reportable range:

Not applicable. The assay is intended for qualitative use.

  • c. Traceability, Stability, Expected values (controls, calibrators, or methods):
    No calibrators are required. The device is calibrated during the manufacturing process.

Controls are required but are not specified in the labeling. Users are instructed to follow federal. state and local guidelines concerning QC practices.

Traceability is not covered in the submission.

  • d. Detection Limit:
    This information appears in the precision section above.

  • e. Analytical specificity:

Cross-Reactivity:

The following metabolites and compounds were initially dissolved in appropriate solvents and then added at varying concentrations to drug-free urine for evaluation. Samples were evaluated in triplicate by in-house operators. Results are expressed as the minimum concentration of metabolite

{4}------------------------------------------------

or compound required to produce a positive test result. Percent cross reactivity of a compound is calculated by dividing the cutoff concentration by the minimum concentration required to obtain a positive result and then multiplying by 100%.

Oxycodone, cutoff = 100 ng/mLResult% Cross-Reactivity
Positive at 100 ng/mL100%
ApomorphineNegative at 100,000 ng/mL< 1%
CodeinePositive at 5,000 ng/mL2%
DiacetylmorphineNegative at 100,000 ng/mL< 1%
DihydrocodeinePositive at 10,000 ng/mL1%
EthylmorphinePositive at 5,000 ng/mL2%
HydrocodonePositive at 75,000 ng/mL< 1%
HydromorphonePositive at 50,000 ng/mL< 1%
LevorphanolNegative at 50,000 ng/mL< 1%
MorphinePositive at 50,000 ng/mL< 1%
6-MonoacetylmorphineNegative at 100,000 ng/mL< 1%
Morphine 3-B-D-GlucuronideNegative at 100,000 ng/mL< 1%
Morphine 6-B-D-GlucuronideNegative at 100,000 ng/mL< 1%
NalorphineNegative at 100,000 ng/mL< 1%
NaloxonePositive at 50,000 ng/mL< 1%
NaltrexoneNegative at 100,000 ng/mL< 1%
NorcodeinePositive at 100,000 ng/mL< 1%
OxymorphonePositive at 200 ng/mL50%
ThebaineNegative at 100,000 ng/mL< 1%

Non Cross-reactive Endogenous Compounds:

Listed compounds were initially dissolved in appropriate solvents and then added to drug-free urine for evaluation with the MEDTOX® OXYCODONE test. Most of the compounds were evaluated for reactivity with the test at 100 ug/mL (albumin was evaluated at 20 mg/mL and bilirubin was evaluated at 200 µg/mL). Samples were evaluated in triplicate by in-house operators. The listed compounds gave negative results with the MEDTOX® OXYCODONE test.

  • Acetaldehvde Acetone Albumin, Human Bilirubin Cholesterol
    Creatinine Epinephrine ß-Estradiol Estriol Glucose Std. Solution

Hemoglobin, Human Sodium Chloride Tetrahydrocortisone d, 1-Thyroxine Uric Acid

Unrelated Compounds, Prescription and Over-the-Counter Medications:

An extensive list of compounds was evaluated by dissolving each compound in appropriate solvents and adding it to a drug-free urine for evaluation. Most compounds were evaluated for reactivity at 100,000 ng/mL. Samples were evaluated in triplicate by in-house operators and the list of compounds evaluated appears in the package insert. Compounds that demonstrated reactivity are included in the Related and Reactive Compounds section of the insert, as well as the table above.

{5}------------------------------------------------

Interference:

pH and Specific Gravity:

The test was assayed with six negative clinical samples with pH values of 4.0, 5.0, 6.0, 7.0, 8.0 and 9.0 ± 0.1. Each sample was assayed in triplicate. The pH samples were fortified with oxycodone to the concentrations of 25 ng/mL and 150 ng/mL. All the pH levels gave negative results when fortified to 25 ng/mL, and all pH levels gave positive results when fortified to 150 ng/mL

The test was assayed with eight samples with specific gravity values of 1.003, 1.005, 1.010, 1.015, 1.020, 1.025, 1.030 and 1.035 ± 0.001. Each sample was assayed in triplicate. The specific gravity samples were fortified with oxycodone to the concentrations of 25 ng/mL and 150 ng/mL. All the specific gravity levels gave negative results when fortified to 25 ng/mL, and all specific gravity levels gave positive results when fortified to 150 ng/mL.

Common Drugs:

Following the study of M.L. Smith, et. al. (Journal of Analytical Toxicology. Volume 24:7. October 2000, pages 522-529) drug free urine samples were spiked with the targeted drugs to the concentrations of 25% and 150% of the cutoff concentrations. 100 µg/mL of the common drugs were then added to the preparation and assayed by the MEDTOX® OXYCODONE test. Samples were evaluated in triplicate by in-house operators. None of the common progs listed in the following table affected the expected results.

Acetylsalicylic AcidChlorpheniramineIbuprofen
AcetaminophenCocaineMorphine
Brompheniramine maleateDextromethorphanPhenobarbital
CaffeineDiphenylhydantoind-Pseudoephedrine
CarbamazepineDoxylamineSalicylic Acid

COMMON DRUGS EVALUATED WITH MEDTOX® OXYCODONE TESTS

f. Assay cutoff:

The assay cutoff is 100 ng/mL, the same as the predicate device.

