K Number
K060206

Validate with FDA (Live)

Device Name
MISSION COMPLETE
Date Cleared
2006-03-21

(54 days)

Product Code
Regulation Number
862.1660
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Mission Complete™ Linearity Controls are intended to be used for confirming the calibration and linearity of instruments measuring pH, pCO2, pO2, Na *, K *, Cr *, CI , Glucose, and Lactate. For In Vitro Diagnostic Use

Device Description

This control material is provided in five (5) distinct levels of pH, pCO2, pO2, Na *, K , Cl., Ca *, Glucose & Lactate covering the significant range of the instrument performance. It is packaged in sealed glass ampules, each containing 1.8 ml of solution. Ampules are packaged in kits containing four (4) ampules of each level.

AI/ML Overview

This document describes the Mission Complete™ Linearity Controls, a quality control material for blood gas, electrolyte, and metabolite analyzers.

Here's an analysis based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The provided document does not explicitly state quantitative "acceptance criteria" for the performance of the Mission Complete™ device in terms of numerical thresholds for accuracy, precision, or other performance metrics. Instead, the "acceptance" for this device appears to be based on demonstrating substantial equivalence to a predicate device (RNA Medical CVC 123) in terms of its technological characteristics, formulation, and intended use.

The "reported device performance" is primarily qualitative, focusing on similarity to the predicate.

CharacteristicAcceptance Criteria (Implicit)Reported Device Performance (Mission Complete™)
Intended UseFor confirming calibration and linearity of blood gas, electrolyte, and metabolite instruments for listed analytes.Confirms calibration and linearity of blood gas, electrolyte, and metabolite instruments for pH, pCO2, pO2, Na+, K+, Cl-, Ca+, Glucose, and Lactate.
Material CompositionAqueous solution of buffers, electrolytes, glucose, lactate, equilibrated with CO2, O2, N2. No human or animal materials.Aqueous solution of buffers, electrolytes, glucose, lactate, equilibrated with CO2, O2, N2. Contains no human or animal materials.
Levels5 distinct levels covering instrument performance range.5 distinct levels covering significant range of instrument performance.
ContainerGlass ampuleSealed glass ampule
Package SizeSimilar to predicate (20 x 2.5ml for predicate)20 x 1.8 ml (Slight difference from predicate, but within acceptable range for a control material)
ColorClearClear
Storage2° - 8°C2° - 8°C
Expiration/Shelf Life3 years3 years
Substantial EquivalenceEquivalent to RNA Medical CVC 123 in function, safety, and efficacy.Claimed to be equivalent (and affirmed by FDA clearance).

2. Sample Size Used for the Test Set and Data Provenance

The document does not specify a separate "test set" in the context of clinical trials on patients. For this type of device (quality control material), the "testing" involves non-clinical studies to demonstrate its characteristics and performance.

  • Sample Size for Non-clinical Tests: Not explicitly stated as a number of "samples" in the traditional sense. The studies involved general "stability for shelf life claim" and "performance & correlation to equivalent device." These would involve multiple ampules and measurements, but a distinct sample size number is not provided.
  • Data Provenance: Not explicitly stated. Given this is a 510(k) submission from a US company (Diamond Diagnostics, Holliston MA), the studies were likely conducted in the US. The studies are non-clinical (bench testing, stability studies, etc.) rather than involving human subjects, so concepts like "retrospective or prospective" don't directly apply in the same way as clinical trials.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

This information is not applicable and not provided in the document. For a quality control material, "ground truth" is typically established by certified reference materials, analytical standards, or highly precise and accurate reference methods, not by expert human interpretation like radiologists. The purpose of the control material is to verify the performance of an instrument against known values.

4. Adjudication Method for the Test Set

This information is not applicable and not provided. Adjudication methods (like 2+1, 3+1) are used in studies involving human interpretation (e.g., image reading) where there might be disagreement among experts. Since this device is a quality control material and its "ground truth" is analytical, such methods are not relevant.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study is relevant for diagnostic devices that involve human interpretation and aims to assess how a device (e.g., an AI algorithm) impacts human reader performance. The Mission Complete™ Linearity Controls is a quality control material, not a diagnostic device for human interpretation.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

No, a "standalone" performance study in the context of an algorithm or AI is not applicable here. This device is a physical control material, not an algorithm. Its performance is evaluated through its chemical stability, consistency, and ability to produce expected results when analyzed by an instrument, not as an algorithm operating independently.

7. The Type of Ground Truth Used

The "ground truth" for a quality control material like this would be established through:

  • Certified Reference Materials/Standards: Precisely known concentrations of analytes (pH, pCO2, pO2, Na+, K+, Cl-, Ca+, Glucose, Lactate) used to develop and characterize the control material.
  • Reference Methods: Highly accurate and precise analytical methods used to determine the exact values of the analytes within the control material batches.
  • Assayed Values: The Expected Values Chart provided with the control material lists the "assay values" for each analyte and level, which are determined by the manufacturer using reference methods and calibrated instruments.

Therefore, the ground truth is analytical and based on reference methods and certified standards, not expert consensus, pathology, or outcomes data in the usual sense.

8. The Sample Size for the Training Set

This information is not applicable and not provided. This device is a physical quality control material and does not involve AI/machine learning algorithms that require a "training set."

