K Number
K060218

Validate with FDA (Live)

Date Cleared
2006-03-17

(49 days)

Product Code
Regulation Number
866.1645
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The BD Phoenix™ Automated Microbiology System is intended for the rapid identification and in vitro antimicrobial susceptibility testing of isolates from pure culture of most aerobic and facultative anaerobic Gram-negative and Gram-positive bacteria of human origin.

The BD Phoenix™ Automated Microbiology System is intended for in vitro quantitative determination of antimicrobial susceptibility by minimal inhibitory concentration (MIC) of most Gram-negative aerobic and facultative anaerobic bacteria isolates from pure culture for Enterobacteriaceae and Non-Enterobacteriaceae and most Gram-positive bacteria isolates from pure culture belonging to the genera Staphylococcus, Enterococcus, and Streptococcus.

This premarket notification is for additional organism groups and BD Phoenix™ Automated Microbiology System with erythromycin (0.0625-8 µg/mL), quinupristin/dalfopristin (0.25-4 ug/mL), and chloramphenicol (1-32 µg/mL) on ID/AST or AST only Phoenix panels.

Device Description

The BD Phoenix Automated Microbiology System (Phoenix System) is an automated system for the rapid identification (ID) and antimicrobial susceptibility testing (AST) of clinically relevant bacterial isolates. The system includes the following components:

  • BD Phoenix instrument and software.
  • BD Phoenix panels containing biochemicals for organism ID testing and antimicrobial agents for AST determinations.
  • BD Phoenix ID Broth used for performing ID tests and preparing AST Broth inoculum.
  • BD Phoenix AST Broth used for performing AST tests only.
  • BD Phoenix AST Indicator solution added to the AST Broth to aid in bacterial growth determination.
    The Phoenix panel is a sealed and self-inoculating molded polystyrene tray with 136 micro-wells containing dried reagents. Organisms for susceptibility testing must be a pure culture and preliminarily identified as a Gram-negative or Gram-positive isolate. For each isolate, an inoculation equivalent to a 0.5 McFarland standard is prepared in Phoenix ID Broth.

The Phoenix AST method is a broth based microdilution test. The Phoenix System utilizes a redox indicator for the detection of organism growth in the presence of an antimicrobial agent. Measurements of changes to the indicator as well as bacterial turbidity are used in the determination of bacterial growth. Each AST panel configuration contains several antimicrobial agents with a wide range of two-fold doubling dilution concentrations.

The instrument houses the panels where they are continuously incubated at a nominal temperature of 35°C. The instrument takes readings of the panels every 20 minutes. The readings are interpreted to give an identification of the isolate, minimum inhibitory concentration (MIC) values and category interpretations, S, I, or R (sensitive, intermediate, or resistant).

AI/ML Overview

Here's the breakdown of the acceptance criteria and study information for the BD Phoenix™ Automated Microbiology System, based on the provided text:

Acceptance Criteria and Device Performance

The acceptance criteria are implied by the statement: "The BD Phoenix™ Automated Microbiology System demonstrated substantially equivalent performance when compared with the CLSI reference broth microdilution method." The specific quantitative acceptance criteria are that Essential Agreement (EA) and Category Agreement (CA) should be high, with the reported performance values being the evidence of meeting these criteria. The FDA Draft guidance document, "Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA," February 5, 2003, is cited as the standard for evaluation.

AntimicrobialConcentrationAcceptance Criteria (Implied)Reported Essential Agreement (EA) (%)Reported Category Agreement (CA) (%)
Erythromycin0.0625-8 µg/mLHigh Agreement95.094.6
Quinupristin/dalfopristin0.25-4 µg/mLHigh Agreement94.595.5
Chloramphenicol1-32 µg/mLHigh Agreement93.493.4

Note: The document states that intra-site reproducibility was greater than 90% and inter-site reproducibility was greater than 95% for the isolates tested in the reproducibility study. While these are strong performance metrics, the primary performance for regulatory approval seems to hinge on EA and CA compared to the reference method.

Study Details

2. Sample Size and Data Provenance (Test Set)

  • Sample Size (Test Set):
    • Erythromycin: 1395 isolates
    • Quinupristin/dalfopristin: 2019 isolates (for EA), 1500 isolates (for CA)
    • Chloramphenicol: 1447 isolates
  • Data Provenance: Clinical, stock, and challenge isolates were tested. The study was conducted "across multiple geographically diverse sites across the United States," indicating a multi-center, U.S.-based origin. It includes both "clinical isolates" (which are typically prospective or recently retrospective) and "stock and challenge isolates" (which are curated laboratory strains).

