The KOWA nonmyd series, non-mydriatic fundus camera, are intended for use with retina image capturing. The retina image can be stored to an external hard disk drive or transferred in other formats through memory card or serial interface depending on the output interface available for each device.

Device Description

KOWA nonmyd series of camera consist of non-mydriatic fundus cameras with a digital image capture system, which does not require 35mm film or instant film. It maintains the same functionalities of the previous fundus cameras, and by capturing digital images it eliminates film development time, allowing the user to view images immediately.

The KOWA nonmyd series cameras are all fundamentally the same noting exception only to their output interfaces. Type D (KOWA nonmyd a-D) contains an internal CCD camera and exports data via USB interface. Type F (KOWA nonmyd 7) requires a CCD camera to be attached to the device and exports data by both memory card and via USB.

Application software for the KOWA nonmyd series contain a filing system, which can save, display, and manage the digital images.

For detailed observation, the KOWA nonmyd series have an optical variable power feature, which enables the viewer to select a picture angle. By selecting a picture angle it can achieve the same amount of image clarity as a conventional fundus camera.

In addition, the KOWA nonmyd series can be used even in cases where the pupil has not been completely dilated; in addition to the normal pupil diameter 4mm mode, there is also a small pupil diameter 3.7mm mode.

Similar to previous fundus cameras, the KOWA nonmyd series uses infrared light for illumination during alignment to provide a better patient experience. In addition, the devices are equipped with a highly sensitive CCD camera, and are therefore able to capture images at a much lower flash light intensity level than previous fundus cameras, further enhancing the patient experience.

AI/ML Overview

Here's an analysis of the provided text regarding the KOWA nonmyd series, focusing on acceptance criteria and the supporting study:

Acceptance Criteria and Device Performance for KOWA nonmyd series

The KOWA nonmyd series (nonmyd 7 and nonmyd α-D) are non-mydriatic fundus cameras intended for retina image capturing. The device's acceptance criteria are implicitly defined by its substantial equivalence to the predicate device, the Nidek NM-1000, and its conformity to specific international standards.

1. Table of Acceptance Criteria and Reported Device Performance

The provided document does not explicitly state quantitative performance acceptance criteria (e.g., minimum image resolution, sensitivity, or specificity for a specific pathology). Instead, acceptance is demonstrated through substantial equivalence to a predicate device and conformity to relevant safety and performance standards.

The table below summarizes the key comparisons made between the KOWA nonmyd series and the predicate device, which inherently serve as the performance characteristics the new device aims to match or exceed.

Feature / Acceptance Criteria CategoryPredicate Device (Nidek NM-1000) PerformanceKOWA nonmyd series Performance (Type D and F)Comparison / Implied Acceptance
Intended UseRetina and anterior segment image capturingRetina image capturingSubstantially Equivalent (retina)
Image Output FormatNidek format, TIFBMP, JPEGDifferent, but still digital output. Acceptable.
Picture Angle45 degrees45 degrees/20 degreesEquivalent (45 deg), KOWA also offers 20 deg. Acceptable.
Working Distance43.3mm (from camera body to corner)30mmDifferent, but a functional attribute. Acceptable.
Working Distance Detection MethodAnterio (Observation), Fundus (Focusing)Anterio (Observation), Fundus (Focusing)Equivalent
Minimum Pupil Diameter4mmNormal: 4mm; Small: 3.7mmEquivalent (4mm), KOWA also offers 3.7mm. Acceptable.
Observation CCD Camera1/3 inch CCD Camera1/3 inch CCD CameraEquivalent
Photographing CCD Camera1/2 inch CCD Camera, Digital Progressive ScanType D: 1/2 inch 2.1MP CCD; Type F: APS-C 6MPEquivalent (Type D), Superior (Type F). Acceptable.
Observation Display (B/W)6.4 inch LCD Monitor5.6 inch LCD MonitorDifferent size, but functional. Acceptable.
Photographing Display6.4 inch LCD MonitorOuter Monitor Use (Type D, F)Different display method, but functional. Acceptable.
Dioptric CompensationTotal -32D to +40DTotal -33D to +40DSubstantially Equivalent
FocusingManual (motor driven), Split line focusSplit luminous bars coincidenceDifferent method, but functional. Acceptable.
Observation Light SourceHalogen lamp (Max 12V 50W) with IR filterHalogen lamp (Max 12V 100W) with IR filterDifferent wattage, but same type. Acceptable.
Photographing Light SourceXenon flash (Max 25WS)Xenon flash (Max 50WS)Different wattage, but same type. Acceptable.
Internal Fixation NavigationManual leverFixed fixator selectingDifferent method, but functional. Acceptable.
Switching Light PathBeam SplitterSame pathway, no beam splitDifferent design, but functional. Acceptable.
Observation Light AdjustmentVolume adjustment styleVolume adjustment styleEquivalent
Photographing Light AdjustmentStep adjustment style (8 steps)Step adjustment style (5 steps)Different number of steps, but functional. Acceptable.
Camera Stand TypeTabletop; power source built-inTabletop; power source built-inEquivalent
Horizontal MovementForward/Backward: 65mm; Left/Right: 106mmForward/Backward: 40mm; Left/Right: 100mmDifferent ranges, but functional. Acceptable.
Vertical Movement (Camera Stand)30mm30mmEquivalent
Shutter ReleaseJoystick upper buttonJoystick upper buttonEquivalent
Signal OutletUSB, RGB Analog, NTSC Composite VideoType D: USB; Type F: Flash memory cardDifferent options, but includes USB. Acceptable.
Vertical Movement (Chinrest)65mm60mmDifferent range, but functional. Acceptable.
External Fixation TargetsFree-arm style (Option)Free-arm style (Option)Equivalent
Safety Requirements (IEC)EN60601-1IEC60601-1Equivalent (different standard version/name)
EMC (IEC)EN60601-1-2IEC60601-1-2Equivalent (different standard version/name)
Ophthalmic Instrument Req (ISO)ISO15004ISO15004Equivalent
Dimensions360mm(W) x 530mm(D) x 565mm(H)Type F: 310x504x548mm; Type D: 310x504x462mmDifferent, but functional. Acceptable.
Weight26 kgType F: 21 kg; Type D: 21 kgDifferent, but functional. Acceptable.

