K Number
K052911

Validate with FDA (Live)

Manufacturer
Date Cleared
2005-10-31

(14 days)

Product Code
Regulation Number
892.1550
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ACCUVIX XQ ™ system and transducers are intended for diagnostic ultrasound imaging and fluid analysis of the human body. The clinical applications include: Fetal (including infertility monitoring of follicle development), Abdominal, Pediatric, Small Organ, Neonatal Cephalic, Adult Cephalic, Trans-rectal, Trans-vaginal, Transesophageal, Muscular-skeletal, Urology, TCD, Cardiac (Adult, Pediatric), and Peripheral-vascular (Carotid, Arterial, Venous) applications.

Device Description

The ACCUVIX XQ "M is a general purpose, mobile, software controlled, diagnostic ultrasound system. Its function is to acquire ultrasound data and to display the data as 2D mode, M mode, Color Doppler imaging, Power Doppler imaging, Harmonic imaging, and PW Spectral Doppler mode on the CRT display. It also provides the 3D imaging mode using the 3D probe in the Mechanical scan mode. The system also provides for the measurement of anatomical structures and for analysis packages that provide information used for clinical diagnostic purposes by competent health care professionals.

AI/ML Overview

The provided document is a 510(k) summary for the ACCUVIX XQ™ Diagnostic Ultrasound System. It primarily focuses on demonstrating substantial equivalence to previously cleared predicate devices and outlining the intended uses and technical characteristics.

Crucially, this document does NOT contain information about specific clinical studies or performance metrics (like sensitivity, specificity, etc.) that would typically define acceptance criteria and prove a device meets those criteria for image analysis software or AI algorithms.

The information provided describes a diagnostic ultrasound system and its transducers, which are hardware devices. The "performance" being discussed relates to their ability to produce various types of ultrasound images and measurements, and their compliance with safety standards. It does not describe a device that interprets images or performs a specific diagnostic task for which concrete performance metrics would be required.

Therefore, many of the requested points cannot be extracted from this document.

Here's an attempt to answer the questions based only on the provided text, highlighting what is missing:


Acceptance Criteria and Device Performance (Based on available information)

The ultimate "acceptance criterion" for this type of device (a diagnostic ultrasound system) in a 510(k) submission is typically demonstrating substantial equivalence to existing legally marketed predicate devices, and adherence to relevant safety and performance standards. The device's performance is implicitly tied to its ability to function as intended across various clinical applications.

Acceptance Criterion (Implicit)Reported Device Performance
Substantial equivalence to predicate ultrasound systemsACCUVIX XQ™ is substantially equivalent to SONOACE 9900 PLUS (K032329) and SONOACE 9900 (K012867, K002185)
Compliance with electromechanical safety standardsMeets UL 60601-1, C22.2 No. 601.1, EN 60601-1, EN 60601-1-2, IEC 61157, ISO10993
Ability to acquire and display various ultrasound modesAcquires and displays 2D, M, Color Doppler, Power Doppler, Harmonic imaging, PW Spectral Doppler, and 3D imaging modes.
Provision of anatomical measurements and analysis packagesProvides measurement of anatomical structures and analysis packages for clinical diagnostic purposes.
Display of acoustic output indices (Mechanical Index, Thermal Index)Real-time acoustic output display with Mechanical Index and Thermal Index, automatically displayed.
Performance across specified clinical applications with listed transducersDemonstrated for various applications (Fetal, Abdominal, Pediatric, Small Organ, etc.) and specific transducers, including modes of operation (B, M, PWD, CWD, Color Doppler, Combined, Other). Specific transducer models listed.
Hazard analysis with "Moderate" level of concernHazard analysis performed, "Level of Concern" classified as "Moderate".

Study Details (Cannot be fully answered due to nature of document)

  1. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Not specified. This document describes a hardware device (ultrasound system) and its transducers. It discusses general intended uses and compliance with technical standards, not the performance of an AI or image analysis algorithm that would typically require a test set of patient data. The "studies" referenced are conformity to electromechanical safety standards.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not applicable. There is no mention of a test set requiring expert-established ground truth for an AI or image analysis task.
  3. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable. No such test set or adjudication is described.
  4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No. This document describes a diagnostic ultrasound system, not an AI or image analysis software that assists human readers. No MRMC study is mentioned.
  5. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    • No. This is a hardware device. No standalone algorithm performance is discussed.
  6. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    • Not applicable. No ground truth for image interpretation or diagnosis is discussed. The "ground truth" here would relate to whether the machine produces images correctly and within safety limits, which is assessed through engineering tests and inspections against standards.
  7. The sample size for the training set

    • Not applicable. This document does not describe an AI or machine learning algorithm requiring a training set.
  8. How the ground truth for the training set was established

    • Not applicable. No training set is described.

