The 0.35T Signa® Ovation with Excite Magnetic Resonance System is an open, whole body scanner designed to support improved higher resolution imaging and shorter scan times. The Signa 0.35T Signa® Ovation with Excite Magnetic Resonance System is indicated for use as a diagnostic imaging device to produce transverse, sagittal, coronal and oblique images of the internal structures and organs of the entire body, including, but not limited to, the musculocskeletal, vascular, cardiac, and neuro systems. The images produced by the Signa 0.35T Signa® Ovation with Excite Magnetic Resonance System reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. The NMR properties that determine the image appearance are proton density, spin-lattice relaxation time (T1), spin-spin relaxation time (T2) and flow. When interpreted by a trained physician, these images provide information that can be useful in determining a diagnosis.
Due to the 'open' design of the Signa 0.35T Signa® Ovation with Excite may also be used for imaging during interventional procedures when performed with MR compatible devices such as, in-room display and MR safe biopsy needles
The 0.35T Signa® Ovation with Excite Magnetic Resonance System is a modification to the 0.35T Signa® MFO/i Magnetic Resonance System (K033504) which utilizes a permanent magnet to acquire 2D single-slice and multi-slice, and 3D volume images. The 0.35T Signa® Ovation with Excite Magnetic Resonance System features a permanent magnet operating at 0.35T. The system can image in the sagittal, coronal, axial, oblique and double oblique planes, using various pulse sequences. Images are acquired and reconstructed using 2D and 3D Fourier transformation techniques. The system is intended for high-resolution anatomical applications, short scan times, and multinuclear spectroscopy.
The provided text describes a 510(k) premarket notification for a medical device, the GE 0.35T Signa® Ovation with Excite MR System. However, it does not contain the detailed information necessary to answer all parts of your request regarding acceptance criteria and a study proving device performance.
Here's what can be extracted and what is missing:
Description of Acceptance Criteria and Study Proving Device Performance
The submission focuses primarily on demonstrating substantial equivalence to a predicate device (Signa® 0.35T® Ovation with Excite System, K033504) through engineering modifications and adherence to performance standards, rather than a clinical performance study with defined acceptance criteria for diagnostic efficacy.
1. A table of acceptance criteria and the reported device performance:
| Acceptance Criteria Category | Reported Device Performance | Comments |
|---|---|---|
| Safety Standards Adherence | Evaluated to IEC 601-1 (General Safety for Medical Equipment) | This indicates compliance with recognized safety standards for electrical and mechanical safety. It's a foundational acceptance criterion for medical devices. |
| MR Specific Safety & Performance Standards Adherence | Evaluated to IEC 601-2-33 (Particular Requirements for Safety of Magnetic Resonance Equipment for Medical Diagnosis) | This indicates compliance with specific safety and performance standards relevant to MR diagnostic devices, including aspects like magnetic field safety, heating, and acoustic noise. |
| General Performance Standards Adherence | Evaluated to appropriate NEMA performance standards | NEMA (National Electrical Manufacturers Association) performance standards for MR systems typically cover aspects like image quality (spatial resolution, signal-to-noise ratio, uniformity), geometric accuracy, and field strength accuracy. Specific values are not provided in this summary. |
| Substantial Equivalence | Concluded to be substantially equivalent to the predicate device (K033504) | The primary "acceptance criterion" for this 510(k) is demonstrating that the modified device is as safe and effective as the predicate device and does not raise new questions of safety or effectiveness. This is the overarching regulatory acceptance criterion for a 510(k) clearance. |
| No New Potential Hazards | Usage of the device "does not result in any new potential hazards." | This is a key part of demonstrating substantial equivalence and is an acceptance criterion related to safety. |
2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective):
- Not specified. The submission does not describe a clinical test set or data derived from human subjects for performance evaluation in the way you might expect for a diagnostic AI device. The evaluation appears to be based on technical performance and safety testing against established standards, and comparison to existing device specifications.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience):
- Not applicable. No clinical ground truth establishment is described for a diagnostic performance study. The "trained physician" is mentioned in the indications for use as interpreting the images, but not in the context of creating a test set for the device's performance.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:
- Not applicable. No clinical test set or adjudication method is described.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- Not applicable. This device is an MR system, not an AI-assisted diagnostic tool. No MRMC study is mentioned.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
- Not applicable. This is an MR system, not an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- Not applicable for diagnostic accuracy. The "ground truth" for this type of submission relates to compliance with engineering and safety standards (e.g., an MR system must correctly measure field strength, produce images with certain resolution, operate safely). No diagnostic accuracy ground truth is discussed.