2. Comparison studies:

a. Method comparison studies:

A method comparison study was performed at three Point of Care sites Performance was evaluated by assaying a panel of blind coded clinical urine samples containing varying concentrations of oxycodone. Results were then compared to GC/MS results. Samples were obtained from MEDTOX Laboratories, where they were screened using DRI Oxycodone 100 ng/mL cutoff Assay, the predicate device. Samples screened positive were confirmed by GC/MS. Samples screened negative were not confirmed. The GC/MS determination included oxycodone and oxycodone and a weighted

{6}------------------------------------------------

concentration using 100% cross-reactivity for oxycodone and a 50% crossreactivity for oxymorphone was calculated. To obtain samples around and below the cutoff, positive samples were diluted in negative urine. These diluted samples were then assayed by GC/MS and those values used to allocate samples into the categories displayed below. Testing was performed by Point of Care personnel. Test results for MEDTOX® OXYCODONE ven GC-MS are tabulated below.

CandidateDeviceResultsNegative byImmunoassayPredicateDeviceConcentrationof up to thecutoff -50%Near CutoffNegative(Between50% belowthe cutoff andthe cutoffconcentration)Near CutoffPositive(Between thecutoff and50% abovethe cutoffconcentration)High Positive(greater than50% abovethe cutoffconcentration)
Positive022637
Negative1035411

Candidate Device Results vs. stratified GC/MS Values

% Agreement among positives is 96%

% Agreement among negatives is 97%

b. Matrix comparison:

Not applicable. The assay is intended for human urine samples only.

    1. Clinical studies:
      Not applicable. Clinical studies are not typically submitted for this device type and matrix.

Comparison of Predicate Device and MEDTOX® OXYCODONE

The chosen predicate device is DRI Oxycodone Assay with 100 ng/mL cutoff (510 (k) number K040411). This assay is routinely run by MEDTOX Laboratories to screen urine samples for oxycodone, allowing MEDTOX Diagnostics ready access to samples screened with the predicate device. The table below compares characteristics of the DRI Oxycodone Assay to the MEDTOX® OXYCODONE test.

CharacteristicPredicate DeviceDRI Oxycodone AssayMEDTOX®OXYCODONE
Type of AssayCompetitive enzymeimmunoassayCompetitiveimmunochromatographictest strip
ProcedureAssay run on instrument(Roche Modular System)One step screeningassay read visually
Mechanism ofDetectionAntibody binding toenzyme (G6PDH)modified with drugAntibody gold colloidcomplex binding to BSAmodified with drug

{7}------------------------------------------------

CharacteristicPredicate DeviceDRI Oxycodone AssayMEDTOX®OXYCODONE
derivativederivative
Test CutoffConcentration100 ng/mL100 ng/mL
Ability to detect truenegative samples( 100% 100%
Ability to detect truepositive samples(+50% above cut-off)100%98%
Crossreactivity toprimary metaboliteoxymorphone103%50%
Crossreactivity toother opiatecompounds<1%2% or less

While the predicate device is an enzymatic test dependent on an instrument, the basic mechanism of detection of oxycodone is the same. Free drug in the urine sample competes with drug conjugated to protein for binding to a drug specific antibody. This basic mechanism and specified cutoff level supports comparison between these two devices for substantial equivalence. Additionally all positive samples were tested against the reference procedure of GC/MS.

Conclusion

Indication of use and test performance support substantial equivalence for MEDTOX® OXYCODONE to DRI Oxycodone 100ng/mL Assay. Both devices have equivalent Indications for Use. Both devices have the ability to detect true negative samples, samples with oxycodone below the limit of detection for GC/MS. Both devices have the ability to detect true positive samples, samples with levels of oxycodone 50% above the cutoff or greater. Both devices have the ability to detect oxycodone's primary metabolite oxymorphone. And both devices have limited crossreactivity with other opiate compounds. These four characteristics define the desired performance for a rapid screen assay.

{8}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/8/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS). The seal features a stylized eagle with outstretched wings, symbolizing protection and care. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES.USA" are arranged in a circular pattern around the eagle, indicating the department's name and national affiliation. The seal is presented in black and white.

Public Health Service

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Mr. Phillip Hartzog Associate Director, Research and Development MedTox Diagnostics, Inc. 1238 Anthony Road Burlington, NC 27215

MAY 12 2006

Re: K060351

Trade/Device Name: MEDTOX® OXYCODONE test system Regulation Number: 21 CFR§862.3650 Regulation Name: Opiate test system Regulatory Class: Class II Product Code: DJG Dated: February 10, 2006 Received: February13, 2006

Dear Mr. Hartzog:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).

{9}------------------------------------------------

Page 2 -

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240) 276-0484. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Alberto G

Alberto Gutierrez, Ph.D. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{10}------------------------------------------------

Indications for Use

510(k) Number (if known): K060351

MEDTOX® OXYCODONE Device Name:

Indications For Use:

The MEDTOX® OXYCODONE Test System uses immunochromatographic test strips for the rapid, qualitative detection of oxycodone in human urine. It is intended for prescription use.

The test detects oxycodone at concentrations of 100 ng/mL and above.

The MEDTOX® OXYCODONE assay provides only a preliminary analytical test result. A more specific alternate chemical method must be used in order to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method. Clinical consideration and professional judgment should be applied to any drug of abuse test result.

Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Carol Benen

c

lety

K06035/

Page 1 of 1

§ 862.3650 Opiate test system.

(a)
Identification. An opiate test system is a device intended to measure any of the addictive narcotic pain-relieving opiate drugs in blood, serum, urine, gastric contents, and saliva. An opiate is any natural or synthetic drug that has morphine-like pharmocological actions. The opiates include drugs such as morphine, morphine glucoronide, heroin, codeine, nalorphine, and meperedine. Measurements obtained by this device are used in the diagnosis and treatment of opiate use or overdose and in monitoring the levels of opiate administration to ensure appropriate therapy.(b)
Classification. Class II (special controls). An opiate test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (e.g., programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).