K Number
K052918

Validate with FDA (Live)

Date Cleared
2006-01-27

(102 days)

Product Code
Regulation Number
888.3030
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Reprocessed external fixation devices are indicated for use in patients requiring external skeletal fixation and treatment of fractures, osteotomy, arthrodesis, correction of deformities, fracture revision, bone reconstruction procedures, limb lengthening, correction of bony or soft tissue deformities and segmental bory or soft tissue defects.

Device Description

External fixation devices are specially designed frames, clamps, rods, rod-to-rod couplings, pins, posts, fasteners, wire fixations, fixation bolts, washers, nuts, hinges, sockets, connecting bars and screws used for the management of bone fractures and reconstructive, as well as corrective, orthopedic surgery. Materials used include metal alloys, plastic and composites. These materials are chosen to address a wide range of fractures and applications as well as to allow for the appropriate amount of rigidity and stability.

AI/ML Overview

The provided document is a 510(k) summary for the Alliance Medical Corporation's Reprocessed External Fixation Devices. It describes the device, its intended use, and claims substantial equivalence to predicate devices. However, this document does not contain a detailed study proving the device meets specific acceptance criteria in the way a clinical or performance study for a novel AI/medical imaging device would.

Instead, the submission for these reprocessed devices focuses on demonstrating that they perform "as originally intended" and are substantially equivalent to the original predicate devices. This is a crucial distinction. For reprocessed devices, the "acceptance criteria" are primarily centered around ensuring the reprocessing does not compromise the device's functional integrity, sterility, and material properties compared to new devices.

Therefore, the information for some of your requested points is not present in this type of submission. I will answer based on the information provided and note when information is not available due to the nature of this 510(k) for reprocessed external fixation devices.


Description of Acceptance Criteria and Study

The acceptance criteria for the Reprocessed External Fixation Devices are implicit in demonstrating that the reprocessed devices are safe, effective, and substantially equivalent to the predicate devices and perform "as originally intended." The study conducted was a series of bench and laboratory tests designed to validate the reprocessing methods and ensure the device's original performance characteristics are maintained.

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (Implicit)Reported Device Performance
Functional Equivalence: Device performs as originally intended, exhibiting the same mechanical design, materials, shapes, and sizes as predicate devices."The design, materials, and intended use of the Reprocessed External Fixation Devices are identical to the predicate devices. The mechanism of action of the Reprocessed External Fixation Device is identical to the predicate devices in that the same standard mechanical design, materials, shapes and sizes are utilized. There are no changes to the claims, intended use, clinical applications, patient population, performance specifications, or method of operation." "Performance testing demonstrates that Reprocessed External Fixation Devices perform as originally intended."
Material Integrity: Reprocessing does not compromise the material properties.Implied by the statement that "design, materials, and intended use... are identical to the predicate devices" and "performance testing demonstrates that Reprocessed External Fixation Devices perform as originally intended." (No specific material property tests mentioned beyond functional tests for reprocessing)
Sterility: If provided sterile, the reprocessed device meets sterility requirements."Sterilization Validation (for sterile devices)" was conducted.
Safety and Effectiveness: Reprocessed devices maintain safety and effectiveness comparable to predicate devices."Bench and laboratory testing was conducted to demonstrate performance (safety and effectiveness) of Reprocessed External Fixation Devices. This included the following tests: Validation of reprocessing, Sterilization Validation (for sterile devices), Function test(s)." "Alliance Medical Corporation concludes that the modified devices (Reprocessed External Fixation Device) are safe, effective, and substantially equivalent to the predicate devices as described herein."
Reprocessing Validation: The reprocessing procedure effectively cleans and prepares the device for re-use without degradation."Validation of reprocessing" was conducted.

2. Sample size used for the test set and the data provenance:

  • Sample Size: Not specified in the provided document. The 510(k) summary only states "Bench and laboratory testing was conducted," implying that a sample of reprocessed devices would have been tested, but the exact number is not disclosed.
  • Data Provenance: The document does not specify the country of origin of the data. Given it's a US submission, it's highly probable the testing occurred in the USA. The tests were likely conducted prospectively as part of the reprocessing validation process.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • This information is not applicable and not provided. The "ground truth" for proving substantial equivalence of reprocessed mechanical devices primarily relies on objective engineering and scientific measurements from bench and laboratory tests, rather than expert clinical assessment of images or patient outcomes. The "ground truth" would be the engineering specifications and performance characteristics of the original predicate devices.

