K Number
K060348

Validate with FDA (Live)

Date Cleared
2006-02-27

(17 days)

Product Code
Regulation Number
862.1545
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Scantibodies Laboratory, Inc. Total Intact Parathyroid Hormone (PTH) test system is a device intended to measure parathyroid hormone in plasma. Measurements of a crother incone levels are used in the differential diagnosis of hypercalcemia (abnormally high levels of calcium in the blood) and hypocalcemia (abnormally low levels of calcium in the blood) resulting from disorders of calcium metabolism.

Device Description

Total Intact PTH ImmunoChemiluminoMetric (ICMA) Assay

AI/ML Overview

This is an FDA premarket notification (510(k)) letter for an in vitro diagnostic device, the Scantibodies Laboratory, Inc. Total Intact PTH ImmunoChemiluminoMetric (ICMA) Assay. This type of document, particularly from 2006, generally does not contain detailed study information, acceptance criteria, or performance data in the same way that a more recent AI/ML device approval would.

510(k) clearances typically establish substantial equivalence to a legally marketed predicate device, and the detailed performance data, acceptance criteria, and study methodologies are usually filed with the FDA and summarized in publicly available summaries if these were required as part of the 510(k) submission. However, the letter itself primarily focuses on the regulatory decision.

Based on the provided text, I can extract the following:

1. A table of acceptance criteria and the reported device performance:

The provided document does not contain any acceptance criteria or reported device performance. The letter is a regulatory announcement of clearance, not a summary of the underlying validation studies.

2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

The provided document does not specify the sample size for the test set or the data provenance.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

The provided document does not mention the number of experts or their qualifications for establishing ground truth. For an in vitro diagnostic assay, the "ground truth" would typically be established through reference methods or clinical diagnosis, not by experts reviewing images.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

The provided document does not describe any adjudication method.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

This device is an ImmunoChemiluminoMetric (ICMA) Assay, which is a laboratory test for measuring parathyroid hormone. It is not an AI-powered image analysis device or a device that involves human readers. Therefore, an MRMC comparative effectiveness study involving "human readers" and "AI assistance" is not applicable to this type of device.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

This device is a standalone laboratory assay. Its performance is evaluated intrinsically, not in the context of human-in-the-loop performance. Its performance would be evaluated against established analytical performance metrics (e.g., precision, accuracy, limit of detection, linearity), which are not detailed in this clearance letter.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

For this type of in vitro diagnostic assay measuring a biomarker (parathyroid hormone), the "ground truth" in validation studies would typically be established by:

  • Correlation with a reference method or a legally marketed predicate device (as implied by the 510(k) process for substantial equivalence).
  • Correlation with clinical diagnosis of hypercalcemia or hypocalcemia (e.g., from patient records, physician diagnosis).
  • Pathology is not directly applicable here as it's not a tissue-based diagnostic. Outcomes data might be used to assess clinical utility but isn't typically the primary ground truth for an assay's analytical performance.

However, the provided document does not explicitly state the type of ground truth used for its studies.

8. The sample size for the training set:

The provided document does not specify the sample size for any training set. As this is not an AI/ML device, the concept of a "training set" in the machine learning sense is not applicable. For traditional IVD assays, there might be studies for method development and optimization, but they wouldn't be termed "training sets."

9. How the ground truth for the training set was established:

Not applicable, as this is not an AI/ML device with a "training set."

In summary: The provided document is a 510(k) clearance letter from the FDA for an immunoassay. It confirms regulatory approval based on substantial equivalence to a predicate device but does not contain the detailed study design, acceptance criteria, or performance results that would be submitted to the FDA as part of the application. Many of the requested details are specific to AI/ML device evaluations and are not relevant or present for a traditional immunoassay from 2006.

{0}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle-like symbol with three curved lines representing wings or feathers. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the symbol.

FEB 2 7 2006

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Mr. Michael Nordstrom Director, Quality Assurance, Regulatory Affairs Scantibodies Laboratory, Inc. 9336 Abraham Way Santee, CA 92071

K060348 Re:

Trade/Device Name: Total Intact PTH ImmunoChemiluminoMetric (ICMA) Assay Regulation Number: 21 CFR§ 862.1545 Regulation Name: Parathyroid hormone test system Regulatory Class: Class II Product Code: CEW Dated: February 7, 2006 Received: February 10, 2006

Dear Mr. Nordstrom:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).

{1}------------------------------------------------

Page 2 --

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240) 276-0484. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Alberto G. A.

Alberto Gutierrez, Ph.D. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known):

K060348

Device Name: Total Intact PTH ImmunoChemiluminoMetric (ICMA) Assay

Indications For Use:

The Scantibodies Laboratory, Inc. Total Intact Parathyroid Hormone (PTH) test system is a device intended to measure parathyroid hormone in plasma. Measurements of a crother incone levels are used in the differential diagnosis of hypercalcemia (abnormally high levels of calcium in the blood) and hypocalcemia (abnormally low levels of calcium in the blood) resulting from disorders of calcium metabolism.

Prescription Use XX (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Carol Benson

Division Sign-Off

Office of In Vitro Diagnostic Device Evaluation and Safety

Page 1 of

§ 862.1545 Parathyroid hormone test system.

(a)
Identification. A parathyroid hormone test system is a device intended to measure the levels of parathyroid hormone in serum and plasma. Measurements of parathyroid hormone levels are used in the differential diagnosis of hypercalcemia (abnormally high levels of calcium in the blood) and hypocalcemia (abnormally low levels of calcium in the blood) resulting from disorders of calcium metabolism.(b)
Classification. Class II.