Intended Use

The BioSphere QuadraSphere™ Microspheres are intended for embolization of hypervascularized tumors and peripheral vascular arteriovenous malformations (AVMs).

QuadraSphere Microspheres are indicated for embolization of hypervascularized arteriovenous malformations.

Device Description

BioSphere Medical, Inc., QuadraSphere™ Microspheres are sterile, biocompatible, hydrophilic (absorbent), non-resorbable, acrylic copolymer microspheres. They are provided in three (nominal) microsphere sizes: 50 to 100 um, 100 to 150 um, and 150 to 200 um in containers (vials) containing 50 mg of the dry spheres.

AI/ML Overview

The provided text is a 510(k) summary for a medical device (QuadraSphere™ Microspheres), which focuses on establishing substantial equivalence to predicate devices rather than presenting a performance study with detailed acceptance criteria and results as one might find in a clinical trial report.

Therefore, many of the requested details about acceptance criteria, specific performance metrics, sample sizes, expert qualifications, and study designs are not present in this type of regulatory submission. The document states that "In-vitro and in-vivo design verification and validation testing demonstrates that the BioSphere Medical, Inc. QuadraSphere™ Microspheres fulfill design and performance specifications," but it does not elaborate on what those specifications or the detailed results were.

Here's a breakdown of what can be extracted and what cannot, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance:

This information is not provided in the 510(k) summary. A 510(k) summary typically focuses on demonstrating substantial equivalence to existing devices rather than detailing specific numerical acceptance criteria and performance outcomes from a statistically powered study. It states that "In-vitro and in-vivo design verification and validation testing demonstrates that the BioSphere Medical, Inc. QuadraSphere™ Microspheres fulfill design and performance specifications," but doesn't list the specifications or the results.

2. Sample Size Used for the Test Set and Data Provenance:

This information is not provided. The summary mentions "in-vitro and in-vivo" testing but does not specify the sample sizes or the origin of any data (e.g., country of origin, retrospective/prospective). This is common for 510(k) submissions, which often rely on bench testing and animal studies rather than large-scale human clinical trials.

3. Number of Experts Used to Establish Ground Truth and Qualifications:

This information is not provided. As no clinical study is described with a ground truth establishment process, there is no mention of experts, their number, or qualifications.

4. Adjudication Method for the Test Set:

This information is not provided, as no clinical test set adjudication process is described.

5. Multi-reader Multi-case (MRMC) Comparative Effectiveness Study:

This information is not provided. The document makes no mention of an MRMC study or any comparison of human readers with and without AI assistance. The device is an embolization agent, not an AI-powered diagnostic tool.

6. Standalone Performance (Algorithm Only):

This information is not applicable/not provided. The device is a physical medical implant (microspheres) for embolization, not a software algorithm. Therefore, "standalone" performance in the context of an algorithm is not relevant.

7. Type of Ground Truth Used:

This information is not provided. Since no specific human clinical study with a defined ground truth for efficacy or safety is outlined, the type of ground truth is not detailed. The "performance testing" likely refers to engineering, biocompatibility, and animal studies.

8. Sample Size for the Training Set:

This information is not provided. As the device is not an AI algorithm, there is no concept of a "training set" in the context of machine learning.

9. How Ground Truth for the Training Set Was Established:

This information is not applicable/not provided for the same reason as point 8.


Summary of Provided Information Relevant to Performance:

  • Performance Testing: The summary states that "In-vitro and in-vivo design verification and validation testing demonstrates that the BioSphere Medical, Inc. QuadraSphere™ Microspheres fulfill design and performance specifications."
  • Intended Use: Embolization of hypervascularized tumors and peripheral vascular arteriovenous malformations (AVMs).
  • Technological Characteristics: Sterile, biocompatible, hydrophilic (absorbent), non-resorbable, acrylic copolymer microspheres, available in three nominal sizes (50 to 100 um, 100 to 150 um, and 150 to 200 um). Delivered via catheters, mixed with contrast media for visualization.
  • Substantial Equivalence: The claim is based on intended use, indications for use, fundamental technological characteristics, and fundamental operational characteristics being similar to predicate devices (GelSpheres, BeadBlock Compressible Microspheres, Contour SE, Contour Emboli, Embospheres). This implies that the performance is considered acceptable because it is similar to already legally marketed devices.

This 510(k) summary serves to demonstrate that the new device is as safe and effective as a legally marketed predicate device, not to independently prove its efficacy and safety against a set of predefined, quantifiable acceptance criteria in a detailed clinical study.

