K Number
K060150

Validate with FDA (Live)

Date Cleared
2006-02-23

(34 days)

Product Code
Regulation Number
890.5575
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The SmartStep™ System/ SmartStep™ Clinic System is intended to sense the amount of weight applied to the planter surface of the foot during rehabilitation. The System alerts the user and/or therapist with alarm when the weight exceeded the pre-selected value. Furthermore, the SmartStep™ System/ SmartStep™ Clinic System is intended to be used in any situation in which therapist and/or patient would benefit from objectively assessing the amount of weight that is being applied to a lower limb.

Device Description

The SmartStep™ System is a monitoring & biofeedback system for gait. The device is a tool, which is intended to assist the clinicians to accurately assess patients who are undergoing weight bearing restrictions and gait therapy training. The SmartStep™ is used to measure the amount of weight being applied to an affected limb, to teach and promote specific weight bearing skills and to test patient's ability to maintain a weight bearing range. The SmartStep™ system is to be used under the supervision of physician or licensed health care provider such as physiotherapist. The SmartStep™ System is comprised of the following three main subelements: Flexible Insole that is placed in the patient's shoe, acting as a pressure-sensing element. Control Unit, connected to PC Software. PC Software that acts as a patient medical record and patient assessment tool. A Manual Pump is used to inflate the Insole compartments.

AI/ML Overview

This document is a 510(k) Summary for the SmartStep™ System, comparing it to a previously cleared SmartStep™ System (K023161). The core of this submission is to demonstrate substantial equivalence, rather than a de novo performance study against defined acceptance criteria. Therefore, much of the requested information regarding a new study proving the device meets acceptance criteria cannot be extracted from this document.

However, based on the provided text, here's what can be gathered regarding performance and equivalence:

1. A table of acceptance criteria and the reported device performance

The document does not specify quantitative acceptance criteria or a new performance study with results against those criteria. Instead, it states that "Electrical and electromagnetic testing and verification and validation testing of the Software were performed to ensure that the modified SmartStep™ System does not raise any new questions of safety and efficacy." The general performance claim is substantial equivalence to the predicate device.

Acceptance CriteriaReported Device Performance
(Not explicitly defined in the document as new quantitative criteria for the modified device for this 510(k) submission.)"Electrical and electromagnetic testing and verification and validation testing of the Software were performed to ensure that the modified SmartStep™ System does not raise any new questions of safety and efficacy."
Continued compliance with voluntary standards (IEC 60601-1, IEC 60601-1-2, AAMI/ISO 14971-1, FCC requirements).The SmartStep™ System complies with these voluntary standards.
Substantial equivalence in technological characteristics, mode of operation, performance characteristics, and intended use to the predicate SmartStep™ System (K023161).Andante Medical devices Ltd. believes the modified SmartStep™ System is substantially equivalent to the cleared SmartStep™ System, without raising new safety and/or effectiveness issues.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document does not describe a new clinical or performance test set for the modified device with specific sample sizes. The testing mentioned is primarily electrical, electromagnetic, and software verification/validation, indicating laboratory or engineering testing rather than patient-based data. Therefore, details like data provenance (country of origin, retrospective/prospective) are not applicable to the type of testing described here.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

Not applicable. The document does not describe a clinical study or diagnostic performance evaluation requiring expert-established ground truth. The testing detailed relates to device function and safety, without clinical interpretation.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable. There is no mention of a test set requiring adjudication in the context of clinical performance.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. The SmartStep™ System is a monitoring and biofeedback system for gait, not an AI-assisted diagnostic or interpretation tool. The document does not describe any MRMC study.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable in the sense of an "algorithm only" performance for a diagnostic or interpretive AI device. The SmartStep™ System inherently involves human interaction (patient wearing the insole, therapist supervising, clinician using PC software). The "performance data" mentioned refers to the device's functional and safety compliance.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

Not applicable. For the engineering and software testing mentioned, the "ground truth" would be the expected functional behavior and safety specifications, rather than clinical ground truth like pathology or expert consensus.

