K Number
K053333

Validate with FDA (Live)

Manufacturer
Date Cleared
2006-05-02

(152 days)

Product Code
Regulation Number
N/A
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The EagleVision Gellansert is intended to temporarily block tear drainage by the occlusion of the canaliculus in order to:

  • Determine the potential effectiveness of permanent occlusion,
  • Temporarily treat dry eye syndrome, and the dry eye components of various ocular surface diseases,
  • Increase retention of ocular medications
  • Temporarily treat contact lens intolerance secondary to dry eye, and
  • Treat dry eye after surgery.
Device Description

EagleVision Gellanserts are intended to temporarily block tear drainage by occlusion of the canaliculus. EagleVision Gellanserts are made of a hydrophilic, resorbable material, and are designed to expand only laterally after insertion into the canaliculus, so that only one size is needed to occlude most patients' lacrimal systems. The EagleVision Gellanserts are designed to expand and fit virtually all patient anatomies from 0.3mm up to 1.1 mm, so multiple inserts in one canaliculus are generally not necessary. There are five inserts per sterile pouch, so that one EagleVision Gellansert can be placed in each punctum with one additional EagleVision Gellansert in case one is dropped or mishandled.

AI/ML Overview

I am sorry, but the provided text does not contain information about the acceptance criteria or a study proving the device meets acceptance criteria. The document is a 510(k) summary for a medical device (EagleVision Gellanserts), which focuses on demonstrating substantial equivalence to predicate devices rather than presenting performance criteria and study results against those criteria.

Therefore, I cannot fulfill your request to describe the acceptance criteria and the study that proves the device meets them, nor can I provide information regarding sample sizes, data provenance, expert ground truth establishment, adjudication methods, MRMC studies, standalone performance, or training set details.

{0}------------------------------------------------

This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

APPLICANTEagle Vision, Inc.8500 Wolf Lake Drive, Suite 110Memphis Tennessee, 38133 USAMAY 2 2006
OFFICIALCORRESPONDENTJeff CobbVice President, RA/CA/QAEagle Vision, Inc.8500 Wolf Lake Drive, Suite 110Memphis Tennessee, 38133 USATel: (901) 380-7000FAX: (901) 380-7001e-mail: jeff@eaglevis.com
TRADE NAME:EagleVision Gellanserts
COMMON NAME:Temporary Punctum PlugsTemporary Inserts
CLASSIFICATIONNAME:Plug, Punctal
DEVICECLASSIFICATION:Class II per 21 CFR § 886
PRODUCT CODE86 (LZU)
PREDICATEDEVICES:EagleVision Gellanserts are substantially equivalent to:K850649 ABSORBABLE INTRACANALICULAR COLLAGEN IMPLANT K890919 TEMPORARY INTRACANALICULAR COLLAGEN IMPLANT K895342 COLLAGEN IMPLANTS FOR THE LACRIMAL EFFICIENCY TEST K897190 LOOK TEMPORARY INTRACANALICULAR COLLAGEN IMPLANT K946357 COLLAGEN INTRACANALICULAR PLUG K013613 SOFT PLUG ABSORBABLE PLUG-SA K020882 SHARPOINT ULTRAPLUG EXTENDED WEAR PLUG K022043 SMARTPLUG MODEL 500 K030300 OPAQUE HERRICK LACRIMAL PLUG

{1}------------------------------------------------

Description of the Device Subject to Premarket Notification:

EagleVision Gellanserts are intended to temporarily block tear drainage by occlusion of the canaliculus. EagleVision Gellanserts are made of a hydrophilic, resorbable material, and are designed to expand only laterally after insertion into the canaliculus, so that only one size is needed to occlude most patients' lacrimal systems. The EagleVision Gellanserts are designed to expand and fit virtually all patient anatomies from 0.3mm up to 1.1 mm, so multiple inserts in one canaliculus are generally not necessary. There are five inserts per sterile pouch, so that one EagleVision Gellansert can be placed in each punctum with one additional EagleVision Gellansert in case one is dropped or mishandled.

INDICATION FOR USE:

The EagleVision Gellansert® has the same primary intended use as the predicated devices. The EagleVision Gellansert is intended to temporarily block tear drainage by the occlusion of the canaliculus in order to:

  • Determine the potential effectiveness of permanent occlusion, .
  • Temporarily treat dry eye syndrome, and the dry eye components of various . ocular surface diseases,
  • Increase retention of ocular medications .
  • Temporarily treat contact lens intolerance secondary to dry eye, and .
  • Treat dry eye after surgery. .