Summary of what the document implies about "studies":

The "studies" are primarily engineering and safety tests to ensure the device meets specified electromechanical standards (UL, C22.2, EN, IEC, ISO10993) and that its acoustic output is within declared limits. The "conclusion" section states: "The submission contains the results of a hazard analysis and the 'Level of Concern for potential hazards has been classified as 'Moderate'". It also mentions that the device "will be manufactured in accordance with the voluntary standards listed." A post-clearance special report is requested to contain "complete information, including acoustic output measurements based on production line devices."

The document is a regulatory filing for a medical imaging hardware system, not a software or AI device for image analysis. Therefore, the types of clinical studies and performance criteria typically associated with AI/software devices (like sensitivity, specificity, sample size of test/training data, expert adjudication, etc.) are absent.

{0}------------------------------------------------

0C1 3 1 2005

510(K) SUMMARY OF SAFETY AND EFFECTIVENESS

Prepared March 28, 2005

This summary of safety and effectiveness is provided as part of this Premarket Notification in compliance with 21 CFR, Part 807, Subpart E, Section 807.92.

1. Submitter's Information: 21 CFR 807.92(a)(1)

Mr. Kyung-Am, Shim Regulatory Affairs Manager Medison Co. Ltd. 997-10, Daechi-dong, Gangnam-gu, Seoul 135-280, Korea 82.2.2194.1381 Telephone: 82.2.2194.1399 Facsimile: kashim@medison.com Email:

2. Name of the device:

Common/Usual Name:

Diagnostic Ultrasound System and Accessories

Proprietary Name:

ACCUVIX XQ TM Diagnostic Ultrasound System and Transducers

Classification Names:FR NumberProduct Code
Ultrasonic Pulsed Doppler Imaging System892.155090-IYN
Ultrasound Pulsed Echo Imaging System892.156090-IYO
Diagnostic Ultrasound Transducer892.157090-ITX

3. Identification of the predicate or legally marketed device:

Medison Co., Ltd. believes that ACCUVIX XQ ™ ultrasound system is substantially equivalent to the currently marketed SONOACE 9900 PLUS ultrasound system (K032329) and SONOACE 9900 ultrasound system (K012867 and K002185)

4. Device Description:

The ACCUVIX XQ "M is a general purpose, mobile, software controlled, diagnostic ultrasound system. Its function is to acquire ultrasound data and to display the data as 2D mode, M mode, Color Doppler imaging, Power Doppler imaging, Harmonic imaging, and PW Spectral Doppler mode on the CRT display. It also provides the 31) imaging mode using the 3D probe in the Mechanical scan mode.

The ACCUVIX XQ ™ has real time acoustic output display with two basic indices, a mechanical index and a thermal index, which are both automatically displayed. The

{1}------------------------------------------------

system also provides for the measurement of anatomical structures and for analysis packages that provide information used for clinical diagnostic purposes by competent health care professionals.

The ACCUVIX XQ ™ has been designed to meet the following electromechanical safety standards:

  • UL 60601-1, Underwriters Laboratories Standards, Medical Electrical Equipment
  • C22.2 No. 601.1, Canadian Standards Association, Medical Electrical Equipment
  • EN 60601-1, European Norm, Medical Electrical Equipment
  • EN 60601-1-2, European Norm, Collateral Standard: Electromagnetic Compatibility .
  • IEC 61157, International Electrotechnical Commission, Requirements for the । declaration of the acoustic output of medical diagnostic ultrasonic equipment
  • ISO10993, Biological evaluation of medical devices –

3. Intended Uses:

The ACCUVIX XQ ™ system and transducers are intended for diagnostic ultrasound imaging and fluid analysis of the human body.

The clinical applications include:

Fetal (including infertility monitoring of follicle development), Abdominal, Pediatric, Small Organ, Neonatal Cephalic, Adult Cephalic, Trans-rectal, Trans-vaginal, Transesophageal, Muscular-skeletal, Urology, TCD, Cardiac (Adult, Pediatric), and Peripheral-vascular (Carotid, Arterial, Venous) applications.

6. Technological Characteristics:

The ACCUVIX XQ TM is substantially equivalent to the SA-9900 PLUS Diagnostic Ultrasound System, cleared via K032329, and the SA-9900 Diagnostic Ultrasound System, cleared via K002185 and K012867. All systems transmit ultrasonic energy into patients, then perform post processing of received echoes to generate on-screen display of anatomic structures and fluid flow within the body. All system allow for specialized measurements of structures and flow, and calculations.

7. Conclusion:

The ACCUVIX XQ™ 510(k) Pre-Market Notification contains adequate information and data to enable FDA - CDRH to determine substantial equivalence to the predicate device.

ACCUVIX XO™ will be manufactured in accordance with the voluntary standards listed in the enclosed voluntary standard survey.