2. Sample Size for Test Set and Data Provenance

The document does not describe a traditional clinical study with a "test set" of patient images or specific "data provenance" in the context of evaluating a diagnostic algorithm. This submission is for an ophthalmic camera, not an AI/ML diagnostic tool.

The "testing" mentioned refers to engineering and compliance testing against international standards rather than a clinical performance study with patient data. Therefore, the concepts of "sample size" and "data provenance" as typically applied to algorithmic performance evaluation are not applicable here.

3. Number of Experts and Qualifications for Ground Truth

Not applicable. As noted above, this submission is for a medical device (fundus camera), not an AI diagnostic algorithm requiring expert ground truth for image interpretation.

4. Adjudication Method for Test Set

Not applicable. There is no "test set" in the context of image interpretation that would require an adjudication method.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

Not applicable. The device is a fundus camera, a tool for image acquisition, not a diagnostic algorithm designed to assist human readers. Therefore, an MRMC study comparing human readers with and without AI assistance is irrelevant to this submission.

6. Standalone (Algorithm Only) Performance

Not applicable. The KOWA nonmyd series is a hardware device (fundus camera), not a standalone algorithm.

7. Type of Ground Truth Used

Not applicable. As this is a hardware device, "ground truth" in the context of diagnostic accuracy is not relevant to its regulatory submission. The "ground truth" for the device's performance is its ability to meet the technical specifications and safety standards.

8. Sample Size for the Training Set

Not applicable. The KOWA nonmyd series is a hardware device, not an AI/ML algorithm that requires a "training set."

9. How Ground Truth for the Training Set Was Established

Not applicable. No training set is involved for this device.

{0}------------------------------------------------

NOV 1 0 2005

510(k) Notification

K 053026

9. Certification 9.1 Summary for public disclosure

Applicant:Kowa Company, Ltd.4-14, Nihonbashi-honcho 3-ChomeChuo-ku, Tokyo, 103-8433 Japan
Phone:+81-3-3279-7329
Fax:+81-3-3279-7541
Contact:Satohiko Takanashi
e-mail address:s-takana@kowa.co.jp
Date summary prepared:January 11 , 2005
Device trade name:KOWA nonmyd 7 (Type F)KOWA nonmyd a -D (Type D)
Classification name:CAMERA, OPTHALMIC, AC-POWERED
Product code:HKI

Intended use:

The KOWA nonmyd series, non-mydriatic fundus camera, are intended for use with retina image capturing. The retina image can be stored to an external hard disk drive or transferred in other formats through memory card or serial interface depending on the output interface available for each device.

