K Number
K060008

Validate with FDA (Live)

Date Cleared
2006-02-02

(30 days)

Product Code
Regulation Number
876.5860
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Aquadex FlexFlow™ System is indicated for:
• Temporary (up to 8 hours) ultrafiltration treatment of patients with fluid overload who have failed diuretic therapy, and
• Extended (longer than 8 hours) ultrafiltration treatment of patients with fluid overload who have failed diuretic therapy and require hospitalization.
All treatments must be administered by a health care provider, under physician prescription, both of whom having received training in extracorporeal therapies.

Device Description

The Aquadex FlexFlow System removes excess fluid from the patient in fluid overload by ultrafiltration of blood across a hollow-fiber hemofilter at the clinician selected rate. The system is comprised on a console mounted on a cart, proprietary software and accessories (venous access catheters, extensions and a blood pump circuit). Patient access is obtained via either peripheral or central venous veins. This 510(k) is to add an alternate blood pump circuit which has a standard male and female luer adapter just before the filter on the withdrawal line to allow the optional use of a blood chamber cuvette to interface with a stand alone Hct monitor.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the Aquadex FlexFlow™ System, based on the provided 510(k) summary:

This submission is for an alteration to an existing device (K050609), specifically to add an alternate blood pump circuit. Therefore, the focus of the testing is on validating this change, not re-proving the core functionality of the entire system from scratch.


1. Table of Acceptance Criteria and Reported Device Performance:

The provided 510(k) summary does not explicitly state quantitative acceptance criteria (e.g., specific thresholds for flow rates, pressure, or other performance metrics). Instead, it states that "The data demonstrated the Aquadex FlexFlow System continues to be safe and effective."

Acceptance Criteria (Implicit)Reported Device Performance
Safety: Device operates without introducing new hazards or compromising existing safety features."The data demonstrated the Aquadex FlexFlow System continues to be safe..."
Effectiveness: Device maintains its intended functionality for ultrafiltration despite the modification."...and effective."
Functionality of alternate blood pump circuit with standard luer adapters and interface with stand-alone Hct monitor.Bench tests were performed to validate the incorporation of the blood chamber placement in-line on the withdrawal line of the blood circuit.
No software change required due to the modification."No software change was required."

2. Sample Size Used for the Test Set and Data Provenance:

  • Sample Size: Not specified. The summary only mentions "Bench tests were performed." It does not provide the number of units tested or the number of test runs.
  • Data Provenance: Not specified, but by nature of "bench tests," this would be laboratory-generated data, not from human subjects. It would be prospective in the sense that the tests were conducted specifically for this 510(k) submission, but it's not clinical data.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

  • This question is not applicable to this specific 510(k) submission. The validation was based on "bench tests" for engineering performance, not on clinical interpretation or subjective assessment requiring expert review to establish ground truth.

4. Adjudication Method for the Test Set:

  • Not applicable. Bench tests typically involve objective measurements against engineering specifications or established performance characteristics, and thus do not usually require a consensus or adjudication process like clinical studies with subjective endpoints.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done:

  • No, an MRMC comparative effectiveness study was not done. This 510(k) is for a minor modification to an existing device validated through bench testing, not a new diagnostic or therapeutic device requiring a clinical comparison of human reader performance.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

  • Not applicable. This device is a medical system (hardware and software for ultrafiltration) and not an algorithm or AI system subject to "standalone performance" testing in the AI context. While it has proprietary software, the testing described is on the mechanical and functional integration of a new blood circuit.

7. The Type of Ground Truth Used:

  • For the bench testing, the "ground truth" would be engineering specifications, established performance parameters of the original device, and objective measurements of the modified system (e.g., flow rates, pressure integrity, absence of leaks, compatibility with the Hct monitor). It's not expert consensus, pathology, or outcomes data.

8. The Sample Size for the Training Set:

  • Not applicable. This is not an AI/machine learning device that requires a training set in the conventional sense. The "proprietary software" is mentioned, but no changes were made to it, and its development would have preceded this specific 510(k).

9. How the Ground Truth for the Training Set Was Established:

  • Not applicable. As this is not an AI/machine learning device with a training set, the question of how ground truth was established for it is irrelevant in this context.

{0}------------------------------------------------

510(k) Summary (Prepared on December 30, 2005)

This 510(k) Summary is submitted in accordance with 21 CFR 807.92.

Trade Names:Aquadex FlexFlow™ System (trademark pending)
Manufacturer:CHF Solutions, Inc., Suite 170 - 7601 Northland Drive,Brooklyn Park, MN 55428
OfficialContact:Chris ScavottoQA ManagerTelephone: 763-463-4621Fax: 763-463-4606
DeviceGeneric Name:Ultrafiltration (Aquapheresis) System
Classification:High permeability dialysis systems - classified as Class II
PredicateDevices:Aquadex FlexFlow System(K050609)Hemametrics Blood Chamber and BloodChamber tubing adaptor set CRIT-LINEIII TQA Monitor(K982412 & K011741)
DeviceDescription:The Aquadex FlexFlow System removes excess fluid from the patient in fluidoverload by ultrafiltration of blood across a hollow-fiber hemofilter at theclinician selected rate. The system is comprised on a console mounted on acart, proprietary software and accessories (venous access catheters,extensions and a blood pump circuit). Patient access is obtained via eitherperipheral or central venous veins. This 510(k) is to add an alternate bloodpump circuit which has a standard male and female luer adapter just beforethe filter on the withdrawal line to allow the optional use of a blood chambercuvette to interface with a stand alone Hct monitor.
Indicationfor Use:The Aquadex FlexFlow™ System is indicated for:• Temporary (up to 8 hours) ultrafiltration treatment of patients with fluidoverload who have failed diuretic therapy, and• Extended (longer than 8 hours) ultrafiltration treatment of patients withfluid overload who have failed diuretic therapy and require hospitalization.All treatments must be administered by a health care provider, underphysician prescription, both of whom having received training inextracorporeal therapies.
Safety &Performance:Bench tests were performed to validate the incorporation of the bloodchamber placement in-line on the withdrawal line of the blood circuit. Nosoftware change was required. The data demonstrated the AquadexFlexFlow System continues to be safe and effective.
Conclusion:Based on the similar intended use, patient population, technologycharacteristics, and performance as assessed with bench testing the alternateUF500 blood circuit (PN A1700) has been shown to be safe and effective, issubstantially equivalent‡ and considered acceptable for the intended use.