The submission contains the results of a hazard analysis and the "Level of Concern for potential hazards has been classified as "Moderate"

END of 510(K) Summary

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the seal of the Department of Health and Human Services (HHS) of the United States. The seal features a stylized eagle with its wings spread, symbolizing protection and service. The words "DEPARTMENT OF HEALTH AND HUMAN SERVICES. USA" are arranged in a circular pattern around the eagle.

Public Health Service

OCT 3 1 2005

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Medison Co., Ltd. % Mr. Neil E. Devine, Jr. Responsible Third Party Official Intertek Testing Serves NA, Inc. 70 Codman Hill Road BOXBOROUGH MA 01719

Re: K052911

Trade Name: ACCUVIX XQTM Diagnostic Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulation Number: 21 CFR 892.1560 Regulation Name: Ultrasonic pulsed echo imaging system Regulation Number: 21 CFR 892.1570 Regulation Name: Diagnostic ultrasonic transducer Regulatory Class: II Product Code: IYN, IYO, and ITX Dated: October 14, 2005 Received: October 17, 2005

Dear Mr. Devine:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

This determination of substantial equivalence applies to the following transducers intended for use with the ACCUVIX XQ™ Diagnostic Ultrasound System, as described in your premarket notification:

Transducer Model Number

3D3-5EK Curved Array 3D4-7EK Curved Linear Array 3D5-8EK Curved Linear Array C2-61C Curved Array

{3}------------------------------------------------

C3-7IM Curved Array C5-2EL Curved Array CW2.0 Static CW CW4.0 Static CW EC4-9IS Curved Array L5-9EE Linear Array L5-12IM Linear Array L6-12IS Linear Array L8-I5IS Linear Array

MPT4-7AO Multiplane TEE P2-4AC Phased Array P3-5AC Phased Array P3-7AC Phased Array VAW3-5 Curved Array VAW4-7 Curved Array CDW5-8 Curved Array VNA6-12 Linear Array

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This determination of substantial equivalence is granted on the condition that prior to shipping the first device, you submit a postclearance special report. This report should contain complete information, including acoustic output measurements based on production line devices, requested in Appendix G, (enclosed) of the Center's September 30, 1997 "Information for Manufacturers Seeking Marketing Clearance of Diagnostic Ultrasound Systems and Transducers," If the special report is incomplete or contains unacceptable values (e.g., acoustic output greater than approved levels), then the 510(k) clearance may not apply to the production units which as a result may be considered adulterated or misbranded.

The special report should reference the manufacturer's 510(k) number. It should be clearly and prominently marked "ADD-TO-FILE" and should be submitted in duplicate to:

Food and Drug Administration Center for Devices and Radiological Health Document Mail Center (HFZ-401) 9200 Corporate Boulevard Rockville, Maryland 20850

This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.

{4}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html

If you have any questions regarding the content of this letter, please contact Rodrigo C. Perez at (301) 594-1212.

Sincerely yours,

Nancyc Brogdon

Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure(s)

{5}------------------------------------------------

510(k) No .:

Section 4.3 INDICATIONS FOR USE

DIAGNOSTIC ULTRASOUND INDICATIONS STATEMENT

ACCUVIX XQTM Diagnostic Ultrasound System
Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal (See Note 3)NNNNNNote 1Notes 2, 7, 8
AbdominalNNNNNNote 1Notes 2, 7, 8
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal Imaging& OtherLaparoscopic
PediatricNNNNNNote 1Note 2,4,5,6,7,8,9
Small Organ (See Note 5)NNNNNNote 1Note 2,5,6,8,9
Neonatal CephalicNNNNNNote 1
Adult CephalicNNNNNNote 1Note 4. 7
Trans-rectalNNNNNNote 1Note 2, 3, 8
Trans-vaginalNNNNNNote 1Note 2, 3. 8
Trans-urethral
Trans-esoph. (non-Cardiac)NNNNNNote 7. 8
Musculo-skel. (Convent.)NNNNNNote 1Note 2.5.6.8.9
Musculo-skel. (Superfic.)NNNNNNote 1Note 2.5.6.8.9
Intra-luminal
Other (spec.)
CardiacCardiac AdultNNNNNNote 1Note 4. 7
Cardiac PediatricNNNNNNote 1Note 4. 7
Trans-esophageal (Cardiac)NNNNNNote 1Note 7. 8
Other (spec.)
PeripheralVesselPeripheral vesselNNNNNNote 1Note 5. 6, 9
Other (spec.)N

N= new indication: P= previously cleared: E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Color Doppler Interdes Porter (Ampilate) (1997) - 1976)
Note 1 : B/M. B/PW Doppler, B/Color Doppler/PW Doppler/PW Doppler/PW Doppler/PW Doppler, B/Color Doppler/Color M

