K Number
K060035

Validate with FDA (Live)

Device Name
MDXNET
Manufacturer
Date Cleared
2006-03-09

(63 days)

Product Code
Regulation Number
892.2050
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

MDxNet is intended for use under physician's orders and/or oversight and management.

MDxNet is indicated for acquisition of related sets of 2D ultrasound images and 3D reconstruction of ultrasound images in the adult and pediatric patient utilizing existing ultrasound systems. It is intended to acquire, analyze, store and retrieve digital ultrasound images for computerized 2D and 3D image processing. MDxNet is indicated to acquire, digitize, archive and retrieve single or sequences of 2D ultrasound images.

MDxNet is intended as a general purpose digital image processing and archiving tool for use in abdominal, pelvic, fetal, cardiac, peripheral vascular and neurovascular imaging.

MDxNet add-on indications are for the assessment of pulmonary function to acquire, analyze, store and retrieve this assessment in male/female adult and pediatric patients utilizing FDA approved Spirometry. MDxNet additional add-on will provide 12-lead resting electrocardiogram (ECG) which permits the detection of abnormalities in the transmission of the cardiac impulse through the heart muscle and serves as an important aid in the diagnosis of heart ailments through utilizing 12-lead ECG.

Device Description

MDxNet consists of MDxPAC, MDxStation (PC or Notebook, laptop computer), MDxServer, and the proprietary software component called MDxView. MDxPAC consists of limited hardware, that bundle and power FDA approved Commercial Off-The-Shelf devices (COTS). MDxNet features an integrated 2D/3D streaming engine which permits PC's or notebooks to control the server and to review 2D and 3D reconstructions. MDxNet is capable of image review, archiving, data collection, database maintenance, reporting and basic 3D capabilities including color rendering. Upon receipt of FDA approval to market MDxNet, finished devices will be distributed only under direction and/or orders of physicians.

MDxNct inherits and integrates concepts first pioneered in the National Aeronautics and Space Administration (NASA) sponsored Telemedicine Instrument Pack (TIP) project and the Defense Advanced Research Projects Agency (DARPA) sponsored Medical Ultrasound Three-dimensional, Portable with Advanced Communication (MUSTPAC) project, and its predecessors. It expands capability to include a server component, augments data acquisition to include wireless options, redesigns the Graphical User Interface (GUI) for more intuitive interaction and provides improved packaging for better usability.

MDxNet has the capability to record ultrasound transducer spatial position in six degrees of freedom during use. Coordinate tracking is achieved with a miniature magnetic field sensor within a transmitted pulsed magnetic field. This is done by attaching a plastic holding plate to the FDA approved probe of the host Ultrasound system, to which the receiver of an electromagnetic sensor device is attached.

2D ultrasound images are acquired sequentially in a series of steps as the ultrasound transducer is moved across the patient scan site. The resulting set of digitized 2D images is then converted into a 3D data volume.

IQmark Digital ECG and Spirometer both FDA approved devices have been integrated into MDxNet as requested by the end user of the validation study. The Midmark software for each of these devices and indications of use have not been altered, though permit the end user as demonstrated in both clinical studies to capture additional potentially key clinical results within MDxNet for data collection, review, comparison, reporting and archiving.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the CyberMDx MDxNet device, based on the provided FDA 510(k) Premarket Notification:

The provided document K060035 for CyberMDx, Inc. MDxNet is a 510(k) Premarket Notification. It explicitly states that MDxNet is an "Image Processing Server" and a "System, Image Processing, Radiological." This type of device primarily functions as a tool for displaying, processing, archiving, and retrieving medical images. It's important to note that this document does not describe a device that performs AI-driven interpretation or diagnosis, nor does it present "acceptance criteria" in the sense of performance metrics (like sensitivity, specificity, or AUC) that an AI algorithm would need to meet.

Instead, the submission focuses on demonstrating substantial equivalence to existing legally marketed predicate devices by comparing features and intended uses. The "studies" mentioned are primarily about verifying the device's functionality and safety as an image processing system rather than validating the diagnostic accuracy of an AI component.