3. Number of Experts and Qualifications for Ground Truth (Test Set)

  • The document does not explicitly state the number of experts used or their specific qualifications.
  • The ground truth for clinical isolates was established by the "CLSI reference broth microdilution method." This method itself is a standardized laboratory procedure, and implicitly relies on expert microbiologists and technicians to perform and interpret it correctly.
  • For "challenge set isolates," the ground truth was "expected results," which would also be determined by established reference methods or expert consensus from culture collections.

4. Adjudication Method (Test Set)

  • The document does not explicitly describe an adjudication method in the traditional sense (e.g., 2+1 reader consensus).
  • The comparison is made between the Phoenix System results and the "CLSI reference broth microdilution method" results. Discrepancies would likely be investigated by re-testing or thorough review of the reference method results by laboratory personnel. However, a formal multi-reader adjudication process for discrepancies is not detailed.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • No, an MRMC comparative effectiveness study was not done. This study focuses on the standalone performance of the automated system against a reference laboratory method, not on how human readers perform with or without AI assistance.

6. Standalone Performance Study

  • Yes, a standalone study was performed. The entire "Clinical Studies" section describes the performance of the BD Phoenix™ Automated Microbiology System (algorithm only) compared to the CLSI reference broth microdilution method. The reported Essential Agreement and Category Agreement percentages are measures of this standalone performance.

7. Type of Ground Truth Used

  • Reference Method/Expert Consensus (implicit): For clinical isolates, the ground truth was established by the "CLSI reference broth microdilution method." This is considered a gold standard laboratory method. While not explicitly "expert consensus" in the sense of multiple human readers interpreting an image, the CLSI method itself is defined by expert microbiologists and its execution requires trained personnel.
  • Expected Results: For "challenge set isolates," the ground truth was "expected results," which implies pre-defined, known results based on established methods for these control strains.

8. Sample Size for the Training Set

  • The document does not explicitly state the sample size for the training set. The description details studies for validation of the device's performance against reference methods with specific antimicrobial agents. The underlying algorithms for the Phoenix system would have been developed and trained using a separate dataset over time, but details of that specific training set are not provided in this 510(k) summary focused on adding new antimicrobial agents.

9. How the Ground Truth for the Training Set Was Established

  • As the training set details are not provided, the method for establishing its ground truth is also not described in this document. It can be inferred that earlier development of the Phoenix system would have involved extensive data collection and the use of reference methods (like CLSI broth microdilution) to establish ground truth for training bacterial identification and antimicrobial susceptibility algorithms.

{0}------------------------------------------------

510(K) SUMMARY

SUBMITTED BY:Becton, Dickinson and Company 7 Loveton Circle Sparks, MD 21152 Phone: (410) 316 - 4278 Fax: 410-316-4499
CONTACT NAME:Monica E Giguere Regulatory Affairs Specialist
DATE PREPARED:March 7, 2006
DEVICE TRADE NAME:BD Phoenix™ Automated Microbiology System - Erythromycin-0.0625-8 µg/mL, Quinupristin/dalfopristin- 0.25-4 µg/mL, Chloramphenicol-1-32 µg/mL
DEVICE COMMON NAME:Antimicrobial susceptibility test system-short incubation
DEVICE CLASSIFICATION:Fully Automated Short-Term Incubation Cycle Antimicrobial Susceptibility Device, 21 CFR 866.1645
PREDICATE DEVICES:VITEK® System (PMA No. N50510) and BD Phoenix™ Automated Microbiology System with Gatifloxacin (K020321, May 23, 2002, Erythromycin (K033784, February 3, 2004), Quinupristin/dalfopristin (K022172, August 12, 2002), and Chloramphenicol (K033889, February 3, 2004).
INTENDED USE:The BD Phoenix™ Automated Microbiology System is intended for the rapid identification and in vitro antimicrobial susceptibility testing of isolates from pure culture of most aerobic and facultative anaerobic Gram-negative and Gram- positive bacteria of human origin.

DEVICE DESCRIPTION:

The BD Phoenix Automated Microbiology System (Phoenix System) is an automated system for the rapid identification (ID) and antimicrobial susceptibility testing (AST) of clinically relevant bacterial isolates. The system includes the following components:

  • . BD Phoenix instrument and software.
  • BD Phoenix panels containing biochemicals for organism ID testing and antimicrobial agents . for AST determinations.
  • BD Phoenix ID Broth used for performing ID tests and preparing AST Broth inoculum. .
  • BD Phoenix AST Broth used for performing AST tests only. .