{0}------------------------------------------------

K05 2742

510(k) Summary for the BioSphere Medical, Inc. QuadraSphere™ Microspheres

1/2

SUBMITTER/510(K) HOLDER 1.

BioSphere Medical, Inc. 1050 Hingham Street Rockland, MA 02370 USA

NGV - 7 2006

Contact Person: Irina Kulinets Telephone: 781-681-7900

Date Prepared: September 29, 2005

2. DEVICE NAME

Proprietary Name: QuadraSphere™ Microspheres Common/Usual Name: Embolization device Classification Name: Vascular embolization device

3. PREDICATE DEVICES

  • Biocompatibles UK Ltd., GelSpheres, BeadBlock Compressible . Microspheres (K042231)
  • Boston Scientific Corporation, Contour SE (K022427, K032542) .
  • Interventional Therapeutics Corporation, Contour Emboli (K944354) .
  • BioSphere Medical, Inc., Embospheres (K991549) .

DEVICE DESCRIPTION 4.

BioSphere Medical, Inc., QuadraSphere™ Microspheres are sterile, biocompatible, hydrophilic (absorbent), non-resorbable, acrylic copolymer microspheres. They are provided in three (nominal) microsphere sizes: 50 to 100 um, 100 to 150 um, and 150 to 200 um in containers (vials) containing 50 mg of the dry spheres.

{1}------------------------------------------------

Korsher

와는

5. INTENDED USE

The BioSphere QuadraSphere™ Microspheres are intended for embolization of hypervascularized tumors and peripheral vascular arteriovenous malformations (AVMs).

6. TECHNOLOGICAL CHARACTERISTICS AND SUBSTANTIAL EQUIVALENCE

BioSphere Medical, Inc. bases its claim of the substantial equivalence of the QuadraSphere™ Microspheres with the cited predicate devices based on intended use, indications for use, fundamental technological characteristics, and fundamental operational characteristics. In all cases, the embolization agent is delivered to selected sites through catheters with a diameter that is appropriate for the vascular target and the size of the embolization agent. Accurate placement is assured through visualization of the injection process using fluoroscopic imaging. Embolization agents are mixed with the contrast media before injection to optimize visualization of the embolization procedure. Like predicate devices the QuadraSphere™ Microspheres are available in a range of sizes to permit selection of the most appropriate size for target vessels. All cited embolization agents are intended for single use and are provided sterile.

7. PERFORMANCE TESTING

In-vitro and in-vivo design verification and validation testing demonstrates that the BioSphere Medical, Inc. QuadraSphere™ Microspheres fulfill design and performance specifications.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH AND HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing right, arranged in a stacked formation, resembling a bird-like shape.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Biosphere Medical, Inc. % Medical Device Consultants, Inc. Rosina Robinson, RN, Med, RAC Principal Consultant 49 Plain Street North Attleboro, Massachusetts 02760

NOV - 7 2006

Re: K052742

Trade/Device Name: QuadraSphere™ Microspheres Regulation Number: 21 CFR 870.3300 Regulation Name: Vascular embolization device Regulatory Class: II Product Code: KRD, HCG Dated: August 15, 2006 Received: August 16, 2006

Dear Ms. Robinson:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set

{3}------------------------------------------------

Page 2 – Rosina Robinson, RN, Med, RAC

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Mark N. Melkerson
Director
Division of General, Restorative
and Neurological Devices

Rumm
DEA
ARECTOR

Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

K052742

Indications for Use

510(k) Number (if known): K052742_____________________________________________________________________________________________________________________________________

Device Name: _________________________________________________________________________________________________________________________________________________________________

Indications for Use:

QuadraSphere Microspheres are indicated for embolization of hypervascularized arteriovenous malformations.

Prescription UseX (Part 21 CFR 801 Subpart D)
AND/OROver-The-Counter Use(21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER

PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off)

Division of General, Restorative,

and Neurological Devices

510(k) Number4052742
------------------------

§ 870.3300 Vascular embolization device.

(a)
Identification. A vascular embolization device is an intravascular implant intended to control hemorrhaging due to aneurysms, certain types of tumors (e.g., nephroma, hepatoma, uterine fibroids), and arteriovenous malformations. This does not include cyanoacrylates and other embolic agents, which act by polymerization or precipitation. Embolization devices used in neurovascular applications are also not included in this classification, see § 882.5950 of this chapter.(b)
Classification. Class II (special controls.) The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Vascular and Neurovascular Embolization Devices.” For availability of this guidance document, see § 870.1(e).