Note 2: Includes imaging for guidance of biopsy

Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, breast, serotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 8: 31) imaging

Note 9: Panoramic inkaging

Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109)

Davie Reardon

and Radiolo 510(k) Num

{6}------------------------------------------------

510(k) No .:
System:ACCUVIX XQ ™ Diagnostic Ultrasound System
Transducer:3D3-5EK Curved Linear Array
Intended Use:Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal (See Note 3)NNNNNote 1Notes 2, 7, 8
AbdominalNNNNNote 1Notes 2, 7, 8
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal Imaging& OtherLaparoscopic
Pediatric
Small Organ (See Note 5)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent)
Musculo-skel (Supertic.)
Intra-luminal
Other (spec.)
CardiacCardiac Adult
Cardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vessel
Other (spec.)

N= new indication: P= previously cleared: E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B/M, B/PW Doppler, B/Color Doppler/PW Doppler/PW Doppler/PW Doppler/PW Doppler/ B/Color Doppler/Color M

Note 2: Includes imaging for guidance of biopsy

Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: theroid, parathyroid, breast, scrotum and penis in adult, pediatrie and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 8: 31) imaging

Note 9: Panoramic imaging

Nancy broadon

(Division Sign-Off) ()
Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number K052911

{7}------------------------------------------------

510(k) No.:
System:ACCUVIX XQ ™ Diagnostic Ultrasound System
Transducer:3D4-7EK Curved Linear Array
Intended Use:Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track 1 only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
Ophthalmic
Fetal (See Note 3)NNNNNote INotes 2, 7, 8
AbdominalNNNNNote 1Notes 2, 7, 8
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal Imaging& OtherLaparoscopic
Pediatric
Small Organ (See Note 5)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)
Musculo-skel. (Superfic.)
Intra-luminal
Other (spec.)
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vessel
Other (spec.)

N= new indication; P= previously cleared: E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1 : B/M. B/I/W Doppler, B/Color Doppler/PW Doppler, B/Power Doppler/PW Doppler/PW Doppler, B/Color Doppler/Color M

Note 2: Includes imaging for guidance of biopsy

Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, breast, scrotum and penis in adult, pediative and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 8: 31) imaging

Note 9: Panoramic imaging

Nansy rogo

(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices 1052011

Radiological Devices
510(k) Number K0529

{8}------------------------------------------------

510(k) No.:
System:ACCUVIX XQ ™ Diagnostic Ultrasound System
Transducer:3D5-8EK Curved Linear Array
Intended Use:Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track 1 only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
Ophthalmic
Fetal (See Note 3)
Abdominal
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal Imaging& OtherLaparoscopic
Pediatric
Small Organ (See Note 5)
Neonatal Cephalic
Adult Cephalic
Trans-rectalNNNNNote 1Notes 2. 8
Trans-vaginalNNNNNote 1Notes 2. 8
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)
Musculo-skel. (Superfic.)
Intra-luminal
Other (spec.)
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralPeripheral vessel
VesselOther (spec.)

N= new indication; P= previously cleared; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1 : B/M, B/PW Doppler, B/Color Doppler/PW Doppler/PW Doppler, B/Power Doppler/PW Doppler. B/Color Doppler/Color M

Note 2: Includes imaging for guidance of biopsy

Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example; thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 8: 31) imaging

Note 9: Panoramic imaging

Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109)

and Radiological Devices

510(k) Number

Nancyc Beagdon


(Division Sign-Off)
Division of Reproductive, Abdominal,

{9}------------------------------------------------

510(k) No.:
System:ACCUVIX XQ ™ Diagnostic Ultrasound System
Transducer:C2-6IC Curved Array
Intended Use:Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal (See Note 3)PPPPNote 1Notes 2, 7, 8
AbdominalPPPPNote 1Notes 2, 7, 8
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal Imaging& OtherLaparoscopic
PediatricPPPPNote 1Notes 2. 7, 8
Small Organ (See Note 5)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)
Musculo-skel. (Superfic.)
Intra-luminal
Other (spec.)
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vessel
Other (spec.)

N= new indication: P= previously cleared: E== added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Nate 1: 13/M, B/PW Doppler, B/Color Doppler/PW Doppler/PW Doppler/PW Doppler/PW Doppler/PW Doppler. B/Color Doppler/Color M

Note 2: Includes imaging for guidance of biopsy

Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatic and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 8: 3D) imaging

Note 9: Panoramic imaging

Nancyc burgdon

(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number 4052911

{10}------------------------------------------------

510(k) No.:
System:ACCUVIX XQ TM Diagnostic Ultrasound System
Transducer:C3-7IM Curved Array
Intended Use:Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal (See Note 3)PPPPPNote 1Notes 2, 7, 8
AbdominalPPPPNote 1Notes 2, 7, 8
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal Imaging& OtherLaparoscopic
PediatricPPPPNote 1Notes 2, 7, 8
Small Organ (See Note 5)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)
Musculo-skel. (Superfic.)
Intra-luminal
Other (spec.)
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vessel
Other (spec.)