4. Adjudication method for the test set:

  • This information is not applicable and not provided. Adjudication methods like 2+1 or 3+1 are typically used in clinical studies or studies involving human interpretation (e.g., radiologists, pathologists) where there might be inter-reader variability. For bench testing of mechanical properties, results are generally objective measurements that do not require clinical adjudication.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • This information is not applicable and not provided. This device is a reprocessed external fixation device, not an AI-powered diagnostic or therapeutic tool that would involve human readers or AI assistance.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

  • This information is not applicable and not provided. This device is a mechanical medical device, not an algorithm or AI system.

7. The type of ground truth used:

  • The ground truth used for these tests would be the original manufacturer's specifications and performance data for the predicate (new) external fixation devices. The reprocessed devices are demonstrated to be substantially equivalent if their performance in the bench and lab tests meet these established specifications.

8. The sample size for the training set:

  • This information is not applicable and not provided. There is no concept of a "training set" for a reprocessed mechanical device as there would be for an AI algorithm.

9. How the ground truth for the training set was established:

  • This information is not applicable and not provided. As explained above, there is no training set for this type of device submission.

{0}------------------------------------------------

K052918 (pg 1 of 2)
JAN 27 2006

PART B: 510(k) SUMMARY

Submitter:Alliance Medical Corporation10232 South 51st StreetPhoenix, Arizona 85044
Contact:Moira BartonRegulatory Affairs Manager(480) 763-5300 (o)(480) 763-5310 (f)mbarton@alliance-medical.com
Date of preparation:August 2, 2005
Name of device:Trade/Proprietary Name: Reprocessed External FixationDevicesCommon or Usual Name: External Fixation Devices, FixationAppliance, Single/Multiple ComponentClassification Name: Single/Multiple Component Metallic BoneFixation Appliances and Accessories and Smooth or ThreadedMetallic Bone Fixation Fastener
Predicate Device510(k) TitleManufacturer
K021933TempFix™ External Fixation System
K875022TempFix™ External Fixation System, MR Safe
K875012DePuy ACE AIM® Titanium Supracondylar Nail
K910986Hoffmann Dynamic Wrist Fixator

Device description: External fixation devices are specially designed frames, clamps, rods, rod-to-rod couplings, pins, posts, fasteners, wire fixations, fixation bolts, washers, nuts, hinges, sockets, connecting bars and screws used for the management of bone fractures and reconstructive, as well as corrective, orthopedic surgery. Materials used include metal alloys, plastic and composites. These materials are chosen to address a wide range of fractures and applications as well as to allow for the appropriate amount of rigidity and stability.

Intended use: External Fixation Devices are intended to be used for the fixation of supracondylar, or condylar fractures of the femur; for fusion of a joint; for surgical procedures that involve cutting the bone, for fixation of bone fractures; bone reconstruction; as a guide pin for insertion of other implants; or may be implanted through the skin so that a pulling force or traction may be applied to the skeletal system; and others may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that pulling force (traction) may be applied to the skeletal system.

{1}------------------------------------------------

K052918 (pg 2 of 2)

  • Indications Reprocessed external fixation devices are indicated for use in statement: patients requiring external skeletal fixation and treatment of fractures, osteotomy, arthrodesis, correction of deformities, fracture revision, bone reconstruction procedures, limb lengthening, correction of bony or soft tissue deformities and segmental bony or soft tissue defects.
  • Technological The design, materials, and intended use of the Reprocessed External Fixation Devices are identical to the predicate devices. characteristics: The mechanism of action of the Reprocessed External Fixation. Device is identical to the predicate devices in that the same standard mechanical design, materials, shapes and sizes are utilized. There are no changes to the claims, intended use, clinical applications, patient population, performance specifications, or method of operation.
  • Performance data: These devices will be provided sterile and non-sterile depending on customer preference.