8. The sample size for the training set:
- Not applicable. This device is an MR system, not an AI model that requires a training set.
9. How the ground truth for the training set was established:
- Not applicable. As above, no training set is relevant for this device.
Summary of the Study Discussed:
The actual "study" described in this 510(k) summary is focused on engineering and safety verification of the device's modifications and its adherence to relevant international and national standards.
- Type of Study: Technical evaluation against established performance and safety standards, and a comparison to a predicate device.
- Standards Referenced:
- NEMA performance standards (specific ones not listed, but implied to be relevant to MR system performance).
- IEC 601-1 (International Medical Equipment Safety standard).
- IEC 601-2-33 (Particular Requirements for Safety of Magnetic Resonance Equipment for Medical Diagnosis).
- Purpose: To demonstrate that the GE 0.35T Signa® Ovation with Excite MR System, with the addition of a motorized control circuit for lateral table movement, is substantially equivalent (in terms of safety and effectiveness) to its predicate device (K033504) and meets applicable safety and performance requirements.
- Conclusion: GE concluded that the device is substantially equivalent to the predicate and introduces no new potential hazards. FDA agreed with this determination.
This submission is typical for modifications to existing medical imaging hardware where the fundamental imaging principles and diagnostic capability remain unchanged, and the focus is on maintaining safety and performance characteristics in the face of minor design alterations.
{0}------------------------------------------------
GE Healthcare
General Electric Company P.O. Box 414. Milwaukee. WI 53201
NOV 1 7 2005
510(k) Summary
This 510(k) summary of safety and effectiveness information is submitted in accordance with the requirements of 21 CFR Part 807.92(c).
| Submitter: | GE HealthcarePO Box 414Milwaukee, WI 53201 |
|---|---|
| Contact Person: | Larry A. Kroger Ph.D.Senior Regulatory ProgramsManager |
| Telephone: | 262- 544-3894 |
| Fax: | 262- 548-4768 |
| Date Prepared: | October 25, 2005 |
Device Name:
GE 0.35T Signa® Ovation with Excite Magnetic Resonance System Magnetic Resonance Diagnostic System, 21 CFR 892.1000, 90-LNH
Marketed Device:
The GE 0.35T Signa® Ovation with Excite MR System is substantially equivalent to the currently marketed Signa® 0.35T® Ovation with Excite System (K033504) with the main differences being the addition of a motorized control circuit for the lateral movement of the table.
Device Description:
The 0.35T Signa® Ovation with Excite Magnetic Resonance System is a modification to the 0.35T Signa® MFO/i Magnetic Resonance System (K033504) which utilizes a permanent magnet to acquire 2D single-slice and multi-slice, and 3D volume images. The 0.35T Signa® Ovation with Excite Magnetic Resonance System features a permanent magnet operating at 0.35T. The system can image in the sagittal, coronal, axial, oblique and double oblique planes, using various pulse sequences. Images are acquired and reconstructed using 2D and 3D Fourier transformation techniques. The system is intended for high-resolution anatomical applications, short scan times, and multinuclear spectroscopy.
Indications for Use:
The 0.35T Signa® Ovation with Excite Magnetic Resonance System is an open, whole body scanner designed to support improved higher resolution imaging and shorter scan times. The 0.35T Signa® Ovation with Excite Magnetic Resonance System is indicated for use as a diagnostic imaging device to produce transverse, sagittal, coronal and oblique images of the internal structures and organs of the entire body, including, but not limited to, the musculocskeletal, vascular, cardiac, and neuro systems. The images produced by the Signa 0.35T Signa® Ovation with Excite Magnetic Resonance System reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. The NMR properties that determine
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" intertwined and surrounded by a circular pattern. The logo is black and white.