N= new indication: P - previously cleared; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Color Doppler, B/Color Doppler, B/Color Doppler/PW Doppler/PW Doppler/PW Doppler/PW Doppler. B/Color Doppler/Color M

Note 2: Includes imaging for guidance of biopsy

Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 8: 31) imaging

Note 9: Panoramic imaging

Nancy C. Hogdon

(Division Sign-Off)
Division of Reproductive, Abdominal,
Radiological Devices `/6R291/

{11}------------------------------------------------

510(k) No.:
System:ACCUVIX XQ ™ Diagnostic Ultrasound System
Transducer:C5-2EL Curved Array
Intended Use:Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal (See Note 3)NNNNNNote 1Notes 2, 7, 8
AbdominalNNNNNote 1Notes 2, 7, 8
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal Imaging& OtherLaparoscopic
PediatricNNNNNote 1Notes 2, 7, 8
Small Organ (See Note 5)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)
Musculo-skel. (Superfic.)
Intra-luminal
Other (spec.)
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vessel
Other (spec.)

N= new indication: P= previously cleared; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1 : B/M, B/PW Doppler, B/Color Doppler, B/Color Doppler/PW Doppler/PW Doppler/PW Doppler/ B/Color Doppler/Color M

Note 2: Includes imaging for guidance of biopsy

Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 8: 3D imaging

Note 9: Panoramic imaging

Nancy C. Brandon

(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices

{12}------------------------------------------------

510(k) No.:
System:ACCUVIX XQ ™ Diagnostic Ultrasound System
Transducer:CW2.0 Static CW
Intended Use:Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationBMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
General(Track I only)Specific(Tracks I & III)
OphthalmicOphthalmic
Fetal (See Note 3)
Abdominal
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal Imaging& OtherLaparoscopic
Pediatric
Small Organ (See Note 5)
Neonatal Cephalic
Adult CephalicP
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)
Musculo-skel. (Superfic.)
Intra-luminal
Other (spec.)
Cardiac AdultP
CardiacCardiac PediatricP
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vesselP
Other (spec.)P

N= new indication; P= previously cleared; E= added under Appendix F.

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B/M, B/PW Doppler, B/Color Doppler/PW Doppler/PW Doppler, B/Power Doppler/PW Doppler, B/Color Doppler/Color M

Note 2: Includes imaging for guidance of biopsy

Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 8: 31) imaging

Note 9: Panoranuc imaging

Nancyc Brandon

(Division S Division of Reproc and Radiological Devices : )

{13}------------------------------------------------

510(k) No.:ACCUVIX XQ ™ Diagnostic Ultrasound System
System:CW4.0 Static CW
Transducer:Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
Intended Use:Clinical ApplicationBMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
General(Track I only)Specific(Tracks I & III)
OphthalmicOphthalmic
Fetal Imaging& OtherFetal (See Note 3)
Abdominal
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Laparoscopic
Pediatric
Small Organ (See Note 5)
Neonatal Cephalic
Adult CephalicN
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)
Musculo-skel. (Superfic.)
Intra-luminal
Other (spec.)
CardiacCardiac AdultN
Cardiac PediatricN
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vesselN
Other (spec.)N

N= new indication; P= previously cleared: E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B/M, B/PW Doppler, B/Color Doppler/PW Doppler/PW Doppler/PW Doppler/PW Doppler. B/Color Doppler/Color M

Note 2: Includes imaging for guidance of biopsy

Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 8: 31) imaging

Note 9: Panoramic imaging

(Division Sid Division of Reproduc ve, A and Radiological Devices 510(k) Number ________________________________________________________________________________________________________________________________________________________________

{14}------------------------------------------------

510(k) No.:
System:ACCUVIX XQ ™ Diagnostic Ultrasound System
Transducer:EC4-9IS Curved Array
Intended Use:Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track 1 only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal (See Note 3)AbdominalIntra-operative (See Note 6)Intra-operative (Neuro.)
Fetal Imaging& OtherLaparoscopic
Pediatric
Small Organ (See Note 5)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginalPPPPNote 1Note 2. 8
Trans-urethralPPPPNote 1Note 2. 8
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)
Musculo-skel. (Superfic.)
Intra-luminal
Other (spec.)
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vessel
Other (spec.)