Device description:

KOWA nonmyd series of camera consist of non-mydriatic fundus cameras with a digital image capture system, which does not require 35mm film or instant film. It maintains the same functionalities of the previous fundus cameras, and by capturing digital images it eliminates film development time, allowing the user to view images immediately.

The KOWA nonmyd series cameras are all fundamentally the same noting exception only to their output interfaces. Type D (KOWA nonmyd a-D) contains an internal CCD camera and exports data via USB interface. Type F (KOWA nonmyd 7) requires a CCD camera to be attached to the device and exports data by both memory card and via USB.

{1}------------------------------------------------

Application software for the KOWA nonmyd series contain a filing system, which can save, display, and manage the digital images.

For detailed observation, the KOWA nonmyd series have an optical variable power feature, which enables the viewer to select a picture angle. By selecting a picture angle it can achieve the same amount of image clarity as a conventional fundus camera.

In addition, the KOWA nonmyd series can be used even in cases where the pupil has not been completely dilated; in addition to the normal pupil diameter 4mm mode, there is also a small pupil diameter 3.7mm mode.

Similar to previous fundus cameras, the KOWA nonmyd series uses infrared light for illumination during alignment to provide a better patient experience. In addition, the devices are equipped with a highly sensitive CCD camera, and are therefore able to capture images at a much lower flash light intensity level than previous fundus cameras, further enhancing the patient experience.

Comparison:

The Nidek NM-1000 was chosen as a substantially equivalent device. The predicate device is a non-mydriatic fundus camera and it is equipped with a high resolution CCD camera so it does not require any film and can display images immediately after image capture. Also because it uses a highly sensitive CCD camera, the flash required for filming is reduced compared to conventional fundus cameras.

Similar to the predicate device, the KOWA nonmyd series cameras are equipped with a highly sensitive CCD so they do not require film and can display images immediately after image capture.

The KOWA nonmyd series, KOWA nonmyd 7 and KOWA nonmyd a-D, and the predicate device also share similar technical and safety characteristics. The result is show in the table below.

Performance testing:

KOWA nonmyd series, KOWA nonmyd 7 and KOWA nonmyd a-D, were tested to the following standards and conform to all specific requirements based on these standards. KOWA nonmyd series are equivalent to the predicate device.

Electrical safety

KOWA nonmyd series were tested in accordance with IEC60601-1: 1988. Amendment 1:1991 and Amendment 2: 1995, and met all requirements of standards.

Electromagnetic compatibility

KOWA nonmyd series were tested in accordance with IEC60601-1-2: 2001. and met all requirements of standard.

{2}------------------------------------------------

Test requirements and test procedure for ophthalmic instruments

KOWA nonmyd series was tested in accordance with ISO15004: 1997, and met all requirements of the standard. Refer to Appendix B for details.

Risk management

KOWA nonmyd series were evaluated in accordance with ISO14971: 2000. The risk management of the devices was deemed satisfactory. Remaining risks will be noted in the user manual, so users will be able to avoid them.

Conclusion:

The KOWA nonmyd series is equipped with the same fundamental technology as the predicate device and maintains the same level of safety performance. Therefore it has been concluded that there are no significant differences in the technical characteristics and safety between KOWA nonmyd series and the predicate device.

Table A: Predicate device

Predicate DeviceManufacturer510(k) No.Date Cleared
Nidek Non-MydriaticFundus CameraModel NM-1000Nidek IncorporatedK014274Apr. 17, 2002

{3}------------------------------------------------

:

Table B: Predicate Device Comparison
--------------------------------------------

:

KOWA nonmyd series: Non-mydriaticfundus camera, type D and FNidek Non-mydriatic Fundus CameraModel NM-1000
Intended useThe KOWA nonmyd series, non-mydriatic fundus cameras areintended for use with retina imagecapturing.The Nidek NM-1000 is an ophthalmic camerathat is intended for use in capturing images ofthe retina and the anterior segment of the eye.
Saved Image FormatBMP, JPEGNidek format, TIF
Picture angle45degree/20degree45degree
Working distance30mm43.3mm (from camera body to corner)
Working distancedetection methodAnterio (Observation)Fundus (Focusing on blight spots)Anterio (Observation)Fundus (Focusing on blight spots)
Minimum diameterof pupilNormal pupil mode: 4mmSmall pupil mode: 3.7mm4mm
CCD camera forobservation1/3 inch CCD Camera1/3 inch CCD Camera
CCD camera forPhotographingType D:1/2inch 2.1mega pixel CCD CameraDigital Progressive Scan (built-in)Type F:APS-C size 6mega pixel CCD CameraDigital Progressive Scan(K9L-BM39 used as Digital camera)1/2inch CCD CameraDigital Progressive Scan (built-in)
Observation Display(B/W)5.6 inch LCD Monitor6.4 inch LCD Monitor
Photographing DisplayType D, F: Outer Monitor Use6.4 inch LCD Monitor
Dioptric compensationTotal -33D to +40DTotal -32D to +40D
FocusingSplit luminous bars coincidenceManual (motor driven)Split line focus on the retina (-10 to +14D)
KOWA nonmyd series: Non-mydriatic funduscamera, Type D and FNidek Non-mydriatic Fundus CameraModel NM-1000
Observation LightSourceHalogen lamp (Max 12V 100W)with infrared filterHalogen lamp (Max 12V 50W)with infrared filter
Photographing LightSourceXenon flash (Max 50WS)Xenon flash (Max. 25WS)
Internal FixationNavigationFixed fixator selectingManual lever
Switching light path ofobservation &photographingSame pathway, no beam splitBeam Splitter
Observation lightadjustmentVolume adjustment styleVolume adjustment style
Photographing lightadjustmentStep adjustment style (5 steps)Step adjustment style (8 steps)
Camera stand (Base)
TypeTabletop; power source built-inTabletop; power source built-in
Horizontal MovementForward/Backward: 40mmLeftward/Rightward: 100mmForward/Backward: 65mmLeftward/Rightward: 106mm
Vertical Movement30mm30mm
Shutter ReleaseJoystick upper buttonJoystick upper button
Signal outletType D: USBType F: Flash memory cardUSB, RGB Analog, NTSC Composite Video
Chinrest
Vertical Movement ofchinrest60mm65mm
External FixationTargetsFree-arm style (Option)Free-arm style (Option)
Compliance with safety standards
Safety requirementsIEC60601-1EN60601-1
EMCIEC60601-1-2EN60601-1-2
Ophthalmic instrumentrequirementsISO15004ISO15004
Dimensions
SizeType F: 310mm(W) x 504mm(D) x 548mm(H)Type D: 310mm(W) x 504mm(D) x 462mm(H)360mm(W) x 530mm(D) x 565mm(H)
WeightType F: 21 kg with the digital cameraType D: 21 kg26 kg

{4}------------------------------------------------

Table C: Predicate Device Comparison

KOW i nomnyd " and KOWA nomnyd a D. Kowa Company, I id.

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image is a black and white circular seal. The seal contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter of the circle. Inside the circle is a stylized image of an eagle with three curved lines representing its wings.

NOV 1 0 2005

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Kowa Company, Ltd. c/o Tamas Borsai Division Manager, Medical Division TUV Rheinland of North America, Inc. 12 Commerce Rd. Newton, CT 06470

Re: K053026

Trade/Device Name: Kowa nonmyd 7 (type F) & nonmyd α-D (type D) Fundus Camera Regulation Number: 21 CFR 886.1120 Regulation Name: Ophthalmic Camera Regulatory Class: Class II Product Code: HKI Dated: October 25, 2005 Received: October 27, 2005

Dear Mr. Borsai:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{6}------------------------------------------------

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 827-8910. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

David M. Whipple

Acting Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{7}------------------------------------------------

510(k) Notification

Indications for Use

510(k) Number (if know): KD5

Device Name: KOWA nonmyd 7 and KOWA nonmyd a-D

Indications for Use:

The KOWA nonmyd series, non-mydriatic fundus camera, are intended for use with retina image capturing. The retina image can be stored to an external hard disk drive or transferred in other formats through memory card or serial interface depending on the output interface available for each device.

Joenette R. Been

Division of Ophthalmic Ear Nose and Throat Dev

510(k) Number

Prescription Use
(Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

Image /page/7/Picture/11 description: The image shows a series of horizontal lines that are crossed by a series of vertical lines. The horizontal lines are thick and dark, and the vertical lines are thinner and lighter. The lines intersect at various points, creating a grid-like pattern. The image is somewhat blurry and difficult to make out.

(PLEASE DO WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

§ 886.1120 Ophthalmic camera.

(a)
Identification. An ophthalmic camera is an AC-powered device intended to take photographs of the eye and the surrounding area.(b)
Classification. Class II (special controls). The device, when it is a photorefractor or a general-use ophthalmic camera, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.