K Number
K060111

Validate with FDA (Live)

Manufacturer
Date Cleared
2006-04-24

(97 days)

Product Code
Regulation Number
892.1170
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

IVA scans are intended for the visualization or quantitative assessment of vertebral bone deformities. IVA also allows the visualization of abdominal aortic calcification, and, if present, clinical correlation may be advised since abdominal aortic calcification may be associated with cardiovascular disease.

Device Description

IVA (MXA -II) software option for the X-Ray bone densitometers.

AI/ML Overview

The provided text is a 510(k) summary for the IVA (MXA-II) software option for X-Ray bone densitometers. It focuses on a modified indication for use, specifically for the visualization of abdominal aortic calcification in addition to vertebral bone deformities. However, it explicitly states that no changes were made to the software's operation or performance specifications for this modification. Therefore, the document does not contain information about acceptance criteria or a study proving that the device meets those criteria for the detection of abdominal aortic calcification, as it relies on the predicate device's established performance for a new indication without altering the core functionality.

The document emphasizes substantial equivalence to predicate devices (Hologic Vertebral Morphometry Software Option and Hologic Vertebral Morphometry Software II Option, and general use x-ray systems like the Toshiba KXO-80G x-ray generator). This suggests that previous studies conducted for the predicate devices likely established their performance for vertebral morphometry, which the IVA (MXA-II) software inherits.

Therefore, based solely on the provided text, I cannot complete the requested information regarding acceptance criteria and a study demonstrating the device meets those criteria for the new indication of abdominal aortic calcification visualization. The document explicitly states: "No changes to the manner in which the MXA or MXA-II software operates will be made for the modification to the intended use, nor will any changes be made in the device's performance specifications."

To fully answer your request for this specific modification, additional documentation detailing the validation of the new indication (abdominal aortic calcification visualization) would be required.

However, I can extract the general information presented in the 510(k) summary:

  • Device Trade Name: IVA (MXA -II) software option for the X-Ray bone densitometers.
  • Device Common Name: X-Ray Bone Densitometer
  • Intended Use: IVA scans are intended for the visualization or quantitative assessment of vertebral bone deformities. IVA also allows the visualization of abdominal aortic calcification, and, if present, clinical correlation may be advised since abdominal aortic calcification may be associated with cardiovascular disease.
  • Predicates: Hologic Vertebral Morphometry Software Option (K941362), Hologic Vertebral Morphometry Software II Option (K992775), General use x-ray systems (e.g., Toshiba KXO-80G x-ray generator; K945668).

Without a study specifically evaluating the IVA (MXA-II) software against acceptance criteria for abdominal aortic calcification visualization, I cannot provide the requested details.

{0}------------------------------------------------

510(k) Summary

APR 2 4 2006

Modified Indication for Use for IVA

Submitter:Hologic, Inc.
Submitter Address:35 Crosby DriveBedford, MA 01730
Contact Person:Jeanette Schier-Pugsley, Regulatory Affairs Manager
Phone Number:(781) 999-7406
Fax:(781) 280-0614
Date of Submission:January 13, 2006
Device Trade Name:IVA (MXA -II) software option for the X-Ray bonedensitometers.
Device Common Name:X-Ray Bone Densitometer
Device Classification:Bone Densitometer21 CFR 892.1170
Predicate Devices:Hologic Vertebral Morphometry Software Option (K941362)Hologic Vertebral Morphometry Software II Option (K992775)General use x-ray systems (e.g., Toshiba KXO-80G x-raygenerator; K945668)
Device Description andIntended Use:IVA scans are intended for the visualization or quantitativeassessment of vertebral bone deformities. IVA also allows thevisualization of abdominal aortic calcification, and, if present,clinical correlation may be advised since abdominal aorticcalcification may be associated with cardiovascular disease.
Substantial Equivalence:No changes to the manner in which the MXA or MXA-IIsoftware operates will be made for the modification to theintended use, nor will any changes be made in the device'sperformance specifications. The modified intended usestatement of the presently commercialized Hologic MXA-IIsoftware option is substantially equivalent to those of theMXA-II and MXA software options (K992775 and K941362,

{1}------------------------------------------------

respectively), and that of general diagnostic x-ray systems, such as the Toshiba KXO-80G x-ray generator (K945668). such as the Toshiba IRTS be be submission support the new claim.

:

:

.

Image /page/1/Picture/3 description: The image shows a sequence of numbers, specifically '000092'. The numbers are printed in a clear, sans-serif font, and they appear to be part of a larger numerical sequence or identifier. The digits are evenly spaced and easily readable, suggesting they are from a digital display or a well-printed document.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo consists of a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is an emblem that resembles an eagle or bird-like figure with three stylized wing segments. The emblem is black, and the text is also in black.

Food and Drug Administration 9200 Corporate Blvd. Rockville MD 20850

APR 2 4 2006

Ms. Jeanette Schier-Pugsley, RAC Regulatory Affairs Manager Hologic, Inc. 35 Crosby Drive BEDFORD MA 01730

Re: K060111

Trade/Device Name: IVA (MXA II Software Option) Regulation Number: 21 CFR 892.1170 Regulation Name: Bone densitometer Regulatory Class: II Product Code: KGI Dated: April 3, 2006 Received: April 4, 2006

Dear Ms. Schier-Pugsley:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Image /page/2/Picture/10 description: The image shows a circular logo with the text "1906 - 2006" at the top and "FDA" in large, bold letters in the center. Below the FDA letters, the word "Centennial" is written in a cursive font. The logo is surrounded by text that reads "Dedicated to Compliance, Protection, and the Norms of Public". The bottom of the image shows the words "ting Public" in a cursive font.

{3}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:

21 CFR 876.xxxx(Gastroenterology/Renal/Urology)240-276-0115
21 CFR 884.xxxx(Obstetrics/Gynecology)240-276-0115
21 CFR 892.xxxx(Radiology)240-276-0120
Other240-276-0100

Also, please note the regulation cntitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html

Sincerely yours,

Nancy C. Brogdon

Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use Statement

510(k) Number (if known): K060111

Device Name: _ IVA (MXA II Software Option)_

Indications for Use:

IVA scans are intended for the visualization or quantitative assessment of vertebral bone I v I seats are intended the visualization of abdominal aortic calcification, and, if present, clinical correlation may be advised since abdominal aortic calcification may be associated with cardiovascular disease.

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use
(Per 21 CFR 801.109)

OR

Over-The-Counter-Use
(Optional Format 1)

Nancy Burton

(Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices 060111 510(k) Number .

000089

§ 892.1170 Bone densitometer.

(a)
Identification. A bone densitometer is a device intended for medical purposes to measure bone density and mineral content by x-ray or gamma ray transmission measurements through the bone and adjacent tissues. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II.