N= new indication; P= previously cleared: E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Color Doppler. B/Color Doppler. B/Color Doppler/PW Doppler. B/Power Doppler/PW Doppler/PW Doppler. B/Color Doppler/Color M

  • Note 2: Includes imaging for guidance of biopsy
    Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 8: 31) imaging

Note 9: Panoramic imaging

Nancy C. Hogdon
(Division Sign-Off)

Division of Reproductive, Abdor m: Padiological Devices the ke Member ________________________________________________________________________________________________________________________________________________________________

{15}------------------------------------------------

510(k) No.:
System:ACCUVIX XQ TM Diagnostic Ultrasound System
Transducer:L5-9EE Linear Array
Intended Use:Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation (*includes simultaneous B-mode)Other(Spec.)
BMPWDCWDColorDoppler*Combined*(Spec.)
General(Track I only)Specific(Tracks I & III)
OphthalmicOphthalmic
Fetal Imaging& OtherFetal (See Note 3)
Abdominal
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Laparoscopic
PediatricNNNNNote 1Note 2, 8, 9
Small Organ (See Note 5)NNNNNote 1Note 2, 8, 9
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)NNNNNote 1Note 2, 8, 9
Musculo-skel. (Superfic.)NNNNNote 1Note 2, 8, 9
Intra-luminal
Other (spec.)
CardiacCardiac Adult
Cardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vessel
Other (spec.)

N= new indication: P= previously cleared: E = added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Color Doppier Includes Poster (Ampliade) Hoppier/PW Doppler/PW Doppler/PW Doppler/PW Doppler. B/Color Doppler/Color M

Note 2: Includes imaging for guidance of biopsy

Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: Color Mentode

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonie Imaging (THI)

Note 8: 3D imaging

Note 9: Panoramic imaging

Nancyc Brogdon

(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
K632911

{16}------------------------------------------------

510(k) No.:
System:ACCUVIX XQ ™ Diagnostic Ultrasound System
Transducer:L5-12IM Linear Array
Intended Use:Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks 1 & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal (See Note 3)AbdominalIntra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal ImagingLaparoscopic
& OtherPediatricPPPPNote 1Note 2, 5, 6. 9
Small Organ (See Note 5)PPPPNote 1Note 2. 5. 6. 9
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)PPPPNote 1Note 2. 5. 6. 9
Musculo-skel. (Superfic.)PPPPNote 1Note 2. 5. 6. 9
Intra-luminal
Other (spec.)
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralPeripheral vesselPPPPNote 1Note 5. 6. 9. 10
VesselOther (spec.)

N= new indication: P= previously cleared: E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Color Dopper Includes Poser (rumplande) Hogiler/PW Doppler/PW Doppler/PW Doppler/PW Doppler/PW Doppler/PW Doppler/ B/Color Doppier/Color M

Note 2: Includes imaging for guidance of biopsy

Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: Eoror without

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 8: 31) imaging

Note 9: Panoramic imaging

Nancyc broadon
(Division Sign-Off)
Division of Reproductive, Abdominal,
Radiological Devices
Ek Number
K052911

{17}------------------------------------------------

510(k) No.:
System:ACCUVIX XQ TM Diagnostic Ultrasound System
Transducer:L6-12IS Linear Array
Intended Use:Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal (See Note 3)
Abdominal
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal Imaging& OtherLaparoscopic
PediatricNNNNNote 1Note 2, 5, 6, 9
Small Organ (See Note 5)NNNNNote 1Note 2, 5, 6, 9
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)NNNNNote 1Note 2. 5. 6. 9
Musculo-skel. (Superfic.)NNNNNote 1Note 2. 5. 6. 9
Intra-luminal
Other (spec.)
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vesselNNNNNote 1Note 5. 6. 9
Other (spec.)

N= new indication; P= previously cleared: E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Color Doppler Internet Princiate) Longiler. B/Color Doppler. B/Power Doppler/PW Doppler/PW Doppler. B/Color Doppler/Color M

Note 2: Includes imaging for guidance of biopsy

Note 3 : Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 8: 3D imaging

Note 9: Panoramic imaging

Nancy C. Hodgson
(Division Sign-Off)

sion of Reproductive, Abdominal,
siological Devices
KU524

{18}------------------------------------------------

510(k) No.:
System:ACCUVIX XQ TM Diagnostic Ultrasound System
Transducer:L8-15IS Linear Array
Intended Use:Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks 1 & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal (See Note 3)
Abdominal
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal ImagingLaparoscopic
& OtherPediatricNNNNNNote 1Note 2, 5, 6, 9
Small Organ (See Note 5)NNNNNNote 1Note 2, 5. 6, 9
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)NNNNNNote 1Note 2. 5. 6, 9
Musculo-skel. (Superfic.)NNNNNNote 1Note 2. 5. 6.9
Intra-luminal
Other (spec.)
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralPeripheral vesselNNNNNNote 1Note 5. 6. 9
VesselOther (spec.)

