K Number
K053529

Validate with FDA (Live)

Manufacturer
Date Cleared
2006-01-17

(29 days)

Product Code
Regulation Number
862.1345
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The OneTouch Ultra 2 Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose in fresh capillary whole blood. The OneTouch Ultra 2 System is intended for use outside the body (in vitro diagnostic use) by people with diabetes at home and/or by healthcare professionals in a clinical setting as an aid to monitor the effectiveness of diabetes control.

The OneTouch Ultra 2 Blood Glucose Monitoring System is specifically indicated for use on the finger, forearm or palm.

Device Description

The OneTouch Ultra 2 Blood Glucose Monitoring System consists of the OneTouch Ultra 2 Meter, OneTouch Ultra Test Strips (provided separately), OneTouch Ultra Control Solution, OneTouch UltraSoft Blood Sampler and OneTouch UltraSoft Sterile Lancets. The predicate OneTouch® Ultra® Meter has been modified to produce the OneTouch Ultra 2 Meter; there are no changes to other system testing components compared to the currently marketed product.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the OneTouch® Ultra® 2 Blood Glucose Monitoring System, based on the provided document:

This 510(k) summary explains that the OneTouch® Ultra® 2 Blood Glucose Monitoring System is a modified version of an existing device (OneTouch® Ultra® Blood Glucose Monitoring System). The submission focuses on demonstrating substantial equivalence to the predicate device, rather than establishing entirely new performance criteria or conducting a comprehensive clinical trial from scratch.

Therefore, the acceptance criteria and performance data provided relate to demonstrating this equivalence.


1. Table of Acceptance Criteria and Reported Device Performance

Acceptance CriterionReported Device Performance
Substantial Equivalence to Predicate DeviceA meter equivalence study demonstrated that the OneTouch® Ultra® 2 Blood Glucose Monitoring System and the currently marketed OneTouch® Ultra® Blood Glucose Monitoring System are substantially equivalent.
Performance, Safety, and Effectiveness (Design Verification)Design verification (including software verification and validation testing) confirmed that the performance, safety and effectiveness of the OneTouch Ultra 2 Blood Glucose Monitoring System was equivalent with that of the predicate device.
Compliance with ISO 15197:2003(E)The modified meter was tested in accordance with ISO 15197:2003(E).
Ease of Use (Human Factors and User Acceptance)A consumer evaluation to assess ease of use (human factors and user acceptance) was conducted.
Labeling ComprehensionReading level assessment of the product labeling was conducted and validated for comprehension.

2. Sample Size Used for the Test Set and Data Provenance

The document does not explicitly state the specific sample size for the "meter equivalence study" or the "consumer evaluation."

  • Sample Size for Test Set: Not explicitly stated.
  • Data Provenance: Not explicitly stated, but generally, such studies for medical devices seeking FDA clearance are conducted with data relevant to the target population (e.g., people with diabetes). It isn't specified whether the data was retrospective or prospective, or the country of origin.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

This information is not provided in the document. For a blood glucose monitoring system, "ground truth" typically refers to reference measurements from a laboratory analyzer, not expert consensus on interpretations.


4. Adjudication Method for the Test Set

Adjudication methods (e.g., 2+1, 3+1) are typically used in studies involving expert interpretation of medical images or complex diagnostic findings. For a blood glucose meter, where the output is a quantitative value, such adjudication methods are not applicable. The ground truth would usually be established by a reference laboratory method.


5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

An MRMC study is not applicable to this device. This type of study assesses how human readers' performance changes with or without AI assistance, which is irrelevant for a standalone blood glucose meter. The document does not mention any human-in-the-loop AI assistance.


6. Standalone Performance (Algorithm Only Without Human-in-the-Loop Performance)

Since this is a blood glucose monitoring system, its primary function is to provide a direct quantitative reading of blood glucose. The performance described (substantial equivalence, ISO 15197 compliance, design verification) inherently relates to its standalone performance as a device measuring glucose levels without human-in-the-loop "interpretation" in the same way an imaging AI would.


7. The Type of Ground Truth Used

The document does not explicitly state the specific "ground truth" method used for the meter equivalence study or ISO 15197 compliance. However, for blood glucose meters, the ground truth is almost universally established by laboratory reference methods (e.g., YSI analyzer) known for their high accuracy and precision, when comparing a point-of-care device.


8. The Sample Size for the Training Set

The document does not mention any training set or machine learning model. The OneTouch® Ultra® 2 is presented as a modification of an existing device, with no indication of new algorithms requiring a separate training process.


9. How the Ground Truth for the Training Set Was Established

As no training set or machine learning model is mentioned, this information is not applicable.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the LifeScan logo, which includes a stylized globe and the company name in bold, sans-serif font. Below the company name, it says "a Johnson & Johnson company" in a smaller, italicized font. The logo has a swoosh design element above the company name, adding a dynamic visual element to the overall design. The logo is in black and white.