{1}------------------------------------------------

  • . BD Phoenix AST Indicator solution added to the AST Broth to aid in bacterial growth determination.
    The Phoenix panel is a sealed and self-inoculating molded polystyrene tray with 136 micro-wells containing dried reagents. Organisms for susceptibility testing must be a pure culture and preliminarily identified as a Gram-negative or Gram-positive isolate. For each isolate, an inoculation equivalent to a 0.5 McFarland standard is prepared in Phoenix ID Broth.

The Phoenix AST method is a broth based microdilution test. The Phoenix System utilizes a redox indicator for the detection of organism growth in the presence of an antimicrobial agent. Measurements of changes to the indicator as well as bacterial turbidity are used in the determination of bacterial growth. Each AST panel configuration contains several antimicrobial agents with a wide range of two-fold doubling dilution concentrations.

The instrument houses the panels where they are continuously incubated at a nominal temperature of 35°C. The instrument takes readings of the panels every 20 minutes. The readings are interpreted to give an identification of the isolate, minimum inhibitory concentration (MIC) values and category interpretations, S, I, or R (sensitive, intermediate, or resistant).

DEVICE COMPARISON:

The BD Phoenix™ Automated Microbiology System demonstrated substantially equivalent performance when compared with the CLSI reference broth microdilution method. This premarket notification provides data for additional organism groups with Erythromycin-0.0625-8 ug/mL, Quinupristin/dalfopristin-0.25-4 ug/mL, and Chloramphenicol-1-32 ug/mL and the BD Phoenix™ Automated Microbiology System.

SUMMARY OF SUBSTANTIAL EQUIVALENCE TESTING:

The BD Phoenix™ Automated Microbiology System has demonstrated substantially equivalent performance when compared to the CLSI reference broth microdilution method (AST panels prepared according to NCCLS M7). The system has been evaluated as defined in the FDA Draft guidance document, "Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA," February 5, 2003.

Site Reproducibility

Intra- and inter-site reproducibility of this antimicrobial agent in the BD Phoenix System was evaluated at three sites using a panel chosen isolates. Each site tested the isolates in triplicate on three different days using one lot of Phoenix panels containing the antimicrobial agents and associated reagents.

The results of the study demonstrate for each antimicrobial agent there was an overall intra-site reproducibility of greater than 90% and an overall inter-site reproducibility greater than 95% for the isolates tested.

{2}------------------------------------------------

Clinical Studies

Clinical, stock and challenge isolates were tested across multiple geographically diverse sites across the United States to demonstrate the performance of the Phoenix antimicrobial susceptibility test with the Phoenix panel formats containing antimicrobial agents. Phoenix System results for Challenge set isolates were compared to the expected results. Phoenix System results for clinical isolates were compared to the results obtained from the CLSI reference broth microdilution method.

The performance of the Phoenix System was assessed by calculating Essential Agreement (EA) and Category Agreement (CA) to expected/reference results for all isolates tested. Essential Agreement (EA) occurs when the BD Phoenix™ Automated Microbiology System agrees exactly or within ± one two-fold dilution to the reference result. Category Agreement (CA) occurs when the BD Phoenix™ Automated Microbiology System agrees with the reference method with respect to the FDA categorical interpretive criteria (susceptible, intermediate, and resistant).

Table 1 summarizes the performance for the isolates tested in this study.

AntimicrobialConcentrationEA (n)EA (%)CA (n)CA (%)
Erythromycin0.0625-8 µg/mL139595.0139594.6
Quinupristin/dalfopristin0.25-4µg/mL201994.5150095.5
Chloramphenicol1-32 µg/mL144793.4144793.4

Table 1: Performance of BD Phoenix System by Drug

Conclusions Drawn from Substantial Equivalence Studies

The data collected from the substantial equivalence studies demonstrate that testing on the BD Phoenix™ Automated Microbiology System with this antimicrobial agent is substantially equivalent as outlined in the FDA draft guidance document, "Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA," February 5, 2003. Technological characteristics of this system are substantially equivalent to those used in the VITEK® system, which received approval by the FDA under PMA number N50510 and BD Phoenix™ Automated Microbiology System with Gatifloxacin (K020321, May 23, 2002), Erythromycin (K033784, February 3, 2004), Quinupristin/dalfopristin (K022172, August 12, 2002) and Chloramphenicol (K033889, February 3, 2004).