N= new indication: P= previously cleared: E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Color Doppler Includes Porce (Ampilate) (106) Per Per Par Poppler/PW Doppler/PW Doppler/PW Doppler/PW Doppler/PW Doppler/PW Doppler/PW Doppler/PW Doppler/PW Doppler/PW Dopple B/Color Doppler/Color M

Note 2: Includes imaging for guidance of biopsy

Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 8: 3D imaging

Note 9: Panoramic imaging

Yauric berger

Division Sign-Off
Division of Reproductive, Abdominal,
and Radiological Devices

K052911

{19}------------------------------------------------

510(k) No .:
System:ACCUVIX XQ TM Diagnostic Ultrasound System
Transducer:MPT4-7AO Multiplane TEE
Intended Use:Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal (See Note 3)AbdominalIntra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal ImagingLaparoscopic
& OtherPediatric
Small Organ (See Note 5)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)NNNNNNote INote 7. 8
Musculo-skel. (Convent.)
Musculo-skel. (Superfic )
Intra-luminal
Other (spec.)
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)NNNNNNote lNote 7. 8
Other (spec.)
PeripheralPeripheral vessel
VesselOther (spec.)

N= new indication: P= previously cleared: E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Color Doppler Internet (Timpleote) (1) ) Doppler/PW Doppler/PW Doppler/PW Doppler/PW Doppler.

B/Color Doppler/Color M

Note 2: Includes imaging for guidance of biopsy

Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, breast, scrounn and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THD

Note 8: 31) imaging

Note 9: Panoramic imaging

Nancy C. Braydon
(Division Sign Off)

Division of Reproductive, Abdominal,
and Radiological Devices
(HHK) Number K052411

{20}------------------------------------------------

510(k) No.:
System:ACCUVIX XQ ™ Diagnostic Ultrasound System
Transducer:P2-4AC Phased Array
Intended Use:Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal (See Note 3)
AbdominalNNNNNNote 1Note 4, 7
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal Imaging& OtherLaparoscopic
Pediatric
Small Organ (See Note 5)
Neonatal Cephalic
Adult CephalicNNNNNNote 1Note 4, 7
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)
Musculo-skel. (Superfic.)
Intra-luminal
Other (spec.)
Cardiac AdultNNNNNNote 1Note 4. 7
CardiacCardiac PediatricNNNNNNote 1Note 4. 7
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vessel
Other (spec.)

N= new indication: P= previously cleared: E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Color Doppler Includes Poller (Anglades) Weler Doppler/PW Doppler/PW Doppler/PW Doppler/PW Doppler. B/Color Doppler/Color M

Note 2: Includes imaging for guidance of biopsy

Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, breast, scrotum and penis in adult, pediatic and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 8: 3D imaging

Note 9: Panoramic imaging

Nancy Beagdon

Division ! Invision of Renrodi P. Redinlogical Devices . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

{21}------------------------------------------------

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal (See Note 3)
AbdominalNNNNNNote 1Note 4, 7
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal Imaging& OtherLaparoscopic
Pediatric
Small Organ (See Note 5)
Neonatal Cephalic
Adult CephalicNNNNNNote 1Note 4, 7
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)
Musculo-skel. (Superfic.)
Intra-luminal
Other (spec.)
Cardiac AdultNNNNNNote 1Note 4, 7, 10
CardiacCardiac PediatricNNNNNNote 1Note 4, 7, 10
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vessel
Other (spec.)

N= new indication: P= previously cleared: E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Color Doppler Includes Porter (Anpacace) Doppler/PW Doppler/PW Doppler/PW Doppler/PW Doppler/PW Doppler/ B/Color Doppler/Color M

Note 2: Includes imaging for guidance of biopsy

Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, breast, scrotum and penis in adult, pediativ and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 8: 3D imaging

Note 9: Panoramic imaging

Nancy Gordon
(Division Sign-Off)

Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number K152911

{22}------------------------------------------------

:

510(k) No.:
System:ACCUVIX XQ TM Diagnostic Ultrasound System
Transducer:P3-7AC Phased Array
Intended Use:Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal (See Note 3)
AbdominalPPPPPNote 1Note 4, 7
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal ImagingLaparoscopic
& OtherPediatricPPPPPNote 1Note 4, 7
Small Organ (See Note 5)
Neonatal Cephalic
Adult CephalicPPPPPNote 1Note 4, 7
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)
Musculo-skel. (Superfic.)
Intra-luminal
Other (spec.)
CardiacCardiac Adult
Cardiac PediatricPPPPPNote 1Note 4. 7. 10
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralPeripheral vessel
VesselOther (spec.)