8. The sample size for the training set

Not applicable. The document does not describe a machine learning or AI algorithm development that would involve a training set.

9. How the ground truth for the training set was established

Not applicable, as there is no training set mentioned.

{0}------------------------------------------------

Kobo150

FEB 2 3 2006

510(K) SUMMARY

Andante Medical Devices LTD.

SmartStep™ System

7.1.1 Applicant's Name:

Andante Medical Devices Ltd. Omer Industrial Park Bldg. 8b. P.O. Box 3023 Omer 84965, ISRAEL Tel: +972-8-6900027 Fax: +972-8-6900319 info@andante.co.il

7.1.2 Contact Person:

Dorit Winitz, Ph. D Biomedical Strategy (2004) Ltd. Moshe Aviv Tower, 34th Floor, 7 Jabotinsky Street Ramat Gan 52520, Israel Tel: +972-3-612-3281 Fax: +972-3-612-3282 dorit@ebms.co.il

7.1.3 Date Prepared:

January, 2006

7.1.4 Trade Name:

SmartStep™ System/ SmartStep™ Clinic System

7.1.5 Classification Name:

Device, Warning, Overload, External

7.1.6 Classification:

Class II; Product Code IRN; Regulation No. 890.5575

7.1.7 Predicate Devices

Andante SmartStep™ System (K023161)

{1}------------------------------------------------

2/5 KAb6150

7.1.8 Device Description:

The SmartStep™ System is a monitoring & biofeedback system for gait the Umar 8000 - " Dy 500" The device is a tool, which is intended to assist the clinicians to accurately assess patients who are undergoing weight bearing restrictions and gait therapy training.

The SmartStep™ is used to measure the amount of weight being applied to an affected limb, to teach and promote specific weight bearing skills and to test patient's ability to maintain a weight bearing range.

The SmartStep™ system is to be used under the supervision of physician or licensed health care provider such as physiotherapist.

The SmartStep™ System is comprised of the following three main subelements:

  • Flexible Insole that is placed in the patient's shoe, acting as a . pressure-sensing element.
  • Control Unit, connected to PC Software. ■
  • PC Software that acts as a patient medical record and patient l assessment tool.

A Manual Pump is used to inflate the Insole compartments.

7.1.9 Intended Use:

The SmartStep™ System is intended to sense the amount of weight applied to the planter surface of the foot during rehabilitation. The System alerts the user and/or therapist with alarm when the weight exceeded the pre-selected value. Furthermore, the SmartStep™ System is intended to be used in any situation in which therapist and/or patient would benefit from objectively assessing the amount of weight that is being applied to a lower limb.

7.1.10 Performance Standards:

No performance standards have been established for such device under Section 514 of the Federal Food, Drug, and Cosmetic Act. However, the SmartStep™ System complies with the voluntary standards such as IEC 60601-1, IEC 60601-1-2, AAMI/ISO 14971-1, and with the FCC requirements of CFR, parts 15, Subparts B & C.

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image contains handwritten text. The text on the left side of the image reads "3/3". To the right of that, the text reads "KOG6150". The text appears to be written in black ink on a white background.

7.1.11 Performance Data & Substantial Equivalence

The SmartStep™ System is substantially equivalent in all aspects, e.g., technological characteristics, mode of operation, performance characteristics, intended use, etc., to the commercially available SmartStep™ System. The principle changes between the devices include:

1. Control Unit

  • 1.1 The design of the Control Unit was changed to improve its ergonomic design and to enable a wireless operated unit.
  • Inflation connectors were added to the Control Unit to enable on-1.2 line monitoring of the inflation status.
  • The operation modes were re-organized and the vibration feedback 1.3 was eliminated.
  • 1.4 Fault-condition indications and battery automatic shut-off capability were added.
  • 1.5 Part of the electronic components of the Control Unit, such as microcontroller, battery and charger, were changed to improve device performance and reliability.

