K Number
K052843

Validate with FDA (Live)

Date Cleared
2006-04-17

(192 days)

Product Code
Regulation Number
880.5570
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Needle Stick Block is a single needle holder to be used after the needle is removed from the patient and until it can be properly disposed, therefore, aiding in the prevention of needle stick injuries. The Needle Stick Block is a safety device designed to help the health care facilities to minimize the risk of needle sticks. Busse Hospital Disposable intent to sell this device individually and as part of safety trays in which needle is involved. The intended use will remain the same whether is sold individually or as part of a tray.

Device Description

Sterile/ Latex Free/ Needle Stick Block. The proposed Needle Stick Block was created to help prevent needle sticks by providing a means of moving used sharps from procedure and hold it temporarily until procedure is completed to then be disposed off in a sharps container.

AI/ML Overview

This document describes the regulatory submission for the "Needle Stick Block" device and the testing performed to demonstrate its substantial equivalence to a predicate device, the "SharpsAway Disposal Cup."

Here's an analysis of the provided information regarding acceptance criteria and the supporting study:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria / Test PerformedReported Device Performance
A) Determination of puncture resistance of the foam insert for both devices.Implicitly, the Needle Stick Block's foam insert was found to have comparable puncture resistance to the predicate device's foam insert. (The document states "Summary of tests performed to prove the substantial equivalence... A) Determination of puncture resistance of the foam insert for both devices.") Specific quantitative results are not provided.
B) Determination of puncture resistance of vessel wall for Needle Stick Block and SharpsAway predicate device.Implicitly, the Needle Stick Block's vessel wall was found to have comparable puncture resistance to the predicate device's vessel wall. (The document states "Summary of tests performed to prove the substantial equivalence... B) Determination of puncture resistance of vessel wall for Needle Stick Block and SharpsAway predicate device.") Specific quantitative results are not provided.
C) Test to show that once inserted, Needle Stick Block with foam insert exerts enough resistance to retain needle for at least one hour under normal circumstances.Implicitly, the Needle Stick Block met this retention criterion. (The document states "Summary of tests performed to prove the substantial equivalence... C) Test to show that once inserted, Needle Stick Block with foam insert exerts enough resistance to retain needle for at least one hour under normal circumstances.") Specific quantitative results are not provided.
D) Simulate use study by experienced medical experts (at least 1 year experience with handling needles using the predicate instrument) to demonstrate: 1. Same intended use 2. Same featuresImplicitly, the simulated use study demonstrated that the Needle Stick Block has the same intended use and features as the predicate device when used by experienced medical experts. (The document states "D) Simulate use study by experienced medical experts (at least 1 year experience with handling needles using the predicate instrument)"). Specific qualitative or quantitative results are not provided beyond the general statement of substantial equivalence.

Note: The document provided is a 510(k) summary, which often provides high-level statements about tests performed but often omits detailed quantitative results. The "reported device performance" is inferred from the statement that these tests were performed "to prove the substantial equivalence."


Study Details:

The overall goal of the study was to demonstrate substantial equivalence of the "Needle Stick Block" to the "SharpsAway Disposal Cup" predicate device.

2. Sample Size Used for the Test Set and Data Provenance:

  • Sample Size (Test Set): Not explicitly stated for any of the tests (A, B, C, D).
  • Data Provenance: Not explicitly stated (e.g., country of origin, retrospective/prospective). However, the company, Busse Hospital Disposables, is located in Hauppauge, NY, United States, suggesting the tests would likely have been conducted in the US. The nature of the tests (material properties and simulated use) suggests they were conducted prospectively.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications:

  • Number of Experts: Not explicitly stated for the "simulate use study" (Test D). It refers to "experienced medical experts."
  • Qualifications of Experts: "at least 1 year experience with handling needles using the predicate instrument."

4. Adjudication Method for the Test Set:

  • Not applicable as the tests described are primarily engineering/material property tests and a simulated use study, not diagnostic performance evaluations requiring adjudication of expert interpretations.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done:

  • No, a multi-reader multi-case (MRMC) comparative effectiveness study was not conducted. The study involves material property testing and a simulated use study, not a comparative effectiveness study of human readers' performance with and without AI assistance.

6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done:

  • This question is not applicable as the device is a physical medical device (a needle holder), not an algorithm or AI software. Therefore, there is no "standalone performance" of an algorithm. The device's performance is inherently physical and involves human interaction for its intended use.

7. The Type of Ground Truth Used:

  • For tests A, B, and C (puncture resistance, needle retention), the "ground truth" would be established by the physical measurement and engineering standards used to quantify these properties.
  • For test D (simulated use study), the "ground truth" was established by the observation and assessment of "experienced medical experts" confirming the device's equivalent intended use and features.