Contraindications:

EagleVision Gellanserts have the same primary contraindications as the predicate devices. EagleVision Gellanserts are contraindicated for use in patients with known gellan sensitivity, infectious conjunctivitis, dacrocycstitis, inflammation of the eyelid, infected eves, or epiphora.

MATERIALS:

Gellan is an anionic polysaccharide that gels in the presence of cations such as Sodium (Na ) and Calcium (Ca *). It is soluble in water, and hydrates rapidly in solution. As the gel hydrates, it also expands (up to 500% or more depending on the concentration of gellan and the strength of the ionic bonds). After hydration, the gellan becomes pliable and malleable to conform to the inside of the volume that constrains it (assuming the volume is less than or equal to the physical size of the gel in its hydrated state). 21 CFR 172.665 states that gellan gum may be safely used in food, in accordance with the following conditions:

    1. The additive is of a high molecular weight, produced from Pseudomonas elodea (P. elodea) by a pure culture fermentation process, and purified by recovery with isopropyl alcohol (IPA).
    1. The strain of P. elodea is non-pathogenic and nontoxic in man and animals

{2}------------------------------------------------

    1. The additive is produced by a process that renders it free of viable cells of P. elodea
    1. The additive meets the following specifications:
    • a. A tough worm-like gel forms when a 1% solution is extruded through a wide bore pipet into a 10% Calcium Chloride bath
    • b. A firm gel forms when 0.50 grams of Sodium Chloride is added (under 80 °C heat for 1 minute) to the 1% solution above.
    • Residual IPA does not exceed 0.075% ﻦ
    1. The additive is intended for use in accordance with good manufacturing practices

Gellan is also listed in:

  • . USP-National Formulary (NF 22) 71010-52-1
  • The Toxic Substances Control Act (TSCA) Chemical Substances Inventory 트
  • I The Canadian Non-Domestic Substances List (NDSL)
  • The European Inventory of existing Commercial Chemical Substances (EINECS) 트

In addition to the above, Toxicological Studies performed by the gellan manufacturer demonstrate the following Safety Data for gellan:

  • Oral dosage - rat LD50: >5,000 mg/kg
  • Inhalation -rat LC50 (4-hr): >6mg/L (nominal) .
  • 트 Eye irritation - rabbit: nonirritating
  • 트 Skin irritation - rabbit: nonirritating
  • No SARA Hazard notifications are applicable for this material 이

DESIGN FEATURES:

The design features of the EagleVision Gellanserts raise no new issues of safety or effectiveness. EagleVision Gellanserts consist of a length of rigid, hydrophilic, resorbable material. Common forceps (jewelers, collagen or otherwise) are used in the insertion of EagleVision Gellanserts. The EagleVision Gellanserts are provided sterile. EagleVision Gellanserts expand to accommodate different patient physiologies and achieve occlusion of the punctum and/or canaliculus. This device is sub-punctal, to limit contact and possible irritation to the eye due to abrasion.

Comparison of the design features of the EagleVision Gellanserts to the predicate devices are given in the table below.

SIMILARITIES AND DIFFERENCES OF EAGLEVISION GELLANSERTS TO PREDICATE DEVICES

A comparison of the similarities and differences between EagleVision Gellanserts to the predicate devices is given in the following table.