N= new indication; P= previously cleared: E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1 : B/M, B/PW Doppler, B/Color Doppler/PW Doppler, B/Power Doppler/PW Doppler/PW Doppler. B/Color Doppler/Color M

Note 2: Includes imaging for guidance of biopsy

Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 8: 3D imaging

Note 9: Panoramic imaging

Nancy C. Hogdon

(Division Sign-Off)

Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number K152911

{23}------------------------------------------------

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal Imaging& OtherFetal (See Note 3)PPPPNote 1Note 2, 7, 8
AbdominalPPPPNote 1Note 2, 7, 8
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Laparoscopic
PediatricPPPPNote 1Note 2, 7, 8
Small Organ (See Note 5)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)
Musculo-skel. (Superfic.)
Intra-luminal
Other (spec.)
CardiacCardiac Adult
Cardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vessel
Other (spec.)

N= new indication: P= previously cleared: E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B/M, B/PW Doppler, B/Color Doppler/PW Doppler, B/Power Doppler/PW Doppler/PW Doppler, B/Color Doppler/Color M

Note 2: Includes imaging for guidance of biopsy

Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatrie and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 8: 31) imaging

Note 9: Panoranne imaging

Nancy C Brogdon

(Division Sign-Off)
Division of Reproductive, Abdominal,
Radiological Devices
K162911
Number

{24}------------------------------------------------

510(k) No .:
System:ACCUVIX XQ TM Diagnostic Ultrasound System
Transducer:VAW4-7 Curved Array
Intended Use:Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
Mode of Operation (*includes simultaneous B-mode)
Clinical ApplicationBMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal ( See Note 3 )PPPPNote 1Note 2, 7, 8
AbdominalPPPPNote 1Note 2, 7, 8
Intra-operative ( See Note 6 )
Intra-operative (Neuro.)
Fetal Imaging& OtherLaparoscopic
PediatricPPPPNote 1Note 2, 7, 8
Small Organ ( See Note 5 )PPPPNote 1Note 2, 7, 8
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)
Musculo-skel. (Superfic.)
Intra-luminal
Other (spec.)
CardiacCardiac Adult
Cardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vessel
Other (spec.)

N= new indication; P= previously cleared; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B/M. B/PW Doppler, B/Color Doppler/PW Doppler/PW Doppler/PW Doppler/PW Doppler/PW Doppler. B/Color Doppler/Color M

Note 2: Includes imaging for guidance of biopsy

Note 3: Includes infertility monitoring of follicle development

Note 4. Color M-mode

Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 8: 31) imaging

Note 9: Panoramic imaging

Nancy C. Snogdon

sion of Reproductive, Abdominal

K05291

{25}------------------------------------------------

510(k) No.:
System:ACCUVIX XQ ™ Diagnostic Ultrasound System
Transducer:VDW5-8 Curved Array
Intended Use:Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColor Doppler*Combined*(Spec.)Other(Spec.)
Ophthalmic
Fetal (See Note 3)
Abdominal
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal Imaging& OtherLaparoscopic
Pediatric
Small Organ (See Note 5)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginalPPPPNote 1Note 2, 8
Trans-urethralPPPPNote 1Note 2, 8
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)
Musculo-skel. (Superfic.)
Intra-luminal
Other (spec.)
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vessel
Other (spec.)

N= new indication: P= previously cleared; E- added under Appendix f:

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B/M, B/PW Doppler, B/Color Doppler/PW Doppler, B/Power Dappler/PW Doppler/PW Doppler. B/Color Doppler/Color M

Note 2: Includes imaging for guidance of biopsy

Note 3: Includes infertility monitoring of follicle development

Note 4 Color M-mode

Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatie and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note S. 3D imaging

Note 9: Panoramic imaging

Nancyi hogdon

Division Sign-Off
Division of Reproductive Abdominal
UC5231

{26}------------------------------------------------

510(k) No.:
System:ACCUVIX XQ TM Diagnostic Ultrasound System
Transducer:VNA6-12 Linear Array
Intended Use:Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal Imaging& OtherFetal (See Note 3)
Abdominal
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Laparoscopic
Pediatric
Small Organ (See Note 5)PPPPNote 1Note 2, 8
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)PPPPNote 1Note 2, 8
Musculo-skel. (Superfic.)PPPPNote 1Note 2, 8
Intra-luminalPPPPNote 1Note 2, 8
Other (spec.)
CardiacCardiac Adult
Cardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vessel
Other (spec.)

N= new indication; P= previously cleared: E+ added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B/M, B/PW Doppler, B/Color Doppler/PW Doppler/PW Doppler/PW Doppler/PW Doppler/PW Doppler, B/Color Doppler/Color M

Note 2: Includes imaging for guidance of biopsy

Note 3: Includes infertifity monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatrie and neanatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 8: 31) imaging

Note 9. Panoramic imaging

Nancy C. Hogdon
Division Sign-Off

Division of Reproductive, Abdominal,
and Radiological Devices
K052911

§ 892.1550 Ultrasonic pulsed doppler imaging system.

(a)
Identification. An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II.