2. Manual Pump

The insole inflation pump was improved to include a Manual Pump rather than a syringe.

  • Software and User Interface క.
  • 3.1 The PC User Interface was improved to include more intuitive, simple and clear GUI and to support multilingual User Interface.
  • 3.2 Several functions and options were added to the PC Software.
  • 3.3 The Software (PC and Control Unit) as a whole has been upgraded.

Electrical and electromagnetic testing and verification and validation testing of the Software were performed to ensure that the modified SmartStep™ System does not raise any new questions of safety and efficacy. Based on these tests results, Andante Medical devices Ltd. believes that the modified SmartStep™ System is substantially equivalent to the cleared SmartStep™ System, without raising new safety and/or effectiveness issues.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH AND HUMAN SERVICES. USA" around the perimeter. Inside the circle is a stylized image of three human figures.

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

FEB 2 3 2006

Andante Medical Devices Ltd. C/O Dorit Winitz, Ph.D Biomedical Strategy (2004) Ltd. Moshe Aviv Tower, 34th floor 7 Jabotinsky Street Ramat, Gan 52520, Israel

Re: K060150

Trade/Device Name: SmartStepTM System/ SmartStep™ Clinic System Regulation Number: 21 CFR 890.5575 Regulation Name: Powered External Limb Overload Warning Device Regulatory Class: II Product Code: IRN Dated: January 15, 2006 Received: January 28, 2006

Dear Dr. Winitz:

We have reviewed your Section 510(k) premarket notification of intent to market the device we nave roviewed your becareer in the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate 10 use stated in the encreativent of the enactment date of the Medical Device Amendments, or to conninered print to may 20, 1978, accordance with the provisions of the Federal Food. Drug, de rices that have been rocksonied in quire approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The r ou may, merely mains of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can may or sabyer to basil as a same as a Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean I toast of actived that I Dr instan that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must or any I vatuation and including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set

{4}------------------------------------------------

Page 2 - Dr. Dorit Winitz

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic forth in the quality systems (Sections 531-542 of the Act); 21 CFR 1000-1050. product radiation control provisions (Doctors over device as described in your Section 5 10(k)
This letter will allow you to begin marketing your device as described in your I ms letter will anow you to begin manketing your arrial equivalence of your device to a legally premarket nothleation. The PDA miding of backandary of your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific advice for your act (240) 276-0120. Also, please note the regulation entitled, a Colliact the Office of Compullios as (21 ) notification" (21CFR Part 807.97). You may obtain Misolanuing by reference to premanter for the Act from the Division of Small other general informational and Consumer Assistance at its toll-free number (800) 638-2041 or Manufacturers, International and Ochess http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Herbert Lewinsohn

Mark N. Melkerson Acting Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

INDICATIONS FOR USE STATEMENT

510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________

SmartStep™ System/ SmartStep™ Clinic System Device Name:

Indications for Use:

The SmartStep™ System/ SmartStep™ Clinic System is intended to sense the amount of weight applied to the planter surface of the foot during rehabilitation. The System alerts the user and/or therapist with alarm when the weight exceeded the pre-selected value. Furthermore, the SmartStep™ System/ SmartStep™ Clinic System is intended to be used in any situation in which therapist and/or patient would benefit from objectively assessing the amount of weight that is being applied to a lower limb.

Prescription Use V (Part 21 C.F.R. 801 Subpart D) AND/OR

Over-The-Counter Use (Part 21 C.F.R. 807 Subpart C)

KO60150

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of General, Restorative,
7-10and Neurological Devices
510(k) Number_

§ 890.5575 Powered external limb overload warning device.

(a)
Identification. A powered external limb overload warning device is a device intended for medical purposes to warn a patient of an overload or an underload in the amount of pressure placed on a leg.(b)
Classification. Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 890.9.