Therefore, the requested information, particularly regarding AI-specific performance metrics, ground truth, expert adjudication, sample sizes for AI training/test sets, and MRMC studies, is largely not applicable (N/A) in the context of this 2006 510(k) for an image processing server. The device predates the widespread adoption and regulatory requirements for AI in medical imaging.

However, I will extract and organize the relevant information that is present in the document.


Acceptance Criteria and Reported Device Performance

Given that this is an image processing server, the "acceptance criteria" discussed are primarily related to its functional capabilities and conformity to design intent and quality standards, rather than diagnostic accuracy metrics. The document emphasizes functional equivalence to predicate devices.

Acceptance Criteria (Functional/Technical)Reported Device Performance (as stated in the document)
Functional Equivalence to Predicates (based on features like 2D/3D imaging, archiving, reformatting, etc.)"Substantial equivalence to the following legally marketed predicate devices with the same or similar indications for use has been demonstrated by a comparison of product features as described in the labeling and promotional literature for the predicate devices." (Section 5.3) Table 5-1 explicitly shows "Yes" for MDxNet across all listed common features (2D Image Review, Multiplanar Reformatting, 3D Volume Rendering, Maximum Intensity Projection, Image Archiving, Image Filming, Image Transfer/Network Connectivity, Examination of 2D image as 3D volume, Comparison of Multiple Scans, General medical imaging indications, Computer Platform, Image Display, Operating System).
Accuracy and Performance to Specification (general system performance)"bench testing was conducted to establish MDxNet's accuracy, performance to specification, as well as testing to accepted industry standards..." (Section 5.3) "Test results to date included elsewhere support the conclusion that the actual device, software and hardware satisfies the design intent. Actual device performance as tested internally and by third parties (including controlled clinical trials) conforms to the system performance standards and has met expectations." (Section 5.6)
Adherence to Quality System Regulations (e.g., FDA 21 CFR 820)"Continual nonclinical testing is being conducted through internal company procedures according the FDA 21CFR820 Quality System and Current Good Manufacturing Practices." (Section 5.6)
Hardware Testing"All hardware testing will be conducted and meet the specified acceptance criteria before the device is marketed." (Section 4, Hardware & Software Information)
Software Level of Concern"The level of concern relative to the software has been determined as minor using the decision tree provided in the revised FDA Software Guidance, May 11, 2005." (Section 4, Hardware & Software Information)

Study Details (Focusing on Functional & Safety Verification)

As noted, this is not an AI performance study. The "clinical studies" mentioned are for system validation, not diagnostic accuracy of an AI algorithm.

  1. Sample size used for the test set and the data provenance:

    • Test Set Sample Size: The document states that "Two MDxNet Systems were developed to acquire potential end user input and to allow validation of the software, hardware, server and integration under Investigational Review Board (IRB) approval in a medical clinic setting." It also mentions "controlled clinical trials." However, no specific sample size (number of patients or images) for any particular "test set" is provided.
    • Data Provenance: "Medical clinic setting" and "third parties (including controlled clinical trials)." No specific countries are mentioned, but the company is US-based (Providence, RI). The studies appear to be prospective as they were conducted for validation.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • N/A. The document does not describe a clinical diagnostic "ground truth" establishment process for imaging analysis that would require experts. The device is a tool, not an AI for diagnosis. The validation focuses on whether the system functions as intended and safely integrates other FDA-approved devices (Spirometry, ECG).
  3. Adjudication method for the test set:

    • N/A. No expert adjudication is described as the device is not performing diagnostic interpretations that would require it.
  4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No. This is not an AI-assisted device. Therefore, no MRMC study comparing human readers with and without AI assistance was conducted or is applicable.
  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • N/A. The MDxNet is a system that allows users to process and view images. It is not a standalone diagnostic algorithm.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • N/A. For the image processing and display features, "ground truth" would relate to the accuracy of image reconstruction, archiving, and retrieval, which are engineering and system performance validations rather than clinical diagnostic ground truth. For the integrated ECG and Spirometry, the "ground truth" would be the direct output of those separate FDA-approved devices. The main focus of the study was about the integration and display of these clinical results, not their diagnostic determination by MDxNet.
  7. The sample size for the training set:

    • N/A. The document does not mention any "training set" in the context of machine learning or AI.
  8. How the ground truth for the training set was established:

    • N/A. No training set for AI is mentioned.