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized caduceus, which is a symbol often associated with healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus.

Public Health Service

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Ms. Monica E. Giguere Regulatory Affairs Specialist BD Diagnostics Systems Becton, Dickinson and Company 7 Loveton Circle Sparks, MD 21152

MAR 1 7 2006

Re: K060218 Trade/Device Name: BD Phoenix™ Automated Microbiology System Erythromycin (0.0625-8 µg/mL), Quinupristin/dalfopristin (0.25-4 ug/mL), and Chloramphenicol (1-32 µg/mL) - ID/AST or AST Regulation Number: 21 CFR 866.1645 Regulation Name: Fully Automated Short-Term Incubation Cycle Antimicrobial Susceptibility Devices Regulatory Class: Class II Product Code: LON Dated: January 26, 2006 Received: January 27, 2006

Dear Ms. Giguere:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).

{4}------------------------------------------------

Page 2 -

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240)276-0484. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html

Sincerely yours,

Sale, a Joy

Sally A. Hojvat, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Page 1 of 2

510(k) Number:

Device Name: BD Phoenix™ Automated Microbiology System for use with the antimicrobial agent Erythromycin (0.0625-8 µg/mL), Quinupristin/dalfopristin (0.25-4 µg/mL), and Chloramphenicol (1-32 ug/mL) on ID/AST or AST only Phoenix panels

Indications for Use:

The BD Phoenix™ Automated Microbiology System is intended for in vitro quantitative determination of antimicrobial susceptibility by minimal inhibitory concentration (MIC) of most Gram-negative aerobic and facultative anaerobic bacteria isolates from pure culture for Enterobacteriaceae and Non-Enterobacteriaceae and most Gram-positive bacteria isolates from pure culture belonging to the genera Staphylococcus, Enterococcus, and Streptococcus.

This premarket notification is for additional organism groups and BD Phoenix™ Automated Microbiology System with erythromycin (0.0625-8 µg/mL), quinupristin/dalfopristin (0.25-4 ug/mL), and chloramphenicol (1-32 µg/mL) on ID/AST or AST only Phoenix panels.

Erythromycin has been shown to be active in vitro against most strains of microorganisms listed below, as described in the FDA-approved package insert for this antimicrobial agent.

Active In Vitro and in Clinical Infections Against:

Staphylococcus aureus

Quinupristin/dalfopristin has been shown to be active in vitro against most strains of microorganisms listed below, as described in the FDA-approved package insert for this antimicrobial agent.

Active In Vitro and in Clinical Infections Against:

Aerobic Gram-positive microorganisms
Enterococcus faecium (vancomycin-resistant and multi-drug resistant strains only)
Staphylococcus aureus (methicillin-susceptible strains only)

Division Sign-Off#### Active In Vitro Against:

Aerobic Gram-positive microorganisms

Staphylococcus aureus (methicillin-resistant strains)

Office of In Vitro Diagnostic Device Evaluation and Safety

Staphylococcus epidermidis (including methicillin-resistant strains) 10(k)____ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _

Synercid is not active against Enterococcus faecalis. Differentiation of enterococcal species is important to avoid misidentification of Enterococcus faecalis as Enterococcus faecium.

BD Diagnostic Systems Becton, Dickinson and Company

{6}------------------------------------------------

Page 2_ of 2

Chloramphenicol has been shown to be active in vitro against most strains of microorganisms listed below, as described in the FDA-approved package insert for this antimicrobial agent.

Active In Vitro and in Clinical Infections Against:

Staphylococcus aureus

Prescription Use イ (Per 21 CFR 801.109)

Over-the-Counter Use

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Luddei h. Poole
Division Sign-Off

Division Sign-Off

Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) K060218

BD Diagnostic Systems Becton, Dickinson and Company

§ 866.1645 Fully automated short-term incubation cycle antimicrobial susceptibility system.

(a)
Identification. A fully automated short-term incubation cycle antimicrobial susceptibility system is a device that incorporates concentrations of antimicrobial agents into a system for the purpose of determining in vitro susceptibility of bacterial pathogens isolated from clinical specimens. Test results obtained from short-term (less than 16 hours) incubation are used to determine the antimicrobial agent of choice to treat bacterial diseases.(b)
Classification. Class II (special controls). The special control for this device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA.”