Bench and laboratory testing was conducted to demonstrate performance (safety and effectiveness) of Reprocessed External Fixation Devices. This included the following tests:

  • Validation of reprocessing ●
  • Sterilization Validation (for sterile devices) .
  • . Function test(s)

Performance testing demonstrates that Reprocessed External Fixation Devices perform as originally intended.

  • Conclusion: Alliance Medical Corporation concludes that the modified devices (Reprocessed External Fixation Device) are safe, effective, and substantially equivalent to the predicate devices as described herein.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the seal of the Department of Health and Human Services (HHS) of the United States. The seal features the department's name encircling a symbol. The symbol is a stylized eagle with three lines representing its wings, and the eagle's head is facing to the left.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JAN 2 7 2006

Ms. Moira Barton Manager, Regulatory Affairs Alliance Medical Corporation 10232 South 51st Street Phoenix, Arizona 85044

Re: K052918

Trade/Device Name: Alliance Medical Corporation Reprocessed External Fixation Devices (See enclosed list) Regulation Number: 21 CFR 888.3030 and 21 CFR 888.3040 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories, Smooth or threaded metallic bone fixation fastener Regulatory Class: II Product Code: KTT, KTW, JEC Dated: December 21, 2005 Received: January 5, 2006

Dear Ms. Barton:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, I'DA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA 's issuance of a substantial equivalence determination docs not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set

{3}------------------------------------------------

Page 2 - Ms. Barton

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0210. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely vours.

Sincerely yours,

Barbara BuchmuD
for

Mark N. Melkerson Acting Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

SystemPartDescription
Ace-Fisher® FixatorSystemFC-10027Composite 2/3 Ring, Small (6" diam.)
FC-10028Composite 1/3 Ring, Small (6" diam.)
FC-10029Composite 2/3 Ring, Medium (8" diam.)
FC-10030Composite 1/3 Ring, Medium (8" diam.)
FC-10031Composite 2/3 Ring, Large (10" diam.)
FC-10032Composite 1/3 Ring, Large (10" diam.)
8180-04-008Ring Foot Medium
8180-04-011Medium Foot Frame
8180-04-013Foot Frame Crossbar
FA-10000-3Long Rod
FA-10000-2Medium Rod
FA-10000-1Short Rod
FA-10000-0Extra Short Rod
FA-103551 Pin Holder
FA-103602 Pin Holder
FA-103653 Pin Holder
FA-10054-290° Pin Holder Adapter
FF-1060590° Adapter Rod
FF-10023Ring Connector
FF-10610Articulation
Hoffman ClassicSterile Wrist KitA13Articulation 8/8 mm
A14Articulation 8/5 mm
A16Compression/Distraction Bar
A18/150Connecting Rod 150mm
A18/200Connecting Rod 200mm
A18/250Connecting Rod 250mm
A18/300Connecting Rod 300mm
A18/350Connecting Rod 350mm
A18/400Connecting Rod 400mm
AC11Ball Joint w/ Rod
AC10Ball Joint
AC16Washer
8082-00-000Classic Sterile Wrist Kit(AMC does not reprocess the pins)

Reprocessed External Fixation Device Models found to be Substantially Equivalent:

: :

{5}------------------------------------------------

2. Indications for Use Statement

510(k) Number (if known): K052918

Device Name: Alliance Medical Corporation Reprocessed External Fixation Devices

Indications For Use: Reprocessed external fixation devices are indicated for use in patients requiring external skeletal fixation and treatment of fractures, osteotomy, arthrodesis, correction of deformities, fracture revision, bone reconstruction procedures, limb lengthening, correction of bony or soft tissue deformities and segmental bory or soft tissue defects. ·

Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR

Over-the-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Tartian Meehim

(Division Sign-Of Division of General, Restorative, and Neurological Devices

Alliance Medical Corporation Reprocessed External Fixation Device Traditional 510(k)

510(k) Number K052918

14

§ 888.3030 Single/multiple component metallic bone fixation appliances and accessories.

(a)
Identification. Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.(b)
Classification. Class II.