K Number
K053327

Validate with FDA (Live)

Device Name
OXICOM CONTROL
Manufacturer
Date Cleared
2006-01-26

(56 days)

Product Code
Regulation Number
862.1660
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Oxicom Control is intended to be used to monitor and evaluate the analytical performance of the Waters Oxicom models 2000, 2100 and 3000 oxygen saturation meters for the measurement of oxygen saturation. The use of quality control materials is indicated as an objective assessment of the precision of methods and techniques in use and is an integral part of good laboratory practice. The three levels of oxygen saturation provided by the controls allow performance monitoring within the clinically important range. For In Vitro Diagnostic Use

Device Description

The solutions in the 3-level Oxicom Control are a suspension of polystyrene beads in an inorganic, aqueous buffer solution with dyes to simulate the absorbance of light equivalent to that of whole blood at a range of oxygen saturation clinically relevant. Oxicom Control provides a convenient method of performing periodic OC checks for laboratories selecting to measure liquid QC material as a part of their quality assurance program. The product is packaged in a glass bottle. Oxicom Control is a non-hazardous liquid control solution containing no human biological materials and requires no reconstitution prior to use.

AI/ML Overview

The provided document is a 510(k) summary for the Bionostics Oxygen Saturation Control device and does not contain the detailed information necessary to complete all sections of your request. This type of document focuses on demonstrating "substantial equivalence" to a predicate device rather than presenting extensive performance studies with specific statistical acceptance criteria and ground truth validation for software or AI-driven devices.

Here's a breakdown of what can be extracted and what is missing:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance CriteriaReported Device Performance
Closed bottle stabilityVerified (details not provided)
Stability after openingVerified (details not provided)
Correlation to predicate deviceVerified (details not provided on quantitative correlation)
Test precision and rangeVerified (details not provided on specific precision values or the range)

2. Sample size used for the test set and the data provenance

The document does not detail specific sample sizes for tests. The tests mentioned (stability, correlation, precision, range) imply testing was performed on the Oxygen Saturation Control device itself, not on patient data in the context of an AI/algorithm.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

Not applicable. This device is a quality control solution, not a diagnostic device that interprets patient data. Therefore, there's no ground truth established by medical experts in the way it would be for an AI diagnostic tool.

4. Adjudication method for the test set

Not applicable. See point 3.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is a quality control material, not an AI or software device used by human readers for diagnosis.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This is a quality control material, not an algorithm.

7. The type of ground truth used

For stability, precision, and range, the "ground truth" would be established by validated reference methods or expected performance characteristics of the control material itself, often using highly accurate laboratory instruments. For correlation to predicate device, the predicate device's established performance would serve as a reference.

8. The sample size for the training set

Not applicable. This is a quality control material, not an AI/ML algorithm that requires a training set.

9. How the ground truth for the training set was established

Not applicable. See point 8.


Summary of Device and Study Context from the Document:

The Bionostics Oxygen Saturation Control is a liquid control solution intended to monitor and evaluate the analytical performance of Waters Oxicom oxygen saturation meters. It contains suspended styrene beads and dyes to simulate the absorbance of light by whole blood at a range of oxygen saturation values.

The "study" referenced in the document is a series of non-clinical tests conducted to verify specific performance requirements:

  • Closed bottle stability
  • Stability after opening
  • Correlation to predicate device
  • Test precision and range

The document explicitly states "Summary of clinical tests submitted with the premarket notification for the device. N/A," indicating no clinical trials were conducted or required for this Class I medical device. The conclusions drawn are based on the comparison of technological characteristics, formulation, and intended use to predicate devices, supporting the claim of substantial equivalence.

{0}------------------------------------------------

BIONOSTICS FRANIS

JAN 2 6 2006

K053327

  • (a) (1) Submitter's name, address Bionostics, Inc. 7 Jackson Road Devens, MA 01434
    510(k) Summary1

Contact Person Kathleen Storro Sr.Director, QA & Regulatory Affairs (978) 772-7070 x 220

Date of preparation of this summary: 30 November 2005

  • Oxygen Saturation Control (2) Device trade or proprietary name: Device common or usual name or classification name:
    Oxygen Saturation Control
PRODUCT NOMENCLATURECLASSIFICATIONNUMBERCLASSPANEL
Control, Single Analyte (Assayed or Unassayed)862.1660 (JJX)IChemistry

I. Substantial Equivalence

Oxygen Saturation Control for Waters Oxicom is substantially equivalent in function, safety and efficacy to other Bionostics products for the quality control evaluation of oxygen saturation.