9. How the Ground Truth for the Training Set Was Established

This information is not applicable and not provided, as there is no "training set" for this type of device.

{0}------------------------------------------------

K060206

510(k) Summary Pg.17 of 45 for Mission Diagnostic pH/Blood Gas ISE, Metabolite Controls for pH/Blood Gas, Electrolyte and Metabolites Analyzers

Submitter's Name & Address 1. Diamond Diagnostics 333 Fiske St Holliston MA 01746 FAX: 508-429-0452

Contact Person:

And 1987 - 1997 - 1997 - 1999 -

Randolph Quinn Product Development & Compliance Manager 508-429-0450

Establishment Registration Number: 3003656721 Date of Preparation of this summary: January 18, 2006

2. Identification of the Device:

Proprietary/Trade name:Mission Complete™
Common or usual name:Quality Control material (assay and unassayed)
Classification name:Multi-Analyte Controls (assay and unassayed)
Device Classification:[
Regulation Number:21 CFR § 862.1660
Panel:Chemistry (75)
Product Code:JJY

3. Substantial Equivalence:

Mission Complete PN DD-92900 is equivalent to RNA Medical CVC 123 with regards to function, safety and efficacy.

Comparison Tables of Characteristics

CharacteristicsMissionOEM Equivalent
PNDD-92900CVC 123
Contents:Any LevelAqueous solution of buffers,electrolytes, glucose,lactate,equilibrated with , CO2, O2,and N2. Contains no human or animalmaterials.Aqueous solution of buffers,electrolytes, glucose,lactate,equilibrated with , CO2,O2 and N2. Contains nohuman or animal materials.
Levels1,2,3,4,51,2,3,4,5
ContainerGlass ampuleGlass ampule
colourclearclear
Package20 x 1.8 ml20 x 2.5 ml
Intended UseFor in-vitro diagnostic use onlyFor in-vitro diagnostic useonly
Storage2° - 8°C2° - 8°C
Expiration/Shelf Life3 years3 years

4. Description of the new device:

This control material is provided in five (5) distinct levels of pH, pCO2, pO2, Na *, K , Cl., Ca *, Glucose & Lactate covering the significant range of the instrument performance. It is packaged in sealed glass ampules, each containing 1.8 ml of solution. Ampules are packaged in kits containing four (4) ampules of each level.

510(k) submission for Mission Diagnostics pH Blood Gas /ISE, Metabolite Controls for pH Blood Gas, Electrolyte and Metabolite Analyzers

{1}------------------------------------------------

5. Intended use of the device:

MISSION COMPLETE™ Linearity Control are assayed materials used for confirming the calibration and linearity of blood gas, electrolyte, and metabolite instruments for the analytes and analyzers listed on the Expected Values Chart.

6. Technological characteristics of the device:

This material consists of a buffered aqueous solution with electrolytes, glucose and lactate equilibrated with CO2, O2 and N2. The product contains no human or animal derived materials. Product is sealed into glass ampules containing 1.8mL and is intended for onetime use.

7. Summary of non-clinical tests submitted with the premarket notification for the device: a) Stability for shelf life claim

b) Performance & Correlation to equivalent device

    1. Summary of clinical tests submitted with the premarket notification for the device: N/A

9. Conclusions drawn from the clinical and non-clinical trials:

Comparison of technological characteristics, formulation and intended use to predicate devices listed in this summary support the claim of substantial equivalence.

510(k) submission for Mission Diagnostics pH Blood Gas /ISE, Metabolite Controls for pH Blood Gas, Electrolyte and Metabolite Analyzers

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, and the words "HEALTH & HUMAN SERVICES - USA" arranged in a circle around the eagle. The eagle is black, and the text is also black. The logo is simple and recognizable.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

MAR 2 1 2006

Mr. Randolph Quinn Development & Compliance Manager Diamond Diagnostics, Inc. 333 Fiske Street Holliston, MA 01746

K060206 Re: Trade/Device Name: Mission Complete™ Linearity Controls Regulation Number: 21 CFR§ 862.1660 Regulation Name: Quality Control (assayed and unassayed) Regulatory Class: Class I Product Code: JJY Dated: January 26, 2006 Received: February 15, 2006

Dear Mr. Quinn:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).

{3}------------------------------------------------

Page 2 -

This letter will allow you to begin marketing your device as described in your Section 510(k) I ins letter will anow you to ough mains of substantial equivalence of your device to a legally promatics notification: "The PD's in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific information about the application of labeling requirements to your device, If you desire specific mission and advertising of your device, please contact the Office of In of questions on the prometive and Safety at (240) 276-0484. Also, please note the viro Diagavitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Tou may of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Alberto Gutt

Alberto Gutierrez, Ph.D. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use

510(k) Number (if known): K060206

Mission Complete Device Name:

Indications For Use:

Mission Complete™ Linearity Controls are intended to be used for confirming the calibration and linearity of instruments measuring pH, pCO2, pO2, Na ', K ', Cr ' , CI , Glucose, and Lactate.

For In Vitro Diagnostic Use

Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use _ (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Carol Benson

Division Sign-

Office of In Vitro Dlag Device Evaluation at

Page 1 of

510(k) K060206

§ 862.1660 Quality control material (assayed and unassayed).

(a)
Identification. A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.(b)
Classification. Class I (general controls). Except when intended for use in donor screening tests, quality control materials (assayed and unassayed) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.