Summary of Device Context:

The CyberMDx MDxNet, as described in this 2006 510(k) submission, is a general-purpose image processing server. Its primary function is to:

  • Acquire, digitize, archive, and retrieve 2D ultrasound images.
  • Perform 3D reconstruction of ultrasound images.
  • Provide tools for image review, multiplanar reformatting, volume rendering, and maximum intensity projection.
  • Integrate data from external FDA-approved devices (like spirometers and 12-lead ECGs) for capture, review, comparison, reporting, and archiving.

The submission demonstrates substantial equivalence to predicate devices by virtue of its functional features as an imaging workstation/server, not by demonstrating AI-driven diagnostic performance. The "studies" mentioned are for systematic validation of the device's ability to perform these functions safely and effectively, and to integrate data from other medical devices, within a clinical setting under IRB approval.

{0}------------------------------------------------

FDA; 510(k) Premarket Notification

Ko60035

CyberMDx, Inc. MDxNet

2006 MAR 9

510(k) Premarket NotificationMDxNetComments of the country of the program and the county of
Summary of Safety and Effectiveness InformationCall Concession Company Company Company Company Company Company Company Company Company Company Company Company Company Company Company Company Company Company Company Compan

Section 5.0: 510K SUMMARY

5.1 Contact Information
Establishment:CyberMDx, Inc.Richmond Square, Suite 166WProvidence, RI 02906
Official Correspondent:John Coleman, PhDChief Technology Officer (CTO)CyberMDx, IncRichmond Square, Suite 166WProvidence, RI 02906Phone: (401) 274-5633FAX: (401) 633-6544
Prepared By:Bonne Farberow, RN, CCRA, CCRPRegulatory Affairs Consultant53 Meetinghouse CtShamong, NJ 08088
Date Summary Prepared:December 24, 2005
Section 5.2 Name of Device
Proprietary:MDxNet
Common/Usual Name:Image Processing Server
Classification:System, Image Processing, RadiologicalClass II
Product Code:90-LLZ

{1}------------------------------------------------

Section 5.3 Substantially Equivalent Devices

Substantial equivalence to the following Icgally marketed predicate devices with the same or similar indications for use has been demonstrated by a comparison of product features as described in the labeling and promotional literature for the predicate devices. In addition, bench testing was conducted to establish MDxNet's accuracy, performance to specification, as well as testing to accepted industry standards under Investigational Review Board (IRB) approval and in the laboratory setting. The predicate devices are as follows

  • 트 AquariusNet Server, Teracon, Inc., K012086
  • . TeraRecon IiVS, TeraRecon, Inc., K994329
  • l Imatron Ultra Access, Acculmage, 972903
  • " EchoTech 3D Imaging Systems, K013088

These devices are substantially equivalent in terms of basic design, features and intended use, though MDxPAC and MDxServer have the additional capability pending end user necds and requirements of integration of FDA approved devices such as the IQmark Spirometry and ECG.

Table 5-1 compares MDxNet with features of its Predicate Devices.