{3}------------------------------------------------

Summary of Safety & Efficac

ElementsEagleVisionGellansertsAbsorbableIntracanalic-ular CollagenImplant(K850649)TemporaryIntracanalic-ular CollagenImplant(K890919)CollagenImplant forthe LacrimalEfficiencyTest(K895342)LookTemporaryIntracanalic-ular CollagenImplant(K897190)CollagenIntracanalic-ular Plug(K946357)Soft PlugAbsorbablePlug-SA(K013613)SharPointUltraPlugExtendedWear Plug(K020882)SmartPlugModel 500(K022043)DissolvableOpaqueHerrickLacrimalPlug(K030300)rrigation orDissolution,acrimasalineprobeGamma (y)Irradiation
IndicationsTemporary punctal or canalicular occlusion may be used to temporarily block tear drainage by the occlusion of the canaliculus in order to:Determine the potential effectiveness of permanent occlusion, Temporarily treat dry eye syndrome, and the dry eye components of various ocular surface diseases, Increase retention of ocular medications Temporarily treat contact lens intolerance secondary to dry eye, and Treat dry eye after surgery.rrigation or_acrimalSalineorobe--------------------------------------------------------------------------------Gamma (yIrradiationContinued and concesso
Design CharacteristicsDissolutionrrigation oLacrimalSalineprobeЗатта (ү)rradiation------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Intracanalic-ular PunctumPlugYesYesYesYesYesYesYesYesYesYes------------------------------------------------------------------------------------------------------------------------------------------------------------------------------Dissolutionrrigation or_acrimaSalineorobeTHE FORCE CONSULTION CONSULTIONAL CONSULTIONAL COGamma (y)Irradiation
IntendedDurationTemporary7 - 10 days7 - 10 days7 - 10 days7 - 10 days7 - 10 daysLess than 3months180 daysPermanent180 daysrrigation orDissolution,_acrimaSalineorobemarily(ANNo) (ANN) (ATTITUTE TGamma (y)rradiationDOCADORESCORDS
MaterialHydrophilic,dissolvablepolymerAnimalCollagenAnimalCollagenAnimalCollagenAnimalCollagenAnimalCollagenPolyglyco-nate suturePCLMonofila-mentSyntheticAbsorbableSutureThermosen-sitiveHydrophobicAcrylicPolymerPCL or PDOor glycolicacid &trimethylenecarbonatecopolymer--------------------------------------------------------------------------------------------------------------Dissolution,irrigation ofLacrimalSalineprobeGamma (y)rradiationConcession Concessional
PackagingTyvek/PolypouchTyvek/polypouchTyvek/polypouchTyvek/polypouchTyvek/polypouchTyvek/polypouchTyvek/polypouchTyvek/polypouchTyvek/polypouchTyvek/polypouchirrigation or------------------------------------------------------------------------------------------------------------------------------------------------------------------------------issolution,LacrimaSalineprobeGamma (y)rradiation---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
ElementsEagleVisionGellansertsAbsorbableIntracanalic-ular CollagenImplant(K850649)TemporaryIntracanalic-ular CollagenImplant(K890919)CollagenImplant forthe LacrimalEfficiencyTest(K895342)LookTemporaryIntracanalic-ular CollagenImplant(K897190)CollagenIntracanalic-ular Plug(K946357)Soft PlugAbsorbablePlug-SA(K013613)SharPointUltraPlugExtendedWear Plug(K020882)SmartPlugModel 500(K022043)OpaqueHerrickLacrimalPlug(K030300)CART FOR CONSULTION CONCELLER THE CONTRACTDissolution,rrigation or_acrimalSalineorobeComments of Contractions And States Concession And------------------------------------------------------------------------------------------------------------------------------------------------------------------------------Gamma (y)rradiation
Method ofInsertionForcepsForcepsForcepsForcepsForcepsForcepsForcepsForcepsForcepsForcepsJissolutionrigation or_acrimaSalineorobe------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------Gamma (yrradiation
Dissolution, Dissolution,followed byrrigation, orrrigation.acrimaSalineorobesalineAddress State of Activities of Active School StateSterilizationEOEtC
would and submit and and the contribution of the country of the course of the course ofMethod ofRemovalMethod ofSterilizatioAND STATE CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION

510(k) Premarket Notification
EagleVision Gellanserts

{4}------------------------------------------------

Kos3333

Summary of Safety & Efficac

CONCLUSION:

Based on the above, the EagleVision Gellanserts are substantially equivalent to the predicate device

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with three wavy lines representing its wings.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MAY 2 2006

Eagle Vision, Inc. c/o Mr. Jeff Cobb 8500 World Lake Drive, Suite 110 Memphis, Tennessee 38133

Re: K053333

Trade/Device Name: EagleVision Gellanserts Regulation Number: Unclassified Regulation Name: Plug, Punctal Regulatory Class: Class II Product Code: LZU Dated: April 18, 2006 Received: April 19, 2006

Dear Mr. Cobb:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{6}------------------------------------------------

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 827-8910. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

M.B. Egelston - M.D.

Malvina B. Eydelman, M.D Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{7}------------------------------------------------

Page _1_of 1

510(k) Number (if known): KO533333

Device Name: EagleVision Gellanserts

Indications for Use:

.

The EagleVision Gellansert™ is intended to temporarily block tear drainage by the occlusion of the canaliculus in order to:

  • . Determine the potential effectiveness of permanent occlusion,
  • . Temporarily treat dry eye syndrome, and the dry eye components of various ocular surface diseases,
  • . Increase retention of ocular medications
  • Temporarily treat contact lens intolerance secondary to dry eye, and ●
  • . Treat dry eye after surgery.

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRHI, Office of Device Evaluation (ODE)
signature
(Division Sign-Off)
Division of Ophthalmic Ear,
Nose and Throat Devices

510(k) NumberK053333
------------------------
Prescription UseXOROver-The-Counter Use
(Per 21 CFR 801.109)(Optional Format 1-2-96)

CONFIDENTIAL Eagle Vision, Inc. November 29, 2005

510(k) Premarket Notification EagleVision Gellanserts