General Electric Comnania P.O. Box 414. Milwaukee. WI 53201
the image appearance are proton density, spin-lattice relaxation time (T1), spin-spin relaxation time (T2) and flow. When interpreted by a trained physician, these images provide information that can be useful in determining a diagnosis.
Due to the 'open' design of the 0.35T Signa® Ovation with Excite may also be used for imaging during interventional procedures when performed with MR compatible devices such as, in-room display and MR safe biopsy needles
Comparison with Predicate Device:
The GE 0.35T Signa® Ovation with Excite MR System is a modification of the Signa® Ovation with Excite system (K033504) with the main differences being the addition of motorized drive control for lateral table movement.
Summary of Studies:
The 0.35T Signa® Ovation with Excite Magnetic Resonance System was evaluated to the appropriate NEMA performance standards as well as the IEC 601-1 International Medical Equipment Safety standard and IEC 601-2-33 Particular Requirements for Safety of Magnetic Resonance Equipment for Medical Diagnosis. The 0.35T Signa® Ovation with Excite Magnetic Resonance System is comparable to the currently marketed Signa® 0.35T Ovation Magnetic Resonance System (K033504).
Conclusion:
It is the opinion of GE that the 0.35T Signa® Ovation with Excite Magnetic Resonance System is substantially equivalent to the 0.35T Signa® Ovation with Excite Magnetic Resonance System. Usage of the 0.35T Signa® Ovation with Excite Magnetic Resonance System does not result in any new potential hazards.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows a black and white seal. The seal is circular and contains the symbol of the United States Department of Health and Human Services (HHS). The HHS symbol is an abstract representation of an eagle with three lines forming its body and wings. The text "HEALTH & HUMAN SERVICES - USA" is written around the perimeter of the circle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
NOV 1 7 2005
Larry A. Kroger, Ph.D. Senior Regulatory Programs Manager GE Healthcare W-400 P.O. Box 414 MILWAUKEE WI 53201
Re: K053009 Trade/Device Name: GE 0.35T Signa® Ovation with Excite MR System Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: II Product Code: LNH Dated: October 25, 2005 Received: October 26, 2005
Dear Dr. Kroger:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If vour device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act s requirements. including, but not limited to registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (0S) regulation (21 CFR Part 820); and if applicable, the clectronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) I mis letter notification. The FDA finding of substantial equivalence of your device to a legally prematicated predicated on one of a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|---|---|---|
| 21 CFR 884.xxxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 892.xxxx | (Radiology) | 240-276-0120 |
| Other | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). You may obtain other general information on your responsibilities under the Act from the 60 : 77 / - 1 St mall Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrl/industry/support/index.html.
Sincerely yours,
Nancy C. Brigdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" intertwined within a circular shape. The letters and the circle are in white, set against a black background, creating a high-contrast design. The logo is a well-known symbol associated with the General Electric company.
GE Healthcare
General Fleriric Compan P.O. Box 414, Mihraukee, W1 53201
STATEMENT OF INTENDED USE
510(k) Number (if known):
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
Device Name: GE 0.35T Signa® Ovation with Excite MR System
Indications for Use
The 0.35T Signa® Ovation with Excite Magnetic Resonance System is an open, whole body scanner designed to support improved higher resolution imaging and shorter scan times. The Signa 0.35T Signa® Ovation with Excite Magnetic Resonance System is indicated for use as a diagnostic imaging device to produce transverse, sagittal, coronal and oblique images of the internal structures and organs of the entire body, including, but not limited to, the musculocskeletal, vascular, cardiac, and neuro systems. The images produced by the Signa 0.35T Signa® Ovation with Excite Magnetic Resonance System reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. The NMR properties that determine the image appearance are proton density, spin-lattice relaxation time (T1), spin-spin relaxation time (T2) and flow. When interpreted by a trained physician, these images provide information that can be useful in determining a diagnosis.
Due to the 'open' design of the Signa 0.35T Signa® Ovation with Excite may also be used for imaging during interventional procedures when performed with MR compatible devices such as, in-room display and MR safe biopsy needles
Daniel A. Lupyan
(Division Sign-Off) Division of Reproductive, Abdomin and Radiological Devices 510(k) Number
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use × (Per 21 CFR 801-109)
OR
Over-The-Counter Use
E-2