{2}------------------------------------------------

FeatureMDxNetSubmission - PendingAquarius NetServer K012086Imatron UltraAccess 972903TeraRecon IiVSK994329EchoTech 3DImaging SystemsK013088
2 D Image ReviewYesYesYesYesYes
Multiplanar reformattingYesYesYesYesYes
3D Volume RenderingYesYesYesYesYes
Maximum IntensityProjectionYesYesYesYesYes
Image ArchivingYesYesYesYesYes
Image FilmingYesYesYesYesYes
Lossy ImageCompressionNoYesYesYesYes
Image Transfer orNetwork ConnectivityYesYesYesYesYes
Examination of 2Dimage as a 3D volumeYesYesYesYesYes
CT and Calcium ScanNoYesYesYesNo
Comparison of MultipleScans or Side by SideComparisonYesYesYesYesYes
Indications of use –general medical imagingunits or sites; acquisition,analysis, store andretrieve digital ultrasoundYesYesYesYesYes
Computer PlatformYesYesYesYesYes
Image DisplayYesYesYesYesYes
Operating SystemYesYesYesYesYes
*Capture and store ECGIndication (s); Assessheart wall and/orfunction forabnormalitiesYesNoNoNoNo
*Capture and storeSpirometry (VC, etc.)Indication (s); AssessPulmonary functionYesNoNoNoNo

Table 5-1: Predicate Devices

*Utilizing FDA approved devices; Terason K992505, IQmark ECG (Registration #2081230) and IQmark Spiromcter K002499

Note - Tempus 2000; K010436 (for integration of COTs)

{3}------------------------------------------------

5.4 Device Description and Function

MDxNet consists of MDxPAC, MDxStation (PC or Notebook, laptop computer), MDxServer, and the proprietary software component called MDxView. MDxPAC consists of limited hardware, that bundle and power FDA approved Commercial Off-The-Shelf devices (COTS). MDxNet features an integrated 2D/3D streaming engine which permits PC's or notebooks to control the server and to review 2D and 3D reconstructions. MDxNet is capable of image review, archiving, data collection, database maintenance, reporting and basic 3D capabilities including color rendering. Upon receipt of FDA approval to market MDxNet, finished devices will be distributed only under direction and/or orders of physicians.

MDxNct inherits and integrates concepts first pioneered in the National Aeronautics and Space Administration (NASA) sponsored Telemedicine Instrument Pack (TIP) project and the Defense Advanced Research Projects Agency (DARPA) sponsored Medical Ultrasound Three-dimensional, Portable with Advanced Communication (MUSTPAC) project, and its predecessors. It expands capability to include a server component, augments data acquisition to include wireless options, redesigns the Graphical User Interface (GUI) for more intuitive interaction and provides improved packaging for better usability.

MDxNet has the capability to record ultrasound transducer spatial position in six degrees of freedom during use. Coordinate tracking is achieved with a miniature magnetic field sensor within a transmitted pulsed magnetic field. This is done by attaching a plastic holding plate to the FDA approved probe of the host Ultrasound system, to which the receiver of an electromagnetic sensor device is attached.

2D ultrasound images are acquired sequentially in a series of steps as the ultrasound transducer is moved across the patient scan site. The resulting set of digitized 2D images is then converted into a 3D data volume.

IQmark Digital ECG and Spirometer both FDA approved devices have been integrated into MDxNet as requested by the end user of the validation study. The Midmark software for each of these devices and indications of use have not been altered, though permit the end user as demonstrated in both clinical studies to capture additional potentially key clinical results within MDxNet for data collection, review, comparison, reporting and archiving.

{4}------------------------------------------------

FDA; 510(k) Premarket Notification

Hardware & Software Information:

The MDxNet utilizes standard "off the shelt" personal computer systems and Bluetooth wireless as its hardware platform. The MDxPAC component includes a Class B magnetic tracking as well as it's own power supply including rechargeable batteries. The software requires the minimum use of Windows XP operating system and a Pentium III - class processor or its equivalent. All hardware testing will be conducted and meet the specified acceptance criteria before the device is marketed.

The level of concern relative to the software has been determined as minor using the decision tree provided in the revised FDA Software Guidance, May 11, 2005.

5.5 Indications/Intended Use:

MDxNct is intended for use under physician's orders and/or oversite and management. MDxNet is indicated for acquisition of related sets of 2D ultrasound images, 3D reconstruction of ultrasound images. MDxNet system is indicated to acquire, digitize, archive and retrieve single and sequences of 2D ultrasound images.

The assessment of pulmonary function in male/female patients through Spirometry and to provide 12-lead resting electrocardiogram (ECG) utilizing FDA approved devices which permits the detection of abnormalities in the transmission of the cardiac impulse through the heart muscle and serves as an important aid in the diagnosis of heart ailments.