8. The Sample Size for the Training Set:

  • This question is not applicable as the device is a physical medical device, not an AI/ML algorithm that requires a training set.

9. How the Ground Truth for the Training Set Was Established:

  • This question is not applicable as the device is a physical medical device, not an AI/ML algorithm that requires a training set.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for "busse Hospital Disposables". The word "busse" is in a bold, sans-serif font, with the letters slightly overlapping each other. Below the word "busse" is the phrase "Hospital Disposables" in a smaller, sans-serif font. The logo is in black and white.

APR 17 2006

SUMMARY

510(k) SUMMARY OF SAFETY AND EFFECTIVENESS NEEDLE STICK BLOCK

Regulatory Affairs Contact:Muhamad AnsariBusse Hospital DisposablesPO Box: 1106775 Arkay Dr.Hauppauge NY 11788
Telephone:631-435-4711 Ext: 254
Fax:631-435-2849
Date Summary prepared:Sept 23, 2005
Date Summary Revised:February 27, 2006
Product Trade Name:Needle Stick Block
Common Name:needle, hypodermic, single lumen.
Classification:Class II, 21 CFR 880.5570, FMI
Predicate Device:SharpsAway Disposal Cup
Description:Sterile/ Latex Free/ Needle Stick BlockThe proposed Needle Stick Block was created tohelp prevent needle sticks by providing a means ofmoving used sharps from procedure and hold ittemporarily until procedure is completed to then bedisposed off in a sharps container.
Intended Use:The Needle Stick Block is a single needle holder tobe used after the needle is removed from the patientand until it can be properly disposed, therefore,aiding in the prevention of needle stick injuries.The Needle Stick Block is a safety device designedto help the health care facilities to minimize the riskof needle sticks.Busse Hospital Disposable intent to sell this deviceindividually and as part of safety trays in whichneedle is involved. The intended use will remain thesame whether is sold individually or as part of atray.

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the logo for Busse Hospital Disposables. The logo features the word "busse" in a bold, sans-serif font, with the letters closely spaced together. Below the word "busse" are the words "Hospital Disposables" in a smaller, less bold font. The logo is simple and professional, and it is likely used on the company's products and marketing materials.

510(k) SUMMARY OF SUBSTANTIAL EQUIVALENCE NEEDLE STICK BLOCK

Summary of tests performed to prove the substantial equivalence of the Needle Stick Block with the predicate device:

  • A) Determination of puncture resistance of the foam insert for both devices.
  • B) Determination of puncture resistance of vessel wall for Needle Stick Block and SharpsAway predicate device.
  • C) Test to show that once inserted, Needle Stick Block with foam insert exerts enough resistance to retain needle for at least one hour under normal circumstances.
  • D) Simulate use study by experienced medical experts (at least 1 year experience with handling needles using the predicate instrument)
      1. Same intended use
      1. Same features

Technological Characteristics [21 CFR 807.92(a)(6)] The subject device has the same Technological Characteristics as a legally marketed predicate device.

Conclusion [21 CFR 807.92(b)(3)]

The above statements are accurate representations of the device Busse intents to market. All data and information submitted in this premarket notification is truthful and accurate and no material fact has been omitted.

Manufacturer:

Busse Hospital Disposables.

Official Correspondent:

Mathurmo Oksic (Signature)

Muhamad Ansari (printed name)
Title: Director of Regulatory Affairs
Date: 312106

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows a circular logo with text around the perimeter and a symbol in the center. The text reads "U.S. Department of Health and Human Services". The symbol in the center appears to be three curved lines that resemble a stylized bird or abstract design.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

APR 1 7 2006

Mr. Muhamad Ansari Director of Regulatory Affairs Busse Hospital Disposables, Incorporated 75 Arkay Drive P.O. Box 11067 Hauppauge, New York 11788

Re: K052843

Trade/Device Name: Needle Stick Block Regulation Number: 880.5570 Regulation Name: Hypodermic single lumen needle Regulatory Class: II Product Code: FMI Dated: February 27, 2006 Received: March 7, 2006

Dear Mr. Ansari:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (IPMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 – Mr. Muhamad Ansari

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4618. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Chiu Lin, Ph.D.

Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

K05284 3

1 જી ।

INDICATIONS FOR USE

510(k) Number (if known): K052843

Device Name: Needle Stick Block.

Indication for Use: The Needle Stick Block is a single needle holder to be used after the needle is removed from the patient and until it can be properly disposed, therefore, aiding in the prevention of needle stick injuries

Prescription Use X (Per 21 CFR 801Subpart D) AND/OR

Over-The-Counter Use (Per 21 CFR 801Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE – CONINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODF.)

Arun Dima

gr. General Hospil.

ps 2843

Page 1 of 1

§ 880.5570 Hypodermic single lumen needle.

(a)
Identification. A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.(b)
Classification. Class II (performance standards).