K053529

IAN 17

OneTouch® Ultra® 2 Blood Glucose Monitoring System

510(k) Summary

SubmitterLifeScan, Inc.1000 Gibraltar DriveMilpitas, CA 95035-6312Contact: Lisa McGrathDate Prepared: December 16, 2005
Device NameOneTouch® Ultra® 2 Blood Glucose Monitoring SystemCommon name: Glucose test systemClassification:(1) OneTouch® Ultra® 2 Blood Glucose Meters andOneTouch® Ultra® Test Strips are Class II devices (21CFR § 862.1345)(2) OneTouch® Ultra® Control Solution is a Class Idevice (21 CFR § 862.1660)(3) OneTouch® UltraSoft® Adjustable Blood Sampler andSterile Lancets are Class I (exempt) devices (21 CFR §

System Description

The OneTouch Ultra 2 Blood Glucose Monitoring System consists of the OneTouch Ultra 2 Meter, OneTouch Ultra Test Strips (provided separately), OneTouch Ultra Control Solution, OneTouch UltraSoft Blood Sampler and OneTouch UltraSoft Sterile Lancets. The predicate OneTouch® Ultra® Meter has been modified to produce the OneTouch Ultra 2 Meter; there are no changes to other system testing components compared to the currently marketed product.

878.4800)

OneTouch Ultra Blood Glucose Monitoring System ( Ko2 4 194 ) Predicate Device

Intended Use

The OneTouch Ultra 2 Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose in fresh capillary whole blood. The OneTouch Ultra 2 System is intended for use outside the body (in vitro diagnostic use) by people with diabetes at home and/or by healthcare professionals in a clinical setting as an aid to monitor the effectiveness of diabetes control. The OneTouch Ultra 2 Blood Glucose Monitoring System is specifically indicated for use on the finger, forearm or palm.

{1}------------------------------------------------

510(K) Summary, Page 2

Comparison to Predicate Device

The modifications to the device encompass meter software/firmware, ergonomic/physical design, hardware and labeling changes. There has been no change to the intended use, operating principle, functionality or material composition of the device.

Technological Characteristics

There has been no change to the fundamental scientific technology.

Summary of Performance Characteristics

There has been no change to the performance characteristics of the system.

A meter equivalence study demonstrated that the OneTouch® Ultra® 2 Blood Glucose Monitoring System and the currently marketed OneTouch® Ultra® Blood Glucose Monitoring System are substantially equivalent.

Design verification (including software verification and validation testing) confirmed that the performance, safety and effectiveness of the OneTouch Ultra 2 Blood Glucose Monitoring System was equivalent with that of the predicate device.

The modified meter was tested in accordance with ISO 15197:2003(E) including a consumer evaluation to assess ease of use (human factors and user acceptance). In addition, reading level assessment of the product labeling was conducted and validated for comprehension.

Conclusion

The modified OneTouch Ultra 2 Blood Glucose Monitoring System is substantially equivalent to the predicate OneTouch Ultra Blood Glucose Monitoring System.

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle. The logo is black and white and appears to be a simple, vectorized design.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

JAN 1 7 2006

Ms. Lisa G. McGrath Regulatory Project Leader LifeScan, Inc. 1000 Gibraltar Drive Milpitas, CA 95035-6312

K053529 Re:

Trade/Device Name: OneTouch® Ultra® 2 Blood Glucose Monitoring System Regulation Number: 21 CFR 862.1345 Regulation Name: Glucose test system Regulatory Class: Class II Product Code: NBW, CGA Dated: December 17, 2005 Received: December 19, 2005

Dear: Ms. McGrath

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).

{3}------------------------------------------------

Page 2 -

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240) 276-0484. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely vours,

Alberto G. A.

Alberto Gutierrez, Ph.D. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use

510(k) Number (if known): K053529

Device Name: OneTouch® Ultra® 2 Blood Glucose Monitoring System

Indications For Use:

The OneTouch Ultra 2 Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose in fresh capillary whole blood. The OneTouch Ultra 2 System is intended for use outside the body (in vitro diagnostic use) by people with diabetes at home and/or by healthcare professionals in a clinical setting as an aid to monitor the effectiveness of diabetes control.

The OneTouch Ultra 2 Blood Glucose Monitoring System is specifically indicated for use on the finger, forearm or palm.

Prescription Use × (Part 21 CFR 801 Subpart D)

Image /page/4/Picture/7 description: The image shows the text "AND/OR" with the "AND" portion of the text circled. The text is written in a simple, sans-serif font. The circle around "AND" is hand-drawn and slightly irregular. The image appears to be a simple, possibly hand-drawn, graphic.

Over-The-Counter Use x (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD) Page 1 of 1

Albert Sevy
Division Sign-Off

Office of 1 a Vitro Diagnostic Device Evaluation and Salery

10(k) ........................................................................................................................................................................

§ 862.1345 Glucose test system.

(a)
Identification. A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.(b)
Classification. Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.