CharacteristicNew DevicePredicate DevicePredicate DevicePredicate Device
Name:Oxygen SaturationControl for WatersOxicomOxicom 2100 QC FiltersOPTI-Check QualityControlMultifunction Blood GasControl
510(k):--K921519K974822K880447
Description:Aqueous solutioncontaining suspendedstyrene beads anddyes to simulate tHband OxygenSaturation1 translucent plastic(QC1) and 2 precisionglass (QC2, QC3) filterswith discreetabsorbance valuesAqueous solutioncontaining suspendedstyrene beads tosimulate tHb andOxygen SaturationAqueous solutioncontaining dyes tosimulate tHb andOxygenSaturation
Intended Use:As a quality controlsolution for use toverify the performanceof the Waters Oxicomsystems formeasurement ofOxygen saturationAs a means to checkthe calibration of theOxicom systemAs a quality controlsolution to verifyperformance of AVLOPTI for measurementincluding oxygensaturation and tHbAs a quality controlsolution for use to verifythe performance of bloodgas instruments formeasurement includingoxygen saturation and tHb
Levels:3333
Analytes:Oxygen Saturation(SO2%)Oxygen Saturation(SO2%)pH, PCO2, PO2, Na, K,Cl, iCa, tHb, SO2%pH, PCO2, PO2, Na, K,Cl, iCa, tHb, SO2%,MetHb%, COHb%

Comparison of Technological Characteristics with Predicate Device

l This summary of safety and effectiveness is submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

{1}------------------------------------------------

II. Description of the new device

The Waters Oxicom family of oxygen saturation meters, (2000, 2100 and 3000) utilize light emitting diodes and a solid state detector to measure the absorbance of light by a whole blood sample contained in an optically clear cuvette at 2 wavelengths. The amount of IR light absorbed by the blood is rclatively independent of the blood's oxygenation. However, RED light absorption or optical density is strongly dependent on the oxygen saturation. The ratio of the optical densities at the two wavelengths is used to calculate the functional oxygen saturation. The Quality Control (QC) filters supplied with the Oxicom systems allows the user to check the calibration of the Oxicom. The set of filters consist of one translucent plastic filter (QCI), and two precision glass filters (QC2 and QC3). The optical densities of these filters are stored in the unit's microprocessor. When inserted into the sample chamber the Oxicom measures the optical density and compares it to the value stored. If the optical densities match, the display will read a %Sat value.

The solutions in the 3-level Oxicom Control are a suspension of polystyrene beads in an inorganic, aqueous buffer solution with dyes to simulate the absorbance of light equivalent to that of whole blood at a range of oxygen saturation clinically relevant.

Oxicom Control provides a convenient method of performing periodic OC checks for laboratories selecting to measure liquid QC material as a part of their quality assurance program. The product is packaged in a glass bottle.

Oxicom Control is a non-hazardous liquid control solution containing no human biological materials and requires no reconstitution prior to use.

(5) Intended use of the device

Oxicom Control is intended for use as a quality control solution to evaluate Oxicom measurement at a range of clinically relevant oxygen saturation values.

Oxicom Control is a three-level, liquid control solution with dyes and polystyrene beads to provide absorbance relative to a range of oxygen saturation values. The concentration of beads and dyes are optimized for the Waters Oxicom Systems to provide an improved range of oxygen saturation values in comparison to the existing, filter systems, and handling and measurement essentially equivalent to the measurement of whole blood.

(b) (1) Summary of non-clinical tests submitted with the premarket notification for the device.

Tests were conducted to verify specific performance requirements:

{2}------------------------------------------------

  • a) Closed bottle stability
  • b) Stability after opening
  • c) Correlation to predicate device
  • d) Test precision and range
  • Summary of clinical tests submitted with the premarket notification for the (b) (2) device.

N/A

  • Conclusions drawn from the clinical and non-clinical trials. (b) (3) Comparison of technological characteristics, formulation and intended use to predicate devices listed in this summary support the claim of substantial equivalence.

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around it. Inside the circle is a stylized symbol that resembles an abstract caduceus or a representation of human figures.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

JAN 2 6 2006

Ms. Kathleen Storro Sr. Director, QA & Regulatory Affairs Bionostics, Inc. 7 Jackson Road Devens, MA 01434

K053327 Re:

Trade/Device Name: Waters Oxicom Control Regulation Number: 21 CFR 862.1660 Regulation Name: Quality control material (assayed and unassayed) Regulatory Class: Class I Product Code: JJX Dated: November 30, 2005 Received: December 1, 2005

Dear. Ms. Storro:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).

{4}------------------------------------------------

Page 2 --

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240) 276-0484. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Alberto Guta

Alberto Gutierrez, Ph.D. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indications for Use

K053327 510(k) Number (if known):

Device Name: Waters Oxicom Control

Indications For Use:

Oxicom Control is intended to be used to monitor and evaluate the analytical performance of the Waters Oxicom models 2000, 2100 and 3000 oxygen saturation meters for the measurement of oxygen saturation. The use of quality control materials is indicated as an objective assessment of the precision of methods and techniques in use and is an integral part of good laboratory practice. The three levels of oxygen saturation provided by the controls allow performance monitoring within the clinically important range.

For In Vitro Diagnostic Use

Prescription Use (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

0

Ann Chappie
Division Sign-Off

Page 1 of 1

Office of In Vitro Diagnostic Device Evaluation and Safety

51 KOS3327

§ 862.1660 Quality control material (assayed and unassayed).

(a)
Identification. A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.(b)
Classification. Class I (general controls). Except when intended for use in donor screening tests, quality control materials (assayed and unassayed) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.