Contraindications:

MDxNet is not intended for long term monitoring. MDxNet is not intended to be used in strong magnetic or electromagnetic fields which are generated for medical purposes, i.c. Magnetic Resonance Imaging (MRI).

Disposable pneumotach mouthpieces are clean though are not sterile and should not be placed over open wounds that are prone to infection. There are no other known medical contraindications other than the physical limitations of the patient.

5.6 Technological Comparison with Predicate Devices

Two MDxNet Systems were developed to acquire potential end user input and to allow validation of the software, hardware, server and integration under Investigational Review Board (IRB) approval in a medical clinic setting.

Clinical testing was conducted according to an IRB approved protocol. Continual nonclinical testing is being conducted through internal company procedures according the FDA 21CFR820 Quality System and Current Good Manufacturing Practices. Additional nonclinical testing will be conducted by a third party laboratory. Test results to date included elsewhere support the conclusion that the actual device, software and hardware satisfies the design intent. Actual device performance as tested internally and by third parties (including controlled clinical trials) conforms to the system performance standards and has met expectations."

{5}------------------------------------------------

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MAR 9 2006

John Coleman, Ph.D. Chief Technology Officer cyberMDx, Inc. 1 Richmond Square, Suite 166W PROVIDENCE RI 02906

Re: K060035 Trade/Device Name: MDxNet Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: December 28, 2005 Received: January 5, 2006

Dear Dr. Coleman:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act 's requirements, including, but not limited to registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.

{6}------------------------------------------------

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:

21 CFR 876.xxxx(Gastroenterology/Renal/Urology)240-276-0115
21 CFR 884.xxxx(Obstetrics/Gynecology)240-276-0115
21 CFR 892.xxxx(Radiology)240-276-0120
Other240-276-0100

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely vours.

Nancy C. Brigton

Naney C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{7}------------------------------------------------

Section 4.0: Indications for Use

510(k) Number (if known): Device Name: MDxNet Indications for Use: MDxNet is intended for use under physician's orders and/or oversight and management.

MDxNet is indicated for acquisition of related sets of 2D ultrasound images and 3D reconstruction of ultrasound images in the adult and pediatric patient utilizing existing ultrasound systems. It is intended to acquire, analyze, store and retrieve digital ultrasound images for computerized 2D and 3D image processing. MDxNet is indicated to acquire, digitize, archive and retrieve single or sequences of 2D ultrasound images.

MDxNet is intended as a general purpose digital image processing and archiving tool for use in abdominal, pelvic, fetal, cardiac, peripheral vascular and neurovascular imaging.

MDxNet add-on indications are for the assessment of pulmonary function to acquire, analyze, store and retrieve this assessment in male/female adult and pediatric patients utilizing FDA approved Spirometry. MDxNet additional add-on will provide 12-lead resting electrocardiogram (ECG) which permits the detection of abnormalities in the transmission of the cardiac impulse through the heart muscle and serves as an important aid in the diagnosis of heart ailments through utilizing 12-lead ECG.

MDxNet intended use is for medical clinics, including though not limited to infirmary, hospital, military, remote and urban clinics.

Contraindications:

MDxNet is not intended to be used in strong magnetic or electromagnetic fields which are generated for medical purposes, i.e. Magnetic Resonance Imaging (MRI).

MDxNet is not intended for long term monitoring. This device is not intended to nor does it sound alarms for any physiological parameters. The ECG is not intended to be utilized on patients with prosthetic limbs.

Disposable pneumotach mouthpieces are clean though are not sterile, and should not be placed over open wounds that are prone to infection. There are no other known medical contraindications other than the physical limitations of the patient.

Prescription UseX
(Part 21 CFR 801 Subpart D)

AND/
OR

Over-The-Counter Use
(21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-

CONTINUE ON ANOTHER PAGE OF NEEDED)

(Division Sign-Off)

ff Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number ________________________________________________________________________________________________________________________________________________________________

N/A