4 This document uses the term "substantial equivalent" as intended in 21 CFR 807.87 and not as defined in Title 36 of the U.S. Code.

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird symbol with three curved lines representing the bird's body and head. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird symbol.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

FEB 2 2006

2006

Mr. Chris Scavotto Quality Assurance Manager CHF Solutions, Inc. 7601 Northland Drive, Suite 170 BROOKLYN PARK MN 55428

Re: K060008

Trade/Device Name: Aquadex FlexFlow™ System Regulation Number: 21 CFR §876.5860 Regulation Name: High permeability hemodialysis system Regulatory Class: II Product Code: KDI Dated: December 30, 2005 Received: January 3, 2006

Dear Mr. Scavotto:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. However, you are responsible to determine that the medical devices you use as components in the kit have either been determined as substantially equivalent under the premarket notification process (Section 510(k) of the act), or were legally on the market prior to May 28, 1976, the enactment date of the Medical Device Amendments. Please note: If you purchase your device components in bulk (i.e., unfinished) and further process (e.g., sterilize) you must submit a new 510(k) before including these components in your kit. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, and labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{2}------------------------------------------------

Page 2 - Mr. Chris Scavotto

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration (21 CFR Part 807); listing (21 CFR Part 807), labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on the labeling regulation, please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll free number (800) 638-2041 or (301) 443-6597, or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely vours,

Daniel R. Lyman

Nancy C.
Director,
Abdomi

Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

INDICATIONS FOR USE STATEMENT (Page 1 of 1)

510(k) Number (if know): __Kf85_Kolg0008

Device Name: Aquadex FlexFlow™ System

FDA's Statement of the Indication For Use for Device:

The Aquadex FlexFlow™ System is indicated for:

  • Temporary (up to 8 hours) ultrafiltration treatment of patients with fluid . overload who have failed diuretic therapy, and
  • Extended (longer than 8 hours) ultrafiltration treatment of patients with . fluid overload who have failed diuretic therapy and require hospitalization.

All treatments must be administered by a health care provider, under physician prescription, both of whom having received training in extracorporeal therapies.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

Concurrence of CDRH, Office of Device Evaluation (ODE)

OR

Daniel C. Lynam

(Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number

Prescription Use
(Per 21 CFR 801.109)

Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________

CHF Solutions, Inc. UF500 with Luer Junction & Hct Adaptor

§ 876.5860 High permeability hemodialysis system.

(a)
Identification. A high permeability hemodialysis system is a device intended for use as an artificial kidney system for the treatment of patients with renal failure, fluid overload, or toxemic conditions by performing such therapies as hemodialysis, hemofiltration, hemoconcentration, and hemodiafiltration. Using a hemodialyzer with a semipermeable membrane that is more permeable to water than the semipermeable membrane of the conventional hemodialysis system (§ 876.5820), the high permeability hemodialysis system removes toxins or excess fluid from the patient's blood using the principles of convection (via a high ultrafiltration rate) and/or diffusion (via a concentration gradient in dialysate). During treatment, blood is circulated from the patient through the hemodialyzer's blood compartment, while the dialysate solution flows countercurrent through the dialysate compartment. In this process, toxins and/or fluid are transferred across the membrane from the blood to the dialysate compartment. The hemodialysis delivery machine controls and monitors the parameters related to this processing, including the rate at which blood and dialysate are pumped through the system, and the rate at which fluid is removed from the patient. The high permeability hemodialysis system consists of the following devices:(1) The hemodialyzer consists of a semipermeable membrane with an in vitro ultrafiltration coefficient (K
uf ) greater than 8 milliliters per hour per conventional millimeter of mercury, as measured with bovine or expired human blood, and is used with either an automated ultrafiltration controller or anther method of ultrafiltration control to prevent fluid imbalance.(2) The hemodialysis delivery machine is similar to the extracorporeal blood system and dialysate delivery system of the hemodialysis system and accessories (§ 876.5820), with the addition of an ultrafiltration controller and mechanisms that monitor and/or control such parameters as fluid balance, dialysate composition, and patient treatment parameters (e.g., blood pressure, hematocrit, urea, etc.).
(3) The high permeability hemodialysis system accessories include, but are not limited to, tubing lines and various treatment related monitors (e.g., dialysate pH, blood pressure, hematocrit, and blood recirculation monitors).
(b)
Classification. Class II. The special controls for this device are FDA's:(1) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Device—Part I: Evaluation and Testing,’ ”
(2) “Guidance for the Content of 510(k)s for Conventional and High Permeability Hemodialyzers,”
(3) “Guidance for Industry and CDRH Reviewers on the Content of Premarket Notifications for Hemodialysis Delivery Systems,”
(4) “Guidance for the Content of Premarket Notifications for Water Purification Components and Systems for Hemodialysis,” and
(5) “Guidance for Hemodialyzer Reuse Labeling.”